K192694 · Microaire Surgical Instruments · QPB · May 14, 2020 · General, Plastic Surgery
Device Facts
Record ID
K192694
Device Name
PAL Single-Use Cannulas
Applicant
Microaire Surgical Instruments
Product Code
QPB · General, Plastic Surgery
Decision Date
May 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MicroAire® PAL® Single-Use Cannulas are indicated for the removal of tissue or fluid from the body during general surgical procedures, including suction lipoplasty for the purpose of aesthetic body contouring.
Device Story
MicroAire PAL Single-Use Cannulas are stainless-steel, blunt-tipped hollow tubes used for tissue/fluid removal during general surgery and suction lipoplasty. The device attaches to a PAL-730 Manual Wand and connects to an external vacuum source via standard tubing. It is a non-powered, non-reciprocating instrument. The device is supplied sterile via gamma irradiation for single-use. It is operated by surgeons in clinical settings. The device facilitates aesthetic body contouring by allowing the surgeon to manually aspirate tissue/fluid. Benefits include reduced risk of cross-contamination compared to multi-use devices due to single-use design and pre-sterilization.
Clinical Evidence
Bench testing only. Testing included sterilization validation (ISO 11137), biocompatibility (ISO 10993: cytotoxicity, sensitization, irritation), mechanical tensile testing (fatigue, axial strength, pullout/pushout force, bend), air leakage, tubing connection force, vacuum pressure, and transportation testing (ASTM D4169). All tests met specifications.
Indicated for removal of tissue or fluid from the body during general surgical procedures, including suction lipoplasty for aesthetic body contouring.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
MicroAire Power Aspiration Device (PAD) System Cannula (K981922)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MicroAire Surgical Instruments % Diane Sudduth Sr. Consultant, Regulatory Barile & Associates, Inc. PO Box 5199 Clinton, New Jersey 08809
June 9, 2021
Re: K192694
Trade/Device Name: PAL Single-Use Cannulas Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Diane Sudduth:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 14, 2020. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue.
May 14, 2020
MicroAire Surgical Instruments % Diane Sudduth Sr. Consultant, Regulatory Barile & Associates, Inc. PO Box 5199 Clinton, New Jersey 08809
# Re: K192694
Trade/Device Name: PAL Single-Use Cannulas Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: April 6, 2020 Received: April 8, 2020
Dear Ms. Sudduth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{3}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K192694
Device Name MicroAire® PAL® Single-Use Cannulas
Indications for Use (Describe)
MicroAire® PAL® Single-Use Cannulas are indicated for the removal of tissue or fluid from the body during general surgical procedures, including suction lipoplasty for the purpose of aesthetic body contouring.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------|--|
| <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510K Summary
# PAL® Single-Use Cannulas K 192694
#### 1. Submission Sponsor
MicroAire Surgical Instruments 3590 Grand Forks Boulevard Charlottesville, Virginia 22911 USA Office Phone: (434) 975-8344 Email: glenn.gerstenfeld@microaire.com Contact: Glenn Gerstenfeld Title: Vice President, QA/RA and Compliance Officer
#### 2. Submission Correspondent
Barile & Associates, Inc. PO Box 5199 Clinton, NJ 08809 Office Phone: (561) 305-5075 Email: drsudduth05@gmail.com Contact: Diane Sudduth, Ph.D. Title: Senior Consultant, RA
#### 3. Date Prepared
September 20, 2019 (amended April 21, 2020)
#### 4. Device Identification
| Trade/Proprietary Name: | PAL® Single-Use Cannulas |
|-------------------------|-------------------------------------|
| Common/Usual Name: | Cannula |
| Classification Name: | Suction Lipoplasty System |
| Regulation Number: | 21 CFR 878.5040 |
| Product Code: | MUU |
| Device Class: | Class II |
| Classification Panel: | GENERAL AND PLASTIC SURGERY DEVICES |
{6}------------------------------------------------
## 5. Legally Marketed Predicate Device(s)
| Predicate Device: | K171286, MicroAire® PAL System, Multi-Use Cannula |
|----------------------|------------------------------------------------------------------|
| Classification Name: | Suction Lipoplasty System |
| Regulation Number: | 21 CFR 878.5040 |
| Product Code: | MUU |
| Classification: | Class II |
| Reference Device: | K981922, MicroAire® Power Aspiration Device (PAD) System Cannula |
| Classification Name: | Suction Lipoplasty System, |
| Regulation Number: | 21 CFR 878.5040 |
| Product Code: | MUU |
| Classification: | Class II |
#### 6. Indication for Use Statement
The MicroAire PAL Single-Use Cannulas are indicated for the removal of tissue or fluid from the body during general surgical procedures, including suction Lipoplasty, for the purpose of aesthetic body contouring.
### 7. Device Description
The MicroAire® PAL liposuction family of instruments is intended to be used for the removal of tissue or fluid from the body during general surgical procedures, including suction Lipoplasty for the purpose of aesthetic body contouring. The MicroAire® PAL Single-Use Cannulas consists of stainless-steel cannula with a plastic hub which attaches to the MicroAire® PAL-730 Manual Wand and has a connection for a standard vacuum (suction) tubing. The MicroAire® Single-Use Cannulas are supplied sterile by the manufacturer through a validated gamma irradiation process. The Single-Use Cannula are offered in various diameters (2.4, 3.0, 4 and 5.0mm diameters), lengths (15, 22, and 30 cm) and with various types of distal fenestrations (Single Port, Mercedes, Bent Flared Mercedes, Double Mercedes, Bent Mercedes, Tri-Port III, Helixed Tri-Port III, Spatula, Multi-Hole, Del Vecchio Track-12).
#### 8. Substantial Equivalence Discussion
The following table compares the MicroAire PAL Single-Use Cannulas to the predicate device MicroAire PAL Multi-Use Cannulas and the reference device MicroAire PAD System Cannula with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{7}------------------------------------------------
| Manufacturer | MicroAire Surgical Instruments | | Comparison |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Trade Name | MicroAire PAL Single-Use<br>Cannula | MicroAire PAL Multi-Use<br>Cannula | |
| 510k Number | K192694 | K171286 | N/A |
| Product Code | MUU | MUU | Same |
| Regulation<br>Number | 21 CFR 878.5040 | 21 CFR 878.5040 | Same |
| Regulation Name | Suction Lipoplasty System | Suction Lipoplasty System | Same |
| Indications for<br>Use | Indicated for the removal of<br>tissue or fluid from the body<br>during general surgical<br>procedures, including suction<br>Lipoplasty, for the purpose of<br>aesthetic body contouring | Indicated for the removal of<br>tissue or fluid from the body<br>during general surgical<br>procedures, including suction<br>Lipoplasty, for the purpose of<br>aesthetic body contouring | Same |
| Performance<br>Characteristics | • Designed to be used with<br>the PAL-730 Manual Wand<br>• Available in 2.4mm, 3mm,<br>4mm and 5mm Diameters<br>• Available in 15cm, 22cm,<br>and 30cm Lengths | • Designed to be used with<br>the PAL-730 Manual Wand<br>• Available in 2.4mm, 3mm,<br>4mm and 5mm Diameters<br>• Available in 15cm, 22cm,<br>30cm and 40cm Lengths | Similar; multi-use<br>available in 40cm<br>length |
| Available<br>Fenestration<br>types | Single Port, Mercedes, Flared<br>Mercedes, Double Mercedes,<br>Tri-port II, Mirrored Tri-port,<br>Tri-port III, Spatula, Track-12,<br>Helixed Tri-Port, Spatula,<br>Multi-Hole, Bent Mercedes,<br>Bent Flared Mercedes | Mercedes, Flared Mercedes,<br>Double Mercedes, Tri-port II,<br>Mirrored Tri-port, Tri-port III,<br>Spatula, Track-12, Helixed Tri-<br>Port, Multi-Hole, Bent<br>Mercedes, Bent Flared<br>Mercedes | Similar; single-use<br>available in single<br>port and spatula<br>fenestration type |
| Function | Used for removal of tissue or<br>fluid from the body during<br>general surgical procedures<br>including suction Lipoplasty<br>for the purpose of aesthetic<br>body contouring | Used for removal of tissue or<br>fluid from the body during<br>general surgical procedures<br>including suction Lipoplasty<br>for the purpose of aesthetic<br>body contouring | Same |
| Mechanism of<br>action | The cannulas attach to the<br>PAL-730 Manual Wand and<br>suction tubing via a cannula<br>hub. This tubing connects to<br>an external suction source<br>(not part of the PAL System). | The cannulas attach to the<br>PAL-730 Manual Wand and<br>suction tubing via a cannula<br>hub. This tubing connects to<br>an external suction source<br>(not part of the PAL System). | Same |
| | | | Table 5A - Comparison of Characteristics with Predicate Device | | | |
|--|--|--|----------------------------------------------------------------|--|--|--|
|--|--|--|----------------------------------------------------------------|--|--|--|
{8}------------------------------------------------
| Manufacturer | MicroAire Surgical Instruments | | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Trade Name | MicroAire PAL Single-Use<br>Cannula | MicroAire PAL Multi-Use<br>Cannula | |
| | Fluid and tissue are removed<br>from the body via the tubing<br>assisted by the external<br>suction source. The<br>handpiece is not powered<br>and does not reciprocate. | Fluid and tissue are removed<br>from the body via the tubing<br>assisted by the external<br>suction source. The<br>handpiece is not powered<br>and does not reciprocate. | |
| Material | Cannula is a blunt tipped<br>hollow tube comprised of 304<br>stainless steel. Cannula Hub<br>is comprised of a Release<br>Tab, made of Dupont Zytel<br>ST801 (nylon), a turbo plug<br>made of Silicon 70 Blue, USP<br>Class VI, and the hub itself<br>made of Polystyrene 825<br>bonded together with<br>cyanoacrylate adhesive | Cannula is a blunt tipped<br>hollow tube comprised of 304<br>stainless steel. Cannula Hub<br>is comprised of a Release<br>Tab, made of Dupont Zytel<br>ST801 (nylon), a Dowel Pin,<br>made of 410 or 416 or 18-8<br>Stainless Steel, and the hub<br>itself made of Vectra MT1310<br>bonded together with<br>cyanoacrylate adhesive | Similar; hub material<br>is polystyrene vs.<br>Vectra MT1310 |
| Sterile | Supplied Sterile; Gamma<br>Irradiation | Supplied non-sterile; Steam<br>Sterilized by end user | Different; single-use<br>is supplied gamma<br>sterilized |
| Single-Use | Yes | No | Different |
| Shelf Life | 1 year | N/A | Different |
| Complies with<br>ISO 10993-1 | Yes | Yes | Same |
# Table 5B – Comparison of Characteristics with Reference Device
| Manufacturer | MicroAire Surgical Instruments | | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | MicroAire PAL Single-Use<br>Cannula | MicroAire PAD System<br>Cannula | |
| 510k Number | K192694 | K981922 | N/A |
| Product Code | MUU | MUU | Same |
| Regulation<br>Number | 21 CFR 878.5040 | 21 CFR 878.5040 | Same |
| Regulation Name | Suction Lipoplasty System | Suction Lipoplasty System | Same |
| Manufacturer | MicroAire Surgical Instruments | | Comparison |
| Trade Name | MicroAire PAL Single-Use<br>Cannula | MicroAire PAD System<br>Cannula | |
| Indications for<br>Use | Indicated for the removal of<br>tissue or fluid from the body<br>during general surgical<br>procedures, including suction<br>lipoplasty, for the purpose of<br>aesthetic body contouring | For the removal of tissue or<br>fluid from the body during<br>general surgical procedures,<br>including suction lipoplasty,<br>for the purpose of aesthetic<br>body contouring | same |
| Function | Used for removal of tissue or<br>fluid from the body during<br>general surgical procedures<br>including suction lipoplasty<br>for the purpose of aesthetic<br>body contouring | Used for removal of tissue or<br>fluid from the body during<br>general surgical procedures<br>including suction lipoplasty<br>for the purpose of aesthetic<br>body contouring | Same |
| Material | Cannula is a blunt tipped<br>hollow tube comprised of 304<br>stainless steel. Cannula Hub<br>is comprised of a Release<br>Tab, made of Dupont Zytel<br>ST801 (nylon), a turbo plug<br>made of Silicon 70 Blue, USP<br>Class VI, and the hub itself<br>made of Polystyrene 825<br>bonded together with<br>cyanoacrylate adhesive | Cannula is a blunt tipped<br>hollow tube comprised of 304<br>stainless steel with an<br>attached PVC suction tube<br>and optional syringe. The<br>cannula hub is comprised of a<br>Release Tab, made of Dupont<br>Zytel ST801 (nylon) and the<br>hub is made of Polystyrene<br>825. | same |
| Sterile | Supplied Sterile; Gamma<br>Irradiation | Supplied sterile; gamma<br>irradiation | same |
| Single-Use | Yes | Yes | Same |
| Shelf Life | 1 year | 4 years | Similar; testing is<br>ongoing to support a<br>4-year shelf life for<br>the Single-Use<br>Cannula, however,<br>was not available at<br>the time of<br>submission |
| Complies with<br>ISO 10993-1 | Yes | Yes | Same |
{9}------------------------------------------------
The PAL Single-Use Cannulas share the same indications for use, and same or similar device operation, overall technical and functional capabilities, and therefore are substantially equivalent to the predicate and reference devices. The PAL Single-Use Cannulas are similar in design and function to
{10}------------------------------------------------
the predicate and reference devices for the modes of operation and use. The PAL Single-Use Cannulas share the same indications for use, and same or similar device operation, overall technical and functional capabilities, and follow those of the reference device. The PAL Single-Use Cannula is sterilized by gamma irradiation, equal to the reference device.
## 9. Non-Clinical Performance Data
The following tests were performed to demonstrate that the proposed MicroAire PAL Single-Use Cannulas met the applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO and other international standards.
- . Sterilization validation per ISO 11137 - demonstrates SAL of 10°
- . Biocompatibility testing – Cytotoxicity, Sensitization, Irritation, per ISO 10993 - PASSED
In addition to the guidance and standards testing, the following testing was performed:
- . Mechanical Tensile Testing (fatigue, axial strength, pullout force, pushout force, bend) – Met specifications
- Air leakage and tubing connection force test - Met specifications
- Vacuum pressure testing - Met specification
- Transportation Testing per ASTM D4169 – demonstrates package integrity maintained
#### 10. Statement of Substantial Equivalence – Conclusions Drawn
The MicroAire PAL® Single-Use Cannulas, as designed and manufactured, are determined to be substantially equivalent to the predicate and reference devices. Nonclinical testing demonstrates the subject device is substantially equivalent to the legally marked predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.