← Product Code QSY · K192667

# gel-e Flex+ gel OTC (K192667)

_Gel-E, Inc. · QSY · Oct 25, 2019 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192667

## Device Facts

- **Applicant:** Gel-E, Inc.
- **Product Code:** QSY
- **Decision Date:** Oct 25, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

gel-e Flex+ gel OTC is indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

## Device Story

gel-e Flex+ gel OTC is a topical chitosan-based gel designed to promote rapid hemostasis (30-60 seconds) in bleeding wounds. The device consists of a viscous gel (palmitoyl-N-acetylglucosamine in lactic acid) dispensed from a syringe. It is intended for over-the-counter use by patients or caregivers for minor lacerations and abrasions. The gel is applied directly to the wound site, optionally with gauze. By creating a physical barrier and promoting clotting, it aids in bleeding control. The device is non-sterile and has a 2-year shelf life.

## Clinical Evidence

No human clinical data provided. Evidence includes in vivo preclinical studies in a controlled acute swine skin laceration model, demonstrating hemostatic performance equivalent to the predicate. Bench testing included pH, viscosity, and biocompatibility (ISO 10993-1: cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity). Preservative effectiveness was validated per USP <51>.

## Technological Characteristics

Composition: palmitoyl-N-acetylglucosamine (chitosan) in 0.1M lactic acid. Form factor: 5 mL and 10 mL syringes. Principle: topical hemostatic gel. Sterilization: non-sterile. Shelf life: 2 years. Performance standards: USP <51> for preservative effectiveness; ISO 10993-1 for biocompatibility.

## Regulatory Identification

To temporarily control bleeding and cover external wounds.

## Predicate Devices

- gel-e Flex+ ([K182811](/device/K182811.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.

April 21, 2023

gel-e, Inc. Elsa Abruzzo Head of Regulatory 387 Technology Dr., Suite 3110B College Park, Maryland 20742

Re: K192667 Trade/Device Name: gel-e Flex+ gel OTC Regulatory Class: Unclassified Product Code: QSY

Dear Elsa Abruzzo:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 25, 2019. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

October 25, 2019

gel-e, Inc. Elsa Abruzzo Head of Regulatory 387 Technology Dr., Suite 3110B College Park, Maryland 20742

Re: K192667

Trade/Device Name: gel-e Flex+ gel OTC Regulatory Class: Unclassified Product Code: FRO Dated: September 24, 2019 Received: September 25, 2019

Dear Elsa Abruzzo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{2}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Lixin Liu -S

For Cynthia J. Chang, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192667

Device Name gel-e Flex+ gel OTC

#### Indications for Use (Describe)

gel-e Flex+ gel OTC is indicated for the local management of bleeding wounds such as minor lacerations and minor abrasions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{4}------------------------------------------------

# GEL-E FLEX+ GEL OTC 510(k) Summary

# 510(k) Summary

| A. Name and Address of Applicant : | gel-e, Inc.<br>387 Technology Dr., Suite 3110B<br>College Park, MD 20742                         |
|------------------------------------|--------------------------------------------------------------------------------------------------|
| B. Contact Person:                 | Matthew Dowling, PhD<br>Chief Scientific Officer<br>Phone: (301) 405-3585<br>Fax: (301) 314-9592 |
| C. Date of Submission :            | Sept 24, 2019                                                                                    |
| D. Device Trade Name:              | gel-eFlex+ gel OTC                                                                               |
| E. Device Common Name:             | Dressing, Wound, Drug                                                                            |
| F. Device Classification:          | Unclassified Device (pre-amendment)                                                              |
| G. Classification Name:            | Unclassified                                                                                     |
| H. Product Code:                   | FRO                                                                                              |
| I. Predicate Device:               | gel-e's gel-e Flex+ (K182811)                                                                    |

- J. Intended Use:
Over-The-Counter Use (21 CFR 801 Subpart C): gel-e Flex+ gel OTC is indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

- K. Device Description:
gel-e Flex+ gel OTC is a laboratory tested non-irritating topical chitosan-based gel designed to rapidly (within 30 to 60 seconds) promote hemostasis when in contact with a bleeding wound. Gel-e Flex+ gel OTC is a clear, convenient, and easy to apply gel that can be used with commercially available gauze. In vitro preservative effectiveness testing based on USP<51> has demonstrated that gel-e Flex+ gel OTC remains effectively preserved for a period of 2 years by meeting the acceptance criteria for 5 strains of microbial challenge organisms.

{5}------------------------------------------------

## L. Performance Data

#### Animal Studies

In vivo preclinical studies were conducted in a controlled acute swine model of skin lacteration to evaluate the performance of both the predicate device (gel-e Flex+: K182811) and the subject device. The swine model was selected based on published comparisons evaluating the effectiveness of hemostatic devices and agents. Both the predicate and subject devices operate by the same mechanism of action using the same materials. In all instances, the gel-e Flex+ gel OTC functioned as intended and the control of bleeding observed was as expected.

## Preservative Effectiveness Testing

The preservative effectiveness of gel-e Flex+ gel OTC to prevent growth of microorganisms within the dressing has been established in accordance with the requirements of USP <51> (Preservative Effectiveness Testing).

#### Biocompatibility and Bench Testing

Representative samples of the device underwent testing including bench testing (pH, viscosity), biocompatibility testing per ISO 10993-1 (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity), packaging testing (burst pressure and dye penetration testing), shelf-life, and use-life stability testing. The performance of gel-e Flex+ gel OTC was substantially equivalent to the predicate device, gel-e Flex+.

#### M. Summary of Device Modification and Substantial Equivalence:

Gel-e has submitted information on modification of gel-Flex+ gel OTC relative to the predicate device. Those changes are: (1) non-sterile processing and (2) extension of use life (i.e. life after initially opening the product). Gel-e has also submitted information on indication for use, design and principle of operation, biocompatibility and performance characteristics to establish that gel-Flex+ gel OTC is substantially equivalent to the currently marketed predicate device, gel-e Flex+. gel-e Flex+ gel OTC has the same intended use/indication for use as the predicate device.

{6}------------------------------------------------

|                          | Gel-e Flex+ gel OTC                                                                                                                                 | Predicate:<br>Gel-e Flex+ (Gel)<br>(K182811)                                                  |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Manufacturer             | GEL-E, Inc.                                                                                                                                         | GEL-E, Inc.                                                                                   |
| Classification           | Unclassified                                                                                                                                        | Same                                                                                          |
| Product Code             | FRO                                                                                                                                                 | Same                                                                                          |
| Indications for Use      | OTC: gel-e Flex+ gel OTC is indicated for the local<br>management of bleeding wounds such as minor cuts,<br>minor lacerations and minor abrasions.  | OTC: Same                                                                                     |
| Device Design            | Viscous gel dispensed from a syringe that may be<br>used with gauze, bandage or by itself                                                           | Same                                                                                          |
| Material                 | Gel-e Flex+ gel OTC is composed of a semi-<br>transparent gel of palmitoyl-N-acetylglucomasine<br>(chitosan) dissolved in 0.1M lactic acid in water | Same                                                                                          |
| Sizes                    | 5 mL and 10 mL syringe                                                                                                                              | Same                                                                                          |
| Weight                   | 10 g / 20 g                                                                                                                                         | Same                                                                                          |
| Sterility                | Not provided sterile                                                                                                                                | 10-6 SAL – Terminally sterilized with gamma radiation,<br>for single patient use within 24 h. |
| Use Life                 | 2 years                                                                                                                                             | 28 days                                                                                       |
| Performance<br>Standards | pH, viscosity, biocompatibility, animal efficacy testing<br>and preservative effectiveness testing                                                  | Same                                                                                          |

# Comparison of gel-e Flex+ and Predicate Device

---

**Source:** [https://fda.innolitics.com/device/K192667](https://fda.innolitics.com/device/K192667)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
