ExSpiron 2Xi is indicated for use by healthcare professionals in healthcare facilities, such as postoperative care and critical care units, to monitor breathing in patients at least one year of age. ExSpiron 2Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of: • Minute Ventilation (MV) • Tidal volume (TV) • Respiratory rate (RR) ExSpiron 2Xi measurements are used as an adjunct to other clinical information.
Device Story
ExSpiron 2Xi is a non-invasive respiratory monitor for use in healthcare facilities by professionals. It uses thoracic bioimpedance via a single-patient-use electrode padset placed on the patient's torso to measure respiratory signals. A stabilized high-frequency (50 kHz) current generator delivers current, and an adaptive circuit conditions the resulting voltage signal. A tablet PC processes these signals using a proprietary non-linear regressive model to calculate and display minute ventilation, tidal volume, and respiratory rate. The device provides a graphical trace of lung volume over time. It serves as an adjunct to other clinical information, aiding healthcare providers in assessing patient respiratory status. The system includes a nurse call relay for clinical alerts. The device is intended to benefit patients by providing continuous, non-invasive respiratory monitoring.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing, including software verification, accuracy testing (MV 1.57%, TV -2.28%, RR 1.55%), system power/battery verification, nurse call functionality, and usability testing per IEC 60601-1-6.
Technological Characteristics
Thoracic bioimpedance sensing; 50 kHz current signal. ABS plastic enclosure. Digital data acquisition board; Linux tablet PC. Connectivity includes integrated nurse call relay. Standards: ANSI AAMI IEC 60601-1, 60601-1-6, 62304, 62366-1, ISO 14971, IEC 62133-2.
Indications for Use
Indicated for patients at least 1 year of age in healthcare facilities (e.g., post-operative, critical care) for non-invasive monitoring of breathing parameters.
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2019
Respiratory Motion, Inc. Susan Hamann Director of Quality and Regulatory 80 Coolidge Hill Road Watertown, Massachusetts 02472
Re: K192595
Trade/Device Name: ExSpiron 2Xi Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: Class II Product Code: BZK, BZO Dated: December 2, 2019 Received: December 3, 2019
Dear Susan Hamann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192595
Device Name ExSpiron 2Xi
Indications for Use (Describe)
ExSpiron 2Xi is indicated for use by healthcare professionals in healthcare facilities, such as post-operative care and critical care units, to monitor breathing in patients at least 1 - year of age.
ExSpiron 2Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:
- Minute Ventilation (MV)
- Tidal volume (TV)
- Respiratory Rate (RR)
ExSpiron 2Xi measurements are used as an adjunct to other clinical information.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – Special 510(k) K192595
| 510(k) Owner: | Respiratory Motion, Inc. |
|--------------------------------------------------|--------------------------------------------------|
| Address: | 80 Coolidge Hill Road<br>Watertown, MA USA 02472 |
| Phone: | 781-373-1636 |
| Fax: | 781-373-1653 |
| Contact person: | Jenny Freeman, MD |
| Date 510(k) Summary prepared and type of 510(k): | December 12, 2019 Special 510(k) |
| Trade name: | ExSpiron 2Xi |
| Common name: | Respiratory Monitoring System |
| Classifications: | Regulation Number: 21 CFR 868.1850 |
{4}------------------------------------------------
| Regulation Name: | Monitoring Spirometer |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Class: | II |
| Product Code: | BZK, BZQ |
| Predicate devices: | ExSpiron 1Xi Respiratory Monitor, marketed by Respiratory Motion, Inc, Watertown, MA (K173181). |
| Device Description: | The ExSpiron 2Xi Respiratory Monitor System consists of:<br>Monitor: The Monitor contains a bioimpedance measurement system and a tablet PC housed within a single enclosure.<br>Bioimpedance measurement system: The Monitor incorporates a stabilized high frequency current generator and an adaptive circuit that conditions the resulting voltage signal and converts it to digital form. Firmware within the Monitor performs signal acquisition and relays data to the tablet PC.<br>Computer: A tablet PC performs signal processing and calibration and runs the graphical user interface (GUI). The PC takes user input from a touch screen through a virtual keyboard and mouse. The GUI is used for recording patient data and displaying the respiratory trace as well as scalar values and trends for minute ventilation, tidal volume, and respiratory rate.<br>Patient Cables and Electrode Padsets that are also included in the system but were cleared in previous 510(k)s: K130170, K162131, K173181<br>ExSpiron Patient Cable: A reusable cable that connects the ExSpiron 2Xi Monitor to the Electrode PadSet.<br>Single patient use ExSpiron Electrode PadSet: A Single-Patient Use Electrode PadSet is placed on the skin of the patient's torso. It delivers current to, and records impedance measurements from, the skin. |
| Intended use: | ExSpiron 2Xi is indicated for use by healthcare professionals in healthcare facilities, such as postoperative care and critical care units, to monitor breathing in patients at least one year of age. ExSpiron 2Xi is a non-invasive monitor that graphically displays lung volume against time and reports an approximate value of:<br>• Minute Ventilation (MV)<br>• Tidal volume (TV)<br>• Respiratory rate (RR)<br>ExSpiron 2Xi measurements are used as an adjunct to other clinical information |
{5}------------------------------------------------
### Special 510(k) Summary Continued
## Comparison to cleared device
| | ExSpiron® 2Xi | ExSpiron® 1Xi (Predicate) | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Proposed Device | (K173181) | Comment |
| Intended Use | ExSpiron 2Xi is indicated for use by<br>healthcare professionals in<br>healthcare facilities, such as<br>postoperative care and critical care<br>units, to monitor breathing in<br>patients at least one year of age.<br>ExSpiron 2Xi is a non-invasive<br>monitor that graphically displays<br>lung volume against time and<br>reports an approximate value of:<br>• Minute Ventilation (MV)<br>• Tidal volume (TV)<br>• Respiratory rate (RR)<br>ExSpiron 2Xi measurements are<br>used as an adjunct to other clinical<br>information. | ExSpiron 1Xi is indicated for use by<br>healthcare professionals in<br>healthcare facilities, such as<br>postoperative care and critical care<br>units, to monitor breathing in<br>patients at least one year of age.<br>ExSpiron 1Xi is a non-invasive<br>monitor that graphically displays<br>lung volume against time and<br>reports an approximate value of:<br>• Minute Ventilation (MV)<br>• Tidal volume (TV)<br>• Respiratory rate (RR)<br>ExSpiron 1Xi measurements are<br>used as an adjunct to other clinical<br>information. | Both the proposed device and the<br>predicate have identical Intended<br>Use |
| Technology | Measurement is by thoracic<br>bioimpedance. | Measurement is by thoracic<br>bioimpedance. | Both the proposed device and the<br>predicate have identical fundamental<br>scientific technology. |
| Volume<br>Measurements | Tidal volume<br>Minute volume<br>Volume vs. time chart | Tidal volume<br>Minute volume<br>Volume vs. time chart | Both the proposed device and the<br>predicate measure the same<br>respiratory volumes. |
| Rate<br>Measurements | Respiratory rate (breaths/min) | Respiratory rate (breaths/min) | Both the proposed device and the<br>predicate measure respiratory rate. |
| Safety | IEC 60601-1, including electrical<br>and mechanical safety | IEC 60601-1, including electrical<br>and mechanical safety | Both the proposed device and the<br>predicate have an equivalent<br>electrical safety profile and are<br>compliant with IEC 60601-1. |
| Energy Source | The bioimpedance output signal is<br>generated internally to the<br>ExSpiron Monitor. It is a low<br>current signal at approximately 50<br>kHz. | The bioimpedance output signal is<br>generated internally to the<br>ExSpiron Monitor. It is a low<br>current signal at approximately 50<br>kHz. | Both the proposed device and the<br>predicate have equivalent energy<br>source. |
| Power<br>Management<br>Board | Board is external to the tablet<br>computer. | Board is internal to the tablet<br>computer. | Functionality of Power Management<br>Board remains unchanged. The two<br>boards are substantially equivalent. |
| Data Acquisition<br>Board | Digital Data Acquisition Board. | Analog Data Acquisition Board. | Updated Data Acquisition Board<br>performs the same core functions as<br>the predicate along with built in Nurse<br>Call Circuitry and additional electric<br>isolation. Safety and effectiveness of<br>differences have been addressed |
| | | | through relevant performance testing. |
| Algorithm | A software algorithm calculates<br>respiratory parameters using a<br>proprietary, non-linear,<br>regressive model trained<br>with a dataset containing<br>respiratory data from a<br>large historical cohort. | A software algorithm calculates<br>respiratory parameters using a<br>proprietary, non-linear, regressive<br>model trained with a dataset<br>containing respiratory data from a<br>large historical cohort. | Both the proposed device and the<br>predicate have the same<br>fundamental algorithm |
| Enclosure | Injection molded plastic (ABS)<br>housing, flame-retardant. | Formed plastic (ABS) housing,<br>flame-retardant. | Both the proposed device and the<br>predicate have equivalent enclosure<br>material. |
| Graphical User<br>Interface | Graphical User Interface allowed<br>input of demographic patients at<br>least 1 year of age. | Graphical User Interface allowed<br>input of demographic patients at<br>least 1 year of age. | The Graphical User Interface is<br>different in the proposed device<br>owing to the different size of tablets<br>used in the devices, however the<br>content is equivalent. |
| EMC/EMI<br>Compliance | IEC 60601-1-2 compliant | IEC 60601-1-2 compliant | Both the proposed device and the<br>predicate are 60601-1-2 compliant. |
| Electrode PadSet | Single-patient use,<br>biocompatible, printed padset. | Single-patient use, biocompatible,<br>printed padset. | Both the proposed device and the<br>predicate use the same electrode<br>padset. |
| Impedance<br>Measurement<br>Range | 15 Ohms to 180 Ohms | 15 Ohms to 180 Ohms | Both the proposed device and the<br>predicate have the same<br>measurement range. |
| Tablet Computer | Linux tablet computer. | Windows tablet computer. | The proposed device and the<br>predicate use different tablet<br>computers. The functions of the<br>tablet remain the same. |
| Nurse Call Relay | Nurse call relay is contained<br>within the monitor. | Nurse call relay module is an<br>accessory to the monitor. | A nurse call relay has been integrated<br>into the monitor in the proposed<br>device. Performance testing was<br>conducted to demonstrate<br>substantial equivalence. |
| Biocompatibility | Neither the Monitor nor the<br>Patient Cable are intended for<br>patient contact. The Electrode<br>PadSet is biocompatible. | Neither the Monitor nor the Patient<br>Cable are intended for patient<br>contact. The Electrode PadSet is<br>biocompatible. | Both the proposed device and the<br>predicate are biocompatible. |
| Usability | IEC 60601-1-6 compliant | IEC 60601-1-6 compliant | Usability Testing was performed to<br>demonstrate that the 2Xi is equivalent<br>to the 1Xi when used by intended<br>users in the intended environment. |
| Bench Accuracy | | | |
| | MV 1.57%<br>TV - 2.28%<br>RR - 1.55% | MV 1.78%<br>TV - 2.39%<br>RR - 1.38% | The accuracy of the proposed device<br>is substantially equivalent to the<br>predicate. |
{6}------------------------------------------------
{7}------------------------------------------------
### Additional Information:
Summary of Performance and Bench Testing Performed – All the testing passed the verification and validation requirements. The list of tests performed is as follows:
- 1. Software Test Protocol
- 2. Accuracy Test Protocol
- 3. System Power Verification Test
- 4. System Battery Verification Test
- 5. Internal Usability Testing
- 6. Nurse Call Test
- 7. User Manual Verification Test
- 8. Do by Analysis Test This tests requirements that need static analysis, calculations, and visual inspection.
Summary of Clinical Testing performed –Design changes to device were confirmed and validated by performance and bench testing. Clinical testing was not required.
Summary of Biocompatibility testing - Additional testing was not required for the ExSpiron 2Xi. The Padset and cables are identical to previous submissions for the ExSpiron 1Xi. The monitor and patient cables are not intended for patient contact.
Conformance to Standards applicable to the device – The ExSpiron 2Xi met the applicable requirements specified in the standards. The standards used are:
- 1. ANSI AAMI IEC 60601-1:2005/(R)2012 And A1:2012 (Consolidated Text)
- 2.
- 3.
- 4. ANSI AAMI IEC 60601-1-6:2010/AMD1:2013
- 5. ANSI AAMI IEC 62304:2006/A1:2016
- 6. ANSI AAMI IEC 62366-1:2015
- 7. ANSI AAMI ISO 14971:2007/(R)2010 (Corrected 4 October 2007)
- ANSI AAMI IEC 62133-2:2017 8.
Conclusion: Based on the testing and evaluations performed on the ExSpiron 2Xi, the device is as safe, as effective, and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.