← Product Code [NXC](/productcode/NXC) · K192596

# ULab Systems Dental Aligner Kit (K192596)

_Ulab Systems, Inc. · NXC · Feb 11, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192596

## Device Facts

- **Applicant:** Ulab Systems, Inc.
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Feb 11, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The uLab Systems Dental Aligner is indicated for the alignment of permanent teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.

## Device Story

System consists of custom-made, removable, clear plastic aligners; used for orthodontic treatment of malocclusions. Dental professional scans or takes physical impressions of patient teeth; digital data processed via uLab Systems uDesign software to generate treatment plan and intermediate tooth states. Practitioner reviews and approves plan; software converts files to 3D models for thermoforming aligners. Patient wears aligners in sequence as prescribed; continuous gentle forces move teeth to target positions. Benefits include gradual, aesthetic tooth alignment. Used in dental offices; operated by dentists or orthodontists.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 (Parts 5 and 10) for materials and finished devices, confirming safety. Design verification, process validation, user validation, and packaging/shelf-life testing conducted to confirm device meets requirements and user needs.

## Technological Characteristics

Clear, removable, thermoformed thermoplastic aligners (Zendura A polyurethane or Zendura FLX copolyester/polyurethane composite). Principle of operation: sequential programmed displacement via continuous gentle force. Non-sterile. Software-driven design process using digital dental scans.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Orthocaps Twinaligner ([K180241](/device/K180241.md))

## Reference Devices

- Derby Dental Laboratory Custom Clear Aligner System ([K173785](/device/K173785.md))
- Sybron Dental Specialties Ormco Spark Aligner System ([K182826](/device/K182826.md))
- uLab Systems uDesign ([K171295](/device/K171295.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features an abstract design of a human figure. To the right of this is the FDA logo, with the letters FDA in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 11, 2020

uLab Systems, Inc. % Sylvia Erickson Regulatory Consultant Sylvia Erickson Consulting 157 Ruby Avenue San Carlos, California 94070

Re: K192596

Trade/Device Name: ULab Systems Dental Aligner Kit Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: January 3, 2020 Received: January 3, 2020

Dear Sylvia Erickson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192596

Device Name uLab Systems Dental Aligner

Indications for Use (Describe)

The uLab Systems Dental Aligner is indicated for the alignment of permanent teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  |

| X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K192596

## 510(k) Summary

This summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### Applicant Information:

uLab Systems, Inc. 3 Lagoon Drive Suite 180 Redwood City, CA 94065

Contact Person: Charlie Wen Phone: 650-804-1397

### Submission Correspondent:

Sylvia Erickson Principal, Sylvia Erickson Consulting

### Device Information:

| Trade Name:                | uLab Systems Dental Aligner Kit |
|----------------------------|---------------------------------|
| Common Name:               | Sequential Aligner              |
| Classification Name:       | Orthodontic Plastic Bracket     |
| Classification Regulation: | 21CFR 872.5470                  |
| Device Class:              | II                              |
| Product Code:              | NXC                             |

### Primary Predicate:

Ortho Caps GmBH Orthocaps Twinaligner, K180241

### Reference Predicates:

Derby Dental Laboratory Custom Clear Aligner System, K173785 Sybron Dental Specialties Ormco Spark Aligner System, K182826 uLab Systems uDesign, K171295

#### Date Prepared:

February 10, 2020

#### Device Description:

{4}------------------------------------------------

The uLab Systems Dental Aligner Kit is an orthodontic treatment system, that consists of doctorprescribed, clear, thin, plastic removable aligners. Patients are seen in a dental office where a dental professional (dentist/orthodontist) scans or takes a physical impression of the teeth, which is then used to fabricate the aligners. The aligners are designed and custom-made by dental professionals using uLab Systems proprietary technology. The aligners are made of aesthetic and medical grade materials chosen by doctors for their patients. During the orthodontic treatment, each preformed plastic aligner is worn in sequence by the patient as prescribed by the dental practitioner, moving the patient's teeth gradually to the ideal position.

### Indications for Use:

The uLab Systems Dental Aligner is indicated for the alignment teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.

## Comparison of Intended Use and Technological Characteristics with the Predicate Device:

The subject and predicate devices share intended use for alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.

The subject and predicate device are based on the following same technological elements:

- Both devices are clear plastic sequential aligners.
- . Both devices have the same intended use for alignment of teeth during orthodontic treatment of malocclusions by way of continuous gentle forces.
- . Both devices have the same indications for use.
- Both devices have the same principle of operation.
- Both devices are fabricated based on an orthodontic treatment plan designed by the aligner manufacturer's technicians from digital scans of the patient's teeth in an untreated state.
- Both devices are made of thermoplastic.

The only differences between the subject and device are the following:

- . The subject device may alternatively be fabricated based on an orthodontic treatment plan designed by a dental practitioner.
- . The subject device materials are the same material used in the reference predicates, whereas the primary predicate device material is an unspecified thermoplastic. The subject and primary predicate devices otherwise share the same technological characteristics.
- The subject device is not designated for use during a specific time of day.

{5}------------------------------------------------

| Attribute                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | uLab Systems Dental Aligner (uLab<br>Systems)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthocaps Twinaligner (Ortho Caps<br>GmBH)<br>K180241                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications<br>for Use       | The ULab System Dental Aligner is<br>indicated for the alignment of permanent<br>teeth during orthodontic treatment of<br>malocclusions by way of continuous<br>gentle forces.                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Orthocaps TwinAligner® System<br>is indicated for the alignment of<br>teeth during orthodontic treatment<br>of malocclusions by way of<br>continuous gentle forces.                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device<br>Description        | A series of custom-made removable clear<br>plastic orthodontic aligners that<br>sequentially position teeth by way of<br>continuous gentle force.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | A series of custom-made removable<br>clear plastic orthodontic aligners that<br>sequentially position teeth by way of<br>continuous gentle force.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Principles of<br>Operation   | Each preformed plastic tray is worn in<br>sequence by the patient as prescribed by<br>the dental practitioner.<br><br>Orthodontic movement occurs through<br>continuous gentle forces applied to the<br>dentition as each tooth follows the<br>programmed displacement based on a<br>doctor's prescription.                                                                                                                                                                                                                                                                                                                                       | Each preformed plastic tray is worn in<br>sequence by the patient as<br>prescribed by the dental practitioner.<br><br>Orthodontic movement occurs<br>through continuous gentle forces<br>applied to the dentition as each tooth<br>follows the programmed<br>displacement based on a doctor's<br>prescription.                                                                                                                                                                                                                                                                                                      |
| Aligner<br>Design<br>Process | Dental software, the uLab Systems<br>uDesign, K171295, for tooth alignment<br>uses digital scan (untreated state) to<br>generate the image of a final, provisional<br>treated state and then interprets a series<br>of images that represent intermediate<br>teeth states. The dental practitioner then<br>reviews these images and has the option<br>to reject, make or request modifications<br>to the set-up prior to approving it for<br>aligner fabrication. Once the dental<br>practitioner approves the treatment plan,<br>the software converts the files to<br>produce the series of 3D models used to<br>produce thermoformed aligners. | Standard dental software for tooth<br>alignment uses digital scan<br>(untreated state) to generate the<br>image of a final, provisional treated<br>state and then interprets a series of<br>images that represent intermediate<br>teeth states. The dental practitioner<br>then reviews these images and has<br>the option to reject or request<br>modifications to the set-up prior to<br>approving it for aligner fabrication.<br>Once the dental practitioner<br>approves the treatment plan, the<br>software converts the files to<br>produce the series of 3D models used<br>to produce thermoformed aligners. |
| Material                     | Zendura A (thin thermoformed<br>polyurethane) or Zendura FLX<br>(copolyester and polyurethane<br>composite)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Thermoplastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Features                     | Aligners made from different materials as<br>prescribed by physician                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Aligners made from different<br>materials as prescribed by physician<br>for daytime or nighttime use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| OTC or Rx                    | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Attribute                    | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                              | uLab Systems Dental Aligner (uLab<br>Systems)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthocaps Twinaligner (Ortho Caps<br>GmBH)<br>K180241                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterilization                | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

{6}------------------------------------------------

## Performance Data:

The following performance data were provided to demonstrate safety and efficacy in support of substantial equivalence determination:

- Biocompatibility testing for the aligner materials was completed in accordance with ISO 10993 per the following:
	- O Part 5: Cytotoxicity Elution -MEM
	- Part 10: Intracutaneous/Intradermal Reactivity O
	- Part 10: Maximimzation for Delayed-Type Hypersensitivity O
	- Part 10: Oral Mucosa Irritation Test O

The test results demonstrate that the material is biocompatible for the intended use.

Additionally, biocompatibility testing for the finished device was completed in accordance with ISO 10993 per the following:

- Part 5: Cytotoxicity Elution -MEM о
- The test results demonstrate the finished device is biocompatible for the intended use.
- Design verification and process validation testing was completed, demonstrating that the manufacturing process and finished device meet product requirement specifications.
- . User validation testing was completed, demonstrating that the device confirms to the user needs and intended use.
- . Packaging verification and shelf life testing was completed, supporting the labeled shelf life of the device.
- Physical properties testing has been provided by the material manufacturer.

### Summary:

The uLab Systems Dental Aligner Kit has the same intended use as the predicate device. In addition, it has similar technological characteristics; performance data demonstrates that any differences in technological characteristics do not raise different questions of safety or effectiveness. Therefore, the uLab Systems Dental Aligner is substantially equivalent to the cleared predicate device.

---

**Source:** [https://fda.innolitics.com/device/K192596](https://fda.innolitics.com/device/K192596)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
