Aneroid sphygmomanometer with stethoscope, Aneroid sphygmomanometer
Applicant
Honsun (Nantong) Co., Ltd.
Product Code
DXQ · Cardiovascular
Decision Date
Apr 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
Device Story
Manual non-invasive aneroid sphygmomanometer; used for blood pressure measurement via auscultatory Korotkoff sounds method. Components: inflation cuff (nylon/cotton), inflatable bladder (PVC/latex), manual inflation bulb, manometer (aluminum/stainless steel), and optional stethoscope. Operated by medical professionals or patients (home use). User wraps cuff around upper arm (or leg for children), manually inflates/deflates using bulb and valve, and listens for Korotkoff sounds via stethoscope to determine systolic/diastolic pressure on the aneroid gauge. Benefits: provides non-invasive, manual blood pressure monitoring.
Clinical Evidence
No clinical data. Bench testing only. Performance verified for compliance with ISO 81060-1 (pressure accuracy ±3mmHg). Biocompatibility testing performed per ISO 10993-1, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
Manual aneroid sphygmomanometer. Materials: aluminum/stainless steel manometer, PVC/latex tubing/bulb/bladder, nylon/cotton cuff. Scale: 0-300mmHg, 2mmHg intervals. Accuracy: ±3mmHg. Biocompatibility: ISO 10993-1/5/10. Performance: ISO 81060-1. No software or electronic components.
Indications for Use
Indicated for measurement of systolic and diastolic blood pressure in newborns, infants, children, young adults, and adults. Used in hospitals, offices, homes, and ambulances. Contraindications: None stated.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
Predicate Devices
RH non-Automated Blood Pressure Meter, MODEL Max02 and RH non-Automated Blood Pressure Meter with stethoscope, MODEL Max0201, Max0202, Max0203 (K190902)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 6, 2020
Honsun (Nantong) Co., Ltd Iris Du RA manager No.8, Tongxing Road, Nantong Economic & Technology Development Area Nantong, 226009 CHINA
Re: K192500
Trade/Device Name: Aneroid Sphygmomanometer Model HS-20A, HS-20D, HS-201W, HS-201Y, HS-201C1, HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C, HS-50B, HS-50D Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ, LDE Dated: March 6, 2020 Received: March 11, 2020
Dear Iris Du:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192500
#### Device Name
Aneroid Sphygmomanometer Model HS-20A, HS-20D, HS-201Y, HS-201Y, HS-201C1, HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C, HS-50B, HS-50D
#### Indications for Use (Describe)
The device is intended to be used by medical professionals or in the measurement of systolic and diastolic pressure. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-201W, HS-201Y,HS-201C1,HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
## Chapter 6. 510(K) Summary
## 1.submitter
- Company name: HONSUN(NANTONG)Co.,Ltd
- Address:No.8, Tongxing Road, Nantong Economic & Technological development Area,Jiangsu, 226009, China
- Postal Code:226010
- TEL:86-0513-80580127
- Contact person: Xinhua Pan (General Manager)
Iris Du(RA Manager)
- E-mail:dulh@lordmed.com
## 2.Subject Device Information
- Model HS-20A,HS-20D, HS-50C, HS-50B, HS-50D, HS-201Q3,HS-201C1,HS-201Y, HS-201W
- Common name: Blood Pressure Kit(Blood Pressure Cuff)
- Trade name: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3/ Aneroid sphygmomanometer with stethoscope, Model HS-50C,HS-50B,HS-50D
- Product Code: DXQ,LDE
- Regulation name: 21CFR 870.1120/21CFR 870.1875
- Regulation class: II
- Review Panel: Cardiovascular
## 3.Predicate device
- RH non-Automated Blood Pressure Meter, MODEL Max02 and RH non-Automated Blood Pressure Meter with stethoscope.MODEL
{4}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
## Max0201,Max0202,Max0203,(K190902)
## 4.Intended use
The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.
## 5.Device description
Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1, HS-201Q3/ Aneroid sphygmomanometer with stethoscope, Model HS-50C,HS-50B, HS-50D are a manual non-invasive aneroid sphygmomanometer which respectively uses an inflation cuff wrapped around the upper arm. The cuff is inflated and deflated by a manual inflation bulb. Besides a manometer (Aneroid gauge), the accessories include cuff, inflation bulb, instruction manual and carrying case. Model HS-50C,HS-50D,HS-50B also include a stethoscope. It is conjunction with stethoscope when use.
The subject devices are same as the predicate device(K190902) in terms of PVC or latex inflation bulb and the PVC or latex bladder. Besides, the material of added the optional various size of cuff (nylon or cotton cuff for this application)and the aluminum manometer are identical to the predicate device (K190902).
## 6. Substantial equivalence comparison
It is substantially equivalence to the predicate device (K190902) RH non-Automated Blood Pressure Meter, MODEL Max02 and RH non-Automated Blood Pressure Meter with stethoscope MODEL Max0201,Max0202,Max0203 with respect to indication for use, device description, and technical description.
All comparison table for applied device are as following, and the substantial equivalence determination is based on the 510(K) Substantial Equivalence
{5}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
Decision-Making Process Flowchart which includes the comparison and discussion of indication for use, technology, and performance specifications.
The new device column includes the following device: Aneroid sphymomanometer Model HS-20A, HS-20D, HS-201W, HS-201Y, HS-201C1, HS-201Q3/ Aneroid sphymomanometer with stethoscope, Model HS-50C, HS-50B, HS-50D.
| Comparison item | Subject device: | Predicate device : | Result of comparison |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Applicant | HONSUN(NANTONG)Co.,Ltd | Wenzhou Renhua Instruments Co.,Ltd | |
| 510(K) number | Applying | K190902 | - |
| Regulation number | 21 CFR 870.1120 | 21 CFR 870.1120 | Same |
| Product code | DXQ.LDE | DXQ.LDE | Same |
| Classification | Class II | Class II | Same |
| Intended use | The device is intended to be used by medical professional or at home for the measurement of systolic and diastolic pressure by detecting korotkoff sounds. | The device is intended to be used by medical professional or at home for the measurement of systolic and diastolic pressure by detecting korotkoff sounds. | Same |
| | Aneroid sphymomanometer<br>Model HS-20A,<br>HS-20D,HS-201W, HS-201Y,<br>HS-201C1, HS-201Q3/<br>Aneroid sphymomanometer<br>with stethoscope, Model<br>HS-50C, HS-50B, HS-50D | RH non-Automated Blood<br>pressure meter Model Max<br>02,RH non-Automated<br>Blood pressure meter<br>with stethoscope, Max0201,<br>Max0202,Max0203 | |
{6}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
# Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3
Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
| Over-the-co<br>unter use | Yes | Yes | Same |
|--------------------------|-----------------------------------------------------------|-----------------------------------------------------------|------|
| Target<br>population | New born, Infants, children,<br>young adults, adults | New born, Infant, children,<br>young adults, adults | Same |
| Where used | Hospital, home, office,<br>and ambulance, etc. | Hospital, home, office,<br>and ambulance, etc. | Same |
| Anatomical<br>sites | Upper arm(leg for child) | Upper arm(leg for child) | Same |
| Measureme<br>nt Method | Ausculatory Korotkoff<br>sounds method | Ausculatory Korotkoff<br>sounds method | Same |
| Inflation | Manual by inflation bulb | Manual by inflation bulb | Same |
| Deflation | Manual deflation via valve | Manual deflation via valve | Same |
| Display | Aneroid Manometer | Aneroid Manometer | Same |
| The monitor<br>scale | From 0 to 300mmHg with a<br>minimum interval of 2<br>mmHg | From 0 to 300mmHg with a<br>minimum interval of 2<br>mmHg | Same |
{7}------------------------------------------------
# Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-201W, HS-201Y,HS-201C1,HS-201Q3
Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
| Design | The device comprised<br>tubing attached to a cuff<br>with an integrated<br>inflatable bladder that is<br>wrapped around the<br>patient's limb and secured<br>by hook and loop closure. | The device comprised<br>tubing attached to a cuff<br>with an integrated<br>inflatable bladder that is<br>wrapped around the<br>patient's limb and secured<br>by hook and loop closure. | Same |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Design of<br>stethoscope | Three types option:<br>Single head<br>Dual head<br>Rappaport | Three types option:<br>Single head<br>Dual head<br>Rappaport | Same |
| Materials | The manometer<br>:aluminum and stainless<br>steel materials.<br>The tubing ,inflation<br>bulb:PVC or latex.<br>Cuff:Nylon cloth or cotton<br>cloth for outside layer.<br>Cuff bladder:PVC or nature<br>latex | The manometer<br>:aluminum and stainless<br>steel materials.<br>The tubing ,inflation<br>bulb:PVC or latex.<br>Cuff:Nylon cloth or cotton<br>cloth for outside layer.<br>Cuff bladder:PVC or nature<br>latex | Same |
| Accuracy | Pressure : ±3mmHg of<br>reading | Pressure : ±3mmHg of<br>reading | Same |
| Compatibility<br>with<br>environment | It can be used from 50°F to<br>104°F (10°C to 40°C) and<br>15%~85% RH humidity. | It can be used from 50°F to<br>104°F (10°C to 40°C) and<br>15%~85% RH humidity. | Same |
{8}------------------------------------------------
# Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-201W, HS-201Y,HS-201C1,HS-201Q3
Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
| Cuff size | 185mm*55mm<br>255mm*75mm<br>345mm*110mm<br>500mm*140mm<br>540mm*145mm<br>615mm*175mm<br>780mm*220mm | 20"×5.5"(510mm*140mm)<br>21.7"×6.3"(550mm*160mm)<br>24.4"×6.9"(620mm*175mm)<br>28.3"×8.3"(720mm*210mm)<br>13.4"×4.15"(340mm*105mm)<br>10.2"×3"(260mm*75mm) | Similar |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Cuff<br>circumference | Fits arm circumferences<br>100mm-660mm,the<br>standard cuff should be<br>available for use in<br>measuring a child's leg<br>blood pressure and for<br>children with larger arms. | Fits arm circumferences<br>8.7" to 17.3"<br>(220mm-440cm),the<br>standard cuff should be<br>available for use in<br>measuring a child's leg<br>blood pressure and for<br>children with larger arms. | Similar |
| Cuff bladder<br>Size | 40mm*80mm<br>60mm*120mm<br>80mm*150mm<br>120mm*220mm<br>145mm*315mm<br>180mm*370mm | 8.7"×4.7"(220mm*120mm)<br>11.8"×5.9"(300mm*150mm)<br>13.4"×6.7"(340mm*170mm)<br>6.9"×3.3"(175mm*85mm)<br>7.5"×2"(190mm*50mm) | Similar |
| Contents(with<br>accessories) | Aneroid gauge,Arm cuff,<br>inflation bulb, and<br>instruction manual,<br>stethoscope(option) | Aneroid gauge,Arm cuff,<br>inflation bulb, and<br>instruction manual,<br>stethoscope(option) | Same |
| Biocompatibility | Biocompatible as requirement of<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Biocompatible as requirement of<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| performance | Compatible as requirement of ISO 81060-1 | Compatible as requirement of ISO 81060-1 | Same |
{9}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
## Discussion:
## Same in details:
- Compared with the predicate device(K190902) RH non-Automated Blood pressure meter Model Max 02,RH non-Automated Blood pressure meter with stethoscope, Max0201, Max0202, Max0203, the subject device Aneroid sphymomanometer Model HS-20A, HS-20D, HS-201W, HS-201Y, HS-201C1, HS-201Q3/ Aneroid sphymomanometer with stethoscope, Model HS-50C, HS-50B, HS-50D are same as the predicate device(K190902) in terms of material s of the inflation bulb,tubing, cuff and manometer.
- Predicate device (K190902) and subject device HS-20A, HS-20D, HS-50C, HS-50B, HS-50D, HS-201Q3, HS-201C1, HS-201Y, HS-201W have same classification,indication for use, target population, measurement method, product physical design and operation, performance parameter ranges,mechanical safety, anatomical sites, operation principles and etc. The structure of the predicate device (K190902) are same as the subject device ,which all have manometer, bulb, tube, various size of cuff and stethoscope(optional).The predicate device and subject device all have dual head stethoscope, single head stethoscope and rappaport stethoscope.
6-7
{10}------------------------------------------------
## Sponsor:HONSUN(NANTONG)Co.,Ltd
Subjective device: Aneroid sphygmomanometer Model HS-20A,HS-20D,HS-201W, HS-201Y,HS-201C1,HS-201Q3 Aneroid sphygmomanometer with stethoscope Model HS-50C,HS-50D
## Difference in details:
- The different sizes of the cuffs and bladders,cuff circumstance are provided in order to accommodate target population. All performance have been tested in this submission and do not raised any safety or effectiveness issue.All performance specification was verified to comply with the ISO 81060-1 standard.
## 7. Non-clinical testing
Non-clinical testing included biocompatibility and bench testing. The tests listed in the ISO 10993 and ISO 81060-1, were included. Results of testing were acceptable.
## 8. Clinical testing
No clinical study is included in this submission.
## Conclusion
RH non-automated blood pressure meter Model Max 02, non-automated blood pressure meter with stethoscope Model Max0201, Max 0202,Max0203(K190902) and subject device have the same intended use and similar technological characteristics. Moreover, information contained in this submission supplied demonstrates that any difference in their characteristic do not raise any new questions of safety or effectiveness. Thus, aneroid sphygmomanometer Model HS-20A, HS-20D, HS-201W, HS-201Y, HS-201C1, HS-201Q3/ aneroid sphygmomanometer with stethoscope, Model HS-50C, HS-50B, HS-50D are equivalent to the predicate device(K190902)..
6-8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.