← Product Code [OWN](/productcode/OWN) · K192292

# L10 LED Light Source with AIM, L11 LED Light Source with AIM, AIM SafeLight Cable (K192292)

_Stryker · OWN · Nov 19, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192292

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Nov 19, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Upon intravenous administration of SPY AGENTTM GREEN (Indocyanine green for injection, USP), the AIM Light Source and SafeLightTM Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and near-infrared fluorescence imaging. The AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the AIM Light Source and SafeLight Cable is intended for use with standardof-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. Upon interstitial administration of SPY AGENT GREEN (ICG drug product), the AIM Light Source and SafeLightTM Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. The AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

## Device Story

AIM Light Source (L10/L11) and SafeLight Cable provide endoscopic illumination for minimally invasive/open surgery. System inputs: electrical power; outputs: visible white light and near-infrared (NIR) light (806nm for fluorescence, 830nm for transillumination). Operated by surgeons in OR. Light source uses RGB LEDs and NIR laser diode combined via dichroic mirrors; light coupled to endoscope via SafeLight cable. Fluorescence imaging requires intravenous/interstitial SPY AGENT GREEN (ICG). Output displayed on monitor for visual assessment of blood flow, tissue perfusion, biliary ducts, lymphatic system, or ureteral transillumination. Assists surgeons in real-time intraoperative decision-making; enhances visualization of anatomical structures/perfusion; potentially improves surgical outcomes.

## Clinical Evidence

Bench testing only. Performance validated via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), sterilization (ISO 14937), software validation (IEC 62304), and usability (IEC 62366-1). Comparative animal study performed to demonstrate equivalence to predicate.

## Technological Characteristics

Light source: RGB LEDs and NIR laser diode (806nm/830nm). Components: Light source, SafeLight cable, camera system, laparoscopes, IRIS Ureteral Kit. Connectivity: Fiberoptic coupling to endoscope. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, IEC 60825-1, ISO 14937. Software: IEC 62304 compliant.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- PINPOINT Endoscopic Fluorescence Imaging System ([K150956](/device/K150956.md), [K161792](/device/K161792.md), [K182606](/device/K182606.md))
- L10 LED Light Source and SafeLight Cable ([K142310](/device/K142310.md), [K151243](/device/K151243.md), [K173866](/device/K173866.md), [K191046](/device/K191046.md))
- L11 LED Light Source and SafeLight Cable ([K182160](/device/K182160.md), [K191046](/device/K191046.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Stryker April Malmborg Director, Regulatory Affairs 5900 Optical Court San Jose. California 95138

November 19, 2019

## Re: K192292

Trade/Device Name: L10 LED Light Source with AIM, L11 LED Light Source with AIM, AIM SafeLight Cable Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN, GCJ Dated: August 22, 2019 Received: August 23, 2019

#### Dear April Malmborg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192292

Device Name

Stryker L10 and L11 LED Light Source with AIM (Advanced Imaging Modality)

#### Indications for Use (Describe)

Upon intravenous administration of SPY AGENTTM GREEN (Indocyanine green for injection, USP), the AIM Light Source and SafeLightTM Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the AIM Light Source and SafeLight Cable is intended for use with standardof-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

Upon interstitial administration of SPY AGENT GREEN (ICG drug product), the AIM Light Source and SafeLightTM Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

The AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                        |                                                                                                   |                                              |                                                                                                  |                                                                                                  |                                             |                                             |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| <table border="0"><tr><td><table border="0"><tr><td></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table></td><td><table border="0"><tr><td></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table></td></tr></table> | <table border="0"><tr><td></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table> |                                              | Prescription Use (Part 21 CFR 801 Subpart D)                                                     | <table border="0"><tr><td></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> |                                             | Over-The-Counter Use (21 CFR 801 Subpart C) |
| <table border="0"><tr><td></td><td>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table>                                                                                                                                                      |                                                                                                   | Prescription Use (Part 21 CFR 801 Subpart D) | <table border="0"><tr><td></td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> |                                                                                                  | Over-The-Counter Use (21 CFR 801 Subpart C) |                                             |
|                                                                                                                                                                                                                                                        | Prescription Use (Part 21 CFR 801 Subpart D)                                                      |                                              |                                                                                                  |                                                                                                  |                                             |                                             |
|                                                                                                                                                                                                                                                        | Over-The-Counter Use (21 CFR 801 Subpart C)                                                       |                                              |                                                                                                  |                                                                                                  |                                             |                                             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K192292

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).

## Submitter:

| Applicant:      | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138                                                                                  |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg, RAC<br>Director, Regulatory Affairs<br>Phone: (408) 754-2473<br>Facsimile: (408) 754-2598<br>Email: april.malmborg@stryker.com |
| Date Prepared:  | August 22, 2019                                                                                                                                |

## Subject Device:

| Name of Device:         | AIM Light Source and SafeLight™ Cable                                                                                                                                    |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or<br>Usual Name | Light Source, Illuminator                                                                                                                                                |
| Classification<br>Name: | Confocal Optical Imaging1 (21 C.F.R. §876.1500)<br>Fiberoptic light ureteral catheter2 (21 C.F.R. §876.4020)<br>Light Source, Fiberoptic, Routine3 (21 C.F.R. §876.4020) |
| Regulatory Class:       | II                                                                                                                                                                       |
| Product Code:           | OWN1<br>FSC2<br>FCW3                                                                                                                                                     |
| 510(k) Review<br>Panel: | General & Plastic Surgery1<br>Gastroenterology/ Urology2,3                                                                                                               |

1When used for assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), and lymphatic system, including lymphatic vessels and lymph nodes, using near-infrared imaging

2When used to transilluminate the ureter during open or laparoscopic surgical procedures

3When used to provide standard endoscopic visible light to support real-time endoscopic visible imaging.

# Predicate Device(s):

| PINPOINT Endoscopic Fluorescence Imaging System (primary predicate) | K150956, K161792,<br>K182606          |
|---------------------------------------------------------------------|---------------------------------------|
| L10 LED Light Source and SafeLight Cable (secondary predicate)      | K142310, K151243,<br>K173866, K191046 |
| L11 LED Light Source and SafeLight Cable (secondary predicate)      | K182160, K191046                      |

NOTE: The predicate devices have not been subject to a design-related recall.

{4}------------------------------------------------

# K192292

# Device Description:

The AIM Light Source (L10 and L11 LED Light Source) and SafeLight™ Cable are part of the AIM (Advanced Imaging Modality) System. The AIM System is an endoscopic real-time visible white light and near-infrared light illumination and imaging system. Near-infrared illumination is used for both fluorescence imaging using indocyanine green (ICG) and transillumination of the ureters using the IRIS Ureteral Kit during minimally invasive and open surgical procedures, respectively. The AIM Light Source is a light-generating unit designed to illuminate surgical sites in the following applications: visible light, near-infrared fluorescence, and near-infrared transillumination. The SafeLight Cable transmits light from the light source to an endoscope during endoscopic procedures.

# Indications for Use:

Upon intravenous administration of SPY AGENT™ GREEN (Indocyanine green for injection, USP), the AIM Light Source and SafeLight™ Cable is used with SPY AGENT™ GREEN to provide real-time endoscopic visible and near-infrared fluorescence imaging. The AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

Upon interstitial administration of SPY AGENT™ GREEN (ICG drug product), the AIM Light Source and SafeLight™ Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system. including lymphatic vessels and lymph nodes.

The AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

| Item                       | Subject Device                                                                                                                                                                                                                                                                                                               | Predicate Devices                                                                                          |                                                                   |                        |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------|
|                            | AIM Light Source and Safe<br>Light Cable                                                                                                                                                                                                                                                                                     | PINPONT Endoscopic<br>Fluorescence Imaging<br>System (primary<br>predicate)                                | AIM Light Source and<br>Safe Light Cable<br>(secondary predicate) |                        |
| Manufacturer               | Stryker                                                                                                                                                                                                                                                                                                                      | Novadaq Technologies<br>ULC (now a part of<br>Stryker)                                                     | Same as subject device                                            |                        |
| Submission Reference       | Current submission                                                                                                                                                                                                                                                                                                           | K150956, K161792,<br>K182606                                                                               | K142310, K151243,<br>K173866, K182160,<br>K191046                 |                        |
| Item                       | Subject Device                                                                                                                                                                                                                                                                                                               | Predicate Devices                                                                                          |                                                                   |                        |
|                            | AIM Light Source and Safe<br>Light Cable                                                                                                                                                                                                                                                                                     | PINPONT Endoscopic<br>Fluorescence Imaging<br>System (primary<br>predicate)                                | AIM Light Source and<br>Safe Light Cable<br>(secondary predicate) |                        |
| Intended Use               | Endoscopic white light and<br>near-infrared illumination andimaging during endoscopic<br>procedures.                                                                                                                                                                                                                         | Same as subject device.                                                                                    | Same as subject device                                            |                        |
| Indications for Use        | NOTE 1                                                                                                                                                                                                                                                                                                                       | Same as subject device.<br>NOTE 2                                                                          | NOTE 3                                                            |                        |
| Principles of Operation    | An electronic driver controls<br>Red/Green/Blue LEDs and a<br>near-infrared laser diode which<br>are combined through dichroic<br>mirrors and projected onto an<br>output light collimator. A fiber<br>output bundle can be inserted<br>into the light source to couple<br>light to the distal end and into<br>an endoscope. | Same as subject device.                                                                                    | Same as subject device.                                           |                        |
| Components                 | Light Source and SafeLight<br>Cable<br>Camera System<br>Laparoscopes<br>IRIS Ureteral Kit<br>Imaging Agent                                                                                                                                                                                                                   | Video Processor/<br>Illuminator (VPI)<br>Light Guide Cable<br>Camera Head<br>Laparoscopes<br>Imaging Agent | Same as subject device                                            |                        |
| Safety Standards           | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60825-1                                                                                                                                                                                                                                                                | Same as subject device                                                                                     | Same as subject device                                            |                        |
| Light Source/ Laser        | RGB LEDs<br>Infrared Laser                                                                                                                                                                                                                                                                                                   | Same as subject device                                                                                     | Same as subject device                                            |                        |
| Infrared Wavelengths       | 806nm (used for NIR<br>fluorescence)<br>830nm (used for NIR<br>transillumination)                                                                                                                                                                                                                                            | 805nm (used for NIR<br>fluorescence)                                                                       | Same as subject<br>device                                         |                        |
| L10<br>Imaging<br>Modes    | White Light                                                                                                                                                                                                                                                                                                                  | Manual                                                                                                     | Same as subject device                                            | Same as subject device |
|                            | NIR<br>Fluorescence                                                                                                                                                                                                                                                                                                          | ENV                                                                                                        | Contrast<br>Overlay<br>Color Segmented<br>Fluorescence            | Same as subject device |
|                            | NIR Trans-<br>illumination                                                                                                                                                                                                                                                                                                   | IRIS                                                                                                       | None                                                              | Same as subject device |
| L11<br>Imaging<br>Modes    | White Light                                                                                                                                                                                                                                                                                                                  | Manual<br>Autolight                                                                                        | Manual                                                            | Same as subject device |
|                            | NIR<br>Fluorescence                                                                                                                                                                                                                                                                                                          | ENV<br>Contrast<br>Overlay                                                                                 | Contrast<br>Overlay<br>Color Segmented<br>Fluorescence            | Same as subject device |
| Item                       | Subject Device                                                                                                                                                                                                                                                                                                               | Predicate Devices                                                                                          |                                                                   |                        |
|                            | AIM Light Source and Safe<br>Light Cable                                                                                                                                                                                                                                                                                     | PINPONT Endoscopic<br>Fluorescence Imaging<br>System (primary<br>predicate)                                | AIM Light Source and<br>Safe Light Cable<br>(secondary predicate) |                        |
| NIR Trans-<br>illumination | IRIS                                                                                                                                                                                                                                                                                                                         | None                                                                                                       | Same as subject device                                            |                        |
| SafeLight<br>Cable         | Single-Use/<br>Reusable                                                                                                                                                                                                                                                                                                      | Reusable                                                                                                   | Same as subject device                                            | Same as subject device |

# Comparison of Technological Characteristics with the Predicate Device:

{5}------------------------------------------------

{6}------------------------------------------------

NOTE 1: Upon intravenous administration of SPY AGENT™ GREEN (Indocyanine green for injection, USP), the AIM Light Source and SafeLight™ Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and near-infrared fluorescence imaging. The AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for biliary duct visualization. Upon interstital administration of SPY AGENT GREEN (ICG drug product), the AIM Light Source and SafeLight Cable is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. The AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

NOTE 2: Upon intravenous administration of SPY AGENT™ GREEN (Indocyanine green for injection, USP), the PINPOINT Endoscopic Fluorescence Imaging System is used with SPY AGENT GREEN to perform intraoperative fluorescence angiography, and it is also indicated for use in fluorescence imaging of biliary ducts, and when intraoperative cholangiography. The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct or common hepatic duct), using near-infraging. Fluorescence imaging of biliary ducts with the PINPOINT System in intended for use with standard of care white light, and when indicated intraoperative cholangiography. The device is not intended for biliary duct visualization. Upon interstital administration of SPY AGENT GREEN (ICG drug product), the PINPOINT System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

NOTE 3: Upon intravenous administration of SPY AGENT™ GREEN (Indocyanine green for injection, USP), the AIM Light Source and SafeLight™ Cable is used with SPY AGENT GREEN to provide real-time endoscopic visible and near-infrared fluorescence imaging. The AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperaphy. The devices are not intended for standalone use for biliary duct visualization. The AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

#### Performance Data:

| Test              | Method                                                                        | Result |
|-------------------|-------------------------------------------------------------------------------|--------|
| Electrical Safety | ANSI IEC 60601-1:2005 + A1:2012;<br>IEC 60601-2-18:2009<br>IEC 60601-1-6:2013 | PASS   |
| EMC Testing       | IEC 60601-1-2:2014                                                            | PASS   |
| Laser Safety      | IEC 60825-1:2014                                                              | PASS   |
| Sterilization     | ISO 14937:2009                                                                | PASS   |

Testing was completed in accordance with the following:

{7}------------------------------------------------

| Test                                  | Method                                                                                                                          | Result              |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Software Validation &<br>Verification | IEC 62304:2006                                                                                                                  | PASS                |
| Usability                             | IEC 62366-1:2015                                                                                                                | PASS                |
| Performance - Bench                   | In accordance with device input specifications                                                                                  | PASS                |
| Performance - Animal                  | In accordance with device user needs;<br>Comparative testing to currently legally<br>marketed device in accordance with 21CFR58 | PASS;<br>Equivalent |

NOTE: The AIM Light Source and SafeLight Cable do not require clinical studies to support the determination of substantial equivalence.

# Conclusions:

Based on the information presented, the AIM Light Source and SafeLight Cable is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. The nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legal marketed predicate device, the PINPOINT System. There are no new issues of safety and/or effectiveness introduced by the AIM Light Source and SafeLight Cable for intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes using SPY AGENT™ GREEN the when used as instructed.

---

**Source:** [https://fda.innolitics.com/device/K192292](https://fda.innolitics.com/device/K192292)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
