VariAx Foot
K192675 · Stryker GmbH · HRS · Nov 22, 2019 · Orthopedic
Device Facts
| Record ID | K192675 |
| Device Name | VariAx Foot |
| Applicant | Stryker GmbH |
| Product Code | HRS · Orthopedic |
| Decision Date | Nov 22, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The VariAx Foot system is intended for use in internal fixation or arthrodesis of small bones including the fore, mid- and hind foot and ankle. Examples of these procedures include but are not limited to: - Replantation - Lag screw techniques - Joint fusions - Corrective osteotomies - Treatment of fractures
Device Story
VariAx Foot is an orthopedic internal fixation system comprising bone plates, washers, and self-tapping locking/non-locking screws. Used by orthopedic surgeons in clinical settings for small bone procedures (foot/ankle). Modifications to the predicate system include adding a cutting flute to 2.7mm and 3.5mm screws and updating the Tox head design on 2.7mm screws. The device provides mechanical stabilization of bone segments to facilitate healing or fusion. Surgeons select appropriate screw lengths (8-70mm) and plate configurations based on patient anatomy and fracture/deformity type. The system functions via standard orthopedic surgical techniques for bone fixation.
Clinical Evidence
No clinical data. Substantial equivalence supported by engineering rationale regarding mechanical performance (insertion torque, pull-out force) and LAL testing for endotoxin limits (<20 EU/device).
Technological Characteristics
Metallic bone fixation system (plates, screws, washers). Features self-tapping screws with cutting flutes and Tox head interfaces. Screw diameters: 2.7mm and 3.5mm; lengths: 8-70mm. Mechanical fixation principle. No software or electronic components.
Indications for Use
Indicated for patients requiring internal fixation, reconstruction, or arthrodesis of small bones in the fore, mid-, and hind foot and ankle, including fracture treatment, joint fusions, and corrective osteotomies.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Stryker Foot Plating System (K063875)
Reference Devices
- VariAx Elbow System (K101056)
Related Devices
- K141992 — VARIAX 2 SYSTEM · Stryker Trauma AG · Aug 7, 2014
- K170771 — VariAx 2 Foot System · Stryker GmbH · May 8, 2017
- K161913 — AOS Small Fragment Plating System · Advanced Orthopaedic Solutions, Inc. (Aos) · Nov 4, 2016
- K151879 — VariAx 2 One-Third Tubular Plating System · Stryker GmbH · Aug 6, 2015
- K151769 — Anchorage 2 CP System · Stryker GmbH · Oct 27, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2019
Stryker GmbH Jemin Dedania Associate Manager Regulatory Affairs 325 Corporate Drive Mahwah, New Jersey 07430
Re: K192675
Trade/Device Name: VariAx Foot Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 24, 2019 Received: September 26, 2019
Dear Jemin Dedania:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192675
Device Name VariAx Foot
Indications for Use (Describe)
The VariAx Foot system is intended for use in internal fixation or arthrodesis of small bones including the fore, mid- and hind foot and ankle. Examples of these procedures include but are not limited to:
- · Replantation
- · Lag screw techniques
- · Joint fusions
- · Corrective osteotomies
- Treatment of fractures
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | VariAx Foot |
| Common Name: | Screw, Fixation, Bone<br>Plate, Fixation, Bone |
| Regulation Description: | Smooth or threaded metallic bone fixation fastener<br>Single/Multiple component metallic bone fixation<br>appliances and accessories |
| Regulation Number: | 21 CFR 888.3030, 21 CFR 888.3040 |
| Product Code: | HWC (Screw, Fixation, Bone/Washer)<br>HRS (Plate Fixation, Bone) |
| Device Class: | Class II |
| Sponsor: | Stryker GMBH<br>Bohnackerweg 1<br>2545 Selzach / Switzerland |
| Contact Person: | Jemin Dedania<br>Associate Manager Regulatory Affairs<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201)-831-6461<br>Fax: (201) - 831-6500 |
| Additional Contact Person: | Christine Adib<br>Regulatory Affairs Intern |
| Date Prepared: | 08/26/2019 |
| Primary Predicate: | Stryker Foot Plating System (K063875) |
| Reference Predicate Device: | VariAx Elbow System (K101056) |
#### Description
VariAx Foot is a system used for internal fixation applications. VariAx Foot consists of self-fapping locking and non-locking screws, as well as corresponding washers and compatible bone plates. Since its original submission (K063875), the screws of the system have been modified. The modification includes the addition of a cutting flute to the 2.7mm and 3.5mm locking and nonlocking screws and a change in the Tox head of the 2.7mm screws. The lengths of the 2.7mm
{4}------------------------------------------------
diameter screws range from 8-50mm. The lengths of the 3.5mm diameter screws range from 10-70mm.
# Indications for Use
VariAx Foot is intended for use in internal fixation, reconstruction or anthrodesis of small bones including the fore, mid- and hind foot and ankle. Examples of these procedures may include but are not limited to:
- . Replantation
- lag screw techniques
- joint fusions ●
- . corrective osteotomies
- treatment of fractures.
## Summary of Technologies
A comparison of the systems demonstrated that the subject VariAx Foot is substantially equivalent to the Stryker Foot Plating System (K063875) in regard to intended use, material, design, and operational principles.
## Non-Clinical Testing
An engineering rational has been provided to demonstrate equivalence of the subject device to the predicate with respect to insertion torque and pull out force.
LAL testing was performed to establish that the subject devices meet the less than 20 EU/device limit.
### Clinical Testing
Clinical Testing was not required for this submission.
### Conclusion
The subject VariAx Foot system described in this submission has the same indications, intended use, target patient population, technological characteristics, and materials as the predicate devices. The subject VariAx Foot system is substantially equivalent to the predicate Stryker Foot Plating System (K063875).