# PatCom Single-Use Introducer (K192691)

_H&A Mui Enterprises, Inc. · FED · Feb 21, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192691

## Device Facts

- **Applicant:** H&A Mui Enterprises, Inc.
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** Feb 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Single-Use Introducer is used for patient esophageal intubation to guide the catheters and tubes via the nasal or oral cavity. It is to be used in conjunction with endoscopes to allow visualization of the placement process. The Single-Use Introducer is to be used in a hospital or clinical setting and under the supervision of a qualified healthcare professional who has received professional training in using the equipment. The Single-Use Introducer is provided non-sterile, and is intended to be non-reusable. The Single-Use Introducer is for transient use (under 24 hrs), and will come in direct contact with the patient's mucosal lining.

## Device Story

Single-use introducer; facilitates esophageal intubation for catheter/tube placement. Device consists of 30cm PVC tube with open tip and two-piece polycarbonate cone. Used with endoscope for visualization; endoscope inserted into introducer; assembly guided through nasal/oral passage to pharynx. Visual aid prevents nasopharyngeal/oropharyngeal wall aggravation. Once positioned, endoscope extracted; introducer remains as open channel for catheter/tube insertion. After placement, introducer extracted along catheter length; cone halves pulled apart for removal/disposal. Used in hospital/clinical settings by trained healthcare professionals. Benefits: minimizes pharyngeal wall aggravation during intubation; simplifies placement of catheters lacking internal lumens for guidewires.

## Clinical Evidence

Bench testing only. Verification testing included dimensional verification, visual inspection, kink resistance, and tensile strength. Validation testing confirmed user needs and intended use. No clinical data presented.

## Technological Characteristics

Materials: medical grade polyvinyl chloride (PVC) and polycarbonate (PC). Dimensions: 30cm length, 5-6mm OD, 4-5mm ID. Single lumen, open tip, two-piece cone design. Non-sterile, single-use. No energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Enteroscopy Overtube ([K100081](/device/K100081.md))

## Reference Devices

- Motility Catheters ([K823701](/device/K823701.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 21, 2020

H&A Mui Enterprises Inc. Tammy Mui Operations Manager 145 Traders Blvd. E., Unit #34 L4Z 3L3 Mississauga, Ontario CANADA

Re: K192691

> Trade/Device Name: PatCom Single-Use Introducer Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: December 13, 2019 Received: December 18, 2019

Dear Tammy Mui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shani P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K192691

Device Name PatCom Single-Use Introducer

#### Indications for Use (Describe)

The Single-Use Introducer is used for patient esophageal intubation to guide the catheters and tubes via the nasal or oral cavity. It is to be used in conjunction with endoscopes to allow visualization of the placement process. The Single-Use Introducer is to be used in a hospital or clinical setting and under the supervision of a qualified healthcare professional who has received professional training in using the equipment.

The Single-Use Introducer is provided non-sterile, and is intended to be non-reusable.

The Single-Use Introducer is for transient use (under 24 hrs), and will come in direct contact with the patient's mucosal lining.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Mui Scientific. The logo consists of a circle with a dot in the center and three smaller circles around the perimeter. To the right of the circle is the text "Mui Scientific" in a serif font.

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 www.muiscientific.com Email: mail@muiscientific.com

February 10, 2020

# 510(k) Summary

### RE: PatCom Single-Use Introducer

Summary prepared by:

Contact Person: Tammy Mui Title: Operations Manager Manufacturer: H&A Mui Enterprises Inc., o/a Mui Scientific Address: 145 Traders Blvd. East, Unit #33-34, Mississauga, Ontario, Canada L4Z 3L3 Phone: (905) 890-5525 Fax: (905) 890-3523 Email: tammy.mui@muiscientific.com

Trade name: PatCom Single-Use Introducer Common name: Introducer Classification name: Endoscopes and accessories (as per CFR 876.1500)

This 510(k) Summary is for the PatCom Single-Use Introducer

### Indications for Use:

The Single-Use Introducer is used for patient esophageal intubation to guide the catheters and tubes via the nasal or oral cavity. It is to be used in conjunction with endoscopes to allow visualization of the placement process.

The Single-Use Introducer is to be used in a hospital or clinical setting and under the supervision of a qualified healthcare professional who has received professional training in using the equipment.

The Single-Use Introducer is provided non-sterile, and is intended to be non-reusable.

The Single-Use Introducer is for transient use (under 24 hrs), and will come in direct contact with the patient's mucosal lining.

### Device Description:

The PatCom Single-Use Introducers are made from single lumen, medical grade polyvinyl chloride plastics. The tubing length is 30cm, with an open tip at one end, and a polycarbonate cone made of 2 halves at the other end.

This device is designed to work in conjunction with endoscopes to make the placement of certain catheters and tubes more pleasant for the patient and easier for the end user. The endoscope is placed inside the introducer and both devices together are guided through the

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Mui Scientific. The logo consists of a circular atom-like symbol on the left, followed by the text "Mui Scientific" on the right. There are two horizontal lines above and below the text.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 Website: www.muiscientific.com Email: mail@muiscientific.com

nasal or oral passage into the pharynx, via the visual aid provided by the endoscope. The purpose of the visual aid is to prevent unnecessary aggravation of the nasopharyngeal or oropharvngeal wall. Once the introducer is in its desired position, the endoscope is extracted.

The introducer then functions as an open channel from the mouth or the tip of the nose through to the distal opening end of the introducer, as a quide for the placement of catheters and tubes, especially ones that do not contain their own open lumen for use of guidewires that are standard in the gastrointestinal industry. By feeding the catheters or tubes through the introducer, it again minimizes any possible aggravation of the pharyngeal wall as the catheters and tubes would follow the tunnel created by the introducer.

After the catheter is inserted through the introducer and placed into its proper positioning, the introducer is then extracted out of the nasal or oral cavity along the length of the catheter. Once the entire introducer is exposed, the user grips the 2 halves of the cone and pulls apart the introducer to remove and discard.

We are claiming equivalence to the following predicate device:

| Product Name         | 510(k) Number | Manufacturer                  |
|----------------------|---------------|-------------------------------|
| Enteroscopy Overtube | K100081       | United States Endoscopy Group |

The PatCom Single-Use Introducers are similar to the overtubes manufactured by United States Endoscopy Group in that they have the same intended purpose, where both devices require the use of a scope, are slid over top of the endoscopic device, and are used to aid in the guidance of medical devices to be inserted into the upper gastrointestinal tract.

We would also like to acknowledge a reference device, listed below:

| Product Name       | 510(k) Number | Manufacturer                                    |
|--------------------|---------------|-------------------------------------------------|
| Motility Catheters | K823701       | H&A Mui Enterprises Inc., o/a Mui<br>Scientific |

The PatCom Single-Use Introducers are similar to the motility catheters manufactured by Mui Scientific in that they are also made of the same polyvinyl chloride (PVC) and polycarbonate (PC) material as the motility catheters, and are assembled following similar manufacturing procedures. They are both also intubated into a patient via the same nasal or oral cavity pathway.

Below, please find a comparison table of the PatCom Single-Use Introducer with the predicate device and reference device:

|                | Submission Device                                        | Predicate Device                 | Reference Device                           |
|----------------|----------------------------------------------------------|----------------------------------|--------------------------------------------|
| Trade name     | PatCom Single-Use<br>Introducer                          | Enteroscopy Overtube             | Motility Catheters                         |
| 510K holder    | Submitter: H&A Mui<br>Enterprises, o/a Mui<br>Scientific | United States Endoscopy<br>Group | H&A Mui Enterprises,<br>o/a Mui Scientific |
| 510K number    | K192691                                                  | K100081                          | K823701                                    |
| Indication for | The Single-Use Introducer is used for                    | The Enteroscopy                  | To be inserted via the                     |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Mui Scientific. The logo consists of a circular symbol to the left of the text "Mui Scientific". The word "Scientific" is underlined. 

Division of Enterprises Inc. 145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3 Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523 www.muiscientific.com Email: mail@muiscientific.com Website:

| use              | patient esophageal intubation to<br>guide the catheters and tubes via the<br>nasal or oral cavity. It is to be used in<br>conjunction with endoscopes to allow<br>visualization of the placement<br>process. The Single-Use Introducer is<br>to be used in a hospital or clinical<br>setting and under the supervision of a<br>qualified healthcare professional who<br>has received professional training in<br>using the equipment. The Single-Use<br>Introducer is provided non-sterile, and<br>is intended to be non-reusable. The<br>Single-Use Introducer is for transient<br>use (under 24 hrs), and will come in<br>direct contact with the patient's<br>mucosal lining. | Overtube is indicated for<br>use to aid the insertion,<br>advancement and<br>removal of appropriately<br>sized endoscopes and<br>endoscopic devices<br>during diagnostic and<br>therapeutic endoscopic<br>procedures in the upper<br>gastrointestinal tract,<br>including the small<br>intestine. | nasal passage, orally, or<br>anorectally, to measure<br>the muscle contractions<br>along the<br>gastrointestinal system. |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Device OD        | 5mm-6mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not found in the predicate<br>510(k) Summary                                                                                                                                                                                                                                                      | Various (2.5mm-6.2mm)                                                                                                    |
| Device ID        | 4mm-5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not found in the predicate<br>510(k) Summary                                                                                                                                                                                                                                                      | Various (0.4mm-3.3mm)                                                                                                    |
| Material         | PVC, PC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Not found in the predicate<br>510(k) Summary                                                                                                                                                                                                                                                      | PVC, PC, stainless steel                                                                                                 |
| Length           | 30cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Not found in the predicate<br>510(k) Summary                                                                                                                                                                                                                                                      | Various (100cm<br>standard)                                                                                              |
| Number of uses   | Single-use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Single-use                                                                                                                                                                                                                                                                                        | Single-use or reusable                                                                                                   |
| Sterility status | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Non-sterile                                                                                                                                                                                                                                                                                       | Non-sterile                                                                                                              |

Nonclinical bench testing was carried out on the PatCom Single-Use Introducer to verify and validate its performance and intended use. Individual verification testing was conducted to verify the performance requirements and specifications of the introducer, such as dimensional verification, visual inspection, kink resistance testing, and tensile strength testing. Validation testing was then conducted on the user's needs and intended use. All testing and results were successfully completed, thereby demonstrating that the device is as safe, as effective, and performs as well as or better than the legally marketed device predicate.

---

**Source:** [https://fda.innolitics.com/device/K192691](https://fda.innolitics.com/device/K192691)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).