K192568 · Wuxi Jiajian Medical Instrument Co., Ltd. · GXY · Feb 3, 2021 · Neurology
Device Facts
Record ID
K192568
Device Name
Jiajian Self-Adhesive Electrode
Applicant
Wuxi Jiajian Medical Instrument Co., Ltd.
Product Code
GXY · Neurology
Decision Date
Feb 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Self-adhesive electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.
Device Story
Non-sterile, flexible cutaneous electrodes transmit electrical stimulation current (TENS/EMS) to patient skin. Composed of non-woven fabric tape, polyurethane electrically conductive carbon cloth, and biocompatible conductive hydrogel. Available in various shapes (round, rectangle, elliptical, calabash, butterfly, saddle, rhombus, meniscus, palm) and connection types (lead wire or snap button). Used as an interface between patient skin and nerve/muscle stimulators. Operated by patients or clinicians in indoor/outdoor settings. Output is electrical current delivery; healthcare providers use output to facilitate therapeutic stimulation. Benefits include effective, safe electrical conduction for nerve/muscle stimulation.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: lead wire compliance (AAMI/ANSI ES 60601-1), impedance/conformability/fluid tolerance (IEC 60601-2-2), peel strength, shelf life (ASTM F1980), reuse testing, and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Materials: Non-woven fabric, polyurethane conductive carbon cloth, biocompatible hydrogel (ValueTrode Carbon, K970426). Sensing/Actuation: Passive conductive interface. Energy: Conducts electrical stimulation. Dimensions: Various shapes/sizes (e.g., round Ø20-85mm). Connectivity: Lead wire (2mm/2.5mm female socket) or snap button (2.5-5mm male socket). Sterilization: Non-sterile. Standards: AAMI/ANSI ES 60601-1, IEC 60601-2-2, ASTM F1980, ISO 10993.
Indications for Use
Indicated for patients requiring electrical stimulation (TENS/EMS) via cutaneous electrodes. For single patient, multiple application use. Suitable for OTC and prescription use.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 3, 2021
Wuxi Jiajian Medical Instrument Co., Ltd Caihong Sun Manager NO.35 Baigiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu 214116 China
### Re: K192568
Trade/Device Name: Jaijian Self-Adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 4, 2021 Received: January 7, 2021
### Dear Caihong Sun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192568
Device Name Self-adhesive Electrode
Indications for Use (Describe)
The Self-adhesive electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.
Type of Use (Select one or both, as applicable)
| <div> <span></span>For In-Situ Use in OFF-Site Indoor Settings </div> | <div> <span></span>For In-Situ Use in OFF-Site Outdoor Settings </div> |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------|
|--------------------------------------------------------------------------|---------------------------------------------------------------------------|
|X | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Jiajian. The logo is in a teal color and features the text "JIA JIAN" in a sans-serif font. A check mark is used in place of the space between the two words. Below the text is the company name in Chinese characters.
## 510(k) Summary [As required by 21 CFR 807.92]
## 1. Submission Information:
| 510(k) Number: | K192568 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 31, 2020 |
| Type of 510(k) Submission: | Special |
| Submitter/Manufacturer: | Wuxi Jiajian Medical Instrument Co.,Ltd<br>No.35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China<br>214116<br>E-mail: sch@jjmed.com.cn<br>Tel: 86 510-88745788 |
| Contactor: | Doris Dong (Consultant)<br>Shanghai CV Technology Co., Ltd.<br>Rm. 903, No. 19 Dongbao Rd., Songjiang Area, Shanghai, 201613 China<br>E-mail: doris.d@ceve.org.cn<br>Tel: 86 21-31261348 |
## 2. Device Description:
| Proprietary Name: | Self-adhesive Electrode |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Cutaneous electrode |
| Classification Name: | Cutaneous electrode |
| Product Code: | GXY |
| Device Class: | II |
| Regulation Number: | 21 CFR 882.1320 |
| Review Panel: | Neurology |
| Indications for use: | The Self-adhesive electrode is intended to be used to transmit electrical<br>stimulation current to the patient's skin. Example electrical stimulations<br>for current applications of the electrodes are: TENS and EMS. |
| Device Description: | The Self-adhesive electrode, lead wire type and snap button type, are<br>non-sterile flexible structures, composed of materials commonly used in<br>this application:<br>First layer: Non-woven fabric tape<br>Second layer: Polyurethane (PU) electrically conductive carbon cloth<br>(Hebei Kangshengda Electronic Technology Co., Ltd)<br>Third layer: Biocompatible conductive hydrogel coupling media<br>(ValueTrode Carbon, K970426), which has passed the required skin<br>sensitivity testing criteria as specified in ISO 10993-10 and cytotoxicity<br>testing criteria as specified in ISO ISO 10993-5.<br>The electrodes are designed for single patient / multiple application use. It<br>can be used for low-frequency or medium-frequency nerve or muscle |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Jiajian. The logo is in blue and consists of the word "JIAJIAN" in a sans-serif font, with a check mark above the "I". Below the word "JIAJIAN" is the company name in Chinese characters. The logo also includes the registered trademark symbol.
stimulators, as the conduction film adhered to body skin.
There are nine shapes of round, rectangle, elliptical, calabash, butterfly, saddle shape, rhombus, meniscus shape and palm shape. For the electrical connection, Jiajian provides lead wire type and snap button type:
Lead wire assembly - at least 40mm long wire with 2mm/2.5mm diameter female socket.
Snap button assembly - with 2.5~5mm diameter male socket.
The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR part 898 by testing under ANSI/AAMI ES60601-1, subclause 8.5.2.3.
### 3. Predicate Device Identification
K090198 - JIAJIAN Self-adhesive Electrode
#### 4. Non-Clinical Test Conclusion
Bench tests were conducted on Self-adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following tests were performed on the proposed device:
· Lead wires test per 8.5.2.3 of AAMI/ANSI ES 60601-1;
- Impedance test, Conformability test and Fluid tolerance test per 201.15.101.6 and 201.15.101.7 of IEC 60601-2-2:2009;
- · Impedance Test (Dispersion Test) per FDA's requirement;
- · Peel strength test according to manufacturer's requirement;
- · Shelf life test per ASTM F1980:2016
- · Reuse test per FDA's requirement.
- · Biocompatibility testing according to ISO 10993-5:2009 and ISO 10993-10:2010.
### 5. Substantial Equivalent Based on Assessment of Clinical Performance Data:
Clinical data was not including in this submission
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Jiajian, a medical company. The logo is in blue and features the company name in English and Chinese. The English name is "JIA JIAN" and the Chinese name is "佳健医疗".
| 6. Substantially Equivalent Comparison Conclusion | | | | |
|---------------------------------------------------|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| | Parameters | New Device | Predicate Device | Remark |
| 1 | 510(k) Number | K192568 | K090198 | -- |
| 2 | 510(k) Holder | Wuxi Jiajian Medical Instrument Co., Ltd | Wuxi Jiajian Medical Instrument Co., Ltd | Same |
| 3 | Trade Name | Self-adhesive Electrode | Jiajian Self- adhesive Electrode | -- |
| 4 | Common Name | Cutaneous electrode | Cutaneous electrode | Same |
| 5 | Classification Name | Electrode, cutaneous | Electrode, cutaneous | Same |
| 6 | Product Code | GXY | GXY | Same |
| 7 | Regulation Number | 882.1320 | 882.1320 | Same |
| 8 | Medical Specialty | Neurology | Neurology | Same |
| 9 | Device Class | II | II | Same |
| 10 | Indications for use | The Self-adhesive electrode is intended to be used to<br>transmit electrical stimulation current to the patient's skin.<br>Example electrical stimulations for current applications of<br>the electrodes are: TENS and EMS. | The self-adhesive electrode is intended to be used to<br>apply electrical stimulation current to the patient's skin.<br>Example electrical stimulations for current applications of<br>the electrodes are: TENS and EMS. | Same |
| 11 | Target population | Single patient use and multiple application | Single patient use and multiple application | Same |
| 12 | Type of use | OTC and Prescription use | OTC and Prescription use | Same |
| 13 | Design (shape & connection) | Round, Rectangle, Elliptical, Calabash, Butterfly, Saddle<br>shape, Rhombus, Meniscus shape, Palm shape according to<br>customized specification.<br>Lead wire with female socket, or snap button with male<br>snap connector. | Round, Rectangle, Oval, Gourd, Butterfly, Saddle<br>according to customized specification.<br>Lead wire with female socket. | Similar<br>Note 1 |
| 14 | Materials | - Non-woven fabric tape<br>- Polyurethane (PU) electrically conductive carbon cloth<br>(Hebei Kangshengda Electronic Technology Co., Ltd)<br>- Biocompatible conductive hydrogel coupling media<br>(Value Trode Carbon. K970426) | - Non-woven fabric tape<br>- Electrically conductive carbon cloth (ValueTrode<br>Carbon, K970426)<br>- Biocompatible conductive hydrogel coupling media<br>(Value Trode Carbon. K970426) | Similar<br>Note 2 |
1-5
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for JIA JIAN, a medical company. The logo is in blue and features the company name in English and Chinese. The English name is "JIA JIAN" and the Chinese name is "佳健医疗".
| 15 | Electrode Pad<br>Size | | Round | Min.Ø20mm; Max.Ø85mm | Round | Min.Ø20mm; Max.Ø80mm | Similar<br>Note 1 |
|----|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-------------------|
| | | | Rectangle | Min.26×26mm; Max.190×110mm | Rectangle | Min.3×10mm; Max.80×130mm | |
| | | | Elliptical | Min.85×45mm; Max.240×100mm | Oval | Min.50×120mm; Max.100×240mm | |
| | | | Calabash | Min.90×47mm; Max.200×88mm | Gourd | Min.85×50mm; Max.170×100mm | |
| | | | Butterfly | Min.73×53mm; Max.195×95mm | Butterfly | Min.55×75mm; Max.95×165mm | |
| | | | Saddle shape | Min.82×35mm; Max.150×220mm | Saddle | Min.75×110mm; Max.150×220mm | |
| | | | Rhombus | Min.136×66mm; Max.200×88mm | | | |
| | | | Meniscus shape | Min.115×40mm; Max.200×88mm | | | |
| | | | Palm shape | 74×47mm | | | |
| 16 | Electrode<br>Impedance of<br>Electrode Pad | | Round | 302~577Ω | Round | 290~1000Ω | Similar<br>Note 3 |
| | | | Rectangle | 303~646Ω | Rectangle | | |
| | | | Elliptical | 407~676Ω | Oval | | |
| | | | Calabash | 401~602Ω | Gourd | | |
| | | | Butterfly | 402~626Ω | Butterfly | | |
| | | | Saddle shape | 401~642Ω | Saddle | | |
| | | | Rhombus | 401~648Ω | | | |
| | | | Meniscus shape | 403~655Ω | | | |
| | | | Palm shape | 402~638Ω | | | |
| 17 | --Patient contact<br>area of electrode | | Round | Min.3.14cm²; Max.56.716cm² | Round | Min.3.14cm²; Max.50.24cm² | Similar<br>Note 3 |
| | | | Rectangle | Min.6.76cm²; Max.209cm² | Rectangle | Min.0.3cm²; Max.104cm² | |
| | | | Elliptical | Min.38.25cm²; Max.240cm² | Oval | Min.60cm²; Max.240cm² | |
| | | | Calabash | Min.42.3cm²; Max.176cm² | Gourd | Min.42.5cm²; Max.170cm² | |
| | | | Butterfly | Min.38.69cm²; Max.185.25cm² | Butterfly | Min.41.25cm²; Max.156.75cm² | |
| | | | Saddle shape | Min.14.35cm²; Max.330cm² | Saddle | Min.82.5cm²; Max.330cm² | |
| | | | Rhombus | Min.89.76cm²; Max.176cm² | | | |
| | | | Meniscus shape | Min.46cm²; Max.176cm² | | | |
| | | | Palm shape | | | | |
| 18 | Max. current<br>Density of<br>Electrode (Use<br>IRMS=10mA for<br>calculation ) | Round | 34.78cm²<br>3.18mA/cm² (Ø20mm: 3.14cm²)<br>1.48mA/cm² (26×26mm: 6.76cm²)<br>0.52mA/cm² (85×45mm/2: 19.13cm²)<br>0.24mA/cm² (90×47mm: 42.3cm²)<br>0.52mA/cm² (73×53mm/2: 19.35cm²)<br>0.70mA/cm² (82×35mm/2: 14.35cm²)<br>0.11mA/cm² (136×66mm: 89.76cm²)<br>0.22mA/cm² (115×40mm: 46cm²)<br>0.29mA/cm² (74×47mm: 34.78cm²) | Round<br>Rectangle<br>Oval<br>Gourd<br>Butterfly<br>Saddle | 3.18mA/cm² (Ø20mm: 3.14cm²)<br>33.3mA/cm² (3×10mm: 0.3cm²)<br>0.17mA/cm² (50×120mm: 60cm²)<br>0.24mA/cm² (85×50mm: 42.5cm²)<br>0.48mA/cm² (55×75mm/2: 20.625cm²)<br>0.24mA/cm² (75×110mm/2: 41.25cm²) | Similar<br>Note 3 | |
| 19 | Hydrogel<br>thickness | 35mils ± 5mils (0.89mm ± 0.13mm) | | 35mils ± 5mils (0.89mm ± 0.13mm) | | Same | |
| 20 | Hydrogel pH | 4.2 ± 1.0 | | 4.2 ± 1.0 | | Same | |
| 21 | Hydrogel volume<br>resistivity | 1500 ohm-cm max | | 1500 ohm-cm max | | Same | |
| 22 | Standards meet | • Lead wires test per 8.5.2.3 of AAMI/ANSI ES 60601-1;<br>• Impedance test, Conformability test and Fluid tolerance test per 201.15.101.6<br>and 201.15.101.7 of ANSI AAMI IEC 60601-2-2:2009;<br>• Impedance Test (Dispersion Test) per FDA's requirement;<br>• Peel strength test according to manufacturer's requirement;<br>• Shelf life test per ASTM F1980:2016<br>• Reuse test per FDA's requirement. | | ISO14971;<br>ISO 13485;<br>EN 980;<br>ANSI/AAMI EC12;<br>IEC 60601-1. | | | Similar<br>Note 4 |
| 23 | Biocompatibility | ISO10993-5; ISO10993-10 | | ISO10993-5; ISO10993-10 | | Same | |
| 24 | Sterility Status | Non-sterile | | Non-sterile | | Same | |
| 25 | Electrical safety | Lead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1 | | Lead wire meets IEC 60601-1 | | Same | |
| 26 | Other<br>Performance | Good electrical conductivity, good adhesive property | | Good electrical conductivity, good adhesive property | | Same | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo for JIA JIAN, a medical company. The logo is in a teal color and includes the company name in English and Chinese. The English name, "JIA JIAN," is written in a bold, sans-serif font, with a checkmark symbol integrated into the design. Below the English name, the Chinese characters for "JIA JIAN" are displayed in a smaller font size.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for JIAJIAN. The logo is in blue and consists of the word "JIAJIAN" in a stylized font, with a checkmark incorporated into the letter "J". Below the word "JIAJIAN" are four Chinese characters. To the right of the word "JIAJIAN" is a circled R, indicating that the logo is a registered trademark.
Differences between proposed device and predicate device, reference device:
### Note 1:
The proposed device has two comection methods happ button type, while the predicate device only has one lead wire type. The provosed device ads more electrodes of various shapes and sizes than the predicate device.
- . Regarding the differences in connection method: Both lead wire type electrodes have passed performance tests.
- . Regarding the differences in shape: All shapes of electrodes have passed performance tests.
Regarding the differences in size: We have selected the maximum size electrodes of each shape to carry out the performance tests, and all have passed the tests. . In addition, all test samples were drawn from thatchs, with 3 samples for each specification. Therefore, there differences will not cause any and effectiveness issues.
## Note 2:
The conductive carbon cloth of the proposed device. The final entre electrodes manufactured have passed biocompatibility tests. So, this difference will not cause any safety and effectiveness issues.
## Note 3:
The patient contact area of the is different from the predicate device. Based on analysis of the maximum current density calculation results, this difference will no cause any safety and effectiveness issues.
The maximum current density of the proposed ceve. According to the calculation results, the naxmum current density of all electrodes except for the round shapes does not exceed 2m2 (cle 6060-2-0). The maximum current density of the round electrode of the proposed device is consistent with that of the predicate device. And we added relevant waring item 6 & item 11) and warming symbol (see label) in the user manual. Therefor, these differences will not cause any safety and effectiveness issues.
The impediate of the proposed device is between 302-6762, within the range of 290-10002 of the predice on each electrode is vell-distibution. Therefore, these differences will not cause any safety and effectiveness issues.
Note 4:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for JIA JIAN. The logo is in blue and features the text "JIA JIAN" in a bold, sans-serif font. A check mark is incorporated into the logo. Below the English text is the Chinese text "佳健医疗".
The proposed device has performance tests than the predicate device, including peeling, current distribution testing, acceleated aging testing and simulated use testing etc. All tests comply with the rest results show that the proposed device can meet the intended requirements. Therefore, these differences will not cause any safety and effectiveness issues.
#### The Conclusions:
Based on the successful electrode current distribution test results, Self-adherse Electrode is and effective when used as an interfice between a user's skin and an approved neve and musics. The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as affective, and performs as well as the legally marketed devices in the subject device is substantally equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.