K192325 · Life Warmer, Inc. · LGZ · Oct 18, 2019 · General Hospital
Device Facts
Record ID
K192325
Device Name
Quantum Blood and Fluid Warming System
Applicant
Life Warmer, Inc.
Product Code
LGZ · General Hospital
Decision Date
Oct 18, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Quantum Blood and Fluid Warming System is indicated for warming blood, blood products and intravenous solutions prior to administration in adult and pediatic patients greater than 28 days old of normal birth weight. It is intended for use by healthcare professionals in hospital, clinical, field, and transport environments to help prevent hypothermia. The Quantum is not for use with neonates (birth to 28 days old) or infants of low birth weight.
Device Story
Portable, battery-powered, in-line fluid warming system; consists of controller, battery, charger, and sterile disposable thermal IV tubing (TIS/TTS-B). Warming element integrated into tubing; does not contact fluid directly. Microprocessor-controlled resistive heating; thermistors on tubing provide feedback to maintain 38°C ± 2°C. Used by healthcare professionals (physicians, nurses, EMTs, military medics) in hospitals, clinics, field, and transport environments. Output is warmed fluid delivered to patient; visual/audible alerts for temperature, battery, and flow status. Benefits include hypothermia prevention during fluid/blood administration. Clinicians use device output to manage patient temperature during emergencies or surgeries.
Clinical Evidence
No clinical data; bench testing only. Substantial equivalence supported by risk analysis, design controls, and previously cleared performance data. Pediatric use supported by analysis of fluid volume/flow rate requirements and updated labeling.
Indicated for warming blood, blood products, and IV solutions in adult and pediatric patients >28 days old of normal birth weight. Contraindicated for neonates (birth to 28 days) and infants of low birth weight.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Quantum Blood and IV Fluid Infusion Warmer (K181775)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2019
Life Warmer, Inc. John Pettini Chief Medical Officer, Founder 4813 Keller Springs Rd Addison, Texas 75001
Re: K192325
Trade/Device Name: Quantum Blood and Fluid Warming System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZG, BSB Dated: August 26, 2019 Received: August 27, 2019
Dear John Pettini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Nikhil Thakur Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K192325
Device Name Quantum Blood and Fluid Warming System
Indications for Use (Describe)
The Quantum" Blood and Fluid Warming System is indicated for warming blood, blood products and intravenous solutions prior to administration in adult and pediatic patients greater than 28 days old of normal birth weight. It is intended for use by healthcare professionals in hospital, clinical, field, and transport environments to help prevent hypothermia.
The Quantum is not for use with neonates (birth to 28 days old) or infants of low birth weight.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LifeWarmer. The logo consists of a stylized letter "W" in red, followed by the word "LIFEWARMER" in a light gray sans-serif font. The logo is simple and modern, and the colors are eye-catching.
| Submitted by/<br>Sponsor: | Life Warmer, Inc.<br>4813 Keller Springs Rd<br>Addison, TX 75287 USA<br>972-908-9808 | Contact Person: | John Pettini, DO, FACEP<br>860-204-1711 |
|------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------|-----------------------------------------|
| Date Prepared: | October 18, 2019 | | |
| Trade Name: | Quantum Blood and Fluid Warming System | | |
| Common Name: | Sterile Fluid Path in-line Blood and Fluid Warmer | | |
| Classification Code<br>Name & Reference: | LGZ | Warmer, Thermal, Infusion Fluid 21 CFR §880.5725 | |
| | BSB | Warmer, Blood, Non-electromagnetic radiation | |
| Predicate Device | K181775: Quantum Blood and IV Fluid Infusion Warmer by Life Warmer, Inc. | | |
# Device Description:
The Quantum Blood and Fluid Warming System is a light weight, portable, battery powered, in-line fluid warming system in which specially designed thermal IV administration tubing is integrated with a warming element. The device is prescription only.
Indications for Use: The Quantum Blood and Fluid Warming System is indicated for warming blood, blood products and intravenous solutions prior to administration in adult and pediatic patients greater than 28 days old of normal birth weight. It is intended to be used by healthcare professionals in hospital, field and transport environments to help prevent hypothermia.
The Quantum is not for use with neonates (birth to 28 days old) or infants of low birth weight.
Technological Characteristics: The technological characteristics of the Quantum Blood and Fluid Warming System are unchanged from the 510(k)-cleared device which is also the predicate. The System is a light weight, portable, battery powered, in-line fluid warming system in which specially designed thermal IV administration tubing is integrated with a warming element. The Quantum is comprised of the following components: Controller, Battery, Charger, and Quantum thermal tubing available in two configurations: TIS (Thermal Infusion Set) and TTS-B (Thermal Transfusion Set for blood) with standard IV administration components. The microprocessor and algorithm in the Quantum Controller continually receive and assess temperature input from thermistors located on the tubing segments. In response to this input, the Controller regulates the energy applied to the tubing to reach and/or maintain a pre-set temperature range of 38 °C ± 2 °C at delivery. A double-extrusion process results in tubing configuration with an inner tubing layer and an outer tubing laver. The warming element is in between but does not contact the fluid. Warming begins at the fluid source and continues the length of the tubing until near the point of patient entry. The TIS an TTS-B tubing perform the same function as standard IV tubing, however, when warming is desired, the TIS or TTS-B is simply connected to the Quantum Controller is connected to the Battery. The TIS and TTS-B tubing sets are provided sterile (i.e., fluid path) for single patient use. The Controller, Battery and Battery Chargers are reusable.
### Comparison to Predicate
The design and technological features of the Quantum Blood and Fluid Warming System that is the subject of this Special 510(k) is identical to the Quantum Blood and IV Fluid Infusion Warmer previously cleared under K181775. The only difference is the inclusion of pediatric patients. This difference does not result in any design change, new or increased risk. The following table presents a comparison of the devices' technology and features.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for LifeWarmer. The logo consists of a stylized letter "W" in red, followed by the word "LIFEWARMER" in a lighter shade of red. The logo is simple and modern, and the colors are eye-catching.
| Substantial Equivalence Comparison Table | | | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject device | Primary Predicate | |
| Trade Name | Quantum Blood and Fluid Warming<br>System (K192325) | Quantum Blood and IV Fluid Infusion<br>Warmer (K181775) | |
| Common Name | Sterile Fluid Path in-line Blood and Fluid<br>Warmer | Sterile Fluid Path, In-Line Blood Fluid<br>Warmer | Same |
| Regulation | 21 CFR Part 880.5725 | 21 CFR Part 880.5725 | Same |
| Prod. Code | LGZ, BSB | LGZ, BSB | Same |
| Indication for<br>Use | The Quantum™ Blood and IV Fluid Infusion<br>Warmer is indicated for warming blood,<br>blood products and intravenous solutions<br>prior to administration in adult and pediatric<br>patients greater than 28 days old of normal<br>birth weight. It is intended for use by<br>healthcare professionals in hospital, clinical,<br>field and transport environments to help<br>prevent hypothermia.<br>The Quantum is not for use with neonates<br>(birth to 28 days old) or infants of low birth<br>weight. | The Quantum™ Blood and IV Fluid<br>Infusion Warmer is indicated for warming<br>blood, blood products and intravenous<br>solutions prior to administration in adult<br>patients. It is intended for use by<br>healthcare professionals in hospital,<br>clinical, field and transport environments to<br>help prevent hypothermia. | Different:<br>Pediatric<br>patients are<br>included in the<br>indication for<br>use.<br>Labeling was<br>modified to<br>mitigate new<br>and increased<br>risks to<br>pediatric<br>patients. |
| Intended Use | Medical emergencies or surgeries where<br>warm fluid administration is required to treat<br>the patient. Whenever parenteral<br>introduction of normothermic fluid is desired<br>or indicated. | Medical emergencies or surgeries where<br>warm fluid administration is required to<br>treat the patient. Whenever parenteral<br>introduction of normothermic fluid is<br>desired or indicated. | Same |
| User Population | Healthcare professionals (e.g., physicians,<br>registered nurses, mid-level practitioners,<br>EMT/Paramedic, military medics) | Healthcare professionals (e.g., physicians,<br>registered nurses, mid-level practitioners,<br>EMT/Paramedic, military medics) | Same |
| Use Environment | Hospital, Clinic, Field and Transport | Hospital, Clinic, Field and Transport | Same |
| User Interface | Visual (LED) and audible | Visual (LED) and audible | Same |
| User Feedback<br>Provided | Over temperature, under temperature,<br>battery low, no-flow/poor connection,<br>system error | Over temperature, under temperature,<br>battery low, no-flow/poor connection,<br>system error | Same |
| System<br>Components | Sterile disposable thermal tubing (TIS/TTS-<br>B), Controller (with LEDs), Battery (w/<br>LEDs and audible alert) | Sterile disposable thermal tubing<br>(TIS/TTS-B), Controller (with LEDs),<br>Battery (w/ LEDs and audible alert) | Same |
| Infusion<br>Temperature | $38 °C ± 2°C$ | $38 °C ± 2°C$ | Same |
| Fluid Path | Sterile; direct path with disposable IV<br>administration tubing | Sterile; direct path with disposable IV<br>administration tubing | Same |
| Flow Rate | Gravity<br>2 to 200 mL/min<br>(depending on starting temperatures) | Gravity<br>2 to 200 mL/min<br>(depending on starting temperatures) | Same |
| Heating Method | Resistive heating | Resistive Heating | Same |
| Heating Control | Software | Software | Same |
| Warmer Type | In-line | In-Line | Same |
| Power Source | Rechargeable battery | Rechargeable battery | Same |
| Biocompatibility | Biocompatibility testing demonstrates<br>tubing/fluid path to be biocompatible and<br>non-bioreactive. | Biocompatibility testing demonstrates<br>tubing/fluid path to be biocompatible and<br>non-bioreactive | Same |
| Sterilization | Disposable TIS, TTS-B tubing:<br>(ethylene oxide) | Disposable TIS, TTS-B tubing<br>(ethylene oxide) | Same |
| Product Specific<br>Standards | ASTM 2172: 2002<br>Standard specification for<br>blood/intravenous fluid irrigation warmer | ASTM 2172:2002<br>Standard Specification for<br>blood/intravenous Fluid Irrigation Fluid<br>Warmers | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for LIFEWARMER. The logo consists of a stylized letter "W" in red, followed by the word "LIFEWARMER" in a smaller, sans-serif font, also in red. The logo appears to be for a company or product named LIFEWARMER.
Substantial Equivalence: The Quantum Blood and Fluid Warminq System subject device is identical in design and technological features as the 510(k) cleared, commercially distributed Quantum predicate device (K181775). The only difference is the inclusion of pediatric patients greater than 28 days old of normal birth weight as stated in the indications for use statement. The additional risks from this device to pediatric patients have been mitigated through labeling, by incorporating warnings and precautions specific to pediatrics. No modification to the device, different or additional equipment, set or training is required. No new questions of safety and effectiveness are raised by this modification.
# Sterilization and Shelf-Life
The Quantum TIS and TTS-B tubing assemblies are disposable with a sterile fluid path. Sterilization was achieved by exposure to ethylene oxide (EO) and validated in accordance with ANSI/AAMI/ISO 11135:2014. ANSI/AAMI/ISO 11737-1: 2006/(R) 2011: ANSI/AAMI/ISO 11737-2: 2009/(R) 2014: ANSI/AAMI/ISO 11138-1: 2006/R) 2015: ANSI/AAMI/ISO 11138-2: 2006/(R) 2015: ANSI/AAMI/ISO10993-7: 2008/(R) 2012; ISO 11607-1:2006, ANSI/AAMI/ISO 10993/(R) 2013.
The TIS and TTS-B shelf-life was supported with testing performed in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, ISTA-2A (2011) Partial Simulation Performance Tests. Bacterial Endotoxin testing was performed using the Kinetic-Chromogenic Method.
# Performance Testing Summary
No additional performance testing was required to support the addition of pediatric patients greater than 28 days old of normal birth weight to the patient user group for this device. The Quantum Blood and Fluid Warming System has been subjected to Design Controls and previously tested to appropriate devicespecific standards to support substantial equivalence, and validation to standards necessary to demonstrate the functionality of the device as presented below:
### Risk Analysis:
The risk profile has changed with the addition of the pediation. An updated risk analysis has been performed to support the addition of all pediatric populations, with the exception of neonates and low birth weight infants. The Table below shows of how each residual risk applies to this population.
| System/ Patient | Adult | Adolescents (12-<br>21 years)<br>39.5 kg to 70.3kg | Preadolescent<br>(11-13 years)<br>35.6 kg to 45.3 kg | Children (2-12<br>years) 12.5 kg<br>to 39.5 kg | Infants (29<br>days to 2<br>years) 4.4 kg-<br>12.5 kg |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Users | Healthcare<br>professionals (e.g.,<br>physicians,<br>registered<br>nurses, midlevel<br>practitioners,<br>EMT/Paramedic,<br>military medics | Healthcare<br>professionals (e.g.,<br>physicians,<br>registered nurses,<br>midlevel<br>practitioners,<br>EMT/Paramedic,<br>military medics | Healthcare<br>professionals (e.g.,<br>physicians,<br>registered<br>nurses, midlevel<br>practitioners,<br>EMT/Paramedic,<br>military medics | Healthcare<br>professionals (e.g.,<br>physicians,<br>registered<br>nurses, midlevel<br>practitioners,<br>EMT/Paramedic,<br>military medics | Healthcare<br>professionals (e.g.,<br>physicians,<br>registered nurses,<br>midlevel<br>practitioners,<br>EMT/Paramedic,<br>military medics |
| Tubing Set<br>Components | Y-spike (vented an<br>and non-vented), 3<br>roller clamps, drip<br>chamber, 2<br>injection ports, 2<br>slide clamps, luer<br>adapter | No additional<br>components<br>required | No additional<br>components<br>required | No additional<br>components<br>required | No additional<br>components<br>required |
| Fluids | IV Fluids, Blood,<br>Blood components | IV Fluids, Blood,<br>Blood components | IV Fluids, Blood,<br>Blood components | IV Fluids, Blood,<br>Blood components | IV Fluids, Blood,<br>Blood components |
| Resuscitation<br>Flow Rate<br>(mL/min) with 10<br>mL/kg bolus<br>over 5 minutes | 100 - 200 | 70 - 200 | 71.2 - 90.6 | 25 - 79 | 6.6 - 25 |
| Resuscitation<br>Flow Rate<br>(mL/min) with 20<br>mL/kg bolus<br>over 5 minutes | 100 - 200 | 79 - 200 | 142.2 - 181.2 | 50 - 158 | 17.6 -50 |
| Resuscitation<br>Volume (mL)<br>From AHA<br>Pediatric<br>Advanced Life<br>Support<br>Guidelines | 1000 - 2000 mL | 20 mL/kg bolus up<br>to X3 isotonic<br>crystalloids.<br>10 mL/kg PRBC's<br>for hemorrhagic<br>shock | 20 mL/kg bolus up<br>to X3 isotonic<br>crystalloids.<br>10 mL/kg PRBC's<br>for hemorrhagic<br>shock | 20 mL/kg bolus up<br>to X3 isotonic<br>crystalloids.<br>10 mL/kg PRBC's<br>for hemorrhagic<br>shock | 20 mL/kg bolus up<br>to X3 isotonic<br>crystalloids.<br>10 mL/kg PRBC's<br>for hemorrhagic<br>shock |
| Labelling | Currently available | Additional<br>warnings and<br>precautions<br>regarding<br>hypothermia,<br>contamination,<br>infection,<br>thermoregulation | Additional<br>warnings and<br>precautions<br>regarding<br>hypothermia,<br>contamination,<br>infection,<br>thermoregulation | Additional<br>warnings and<br>precautions<br>regarding<br>hypothermia,<br>contamination,<br>infection,<br>thermoregulation | Additional warning<br>to not use with low<br>birth weight infants<br>Additional<br>warnings and<br>precautions<br>regarding<br>hypothermia,<br>contamination,<br>infection,<br>thermoregulation |
| Biocompatibility | ISO 10993 | No additional<br>requirements | No additional<br>requirements | No additional<br>requirements | No additional<br>requirements |
| Electrical Safety | IEC 60601 | No additional<br>requirements | No additional<br>requirements | No additional<br>requirements | No additional<br>requirements |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for LIFEWARMER. The logo consists of a stylized "W" symbol in red, followed by the word "LIFEWARMER" in a sans-serif font, also in red. The logo is simple and modern, with a focus on the company name.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for LifeWarmer. The logo consists of a stylized letter "W" in red, followed by the text "LIFEWARMER" in a bold, red font. The left side of the text is slightly blurred, while the right side is more sharply focused.
### Risk Analysis Summary:
The predicate K181775 Quantum system is indicated for adult patients. Hypothermia prevention also applies to pediatric patient can achieve its intended use at the resuscitation flow rates indicated for pediatric patients. Additional requirements or safety measures were assessed, including biocompatibility, electrical safety, design, and changes in the users and use scenarios of the device. The system has no additional biocompatibility or electrical safety requirements when used with pediatrics. The design of the patient contacting component, the tubing set, was assessed to address whether any additional risk may arise when with pediatric population. The users are generally the same, however there are differences in use cases when being used with pediatrics. Clinician users of the Quantum who treat pediatric patients would have the same credentials as current Quantum users, and by training be knowledgeable in appropriate fluid volume requirements and flow rates for pediatric patients. Hence, this narrows the focus of the risk assessment down to use. The table above lists the various fluid volume and flow requirements for pediatic patients which can all be achieved with the current Quantum System. The Quantum Use Failure Modes and Effects Analysis (FMEA) and Thermal Tube Set (TTS) Design FMEA has been updated to assess if any additional risks arise when considering a pediatric patient population and labeling was updated to include warnings and precautions specific to the pediatric population in order to mitiqate for the change in risk severity has been increased for risks related to contamination, air embolisms, hypothermia and others, since pediatric patients are at a higher risk for certain failure modes as compared to adult patients.
Considering risks involved with device use with pediatric patients, existing risks were revaluated to confirm whether their severity scores remained the same or were increased due to possibly more serious effects with pediatric patients, and the analysis was reviewed in order to identify any new risks. Existing risks that relate to IV administration and blood warmer use that are inherent to the r severity unchanged when considering pediatric population since these risks were related to operator training and basic knowledge. Some existing risks were identified that could have more serious effects with pediatric patients.
A risk benefit analysis demonstrated that, even with increased risk severity for some risks, the risks were inherent to IV administration sets/sterile products and there is a benefit to providing treatment/warming versus risks associated with potential hypothermia. No new risks were identified when using the Quantum with pediatric patients older than 28 days of age and of normal birth weight that were not identified with adults. The pediatric nurses and physicians have similar training to adult nurses and physicians pertaining to transfusions. Warming at the lower flow rates and volumes can be achieved with the Subject device. Therefore, the Quantum would be beneficial to pediatric patients when fluid warming is desired. Clinician users of the Quantum who treat pediatric patients would have the same credentials as current Quantum users, and by training be knowledgeable in appropriate fluid volume requirements and flow rates for pediatric patients.
#### Conclusions
The Quantum Blood and Fluid Warming System is indicated for use in both adult and pediatric patients greater than 28 days old of normal birthweight and is substantially equivalent to the predicate device, based on the performed risk analysis.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.