K192402 · Siemens Medical Solutions USA, Inc. · JAK · Sep 20, 2019 · Radiology
Device Facts
Record ID
K192402
Device Name
syngo.CT Extended Functionality
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Sep 20, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners, and possibly other medical imaging modalities (e.g. MR scanners). An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT Scanner system. Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
Device Story
syngo.CT Extended Functionality is a software bundle providing advanced visualization and measurement tools for medical images (CT/MR). It functions as a post-processing application for clinical data acquired by imaging scanners. The device includes specific extensions for Interactive Spectral Imaging, vascular assessment, oncology (nodule evaluation), bone mineral density (Osteo), neuro DSA, dual energy ROI, endoscopic views, and HU thresholding. It is operated by technicians or physicians in a clinical setting. The software processes image data to generate visualizations and measurements, which are then reviewed by clinicians to assist in diagnosis. The device supports multiphase data grouping and editing of pre-generated results. It does not acquire images directly but connects to imaging systems via an interface.
Clinical Evidence
Bench testing only. Performance was validated through phantom-based testing for the Interactive Spectral Imaging feature and verification of software modifications. No clinical data was required or provided. Testing confirmed that the device performs as intended and meets safety/performance standards including DICOM (PS 3.1-3.20), IEC 62304, ISO 14971, and IEC 62366-1.
Technological Characteristics
Software-based post-processing application (SOMARIS/8 VB40). Operates on CT/MR image data. Features include spectral imaging, vascular/oncology/osteo/neuro/dual-energy/endoscopic tools. Connectivity via DICOM. Complies with IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366-1 (usability).
Indications for Use
Indicated for use by trained technicians and physicians to provide advanced visualization, qualitative and quantitative measurements, and analysis of medical images (CT and MR) for diagnostic purposes.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2019
Siemens Medical Solutions USA, Inc. % Alaine Medio Regulatory Affairs Specialist 810 Innovation Drive KNOXVILLE TN 37932
Re: K192402
Trade/Device Name: syngo.CT Extended Functionality Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: August 29, 2019 Received: September 3, 2019
Dear Alaine Medio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192402
Device Name syngo.CT Extended Functionality
#### Indications for Use (Describe)
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners, and possibly other medical imaging modalities (e.g. MR scanners).
An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT Scanner system.
Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------|--|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |
| <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY FOR SYNGO.CT EXTENDED FUNCTIONALITY
Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Date Prepared: September 18, 2019
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### I. Submitter
#### Importer/Distributor
Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Establishment Registration Number 2240869
Manufacturing Site Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany
Establishment Registration Number 3004977335
### Contact Person
Alaine Medio Regulatory Affairs Siemens Medical Solutions USA, Inc. Molecular Imaging 810 Innovation Drive Knoxville, TN 37932 Phone: (865)206-0337 Fax: (865) 218-3019 Email: alaine-medio(@siemens-healthineers.com
#### Alternate Contact :
Tabitha Estes
### II. Device Name and Classification
| Product Name: | syngo.CT Extended Functionality |
|-----------------------|----------------------------------|
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
{5}------------------------------------------------
# III. Predicate Device
#### Primary Predicate Device:
| Trade Name: | syngo.CT Clinical Extensions |
|-----------------------|------------------------------------------------|
| 510(k) Number: | K173625 |
| Clearance Date: | 03/08/2018 |
| Classification Name: | System, X-Ray, Tomography, Computed |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | There are currently no recalls for this device |
#### Secondary Predicate Device:
| Trade Name: | syngo.CT Dual Energy |
|-----------------------|------------------------------------------------|
| 510(k) Number: | K150757 |
| Clearance Date: | 08/11/2015 |
| Classification Name: | System, Image Processing, Radiological |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | There are currently no recalls for this device |
# IV. Device Description
syngo.CT Extended Functionality is a software bundle that offers tools to support special clinical evaluations. The "tools" are represented by the so-called Extensions. syngo.CT Extended Functionality can be used to create advanced visualizations and measurements on clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners or other medical imaging modalities (e.g. MR scanners) by using the Extensions. Advanced visualizations and measurements are listed as follows. In addition, the table below shows which extensions has been changed in the current software version SOMARIS/8 VB40.
| No | Extension | No | Extension |
|----|--------------------------------------|----|-------------------------------|
| 1 | Interactive Spectral Imaging (new) | 5 | Neuro DSA (unmodified) |
| 2 | Vascular/Vessel Extension (modified) | 6 | ROI HU Threshold (unmodified) |
| 3 | Oncology Extension (modified) | 7 | Dual Energy (unmodified) |
| 4 | Osteo (unmodified) | 8 | Endoscopic View (unmodified) |
syngo.CT Extended Functionality is an extension of the previously cleared primary predicate device post-processing application software syngo.CT Clinical Extensions (K173625) and includes the following modifications in comparison to the primary predicate device:
- Support of Interactive Spectral Imaging a.
- Vascular Extension modification: b.
- i. 2D Filter Type Support
- ii. Multiphase Support in Vessel Layout
- Oncology Extension modification: C.
- i. Support of MR Data
- General: Multiphase support for merged 4D series d.
{6}------------------------------------------------
# V. Indications for Use
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners, and possibly other medical imaging modalities (e.g. MR scanners).
An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT Scanner system.
Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
# VI. Comparison of Technological Characteristics with the Predicate Device
The indications for use for the syngo.CT Extended Functionality are identical to the predications for Use (see Table 2). The only difference is the renaming of the feature from syngo.CT Clinical Extensions to syngo.CT Extended Functionality. The intended use of this product is identical to the predicates.
| Subject Device | Predicate Device (K173625) | Comparison Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| syngo.CT Extended Functionality | syngo.CT Clinical Extensions | Identical with the exception<br>of the name of the product. |
| syngo.CT Extended Functionality is<br>intended to provide advanced visualization<br>tools to prepare and process medical<br>images for diagnostic purpose. The<br>software package is designed to support<br>technicians and physicians in qualitative<br>and quantitative measurements and in the<br>analysis of clinical data that was acquired<br>and reconstructed by Computed<br>Tomography (CT) scanners, and possibly<br>other medical imaging modalities (e.g. MR<br>scanners).<br>An interface shall enable the connection<br>between the syngo.CT Extended<br>Functionality software package and the<br>interconnected CT Scanner system.<br>Result images created with the syngo.CT<br>Extended Functionality software package<br>can be used to assist trained technicians or<br>physicians in diagnosis. | syngo.CT Clinical Extensions is intended<br>to provide advanced visualization tools to<br>prepare and process medical images for<br>diagnostic purpose. The software package<br>is designed to support technicians and<br>physicians in qualitative and quantitative<br>measurements and in the analysis of<br>clinical data that was acquired and<br>reconstructed by Computed Tomography<br>(CT) scanners, and possibly other medical<br>imaging modalities (e.g. MR scanners).<br>An interface shall enable the connection<br>between the syngo.CT Clinical<br>Extensions software package and the<br>interconnected CT Scanner system.<br>Result images created with the syngo.CT.<br>Clinical Extensions software package can<br>be used to assist trained technicians or<br>physicians in diagnosis. | |
Table 2: Indications for Use / Intended Use Comparison
As with the primary predicate device, the subject device is a bundle software package consisting of previously cleared software applications (unmodified and modified) that provide advanced visualization and measurement tools.
Software version SOMARIS/8 VB40 supports additional post-processing application features for MR datasets, dual energy images, and endoscopic visualization tools. At a high-level, the subject and predicate devices are based on the following same or similar technological characteristics as listed in Table 3 below:
{7}------------------------------------------------
| | | Table 3: Technological Characteristic Comparison | |
|--|--|---------------------------------------------------|--|
|--|--|---------------------------------------------------|--|
| Feature | Subject Device | Predicate Device<br>and Supported Functionality | Comparison Result |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| | yngo.CT Extended Functionality | | |
| Software<br>Version | SOMARIS/8 VB40 | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br>SOMARIS/8 VB30 | Modified to<br>support additional<br>functionality |
| | 1. Interactive Spectral Imaging | | |
| Interactive<br>Spectral<br>Imaging | The Interactive Spectral Imaging<br>functionality allows the user to display<br>different representations of Dual<br>Energy data with an interactive image<br>text. | Secondary Predicate Device<br>K150757, clearance date 08/11/2015<br><br>The user can change the visualization<br>of Dual Energy data generated from<br>Low and High data to different Dual<br>Energy types. | This feature is<br>new in the subject<br>device, but the<br>fundamental<br>technology has<br>been cleared with<br>the secondary<br>predicate device. |
| | 2. Vascular/Vessel Extension | | |
| Vascular/<br>Vessel<br>Extension | The Vascular Tool provides tools and<br>layouts for vascular assessment.<br>Additionally, the subject device<br>Vascular Tool provides a 2D Filter<br>Type Support and Multiphase Support<br>in Vessel Layout. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>The Vascular Tool provides tools and<br>layouts for vascular assessment and<br>CT data. | This modification<br>is a usability<br>improvement. |
| | 3. Oncology Extension | | |
| Oncology<br>Extension | The oncology Tool offers tools for<br>localization and evaluation of nodules.<br>Additionally, the subject device now<br>provides MR Support for Diameter<br>WHO. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>The oncology extension offers tools<br>for localization and evaluation of<br>nodules. | Modified to<br>support MR data<br>for Diameter<br>WHO. |
| | 4. Osteo Extension | | |
| Osteo<br>Extension | The Osteo extension is used for the<br>evaluation of Bone Mineral Density<br>(BMD) values (mg CA-HA/ml) of the<br>lumbar spine based on Osteo CT scans. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>The Osteo extension is used for the<br>evaluation of Bone Mineral Density<br>(BMD) values (mg CA-HA/ml) of the<br>lumbar spine based on Osteo CT<br>scans. | Same – There are<br>no modifications. |
| | 5. Neuro DSA Extension | | |
| Neuro DSA<br>Extension | Bone removal tool from a CT<br>angiography dataset. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>Bone removal tool from a CT<br>angiography dataset. | Same – There are<br>no modifications. |
| | 6. Dual Energy ROI | | |
| Dual<br>Energy ROI | Evaluation of low/high kV images from<br>dual energy data. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>Evaluation of low/high kV images<br>from dual energy data. | Same – There are<br>no modifications. |
| 7. Endoscopic View | | | |
| Endoscopic<br>View Tool | Fly through tubular structures which<br>are either filled by low-intensity (e.g.<br>air-filled) or high-intensity (e.g. blood-<br>filled) material. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br><br>Fly through tubular structures which<br>are either filled by low-intensity (e.g.<br>air-filled) or high-intensity (e.g.<br>blood-filled) material. | Same – There are<br>no modifications. |
{8}------------------------------------------------
| Feature | Subject Device<br>syngo.CT Extended Functionality | Predicate Device<br>and Supported Functionality | Comparison<br>Result |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| 8. ROI HU Threshold | | | |
| ROI HU<br>Threshold | Evaluation of HU value distributions<br>within a user defined region of interest. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br>Evaluation of HU value distributions<br>within a user defined region of<br>interest | Same – There are<br>no modifications. |
| General, Extension-independent Functionality | | | |
| Pre-<br>generated<br>results | Support of pre-generated results and the<br>ability to edit pre-generated results | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br>Support of pre-generated results and<br>the ability to edit pre-generated<br>results | Same – There are<br>no modifications. |
| | | | |
| Multiphase<br>support for<br>merged 4D<br>series | Grouping of multiphase series<br>generated by respiratory gated,<br>perfusion and cardiac datasets. | Primary Predicate Device:<br>K173625, clearance date 03/08/2018<br>Grouping of multiphase series<br>generated by respiratory gated and<br>perfusion datasets. | Usability<br>improvement: The<br>grouping logic has<br>been extended to<br>include cardiac<br>gated datasets. |
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject device, syngo.CT Extended Functionality are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.
# VII. Performance Data
## Software Verification and Validation Testing
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is included as part of this submission.
## Non-Clinical / Clinical Testing Summary
syngo.CT Dual Energy is designed to fulfill the requirements of the following safety and performance standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Publication<br>Date | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|------------------------------------------------------------------------------------------------------------|---------------------|------------------------------------------|
| 12-300 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set; PS 3.1 – 3.20 | 06/27/2016 | NEMA |
| 13-32 | Software | Medical Device Software –Software Life<br>Cycle Processes; 62304:2006 (1st Edition) | 08/20/2012 | AAMI, ANSI,<br>IEC |
| 5-40 | Software/<br>Informatics | Medical devices – Application of risk<br>management to medical devices; 14971<br>Second Edition 2007-03-01 | 08/20/2012 | ISO |
| 5-114 | General I<br>(QS/RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices<br>IEC 62366-1:2015 | 12/23/2016 | IEC |
Performance tests were conducted to test the Interactive Spectral Imaging (ISI) functionality of the syngo.CT Extended Functionality post-processing application. A phantom-based validation "Detailed Description and Bench
{9}------------------------------------------------
Tests for the Plugin "Interactive Spectral Imaging"" has been conducted to show that the feature ISI operates as intended.
Non-clinical tests and a phantom-based bench test have been conducted to establish the proficiency of the features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.
### General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing.
# VIII. Conclusions
syngo.CT Extended Functionality has the same indication for use as the primary predicate device. The technological characteristics such as image visualization, operating platform, and image measurement are the same as the predicate devices. For the subject device, syngo.CT Extended Functionality, Siemens used the same testing with the same workflows as used to clear the primary predicate device. Siemens considers syngo.CT Extended Functionality to be as safe, as effective, and with performance substantially equivalent to the commercially available predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.