Ethos Treatment Management is indicated for use in managing and monitoring radiation therapy treatment plans and sessions. Ethos Treatment Planning is indicated for use in generating and modifying radiation therapy treatment plans. Halcyon and Ethos Radiotherapy System are indicated for the delivery of stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated for adults and pediativ patients.
Device Story
System comprises Ethos Treatment Management, Ethos Treatment Planning, and Ethos Radiotherapy System/Halcyon linacs. Workflow: Management software tracks treatment sessions; Planning software generates/modifies plans, including adaptive radiotherapy; Radiotherapy system delivers 6 MV FFF photon beams via IMRT/VMAT. System uses kV imaging for IGRT and patient motion monitoring. Operated by radiation therapy professionals in clinical settings. Output includes treatment plans and delivered radiation doses. Benefits include precision radiotherapy and adaptive capabilities to account for patient anatomy changes during treatment course.
Clinical Evidence
No clinical data or animal studies included. Evidence based on bench testing, including hardware/software verification and validation, electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
Medical linear accelerator (6 MV FFF, 800 MU/min) with MLC, kV imager, and patient support. Software components for management and planning. Standards: IEC 62304, 61217, 62274, 62083, 82304-1, 62366-1, 60601 series. Software class: Major level of concern. Connectivity: Networked system with treatment/imaging consoles.
Indications for Use
Indicated for adults and pediatric patients requiring stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 10, 2020
Varian Medical Systems, Inc. % Mr. Peter J. Coronado Senior Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
# Re: K192377
Trade/Device Name: Ethos Treatment Management, Ethos Treatment Planning, Ethos Radiotherapy System, and Halcyon Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE, MUJ Dated: January 10, 2020 Received: January 13, 2020
# Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192377
Device Name Ethos Treatment Management, Ethos Treatment Planning, Ethos Radiotherapy System, and Halcyon
Indications for Use (Describe)
Ethos Treatment Management is indicated for use in managing and monitoring radiation therapy treatment plans and sessions.
Ethos Treatment Planning is indicated for use in generating and modifying radiation therapy treatment plans.
Halcyon and Ethos Radiotherapy System are indicated for the delivery of stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated for adults and pediativ patients.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# PREMARKET NOTIFICATION
# 510(k) Summary
# Ethos Treatment Management 2.0, Ethos Treatment Planning 1.0, Ethos Radiotherapy System 3.0 & Halcyon 3.0
The following information is provided as required by 21 CFR 807.92.
| Submitter's Name: | Varian Medical Systems<br>3100 Hansen Way, m/s E110<br>Palo Alto, CA 94304 | | | | | | | | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|
| | Contact Name: Peter J. Coronado, Senior Director Regulatory Affairs<br>Phone: 650-424-6320 Fax: 650-646-9200<br>E-mail: submissions.support@varian.com<br>Date Prepared: 29 August 2019 | | | | | | | | |
| Proprietary Names: | Ethos Treatment Management<br>Ethos Treatment Planning<br>Ethos Radiotherapy System<br>Halcyon | | | | | | | | |
| Classification Name: | Medical charged-particle radiation therapy system<br>21 CFR 892.5050 | | | | | | | | |
| Common/Usual Names: | Ethos Treatment Management Treatment Plan and Image Management Application Ethos Treatment Planning Treatment Planning System Ethos Radiotherapy System Medical Linear Accelerator Halcyon Medical Linear Accelerator | | | | | | | | |
| Predicate Devices: | Ethos Treatment Management ARIA Radiation Therapy Management (K173838) Ethos Treatment Planning Eclipse Treatment Planning System (K181145) Ethos Radiotherapy System Halcyon (K181032) Halcyon Halcyon (K181032) | | | | | | | | |
| Device Description: | Ethos Treatment Management Ethos Treatment Management is software designed<br>for radiation therapy medical professionals to<br>support them in managing radiation treatments for<br>patients. Ethos Treatment Planning Ethos Treatment Planning is software that is<br>designed generate treatment plans, modify<br>treatment plans, and guide users within adaptive<br>treatment sessions. Ethos Radiotherapy System Halcyon and Ethos Radiotherapy System are single<br>energy linacs designed to deliver Image Guided<br>Radiation Therapy and radiosurgery, using Intensity<br>Modulated and Volumetric Modulated Arc Therapy Halcyon techniques. They consist of an accelerator and<br>patient support within a radiation shielded<br>treatment room and a control console outside the<br>treatment room. | | | | | | | | |
{4}------------------------------------------------
| Ethos Treatment Management | Ethos Treatment Management is used to manage and monitor radiation therapy treatment plans and sessions; it is intended to be used with a treatment planning system. |
|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ethos Treatment Planning | Ethos Treatment Planning is used to generate and modify radiation therapy treatment plans. |
| Ethos Radiotherapy System<br>(The intended use statement is<br>identical to the predicate) | Halcyon and Ethos Radiotherapy System are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. |
| Halcyon<br>(The intended use statement is<br>identical to the predicate) | |
#### Indications for Use:
| Ethos Treatment Management | Ethos Treatment Management is indicated for use in<br>managing and monitoring radiation therapy<br>treatment plans and sessions. |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ethos Treatment Planning | Ethos Treatment Planning is indicated for use in<br>generating and modifying radiation therapy<br>treatment plans. |
| Ethos Radiotherapy System | Halcyon and Ethos Radiotherapy System are<br>indicated for the delivery of stereotactic radiosurgery<br>and precision radiotherapy for lesions, tumors and<br>conditions anywhere in the body where radiation is |
| Halcyon | indicated for adults and pediatric patients. |
#### Relationships Between the Devices:
- 1. The three devices of Ethos Treatment Management, Ethos Treatment Planning, and Ethos Radiotherapy System are used together within a radiation therapy treatment workflow. The overall clinical workflow involves each device performing a designated set of functions for radiotherapy treatment related to management, planning, and delivery which are best understood and described together.
- Ethos Radiotherapy System and Halcyon are similar device models within a generic device type that share 2. significant technological and functional characteristics.
- 3. For this Traditional 510(k) submission, these devices have been bundled together such that they can all be addressed during one review, given the shared scientific and regulatory considerations these devices raise.
#### Comparison of Technological Characteristics with the Predicate Device:
At a high level, each subject device and its corresponding predicate are similar based on the following characteristics:
#### Ethos Treatment Management
- Stand-Alone Software for Radiotherapy Management
- Support for Management and Monitoring Tools related to Radiation Treatment
- Support for Recording and Storing Treatment Parameters and Treatment Course Progression ●
#### Ethos Treatment Planning
- Stand-Alone Software for Radiotherapy Treatment Planning
- Support for Radiotherapy Treatment Planning with an External Beam Photon Delivery System
- Support for IGRT, IMRT, IMRS, VMAT Treatment Planning Techniques ●
{5}------------------------------------------------
### Ethos Radiotherapy System
- Medical Linear Accelerator with Patient Support Couch ●
- . Support for Radiotherapy Treatment using External Beam Photon Delivery
- . Treatment Energy: 6 MV Flattening Filter Free
- Dose rate: Up to 800 MU/min
- Integrated Components: Treatment and Imaging Consoles, MLC, kV imager, Patient Camera, and others
#### Halcyon
- Medical Linear Accelerator with Patient Support Couch ●
- Support for Radiotherapy Treatment using External Beam Photon Delivery
- Treatment Energy: 6 MV Flattening Filter Free ●
- Dose rate: Up to 800 MU/min
- Integrated Components: Treatment and Imaging Consoles, MLC, kV imager, Patient Camera, and others ●
### Significant Differences:
At a high level, each subject device and its corresponding predicate are different based on the following features:
### Ethos Treatment Management
- Enhanced Work Areas and Tools for Users to specify Radiation Treatment Intent
- . Enhanced Work Areas and Tools for Monitoring over the course of Radiation Treatment
- Compatibility with Ethos Treatment Planning and Ethos Radiotherapy System
### Ethos Treatment Planning
- Support for Adaptive Radiotherapy Treatment Planning and Automated Plan Generation
- Support for Treatment Session Management with Guided Workflows and Work Areas .
- Compatibility with Ethos Treatment Management and Ethos Radiotherapy System
### Ethos Radiotherapy System
- Support for Adaptive Radiotherapy Treatment Delivery ●
- Compatibility with Ethos Treatment Management and Ethos Treatment Planning
- Patient Motion Monitoring & Extended Treatment Field Area
#### Halcyon
- Patient Motion Monitoring & Extended Treatment Field Area (same as in Ethos Radiotherapy System) ●
# Performance Data (Non-Clinical Testing):
Hardware and software verification and validation testing was conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards listed below.
Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
The software for these devices was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Electrical safety and electromagnetic compatibility (EMC) testing were conducted for the linear accelerators. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
No animal studies or clinical tests have been included in this pre-market submission.
{6}------------------------------------------------
## Standards Conformance:
The devices conform in whole or in part to the following recognised standards.
| Applied Standard | Device Name | | |
|------------------|----------------------------------------|--------------------------|---------------------------------------|
| IEC 62304 | Ethos Treatment Management | Ethos Treatment Planning | Ethos Radiotherapy System and Halcyon |
| IEC 61217 | | Ethos Treatment Planning | Ethos Radiotherapy System and Halcyon |
| IEC 62274 | Ethos Treatment Management | | Ethos Radiotherapy System and Halcyon |
| IEC 62083 | Ethos Treatment Management | Ethos Treatment Planning | |
| IEC 82304-1 | Ethos Treatment Management | Ethos Treatment Planning | |
| IEC 62366-1 | Ethos Treatment Management | Ethos Treatment Planning | |
| IEC 60825-1 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60976 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-1 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-1-2 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-1-3 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-1-6 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-2-1 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
| IEC 60601-2-68 | Not applicable to the software devices | | Ethos Radiotherapy System and Halcyon |
Additional general (non-device specific) standards applied include:
- EN ISO 13485:2016 Edition 3 2016-03
- O Medical devices – Quality management systems – Requirements for regulatory purposes
- EN ISO 14971:2012 2012-07
- Medical devices Application of risk management to medical devices o
- ISO 15223-1 Third Edition 2016-11-01
- O Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements
# Argument for Substantial Equivalence to the Predicate Device
A subset of technological characteristics and features of each of the current devices are different to their corresponding predicates. These differences are all considered by Varian to be enhancements of the predicate.
The Intended Use and indications for use are not significantly different. There are no changes in the principle of operation of the devices. The Verification and Validation demonstrates that the device is as safe and effective as the predicate. Varian therefore believes that Ethos Treatment, Ethos Treatment Planning, Ethos Radiotherapy System, and Halcyon are substantially equivalent to their corresponding predicates.
#### Conclusion:
The assessment following the outcomes observed in the performance testing and software verification and validation was that these products conformed to the defined user needs and intended uses. There were no DRs (discrepancy reports) remaining which had a priority of Safety Intolerable. Varian therefore considers Ethos Treatment Management, Ethos Treatment Planning, Ethos Radiotherapy System, and Halcyon to be safe and effective and to perform at least as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.