← Product Code [OAP](/productcode/OAP) · K192585

# Bosley Revitalizer 96 Laser Cap, Bosley Revitalizer 96 XL Laser Cap (K192585)

_Pjm Worldwide, LLC D/B/A Phoenix Medical Technology, LLC · OAP · Jan 29, 2020 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192585

## Device Facts

- **Applicant:** Pjm Worldwide, LLC D/B/A Phoenix Medical Technology, LLC
- **Product Code:** [OAP](/productcode/OAP.md)
- **Decision Date:** Jan 29, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Bosley Revitalizer 96 Laser Cap is indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

## Device Story

Device is a wearable laser cap for home use; treats Androgenic Alopecia via Low-Level Laser Therapy (LLLT). Input: electrical power from rechargeable Li-ion battery. Transformation: 96 red diode lasers (650 nm) emit light onto scalp; device includes sensors to detect correct head position, automatically pausing therapy if misaligned. Output: visible red light energy. Healthcare provider/patient use: patient self-administers treatment for max 30 minutes every other day. Benefit: promotes hair growth; treatment must be continued indefinitely to maintain results. Output informs patient of active therapy status via position-sensing safety interface.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing verified laser diode output (≤ 5mW), wavelength (650 nm), and safety interface functionality. Compliance with IEC 60601-1, IEC 60601-1-11, IEC 60601-2-22, and IEC 60825-1 confirmed.

## Technological Characteristics

Infrared lamp (OAP); 96 red diode lasers (650 nm); Class 3R laser; rechargeable Li-ion battery; helmet form factor; position-sensing safety interface; standalone operation; IEC 60601-1, IEC 60601-1-11, IEC 60601-2-22, IEC 60825-1, IEC 61959 compliant.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- Illumiflow Laser Cap ([K162071](/device/K162071.md))
- Theradome LH80 PRO ([K122950](/device/K122950.md))
- Capillus 272 ([K143199](/device/K143199.md))
- Capillus 202 ([K150578](/device/K150578.md))
- Capillus 82 ([K151516](/device/K151516.md))

## Reference Devices

- iGrow-II Hair Growth System ([K122248](/device/K122248.md), [K140931](/device/K140931.md), [K141567](/device/K141567.md))
- HairMax LaserComb 82 ([K142573](/device/K142573.md))
- Bosley Revitalizer 272 Laser Cap ([K181253](/device/K181253.md))
- Bosley Revitalizer 164 Laser Cap ([K181253](/device/K181253.md))
- Grivamax Hair Growth System ([K171895](/device/K171895.md))
- Ihelmet Hair Growth System ([K162782](/device/K162782.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

PJM Worldwide, LLC d/b/a Phoenix Medical Technology, LLC % Ronald Berglund Member/Manager Grace Consulting, LLC 6615 Lake Shore Drive, Ste 605 Minneapolis, Minnesota 55423

#### Re: K192585

Trade/Device Name: Bosley Revitalizer 96 Laser Cap, Bosley Revitalizer 96 XL Laser Cap Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: December 27, 2019 Received: January 3, 2020

### Dear Ronald Berglund:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192585

#### Device Name

Bosley Revitalizer 96 Laser Cap

#### Indications for Use (Describe)

The Bosley Revitalizer 96 Laser Cap is indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

Type of Use (Select one or both, as applicable)

| ☐ Registration (Part 21 CFR 801, Subpart D) | ☑ Over-The-Counter Use (21 CFR 801, Subpart G) |
|---------------------------------------------|------------------------------------------------|
|---------------------------------------------|------------------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Special 510(k) Summary

## Date: December 27, 2019

## Submitter's Contact Information:

Name: Ronald E. Berglund, GRACE Consulting, LLC

Address: 6615 Lake Shore Drive #605, Richfield, MN 55423

Establishment Registration #: 3008855505 Owner/Operator #: 10059366

Telephone: (952)220-3014 Facsimile: (952)888-8282

### Name of Device and Name / Address of Sponsor:

| Trade Name:           | Bosley Revitalizer 96 Laser Cap                 |
|-----------------------|-------------------------------------------------|
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth |
| Classification Name:  | Infrared lamp per 21 CFR 890.5500               |
| Classification Code:  | OAP (Laser, Comb, Hair)                         |

Sponsor Contact Information: PJM Worldwide, LLC d/b/a Phoenix Medical Technology, LLC  
1499 Northwest 79th Av., Miami, FL 33126Telephone: (305)477-2515

## Predicate Devices:

Illumiflow Laser Cap (K162071)—Cleared for OTC use; Theradome LH80 PRO (K122950)-Cleared for OTC use; Capillus 272, 202 and 82 LLLT Devices (K143199; K150578; K151516; K153618; K163170 and K160285)

## Reference Devices:

iGrow-II Hair Growth System (K122248; K140931, additional indication for use in females; K141567, original clearance for use in males) HairMax LaserComb 82 (now sold as the HairMax LaserBand) (K142573, hands-free version of the HairMax LaserComb for use in males and females)

{4}------------------------------------------------

#### Intended Use / Indications for Use:

The Bosley Revitalizer 96 Laser Cap is indicated to treat Androgenic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of Ila to V patterns of hair loss and to treat Androgenic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-1 to I-4, II-1, II-2, or frontal; both with Fitzpatrick Skin Types I to IV.

#### Technological Characteristics

The Bosley Revitalizer 96 Laser Cap incorporates the same technology as the Bosley Revitalizer 272 and 164 Laser Caps in every way except the following:

The Bosley Revitalizer 96 Laser Cap has fewer diodes - 96 compared to 272 and 164 for the Bosley Revitalizer 272 and 164 Laser Caps

The Bosley Revitalizer 272 Laser Cap consists of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer cap and protective inner liner, and configured for portable use with rechargeable battery and adapter. The Bosley Revitalizer 164 Laser Cap consists of 164 red, visible light, diode lasers operating at 650 nanometers, configured within an outer cap and protective inner liner, and configured for portable use with rechargeable battery and adapter. The Bosley Revitalizer 96 Laser Cap consists of 96 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner liner, and configured for portable use with rechargeable battery and adapter. Just as with the Bosley Revitalizer 272 and 164 Laser Caps, the 96 Laser Cap system can detect whether or not the device is in the correct position on the user's head and will automatically pause therapy if it is not. The device will resume therapy when the correct head position is re-established. The system is powered by rechargeable Lithium-ion battery cells assembled into a proprietary battery pack. Both the battery pack and charger are fully compliant with recognized, international standards.

{5}------------------------------------------------

#### Design Control Activities:

The Bosley Revitalizer 96 Laser Cap design and development team followed ANSI/AAMI/ISO 14971:2007/(R)2010 Risk Management: Medical devices – Application of risk management to medical devices. Significant changes include assessing the effect a reduction of lasers would have on the end product, both for safety and for efficacy. All residual risks are found acceptable and risk management file and summary are available for review. Based on the Bosley Revitalizer 272 and 164 Laser Caps, the verification and validation activities remain the same (e.g. output of each laser, operation of safety interface, and output of power pack), and were performed and assessed by designated personnel qualified to perform such activities. All methods, tests, and acceptance criteria are stipulated on the verification and validation reports. All documents are available for review.

Safety ratings remained the same; there was no change to the output of the individual lasers (≤ 5mw); however, the "dose" (total delivered energy/cm2) is reduced due to reduced number of lasers. This may affect efficacy; this observation is made based on the performance of other devices similar to Bosley Revitalizer Laser Caps that are currently cleared under device code OAP and utilize much fewer than 272 lasers.

Instructions for use with the Bosley Revitalizer 96 Laser Cap do not change from the instructions provided with the Bosley Revitalizer 272 and 164 Laser Caps. Indications for use and the dose schedule remain exactly the same as for the Bosley Revitalizer 272 and 164 Laser Caps and as for other devices in the same category (product code OAP); which is for maximum 30 minutes, every other day. The output of the laser diodes was verified to remain the same (each diode) as for the previous model. Design control activities were followed per 21 CFR 820.30:

{6}------------------------------------------------

- a) General: The Beijing Toplaser Technology Co., Ltd. Quality Management System is compliant to the requirement of the quality system regulation and specifically to design controls as stipulated by 21 CFR 820.30. The main design activities included the following: 1.) identification of product technical solutions; 2.) risk analysis; and 3.) product validation.
- b) Design and development planning: During project planning activities, the risks were identified and assessed with respect to the proposed design change. The proposed design change was minimal and affected only the number of diodes in the PCB array. The power pack remains exactly the same. Review of the risk assessment from the previous models revealed no significant changes or risks identified for the new model Bosley Revitalizer 96 Laser Cap. Vigilance activities for the Bosley Revitalizer 96 Laser Cap indicated that there were no significant problems identified (no adverse events reported).
- c) Design input: Design requirements for the Bosley Revitalizer 272 and 164 Laser Caps are transferred to the Bosley Revitalizer 96 Laser Cap and an input/output matrix was completed for the project.
- d) Design output: Design requirements for the Bosley Revitalizer 272 and 164 Laser Caps are transferred to the Bosley Revitalizer 96 Laser Cap. Design outputs are verified per performance tests.
- e) Design review: Appropriate design reviews were conducted. Due to the minimal modifications made (i.e., only the reduced number of laser diodes), a technical evaluation and a verification of inputs and outputs (physical values) were completed.
- f) Design verification: All verification activities to assess the impact of the modification (reduction of number of laser diodes) were completed. It was determined that the laser

{7}------------------------------------------------

diode output remained the same, but the total dosage is reduced due to the lower number of laser diodes used with the new model. Since the cap is intended to be used indefinitely (if treatment is stopped the results are lost), the reduced number of laser diodes will likely affect only the time required for users to experience the same results as achieved with the Bosley Revitalizer 272 and 164 Laser Caps. The Bosley Revitalizer 96 Laser Cap is offered as a lower-cost alternative to the original models.

- g) Design validation: Feedback from hair restoration specialists and consumers indicate that a lower-cost alternative is desired. The modestly lower efficacy is accepted by users, with the understanding that treatment is intended to be continued indefinitely.
- h) Design Transfer: The Bosley Revitalizer Laser Cap family of products are manufactured on-site in the facility of the manufacturer in Beijing, China. Design transfer was minimal; training requirement to the new PCB was minimal, and all other components (including overall build) remained the same.
- Design changes: The design of all models follows the manufacturer's established design i) control, change control, and document control procedures.
- i) Design history: The design history file for the Bosley Revitalizer 96 Laser Cap is available for review.

## Performance Data:

All verification and validation activities were performed by designated individuals.

The product design change (reduction of number of laser diodes) was carried out in accordance with the control requirements of design and development activities as detailed in ISO13485 and CFR820, and records of all activities have been maintained.

{8}------------------------------------------------

The Bosley Revitalizer 96 Laser Cap is a new model modified on the basis of the Bosley Revitalizer 272 and 164 Laser Caps. With the new 96 model the number of laser diodes has been reduced from either 272 or 164 to 96. The parameters of the laser such as wavelength, and output power remain unchanged. The control method, software and usage method also remain the same as with the Bosley Revitalizer 272 and 164 Laser Caps. The risk analysis method utilized was failure modes and effects analysis (FMEA)-- a step-by-step approach for identifying all possible failures in a design, a manufacturing or assembly process, or a product or service. FMEA incorporates each of the following risk analysis components:

- "Failure modes" (the ways, or modes, in which something might fail. Failures are any errors or . defects, especially ones that affect the customer, and can be potential or actual);
- · "Effects analysis" (refers to studying the consequences of those failures). Failures are prioritized according to how serious their consequences are, how frequently they occur, and how easily they can be detected. The purpose of FMEA is to take actions to eliminate or reduce failures, starting with the highest-priority failures. Failure modes and effects analysis also documents current knowledge and actions about the risks of failures, for use in continuous improvement. FMEA is initially used during design to prevent failures, but after production has begun it is used for control, before and during ongoing operation of the process.

Based on the risk analysis report (Documents No.: FRZ/TZ-YA-12) produced for the Bosley Revitalizer 272 and 164 Laser Caps, the risks caused by the laser diode reduction of this change have been analyzed one by one. The conclusion is that there is no new unacceptable risk due to the reduced number of laser diodes.

{9}------------------------------------------------

New product samples of the 96-laser diode model have been tested to meet the requirement and the test methods utilized were identical to the test methods utilized for the Bosley Revitalizer 272 and 164 Laser Caps (Test report number: FRZ/JL-T-028). Due to changes in product specifications, the identification label has been modified and new models have been added to the operation manual.

Due to the reduction of the laser diodes, the corresponding product BOM has also been modified. There is, however, no change in the manufacturing process flow. The results of the verification and validation activities (available for review) demonstrate that the predetermined acceptance criteria were met.

Performance testing was conducted to confirm compliance to design specifications. All functions of the modified product have been verified to operate as designed, and all acceptance criteria were met by the new device. The Bosley Revitalizer 96 Laser Cap conforms to the following recognized standards: IEC-60601-1, IEC-60601-11, IEC-60601-2-22 (2007/Third Ed.), and IEC60825-1 (Ed. 3). These IEC standards are recognized and accepted standards by the FDA. The guidance document for these standards is found in the Federal register, July 26, 2001 (volume 66, Number 144). This report validates for the Bosley Revitalizer 96 Laser Cap the laser class of 3R which establishes the AEL (accessible emission limits) as 5 milli Watts maximum. The charger conforms to IEC 61959. The performance data -- available for review-- demonstrates that the Bosley Revitalizer 96 Laser Cap has exactly the same laser wavelength, output power (per diode), output beam, energy type, laser field, and treatment area as the Bosley Revitalizer 272 and 164 Laser Caps cleared under K181253. Total "dose" – delivered energy over time – (J/cm ²) is reduced due to the reduced number of laser diodes. However, since the cap is intended to be used indefinitely (if treatment is stopped, results are lost ), it is believed that the reduced number of laser diodes will 

{10}------------------------------------------------

only affect the time required to witness the same result as achieved with the Bosley Revitalizer 272 and 164 Laser Caps. Just as for the Bosley Revitalizer 272 and 164 Laser Caps (K181253) and reference devices, there have been no reported adverse events for these products.

## Substantial Equivalence

The Bosley Revitalizer 96 Laser Cap is the same technology used by Bosley Revitalizer 272 and

164 Laser Caps (K181253) and other reference devices; specifically, iGrow-II Hair Growth

System (K122248 and K140931). The Bosley Revitalizer 96 Laser Cap is as safe and

(commensurate to number of diodes) effective as the Bosley Revitalizer 272 and 164 Laser Caps,

as well as other reference devices in its class, such as the Hairmax Lasercomb.

| Bosley Revitalizer<br>96 Laser Cap                                                | Bosley Revitalizer<br>272/164 Laser<br>Caps                                         | Illumiflow<br>Laser Cap                            | Grivamax<br>Hair Growth<br>System                  | Capillus272,<br>Capillus202,<br>Capillus82                      | Ihelmet Hair<br>Growth System                                   | Theradome                                                    | 200 mm X 200 mm X 112 mm (L X W X H)                                                                                                     | 180 mm X 180 mm X 95 mm (L X W X H)                                                                                                                              | OAP Classification                                         | OAP Classification                                         | OAP Classification                                                                                                                                                                                                            | OAP Classification                                                                                          | OAP Classification                                          |
|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| K192585                                                                           | K181253                                                                             | K162071                                            | K171895                                            | K163170,<br>K153618,<br>K160285                                 | K162782                                                         | K122950                                                      | OAP Classification                                                                                                                       | OAP Classification                                                                                                                                               | Classification Name - Infrared Lamp                        | Classification Name - Infrared Lamp                        | Classification Name - Infrared Lamp                                                                                                                                                                                           | Classification Name - Infrared Lamp                                                                         | Classification Name - Infrared Lamp                         |
| LLLT Device Type                                                                  | LLLT Device Type                                                                    | LLLT Device<br>Type                                | LLLT Device<br>Type                                | LLLT Device Type                                                | LLLT Device Type                                                | LLLT Device Type                                             | Classification Name -Infrared Lamp                                                                                                       | Classification Name -Infrared Lamp                                                                                                                               | Common Usage Name - Lamp, Non-Heating                      | Common Usage Name - Lamp, Non-Heating                      | Common Usage Name - Lamp, Non-Heating                                                                                                                                                                                         | Common Usage Name - Lamp, Non-Heating                                                                       | Common Usage Name - Lamp, Non-Heating                       |
| OTC Use                                                                           | OTC Use                                                                             | OTC Use                                            | OTC Use                                            | OTC Use                                                         | OTC Use                                                         | OTC Use                                                      | Common Usage Name - Lamp, Non-Heating                                                                                                    | Common Usage Name - Lamp, Non-Heating                                                                                                                            | General & Plastic Surgery Committee                        | General & Plastic Surgery Committee                        | General & Plastic Surgery Committee                                                                                                                                                                                           | General & Plastic Surgery Committee                                                                         | General & Plastic Surgery Committee                         |
| Intended Use -<br>Androgenic<br>Alopecia+promote<br>hair growth                   | Intended Use -<br>Androgenic<br>Alopecia+promote<br>hair growth                     | Intended Use<br>- Androgenic<br>Alopecia           | Intended Use<br>- Androgenic<br>Alopecia           | Intended Use -<br>Androgenic<br>Alopecia+promote<br>hair growth | Intended Use -<br>Androgenic<br>Alopecia+promote<br>hair growth | Intended Use -<br>Androgenic<br>Alopecia in adult<br>females | General & Plastic Surgery Committee                                                                                                      | General & Plastic Surgery Committee                                                                                                                              | Fitzpatrick Skin Photo-types - I- IV                       | Fitzpatrick Skin Photo-types - I- IV                       | Fitzpatrick Skin Phototypes - I- IV                                                                                                                                                                                           | Fitzpatrick Skin Photo-types - I- IV                                                                        | Fitzpatrick Skin Phototypes - I- IV                         |
| Contain Laser<br>Diodes- Class 3R                                                 | Contain Laser<br>Diodes- Class 3R                                                   | Contain Laser<br>Diodes- Class<br>3R               | Contain Laser<br>Diodes- Class<br>3R               | Contain Laser<br>Diodes- Class 3R                               | Contain Laser<br>Diodes- Class 3R                               | Contain Laser<br>Diodes- Class 3R                            | Fitzpatrick Skin Photo-types - I- IV                                                                                                     | Fitzpatrick Skin Photo-types - I- IV                                                                                                                             | Ludwig-Savin I-II (females) Norwood Hamilton IIA-V (males) | Ludwig-Savin I-II (females) Norwood Hamilton IIA-V (males) | Ludwig-Savin I1 to I-4, II-1, II-2, or frontal (females) + Norwood Hamilton Ila-V (males)                                                                                                                                     | Ludwig-Savin 11 to I-4, II-1, II-2, or frontal (females) + Norwood Hamilton Ila-V (males)                   | Ludwig-Savin I-II (females)                                 |
| Helmet/Cap<br>Design                                                              | Helmet/Cap<br>Design                                                                | Helmet/Cap<br>Design                               | Helmet/Cap<br>Design                               | Helmet/Cap<br>Design                                            | Helmet/Cap<br>Design                                            | Helmet/Cap<br>Design                                         | Ludwig-Savin I-1 to I-4, II-1, II-2 or frontal (females) Norwood Hamilton IIA-V (males)                                                  | Ludwig-Savin I-1 to 1-4, II-1, II-2 or frontal (females) Norwood Hamilton IIA-V (males)                                                                          | Efficacy Rates - High Compared to Placebo                  | Efficacy Rates - High Compared to Placebo                  | Efficacy Rates - High Compared to Placebo                                                                                                                                                                                     |                                                                                                             |                                                             |
| 650 nm                                                                            | 650 nm                                                                              | 650 nm                                             | 650 nm                                             | 650 nm                                                          | 650 nm                                                          | 678 nm                                                       | Efficacy Rates - High Compared to Placebo                                                                                                | Efficacy Rates - High Compared to Placebo                                                                                                                        | Treatment- 16 weeks, every other day (30 min. per session) | Treatment- 16 weeks, every other day (30 min. per session) | Treatment- 17 weeks, every other day (indefinite)                                                                                                                                                                             | Treatment- 16 weeks, every other day (20 – 35 min. per session)                                             | Treatment-20 minutes, twice per week (non-consecutive days) |
| Marketing<br>clearance for<br>males and<br>females                                | Marketing<br>clearance for<br>males and<br>females                                  | Marketing<br>clearance for<br>males and<br>females | Marketing<br>clearance for<br>males and<br>females | Marketing<br>clearance for<br>males and<br>females              | Marketing<br>clearance for<br>males and<br>females              | Marketing<br>clearance for<br>females                        | Treatment- 16 weeks, every other day (indefinite)                                                                                        | Treatment-16 weeks, every other day (indefinite)                                                                                                                 | Treatment Area:                                            |                                                            | Treatment Area:             Capillus272:             495.37 cm2             Capillus202:             449.51 cm2             Capillus82: 194.42 cm2             Mathematically             Max. Derived                        | Treatment Area:             424.93 cm2             Mathematically             Max. Derived                  |                                                             |
| Passive Use-<br>Hands Free                                                        | Passive Use-<br>Hands Free                                                          | Passive Use-<br>Hands Free                         | Passive Use-<br>Hands Free                         | Passive Use-<br>Hands Free                                      | Passive Use-<br>Hands Free                                      | Passive Use-<br>Hands Free                                   | Treatment Area:             Bosley 96: 218 cm2             Bosley 96XL: 240cm2             Mathematically             Max. Derived       | Treatment Area:             Bosley272: 509 cm2             Bosley164: 328 cm2             Mathematically             Max. Derived                                | Irradiance (power per area):                               |                                                            | Irradiance (power per area):             Capillus272:             2.7454 mW/ cm2             Capillus202:             2.2469 mW/cm2             Capillus82: 2.1088 mW/cm2             Mathematically             Max. Derived | Irradiance (power per area):             2.3533 mW/ cm2             Mathematically             Max. Derived |                                                             |
| 96 Laser Diodes<br>(5 mW ea.)                                                     | 272 and 164 Laser<br>Diodes (5 mW ea.)                                              | 272 Laser<br>Diodes (5 mW ea.)                     | 272 Laser<br>Diodes (5 mW ea.)                     | 272, 202 and 82<br>Laser Diodes                                 | 200 Laser Diodes                                                | 80 laser diodes                                              | Irradiance (power per area): Bosley 96: 2.2 mW/ cm2 Bosley 96XL: 2.0 mW/ cm2 Mathematically Max. Derived                                 | Irradiance (power per area):             Bosley272:             2.67mW/cm2             Bosley164: 2.5 mW/cm2             Mathematically             Max. Derived | Energy Fluence:                                            |                                                            | Energy Fluence:             Capillus272:             4.9417 J/cm2             Capillus202: 4.044 J/cm2             Capillus82: 3.792 J/cm2             Mathematically             Max. Derived                                | Energy Fluence:             4.9420 J/cm2             Mathematically             Max. Derived                |                                                             |
| Dimensions:<br>Bosley 96: 180<br>mmX180 mm X 95<br>mm (L X W X H)<br>Bosley 96Xl: | Dimensions:<br>Bosley 272: 180<br>mm X 180 mm X<br>95 mm (L X W X H)<br>Bosley 164: |                                                    |                                                    |                                                                 | Dimensions: 266<br>mm X 196 mm X<br>135 mm (L X W X<br>H)       |                                                              | Energy Fluence:             Bosley 96: 3.96 J/cm2             Bosley 96XL: 3.6 J/cm2             Mathematically             Max. Derived | Energy Fluence:             Bosley272: 4.8 J/cm2             Bosley 164: 4.5 J/cm2             Mathematically             Max. Derived                           | Device Class II                                            | Device Class II                                            | Device Class II                                                                                                                                                                                                               | Device Class II                                                                                             | Device Class II                                             |
| Device Class II                                                                   | Device Class II                                                                     |                                                    | ll                                                 |                                                                 |                                                                 |                                                              |                                                                                                                                          |                                                                                                                                                                  |                                                            |                                                            |                                                                                                                                                                                                                               |                                                                                                             |                                                             |

# Predicate Comparison Table

{11}------------------------------------------------

{12}------------------------------------------------

The submitter believes that with the exception of the number of laser diodes and the configuration of the optical elements, the Bosley Revitalizer 96 Laser Cap is the same device in form, function, safety, and efficacy as the previous versions and the predicate device(s). The Hairmax Lasercomb, offered as a reference, is proof of the functionality and acceptability of devices with fewer laser diodes cleared by the FDA in the category of OAP, both technically and clinically. The submitter believes that the difference in the physical appearance, number of diodes, or in the method of delivering the radiant energy of the systems is of no consequence and does not affect the therapeutic value or the safety profile.

All compliant LLLT systems which use red light diode lasers are classified as class 3R laser systems by the IEC standard for allowable emission levels, which is a recognized standard by the FDA as well, and the adverse event profile is the same.

For these reasons, the Bosley Revitalizer 96 Laser Cap satisfies the FDA's requirements (for device modification notification) with respect to intended use, and technological and design characteristics. Additionally, no new safety or efficacy concerns are raised due to the minor differences between these devices.

---

**Source:** [https://fda.innolitics.com/device/K192585](https://fda.innolitics.com/device/K192585)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
