SciPharm's DFV Desensitizing Varnish is a fluoride-containing resin preparation for the treatment of dentinal hypersensitivity and for the reduction of post-operative sensitivity.
Device Story
DFV Desensitizing Varnish is a topical, flavored dental varnish containing 5% sodium fluoride in a synthetic rosin base. Applied by professional clinicians to prepared tooth cavities using a brush; the varnish forms a film that occludes open dentinal tubules. Ethanol solvent facilitates application and quick drying. The device acts as a barrier to prevent restorative material penetration into dental tissue. Benefits include immediate and sustained relief of dental pain and sensitivity. The device is supplied in unit-dose trays.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-5 and ISO 7405 (cytotoxicity) showed results consistent with predicate devices. SEM imaging of human dentine specimens confirmed the varnish forms a smooth, uniform film that occludes dentinal tubules, performing comparably to the predicate. Physical testing (viscosity, fluoride content, stability) confirmed product specifications.
Technological Characteristics
Synthetic rosin-based varnish containing 5% sodium fluoride, ethanol, polyamide, carbamide peroxide, xylitol, sucralose, and silica filler. Form factor: viscous liquid in unit-dose PET/PVC trays. Application: manual brush. Biocompatibility: ISO 10993-1 compliant. No electronic components or software.
Indications for Use
Indicated for the treatment of dentinal hypersensitivity and reduction of post-operative sensitivity in patients requiring professional dental care. Contraindicated in patients with allergies or hypersensitivity to natural resin or colophony, and patients with stomatitis or ulcerative gingivitis.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
Predicate Devices
Young Dental Manufacturing: Oral-B 5% Sodium Fluoride Varnish (K153334)
Reference Devices
3M: Vanish 5% Sodium Fluoride White Varnish (K090519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 18, 2020
Scientific Pharmaceuticals Thomas Kelly Regulatory Affairs/Quality Assurance Manager 3221 Producer Way Pomona, California 91768
Re: K192381
Trade/Device Name: DFV Desensitizing Varnish-Strawberry Flavor, DFV Desensitizing Varnish-Bubblegum (Tutti-Fruity) Flavor, DFV Desensitizing Varnish-Spearmint Flavor, DFV Desensitizing Varnish-Chocolate Flavor, DFV Desensitizing Varnish-Marshmallow Flavor
Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: November 20, 2019 Received: November 20, 2019
Dear Thomas Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K192381
Device Name DFV Desensitizing Varnish
#### Indications for Use (Describe)
SciPharm's DFV Desensitizing Varnish is a fluoride-containing resin preparation for the treatment of dentinal hypersensitivity and for the reduction of post-operative sensitivity.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue Erlenmeyer flask on the left, followed by the text "SCI-PHARM" in blue. To the right of the text is a mortar and pestle, also in blue, with the registered trademark symbol above it.
# 510(k) Summary — K192381
| 510(k) Submission: | Traditional |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Device Type: | Varnish, Cavity |
| Submitter: | Scientific Pharmaceuticals, Inc.<br>3221 Producer Way<br>Pomona, CA 91768 |
| Contact Person: | Thomas Kelly<br>Regulatory Affairs/Quality Assurance Manager<br>Regulatory@scipharm.com<br>Telephone: 909-595-9922 ext. 237 |
| Alternate Contact Person: | David Butler<br>President<br>Dave@scipharm.com<br>Telephone: 909-595-9922 ext. 225 |
| Date: | February 18, 2020 |
| Confidentiality: | According to 21 CFR 807.95 |
| Trade Name: | Sci-Pharm DFV Desensitizing Varnish |
| Common Name: | Cavity Varnish or Topical Varnish |
| Classification Name<br>and References: | Cavity Varnish<br>21CFR 872.3260 |
| Product Code: | LBH |
| Classification Panel: | Dental |
| Classification: | Medical Device, Class II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the address of Scientific Pharmaceuticals. The address is 3221 Producer Way, Pomona, CA 91768. The text is black and the background is white.
Image /page/4/Picture/1 description: The image shows the logo for SCI-PHARM. The logo is blue and features a graphic of two beakers on the left, the text "SCI-PHARM" in the center, and a mortar and pestle on the right. The logo also includes a registered trademark symbol.
Purpose and Description:
SciPharm's DFV Desensitizing Varnish is a device that consists of a synthetic compound from a natural rosin that is intended to coat a prepared tooth cavity prior to the insertion of a restorative material by forming a film that facilitates occlusion of compromised surfaces, including open dentinal tubules. DFV Desensitizing Varnish and the predicate device consist of a viscous liquid that can be applied to the teeth using a brush or similar applicator. The purpose of the 5% sodium fluoride as the source of fluoride ions for the formation of calcium fluoride, and along with the rosin, occludes the dentin tubules. Both use denatured alcohol (ethanol) as a significant component, providing fluidity for application and to promote quick drying on the tooth surface.
SciPharm's DFV Desensitizing Varnish is a topically applied, flavored (Strawberry, Bubblegum, Spearmint, Chocolate, and Marshmallow) topical varnish containing sodium fluoride on a synthetic rosin based preparation from a natural rosin. Only a professional clinician may apply DFV Desensitizing Varnish. The varnish and applicator are packaged in a thermoformed PET/PVC opaque unit dose tray, and sealed with foil.
The technological difference between DFV Desensitizing Varnish and the other six (6) varnishes that Scientific Pharmaceutical's manufactures is the formulation of components. The components are commonly used in the dental materials industry. The DFV Desensitizing Varnish uses a synthetic resin as the base material instead of the natural rosin used in other formulations. This synthetic rosin is the same used in the predicate device. DFV Desensitizing Varnish also includes additional base material, sweeteners, bitterness blockers, and flavoring agents.
The device is intended to prevent the penetration of restorative materials, such as amalgam, into dental tissue.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue Erlenmeyer flask on the left, followed by the text "SCI-PHARM" in blue. On the right, there is a blue mortar and pestle with a registered trademark symbol above it. The logo is simple and professional, and it conveys the company's focus on science and pharmaceuticals.
# Chemical Composition:
| | Raw Material<br>(Ingredient) | Purpose of Ingredient | Concentration of<br>Ingredient |
|-----|----------------------------------------------------------------------------------------|-----------------------------------------|--------------------------------|
| 1. | Purified Rosin | Film Former | 45.0% - 75.0% |
| 2. | Polyamide | Film Former | 10.0% - 25.0% |
| 3. | Ethanol,<br>200 Proof | Solvent,<br>Surface Wettability | 10.0% - 25.0% |
| 4. | Sodium Fluoride | Reduce Tooth Decay | 4.50% - 5.50% |
| 5. | Carbamide<br>Peroxide | Whitening Agent for<br>color stability | 0.25% - 5.0% |
| 6. | Xylitol | Sweetener and<br>stimulates saliva flow | 0.5% - 4.0% |
| 7. | Sucralose | Sweetener | 0.25% - 5.0% |
| 8. | Bitter Blocker | Flavoring Agent | 0.10% - 3.00 % |
| 9. | Silica Filler | Thickener | 1.0% - 5.0% |
| 10. | Flavorants<br>(Strawberry,<br>Bubblegum,<br>Spearmint,<br>Chocolate, &<br>Marshmallow) | Flavoring Agent | 0.25% - 10.0% |
Contraindications: Patients allergic, or hypersensitive, to natural resin or colophony (kolophonium or resembling agents) should avoid the use of this product. In case of an adverse reaction, the varnish may be removed by brushing with a soft toothbrush and flossing, followed by rinsing with warm water. Direct ingestion should be avoided. Dyspnea, (difficult or labored breathing) associated with asthmatic children, has been reported in rare cases. Patients with sensitive digestive systems may experience nausea, usually after a prolonged treatment. Patients with stomatitis or ulcerative gingivitis should avoid this product. Primary Predicate Device: Young Dental Manufacturing: Oral-B 5% Sodium Fluoride Varnish (K153334). Reference Device: 3M: Vanish 5% Sodium Fluoride White Varnish (K090519) Unit dose container, Nylon Brushes for application, mixing Accessories(s):
well.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of two blue Erlenmeyer flasks, the text "SCI-PHARM" in blue, and a blue mortar and pestle. The registered trademark symbol is located to the right of the text.
## Comparison to Predicate Device
| | DFV Desensitizing<br>Varnish | Primary Predicate<br>Device<br>Oral-B 5% Sodium<br>Fluoride Varnish | Reference Device<br>Vanish Varnish |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(abbreviated) | Treatment of dental<br>hypersensitivity and<br>reduction of post-<br>operative sensitivity | K153334<br>Treatment of dental<br>hypersensitivity and<br>reduction of post-<br>operative sensitivity | K090519<br>Treatment of dental<br>hypersensitivity and<br>reduction of post-<br>operative sensitivity |
| Contraindications | Hypersensitivity to<br>colophony should avoid<br>this product, as well as<br>patients with ulcerative<br>gingivitis and stomatitis | Ulcerative gingivitis and<br>stomatitis | Hypersensitivity to<br>colophony should avoid<br>this product, as well as<br>patients with ulcerative<br>gingivitis and stomatitis |
| Mode of Action | Dentin Tubule Occlusion | Dentin Tubule Occlusion | Dentin Tubule Occlusion |
| Method of<br>Application | Application of a thin coat<br>of varnish of varnish on<br>tooth surfaces only by a<br>professional. Not for<br>home use. | Application of a thin coat<br>of varnish of varnish on<br>tooth surfaces. | Application of a thin coat<br>of varnish of varnish on<br>tooth surfaces only by a<br>professional. Not for<br>home use. |
| Percentage of<br>Sodium Fluoride | 5% | 5% | 5% |
| Packaging | Varnish and applicator<br>brush placed in a<br>thermoformed PET/PVC<br>opaque unit dose tray<br>and closed with a<br>peelable foil seal. | Varnish and applicator<br>brush placed in a molded<br>opaque acclar tray and<br>closed with a foil seal. | Varnish and applicator<br>brush placed in a molded<br>opaque tray and closed<br>with a foil seal. |
| Unit Package | 0.4 mL | 0.4 mL | 0.5 mL |
| Basic Composition | Synthetic Rosin | Synthetic Rosin | Synthetic Rosin |
Technological Characteristics:
The basic composition, fundamental technology, and intended use of DFV Desensitizing Varnish is substantially equivalent to the predicate device and comparable to devices that have been on the market for many decades. These products use components used in the dental industry such as resins, rosins, modified rosins hydroxypropyl cellulose, polyurethane, methacrylates, polyethylene glycol, amides, dimethacrylates, and other film forming ingredients to cover dentinal tubules, which provide relief from dentinal
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue graphic of two Erlenmeyer flasks on the left, the text "SCI-PHARM" in blue in the center, and a blue mortar and pestle on the right. There is a registered trademark symbol next to the mortar and pestle.
hypersensitivity. DFV Desensitizing Varnish is a varnish that contains sodium fluoride, like all of the others, and contains rosin that provides immediate and sustained relief of dental pain or sensitivity. The immediate relief of hypersensitivity is achieved by the hardened resin films that forms immediately after application and occludes the dentin tubules.
Biocompatibility Tests: The chemical components in Sci-Pharm's DFV Desensitizing Varnish have been used in a variety of predicate devices. Rosin and other rosin derivatives are used as adhesive agents in approved dental varnishes.
> Biocompatibility testing is an essential element of developing a medical device. Each medical device used in dentistry shall be subjected to a structured biological evaluation program within a risk management process. In order to comply with the requirements, DFV Desensitizing Varnish was tested for cytotoxicity and to provide evidence that the formula blocks dentinal tubules.
> Cytotoxicity followed the procedure outlined in ISO 10993-5 and ISO 7405, where a smear was extracted in Minimal Essential Medium (MEM) with 10% Fetal Bovine Serum for 24 hours at 37°C. Then L929 cells were exposed to the extracts and the viability of the cells were measured by their ability to uptake a vital dye. The viability percentage of the cells exposed to the extracts from DFV Desensitizing Varnish was 11%, which is below the 70% (Toxicon Report 17-04113-G1). This value is expected as it contains significant ethanol, so an additional study was conducted using an approved dental varnish that is similar, and currently being used on the market. The predicate product is another dental varnish already on the market and compared with the same methodology to determine cytotoxicity from DFV Desensitizing varnish. The viability percentage of the cells exposed to the extracts from an equivalent device was also 11%, which is also below the 70% (Toxicon Report 18-00602-G1). The cytotoxicity for both varnishes are the same, indicating that there is no difference. Based on the chemical content, the residual ethanol will disrupt cellular integrity, the sodium fluoride is a bacterial growth inhibitor, which can also affect the viability of the cells, as can the carbamide peroxide. All of these chemicals had an effect that contributed towards the low cellular viability result, but since these are the very same chemicals used for all approved dental varnishes
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for SCI-PHARM. The logo is blue and features a beaker on the left, the text "SCI-PHARM" in the middle, and a mortar and pestle on the right. The logo is simple and professional.
currently on the market, then they are found acceptable. Additional testing can be requested, but the results will be the same due to the ethanol, not the flavor.
The second test is where dentin is exposed to the varnish and viewed under SEM to verify that the varnish blocks dentinal tubulae. There were two varnishes evaluated, Sci-Pharm's DFV Desensitizing Varnish and an equivalent predicate varnish. Disk-shaped human dentine specimens had the varnishes applied separately using a plastic brush and a single stroke. The surfaces were imaged using reflection optical microscopy (ROM), allowing the identification morphology of the specimen surfaces without the dehydration effects. Then the top surface of the sectioned specimens were examined using a scanning electron microscope (SEM). The ROM images demonstrated a typical granular appearance with dentinal tubule orifices closed by smear plugs. Both varnishes tested showed highly smooth material layer covering all of the dentine and when evaluated by SEM, Sci-Pharm's DFV Desensitizing Varnish remained smooth with minimal surface defects while the other varnish showed subsurface porosity and heterogeneity. The conclusion was that Sci-Pharm's DFV Desensitizing Varnish demonstrated a thick uniform film with fewer defects and protrusions than the competitors varnish. Additional testing is not necessary as the benefits are from sodium fluoride and the rosin mixture, not the flavors.
| | Proposed DFV<br>Desensitizing<br>Varnish | Equivalent Varnish |
|-----------------------|------------------------------------------|------------------------------------|
| Test | | |
| Contamination | No Foreign<br>Particles | No Foreign<br>Particles |
| Color | Match Standard | Match Standard |
| Odor | Match Standard | Match Standard |
| Viscosity | 4500 – 6500 cps | 7000 – 8500 cps |
| Non-<br>Pyrolizable % | 7.5% - 8.7% | 8.4% - 9.5% |
| Fluoride<br>Content | 4.5% - 5.5% NaF<br>(2.0% - 2.5% F) | 4.5% - 5.5% NaF<br>(2.0% - 2.5% F) |
## Physical Tests:
### Shelf Life
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for SCI-PHARM. The logo consists of a blue graphic of two Erlenmeyer flasks on the left, followed by the text "SCI-PHARM" in blue. To the right of the text is a blue mortar and pestle with a registered trademark symbol.
Shelf Life was established following Sci-Pharm's internal procedure, EX-06, where the product being tested is analyzed during intervals within the shelf life, such as Time 0, Time 6 months, Time 12 months, Time 18 months, etc., and continued for up to 4 years. The product is evaluated for any changes that may occur over time. Testing is performed on contamination, color, viscosity, fluoride content, nonpyrolizable percentage, and odor. Color changes may occur over time, but since a small thin layer that is washed off within hours, the color change in the product is not substantial. Report 2019-R-012 provides evidence that the product is stable for at least 24 months.
Substantial Equivalence: In summary, this submission demonstrates that Sci-Pharm's DFV Desensitizing Varnish is substantially equivalent in safety and effectiveness, and performs equivalently or better than the identified primary predicate device and the reference device varnishes for their intended use. Device is biocompatible when used as directed by dental professionals per ISO 10993-1. DFV Desensitizing Varnish does not introduce any new indications for use or potential hazards.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.