K192338 · New England Ortho Lab, Inc. · NXC · Mar 12, 2020 · Dental
Device Facts
Record ID
K192338
Device Name
NEOLab Clear Aligners
Applicant
New England Ortho Lab, Inc.
Product Code
NXC · Dental
Decision Date
Mar 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NEOLab Clear Aligners are indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). The NEOLab Clear Aligners position teeth through the use of continuous gentle force.
Device Story
NEOLab Clear Aligners are customized, removable, thermoformed polyurethane trays; designed to move teeth incrementally from misalignment to aligned state. Input: physical impressions or intraoral scans. Process: impressions scanned to digital files; 3D printing of models; thermoforming of aligners. Used in dental clinic; operated by dental clinician. Clinician distributes sequential aligners to patient; monitors progress via orthodontic exams. Output: mechanical force applied to teeth. Benefit: progressive tooth realignment.
Clinical Evidence
Bench testing only. Mechanical properties (tensile strength, elongation, modulus) verified per ASTM D638 for 5-year shelf life. Biocompatibility testing performed per ISO 10993 (Parts 3, 5, 10, 11), confirming non-cytotoxicity and safety. No animal or human clinical testing required due to material and design equivalence to predicate.
Technological Characteristics
Customized thermoformed polyurethane aligners. Mechanical properties tested per ASTM D638. Biocompatibility per ISO 10993. Manufacturing involves 3D printing of digital models derived from impressions or intraoral scans. Non-sterile.
Indications for Use
Indicated for treatment of tooth malocclusions in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Thursday, March 12, 2020
New England Ortho Lab, Inc. % Patsy Trisler Regulatory Consultant Oserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
Re: K192338
Trade/Device Name: NEOLab Clear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: February 21, 2020 Received: February 24, 2020
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Srivinas 'Nandu' Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192338
Device Name
NEOLab Clear Aligners
Indications for Use (Describe)
NEOLab Clear Aligners are indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). The NEOLab Clear Aligners position teeth through the use of continuous gentle force.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter Name: | New England Ortho Lab, Inc. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 3 Riverside Drive<br>Andover, MA 01810 |
| Phone Number: | 1-800-922-6365 |
| Contact Person: | Ms. Olivia Channon<br>Document Control Coordinator |
| Date Prepared: | February 21, 2020 |
| Device Trade Name: | NEOLab Clear Aligners |
| Common Name | Aligner, Sequential |
| Classification Name<br>Number<br>Product Code<br>Regulatory Class | Orthodontic Plastic Bracket<br>21 CFR 872.5470<br>NXC<br>2 |
| Primary Predicate<br>Device: | K113618, ClearCorrect System, ClearCorrect, LLC |
| Reference Devices: | K152086, Ortho System™, 3Shape A/S<br>K180941, Ortho System™ , 3Shape A/S |
| Indications for Use<br>Statement | NEOLab Clear Aligners are indicated for the treatment of tooth<br>malocclusions in patients with permanent dentition (i.e. all second<br>molars). The NEOLab Clear Aligners position teeth through the<br>use of continuous gentle force. |
| Device Description<br>and Summary of<br>Technological<br>Characteristics | NEOLAB Clear Aligners consist of a series of customized clear<br>plastic removable aligners that are fabricated from a clear, thin<br>thermoformed polyurethane. The aligners are designed to<br>gradually move the patient's teeth incrementally, repositioning<br>them from their original misalignment to a more aligned state.<br><br>The aligners, customized for each patient's prescription, are<br>designed and manufactured based on either standard impressions<br>or intraoral scans taken by the dental clinician and which are then<br>sent to the company. The aligners are then sent back to the<br>dental clinician who then distributes them in sequential stages to<br>the patient and follows up with the patient through orthodontic<br>examinations to check device fit and function.<br><br>During manufacturing, models are made from physical<br>impressions or intraoral scans. Those made from stone or plaster<br>impressions are scanned into the computer and made into digital<br>files. The intraoral scans arrive as digital files. The digital files<br>are 3D printed and are used to produce the aligner series with the<br>thermoplastic polyurethane. |
| The thermoplastic material used for fabrication of the aligners is<br>commonly used in many dental appliances, including the<br>predicate aligners. | |
| Mechanism of<br>Action | Based on the clinician's treatment plan, each aligner is used for a<br>defined period of time to exert gentle force to achieve progressive<br>realignment of the teeth. This occurs over time until the final<br>correction has been achieved. |
| Device Testing | Laboratory/Bench Testing<br>Test data were submitted to:<br>assure the mechanical properties of the aligner material<br>meet specifications for up to 5 years shelf life (according to<br>ASTM D638) - testing included: tensile strength (PSI) elongation (%) tensile @ yield (PSI) elongation @ yield (PSI) tensile modulus (PSI) assure the aligner material packaging retains the required<br>moisture barrier properties; validate the processes used for the design and<br>manufacture of the customized aligners, to ensure<br>consistency between the aligner's design and the<br>manufactured aligners. All testing met the pre-determined acceptance criteria.<br>Biocompatibility<br>The thermoplastic polyurethane used for making the aligner series<br>has been tested according to Good Laboratory Practices for its<br>biocompatibility according to ISO 10993, as follows:<br>Part 3 (Bacterial Mutagenicity – Ames Assay)<br>Part 5 (Cytotoxicity Elution - MEM),<br>Part 10 ((Intracutaneous/Intradermal) Reactivity),<br>Part 10 (Oral Mucosa Irritation),<br>Part 10 (Maximization for Delayed-Type Hypersensitivity),<br>Part 11 (Subacute Systemic Toxicity)<br>Further, the finished customized aligner was tested according to<br>ISO 10993, Part 5, Cytotoxicity. Results show it is non-cytotoxic.<br>All testing showed that the material and manufactured aligner met<br>the requirements of the test methods and are safe and<br>biocompatible for the stated intended use. |
| | Animal Human Testing<br>Neither animal nor human testing are required for this product<br>because it is composed of the same materials, is designed<br>similarly, and is manufactured by a similar method as the predicate<br>device. |
| Comparison to<br>Predicate Device: | There are no notable differences comparing the NEOLab clear<br>aligner to the ClearCorrect predicate device:<br>The intended use is the same. The mechanism of action is similar. The polyurethane material used to make the aligners is the same. The method of manufacture and customizing the aligners is similar. The use of software for planning and manufacturing are similar. Refer to the Substantial Equivalence Comparison table below for a<br>side-by-side comparison of the indications for use, technological<br>characteristics, materials and principles of operation. |
| Substantial<br>Equivalence<br>Conclusion: | As presented in the 510(k) and summarized herein, it can be<br>concluded that NEOLab Clear Aligner is substantially equivalent to<br>the predicate device. |
{4}------------------------------------------------
{5}------------------------------------------------
I
{6}------------------------------------------------
| Trade Name: | NEOLab Aligner System | ClearCorrect System |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K1923338 | K113618 |
| Manufacturer | New England Ortho Lab, Inc. | ClearCorrect, LLC |
| Classification # &<br>Product Code<br>Class | 21 CFR 852.5470<br>NXC<br>2 | 21 CFR 852.5470<br>NXC<br>2 |
| Indications for<br>Use | NEOLab Clear Aligners are<br>indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent dentition (i.e. all<br>second molars). The NEOLab<br>Clear Aligners position teeth by<br>way of continuous gentle force. | The ClearCorrect System is<br>indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent dentition (i.e. all<br>second molars). The Clear<br>Correct System positions teeth by<br>way of continuous gentle force. |
| Mode of Action | Alignment of teeth by<br>application of continuous gentle<br>force, by sequential use of<br>preformed plastic trays. | Alignment of teeth by application<br>of continuous gentle force, by<br>sequential use of preformed<br>plastic trays. |
| Method of Use | Each preformed plastic tray is<br>worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the next<br>sequential aligner tray. | Each preformed plastic tray is<br>worn by the patient as prescribed<br>by the dental practitioner, usually<br>a few weeks prior to using the<br>next sequential aligner tray. |
| Material | Thin thermoformed<br>polyurethane | Thin thermoformed polyurethane |
| Biocompatible | Yes | Yes |
| OTC or Rx | Rx | Rx |
| Software Use | Yes | Yes |
| Sterile | No | No |
### Substantial Equivalence Comparison
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.