FDR SE Lite
K192440 · Fujifilm Corporation · MQB · Nov 1, 2019 · Radiology
Device Facts
| Record ID | K192440 |
| Device Name | FDR SE Lite |
| Applicant | Fujifilm Corporation |
| Product Code | MQB · Radiology |
| Decision Date | Nov 1, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1680 |
| Device Class | Class 2 |
Intended Use
The FDR SE Lite flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film/screen or CR systems may be used. The FDR SE Lite is not intended for mammography, fluoroscopy, and angiography applications, as well as pediatric and neonatal exams.
Device Story
Portable digital flat panel detector system; captures and digitizes X-ray exposures from standard radiographic systems. Detector supports wireless data communication to console. Used in wall bucky, table bucky, or free-cassette configurations. Operated by clinical staff in general radiographic environments. Output displayed as radiographic images for physician review to support clinical decision-making. Benefits include digital image capture replacing conventional film/screen or CR systems.
Clinical Evidence
No clinical study performed. Substantial equivalence demonstrated via non-clinical performance data, including software testing and image quality evaluation compared to the predicate device.
Technological Characteristics
Portable digital flat panel detector (MarsXF). Wireless data communication. Conforms to AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, and DICOM standards. Designed for use in wall/table bucky or free-cassette modes.
Indications for Use
Indicated for capturing radiographic images of human anatomy in general projection radiographic applications for adult patients. Contraindicated for mammography, fluoroscopy, angiography, and pediatric or neonatal exams.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- FDR D-EVO II Flat Panel Detector System (DR-ID 1200) (K142003)
Related Devices
- K162344 — INNOLUX RIC · Innolux Corporation · Dec 20, 2016
- K201796 — 1717SCV, 1717SGV · Rayence Co., Ltd. · Jul 23, 2020
- K192932 — FDR D-EVO III flat panel detector system · Fujifilm Corporation · Nov 12, 2019
- K182551 — Wireless Digital Flat Panel Detector · Iray Technology Taicang , Ltd. · Oct 17, 2018
- K181526 — XRpad2 4343 HWC-M Flat Panel Detector · Varex Imaging Corporation · Jul 3, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2019
Fujifilm Corporation % Kamila Sak Specialist. Regulatory Affairs FuiiFilm Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 LEXINGTON MA 02421
Re: K192440
Trade/Device Name: FDR SE Lite Flat Panel Detector System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: September 5, 2019 Received: September 6, 2019
Dear Kamila Sak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K192440
Device Name FDR SE Lite Flat Panel Detector System
Indications for Use (Describe)
The FDR SE Lite flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film/screen or CR systems may be used. The FDR SE Lite is not intended for mammography, fluoroscopy, and angiography applications, as well as pediatric and neonatal exams.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Fujifilm Medical Systems, USA. The logo consists of the word "FUJIFILM" in large, bold, black letters. There is a red square above the second "I" in "FUJIFILM". Below the word "FUJIFILM" is the text "Fujifilm Medical Systems, USA" in a smaller, black font.
# 510(k) Summary
# FDR SE-Lite Flat Panel Detector System (DR-ID 330)
## Date: September 5, 2019
## Submitter's Information:
FUJIFILM Corporation
798 Miyanodai Kaisei-Machi
Ashigarakami-Gun, Kanagawa, 258-8538, Japan
FDA Establishment Registration Number: 3001722928
### Contact Person:
| Name: | Kamila Sak |
|------------|--------------------------------|
| Title: | Specialist, Regulatory Affairs |
| Telephone: | (347) 577-2309 |
### ldentification of the Device:
Proprietary:
Model Number: DR-ID330 Classification Name: Stationary X-ray system CFR Section: 21 CFR 892.1680 Product Codes: 90 MQB Device Class: Class II Review Panel: Radiologv Common Name: Flat Panel Digital Detector System
### ldentification of the Legally Marketed Device:
FDR D-EVO II Flat Panel Detector System (DR-ID 1200), K142003 cleared 10/21/2014
FDR SE Lite Flat Panel Detector System
#### l. DEVICE DESCRIPTION
Fujifilm's FDR SE Lite (DR-ID 330), is a portable digital detector system that interfaces with, and acquires and digitizes X-ray exposures from, standard radiographic systems. DR-ID 330 is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, not for pediatrics and neonates. The detector models support only wireless data communication between the detector and the console. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams.
#### II. INDICATIONS FOR USE
The FDR SE Lite flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film/screen or CR systems may be used. The FDR SE Lite is not intended for mammography, fluoroscopy, tomography, and angiography applications, as well as pediatric and neonatal exams.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Fujifilm Medical Systems, USA. The word "FUJIFILM" is written in large, bold, black letters. There is a red square on top of the letter "I". Below the word "FUJIFILM" is the text "Fujifilm Medical Systems, USA".
#### SUBSTANTIAL EQUIVALENCE ..............................................................................................................................................................................
Fujifilm FDR SE Lite FPD System (DR-ID 330) is substantially equivalent to the following legally marketed device.
| Legally Marketed Device | 510(k) # | Clearance Date |
|-----------------------------------------------------|----------|----------------|
| FDR D-EVO II Flat Panel Detector System (DR-ID1200) | K142003 | 10/21/2014 |
Both the subject device FDR SE Lite (DR-ID 330) and predicate device FDR D-EVO II (DR-ID 1200) in K142003 are portable digital detector systems that are intended to be used for the same purposes except for pediatric and neonatal use and wired communication. The most detector characteristics remain unchanged for FDR SE Lite, and the image quality is substantially equivalent to the predicate device. The design made for the FDR SE Lite have been successfully tested and validated as summarized below.
#### IV. SUMMARY OF STUDIES
Non-clinical Performance Data: FDR SE Lite FPD System (DR-ID330) conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366 and DICOM. In addition, the FDA's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (September 1, 2016) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (August 14, 2013) was followed to test wireless features. As required by the risk analysis, necessary verification activities were performed including software testing, and the results were satisfactory. Furthermore, the image quality evaluation confirmed that the image quality of the FDR SE Lite system using MarsXF detectors is substantially equivalent to that of the predicate device.
Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
#### > CONCLUSION
Based upon the supporting data summarized above, we concluded FDR SE Lite Flat Panel Detector System (DR-ID 330) is as safe and effective as the legally marketed device DR-ID 1200 (K142003), and do not raise different questions of safety and effectiveness than K142003.