DualX Lumbar Intervertebral Body Fusion Device

K192434 · Amplify Surgical, Inc. · MAX · Oct 15, 2019 · Orthopedic

Device Facts

Record IDK192434
Device NameDualX Lumbar Intervertebral Body Fusion Device
ApplicantAmplify Surgical, Inc.
Product CodeMAX · Orthopedic
Decision DateOct 15, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DualX™ PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ PLIF is indicated for unilateral implantation. The DualX™ LLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to L5, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ TLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

Device Story

Expanding interbody fusion device for minimally invasive lumbar spine surgery; implants inserted in collapsed state, then expanded sequentially in lateral and vertical directions to maximize cortical rim contact and maintain distraction; binary locking mechanism in fully expanded state; used by surgeons in clinical settings; provides structural support for fusion; benefits patients by restoring disc height and facilitating fusion in DDD cases; modification involves sterilization tray layout for non-sterile implants.

Clinical Evidence

No clinical data; bench testing not required due to lack of changes to implants or instruments; substantial equivalence established via comparison to predicate (K181397).

Technological Characteristics

Titanium (ASTM F136); expanding interbody fusion cage; lateral and vertical expansion mechanism; binary locking; non-sterile, end-user sterilized.

Indications for Use

Indicated for intervertebral body fusion of the lumbar spine (L2-S1 for PLIF/TLIF; L2-L5 for LLIF) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis or retrolisthesis. Requires supplemental fixation and bone graft.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 15, 2019 Amplify Surgical Inc. % Mr. Nathan Wright, MS Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K192434 Trade/Device Name: DualX™ Lumbar Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: September 4, 2019 Received: September 5, 2019 Dear Mr. Wright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Ronald P. Jean, PhD Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use #### K192434 Device Name DualX™ Lumbar Intervertebral Body Fusion Device Indications for Use (Describe) The DualX™ PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ PLIF is indicated for unilateral implantation. The DualX™ LLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to L5, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ TLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Type of Use (Select one or both, as applicable) | <span style="font-size: 12px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |--------------------------------------------------------------------------------------| | <span style="font-size: 12px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Page 1 of 1 FORM FDA 3881 (7/17) PSC Publishing Services (301) 443-6740 EF Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 {3}------------------------------------------------ | 510(K) SUMMARY | |----------------| |----------------| | Submitter's Name: | Amplify Surgical Inc. | |----------------------------|--------------------------------------------------------------| | Submitter's Address: | 27071 Cabot Road, Suite 118<br>Laguna Hills, CA 92653 | | Submitter's Telephone: | 765-267-5439 | | Contact Person: | Nathan Wright, MS<br>Empirical Testing Corp.<br>719-351-0248 | | Date Summary was Prepared: | 03-Sept-2019 | | Trade or Proprietary Name: | DualX™ Lumbar Intervertebral Body Fusion Device | | Common or Usual Name: | Intervertebral Fusion Device With Bone Graft, Lumbar | | Classification: | Class II per 21 CFR §888.3080 | | Product Code: | MAX | | Classification Panel: | Division of Orthopedic Devices | The purpose of this submission is to modify the layout of the sterilization trays of the DualX™ Lumbar Intervertebral Body Fusion Device to include non-sterile implants to be sterilized by the end user with no changes to the implants and instruments previously cleared under K181397. # DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The DualX™ Lumbar Intervertebral Body Fusion Device implants are an expanding family of interbody fusion devices that expand sequentially in lateral and then vertical directions. The implants are designed to be used in minimally invasive spine surgery and, in the collapsed configuration, are between 12.3 mm wide (TLIF and PLIF) and 13.3mm wide (LLIF). # INDICATIONS FOR USE The DualX™ PLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The DualX™ PLIF is indicated for unilateral or bilateral implantation. The DualX™ LLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to L5, in skeletally mature patients. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. {4}------------------------------------------------ The DualX™ TLIF is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1. in skeletally mature patients. The device is intended for use at either one or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The indications for use for the DualX™ Lumbar Intervertebral Body Fusion Device are identical to the indications previously cleared DualX™ Lumbar Intervertebral Body Fusion Device under K181397. ### Table 5-1 Predicate Devices | 510k Number | Trade or Proprietary or Model<br>Name | Manufacturer | Designation | |-------------|----------------------------------------------------|------------------|-------------| | K181397 | DualX™ Lumbar Intervertebral<br>Body Fusion Device | Innovasive, Inc. | Primary | ## TECHNICAL CHARACTERISTICS After entering the joint space, the implants expand laterally to maximize the contact area around the cortical rim of the vertebral bodies. The implants then expand vertically to maintain distraction of the vertebral bodies and maximize clearance for exiting nerve roots. The implants are binary, and each implant can only be locked out in the fully expanded state. Specific sizes are included in the system with various expanded heights to optimize the fit for the patient. The DualX™ Lumbar Intervertebral Body Fusion Device is made from titanium that conforms to ASTM F136. The subject and predicate devices have identical technological characteristics; the modification proposed does not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates: - . Indications for Use - Materials of manufacture . - Structural support mechanism - . Size ### PERFORMANCE TESTING SUMMARY In support of this Special 510(k) Device Modification Premarket Notification, mechanical testing was not required to demonstrate substantial equivalent mechanical strength since there are no modifications to the implants and instruments. The substantial equivalence of mechanical {5}------------------------------------------------ strength for the DualX™ Lumbar Intervertebral Body Fusion Device was established in the clearance of K181397. ## CONCLUSION The subject modified DualX™ Lumbar Intervertebral Body Fusion Device is nearly identical to the previously cleared DualX™ Lumbar Intervertebral Body Fusion Device. The subject DualX™ Lumbar Intervertebral Body Fusion has identical intended uses, indications, technological characteristics, and principles of operation as the predicate. The modifications raise no new types of safety or effectiveness questions. The overall technology characteristics lead to the conclusion that the DualX™ Lumbar Intervertebral Body Fusion Device is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...