Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal.
Device Story
Cordless endo motor for root canal preparation; assists dentists/dental surgeons in standard endodontic procedures. Device consists of main unit, charging station, and adapter; requires external FDA-cleared contra-angle handpiece (16:1 ratio) to hold drill bits/files. User controls speed, torque, and auto-reverse/auto-forward modes via buttons on main unit. Powered by rechargeable Li-ion battery. Used in dental clinic settings. Provides mechanical rotation for root canal files; auto-reverse/stop function activates when torque resistance exceeds user-defined settings. Benefits include controlled mechanical preparation of root canals, reducing manual effort and improving procedural consistency.
Clinical Evidence
No clinical data provided. Safety and performance established via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and dental equipment performance (IEC 80601-2-60). Cleaning and disinfection validated per AAMI TIR30 and AAMI TIR12.
Technological Characteristics
Portable cordless endo motor; ABS enclosure. Powered by 3.7Vdc 1050mAh Li-ion battery. Interfaces with 16:1 contra-angle handpiece. Features adjustable speed, torque (0.3-3.0 Ncm), and auto-reverse/stop modes. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 80601-2-60. Not sterile; not autoclavable.
Indications for Use
Indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal by dentists and dental surgeons.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 12, 2021
Foshan CICADA Dental Instrument Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong CHINA
Re: K192649
Trade/Device Name: Endo Motor Model: T-Fine-II Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories Regulatory Class: Class I, reserved Product Code: EKX Dated: February 7, 2021 Received: February 10, 2021
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192649
Device Name Endo Motor Model: T-Fine-II
### Indications for Use (Describe)
Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Foshan CICADA Dental Instrument Co, Ltd Sponsor: Subject Device: Endo Motor, Model: T-Fine-II. 510(k) submission report (V1.3), Chapter 6 File No.: File File File
# Chapter 6. 510(k) Summary
# 510(k) Summary - K192649
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 872.6070.
#### 1. Submitter Information
Sponsor: Foshan CICADA Dental Instrument Co, Ltd .
Address: B5-2F, Guangdong New Light Source Industrial Base, Shihan Town, Nanhai District, Foshan,
Guangdong, China
Contact Person: Juan Liu
Phone: 86-757-85775667
E-mail: 149713771@qq.com
Application Correspondent: Jet Li
Company: Guangzhou KEDA Biological Technology Co., Ltd
E-mail: med-jl @foxmail.com
Phone: 86-18588874857
Address: 6F, No.1 TianTairoad, Science City, LuoGang District, GuangZhou City, China
#### Subject Device Information 2.
- � Type of 510(k) submission: Traditional
- � Common Name: Handpiece, Direct Drive, Ac-Powered
- Trade Name: Endo Motor Model: T-Fine-II �
- Classification Name: Dental Handpiece and accessories �
{4}------------------------------------------------
Subject Device: Endo Motor, Model: T-Fine-II.
File No.: 510(k) submission report (V1.3), Chapter 6
- Review Panel: Dental �
- � Product Code: EKX
- � Regulation Number: 21 CFR 872.4200
- � Regulation Class: 1
#### 3. Primary Predicate Device Information
Manufacturer: W&H Dentalwerk Buermoos GmbH Cordless ENDO-Handpiece "ENTRAN" (W&H) Device: Cordless ENDO-Handpiece "S5 ENDO Motor 510(k) number: K090931
#### Device Description 4.
Cordless endo motor is a professional equipment for root canal preparation. It is a drive for contra angle headpiece which is designed to assist dentists and dental surgeons perform standard endodontic procedures.
The endo motor consists of Cordless main unit, Charger station and Adapter. It is intended to use with FDA cleared contra-angle handpiece (transmission ratio of contra angle handpiece: 16:1) which would hold the drill bit or file. However the device do not included the contra angle handpiece and files. The Endo smart is equipped with rechargeable Li-ion battery, which can be recharged using the provided charge station. By means of the different buttons on the main unit, the user could control various functions in the device, such as On/off, Speed, Torque and auto-reverse/auto forward mode. The Endo motor is not sterile and cannot be autoclaved.
#### 5. IntendedUse/ Indication for Use
Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodonticfiles for mechanical and rotary preparation of the root canal.
#### 6. Test Summary
{5}------------------------------------------------
Subject Device: Endo Motor, Model: T-Fine-II.
File No.: 510(k) submission report (V1.3), Chapter 6
ENDO Motor had been evaluated for its safety and performance by lab bench testing according to the following standards:
#### A Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device, and was found to comply with IEC 60601-1 and 60601-1-2, IEC80601-2-60.
Detail standard lists is as below items:
- -IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 2005+A1:2012
- -IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, 2014
- । IEC 80601-2-60:2012 Medical electrical equipment — Part 2-60: Particular requirements for basic safety and essential performance of dental equipment
*Cleaning and Disinfection vas conducted according to AAMI TIR30 and AAMI TIR12.
#### Software Verification and Validation Testing A
Software verification and validation testing were conducted and its documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was determined to be of "moderate" level of concern.
# 7. Clinical Testing
Clinical data were not required in this submission to support a finding of substantial equivalence.
#### 8. Comparison to Predicate De vice
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise any new questions of safety or effectiveness.
{6}------------------------------------------------
Subject Device: Endo Motor, Model: T-Fine-II.
> File No.: File File File 510(k) submission report(V1.3), Chapter 6
| Item | Subject Device | Predicate Device | Remark |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Name and mode | Endo Motor, T-Fine-II | Cordless ENDO-Handpiece "ENTRAN" (W&H) | -- |
| 510K number | K192649 | K090931 | - |
| Manufacturer | Foshan CICADA Dental Instrument Co., Ltd | W&H Dentalwerk Buermoos GmbH | - |
| Intended Use | Endo Motor T-Fine-II is indicated for use in standard endodontic procedures using rotary endodontic files for mechanical and rotary preparation of the root canal. | Modular electrical system for mechanical preparation of the root canal, using a special root canal instrument, which is intended by the manufacturer for use in the mechanical and rotary preparation of the root canal. | Minor difference<br>Note 1: |
| Rx or OTC | Rx | Rx | Same |
| Physical Attributes | | | |
| Weight | 200g | 120g | Minor difference;<br>Note 2 |
| Displayed Parameters | Speed/Ratio, Torque, Volume, Antomatic control mode, | Torque, Speed | Minor Different,<br>Note 3 |
| Other displayed information | Program number, Power display | Power display | Minor Different,<br>Note 3 |
| Electrical Power | | | |
| DC Mains | AC100~240V 50/60Hz<br>Output voltage: 5Vdc<br>Output current: 1A | AC100~240V 50/60Hz<br>Output voltage: 5Vdc<br>Output current: 1A | Minor difference<br>Note 4 |
| Battery | Battery Voltage: 3.7Vdc | Battery Voltage: | |
| Battery capacity:<br>1050mAh | 3.7Vdc<br>Battery capacity:<br>680mAh | | |
| Environmental Operation | | | |
| Temperature | 10~40°C | 0~40°C | Minor difference<br>Note 5 |
| Humidity | <80%RH | 15%~80% | |
| Environmental Storage | | | |
| Temperature | -40℃~55℃ | -40℃~70℃ | |
| Humidity | <80%RH | 8%~80% | |
| Performance | | | |
| transmission<br>ratio of electric<br>contra angle | 16:1 | 16:1 | Same |
| Torque | 0.3-3.0 Ncm | 0.5-4.0 Ncm | Minor difference<br>Note 6 |
| Auto-control<br>mode | Yes | Yes | Same |
| Funtion can be<br>settable | Speed, Torque, and<br>Auto-control setting | Speed, Torque and<br>Auto-control setting | Same |
| Lubricant for<br>motor | N/A | N/A | Same |
| Material | | | |
| Enclosure of<br>Main unit | ABS | ABS | Same |
| Performance standard | | | |
| Standard | AAMI/IEC60601-1<br>IEC60601-1-2<br>IEC 80601-2-<br>60:2012 | IEC60601-1;<br>IEC60601-1-2 | Same |
| Device shape | | | |
| Outlook<br>appearance | Portable | Portable | Same |
{7}------------------------------------------------
Subject Device: Endo Motor, Model: T-Fine-II.
> File No.: File File File 510(k) submission report (V1.3), Chapter 6
# Note for Detailed comparison
1. For the statement of intended use, even there is only minor difference on the words for indented use, but its meaning of the description So it has the same intended use as the predicate device (K090931).
2. There is difference of the weight between T-Fine-II and the predicate device. That is because the appearance of T-Fine-II is different from the predicate device, but the design of subject device comply with electrical safety standard IEC/AAMI 60601-1, so the minor difference would not affect its safety and Report by Foshan CICADA Dental Instrument Co, Ltd Page 17 of 50
{8}------------------------------------------------
Foshan CICADA Dental Instrument Co, Ltd Sponsor: Subject Device: Endo Motor, Model: T-Fine-II. File No.: File File File 510(k) submission report (V1.3), Chapter 6
### effectiveness.
.
3. The displayed parameters of T-Fine-Il are different from the predicated device, but the design of subject device comply with electrical safety standard IEC/AAMI 60601-1 and display function had been validated in software evaluation report, so the minor difference would not affect its safety and effectiveness.
4. There is minor difference on the battery capacity from 1050mAh, but the design of subject device comply with electrical safety standard IEC/AAM 60601-1, so the minor difference would not affect its safety and effectiveness.
5. There is minor difference on storage temperature and humidity of T-Fine-II, comparing with the predicate device, the device shall be stored according to the specified environment, so that it do not affect the function of the device. So the minor difference would not affect its safety and effectiveness.
6. Even there is minor difference on the torque setting range, however it is adjustable in the main unit by dentist operator, the device provide with auto reverse and stop function when motor bear the torque resistance higher than the setting torque: and the device comply with IEC80601-2-60 Requirement. So such minor difference would not affect its safety and effectiveness
# Conclusion
The subject device Endo motor has all features of the predicate device for intended use. Thus, the subject device is substantially equivalent to the predicate device.
# 9. Summary Prepared Date
10 Mar 2021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.