← Product Code [DZJ](/productcode/DZJ) · K192282

# MedCAD® AccuPlan® System (K192282)

_Medcad · DZJ · Oct 29, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192282

## Device Facts

- **Applicant:** Medcad
- **Product Code:** [DZJ](/productcode/DZJ.md)
- **Decision Date:** Oct 29, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4120
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Software as a Medical Device

## Indications for Use

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces including anatomical models, surgical guides, and dental splints for use in maxillofacial guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a preoperative software tool for simulating / evaluating surgical treatment options.

## Device Story

System processes 3D medical images (CT, CBCT) and dental models to create patient-specific anatomical models, surgical guides, and dental splints; used for orthognathic and reconstructive maxillofacial surgery. Operated by trained MedCAD employees using COTS software; physician provides input for model manipulation and reviews digital outputs prior to finalization. System supports preoperative surgical simulation and treatment planning. Physical outputs produced via additive manufacturing; intended for use as positioning and marking aids. Benefits include improved surgical planning and patient-specific anatomical accuracy.

## Clinical Evidence

Bench testing only. Validation included IQ, OQ, and PQ for manufacturing processes. Dimensional analysis confirmed fit. Mechanical performance evaluations (dynamic compressive strength, ligature wire pullout) demonstrated equivalence to predicate. Software verification and validation confirmed subsystem and integrated system requirements. Biocompatibility testing conducted per ISO 10993-1. Sterilization validation performed per ISO 17665 to SAL 1x10^-6.

## Technological Characteristics

Uses additive manufacturing to produce patient-specific devices from biocompatible polymers. Inputs include CT/CBCT scans and dental models. Outputs include anatomical models, surgical guides, and dental splints. System is non-sterile, intended for end-user steam sterilization. Software utilizes COTS packages for 3D image manipulation and segmentation. Complies with ISO 10993-1 (biocompatibility) and ISO 17665 (sterilization).

## Regulatory Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

## Predicate Devices

- VSP® System, Medical Modeling, Inc. ([K120956](/device/K120956.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. The FDA logo is a recognizable symbol of the agency's role in protecting public health.

October 29, 2020

MedCAD % Linda Braddon, Ph.D. President/CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock, Georgia 30188

Re: K192282

Trade/Device Name: MedCAD® AccuPlan® System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: DZJ, LLZ Dated: October 2, 2020 Received: October 2, 2020

Dear Linda Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K192282 Device Name

MedCAD® AccuPlan® System

Indications for Use (Describe)

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces including anatomical models, surgical guides, and dental splints for use in maxillofacial guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a preoperative software tool for simulating / evaluating surgical treatment options.

Type of Use (Select one or bothas applicable)

図 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 |
|----------------------|-------------|----------------------------------------|
|----------------------|-------------|----------------------------------------|

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MedCAD. The text "Med" is in a large, bold, serif font. The text "CAD" is in a smaller, sans-serif font and is set against a blue-green background. The background is a parallelogram shape.

# 510(k) Summary MedCAD® AccuPlan® Svstem K192282

# October 29, 2020

### Sponsor

MedCAD 501 S 2nd Ave, Suite A-1000 Dallas, TX 75226 (214) 453-8864 x305

# Contact

Secure BioMed Evaluations Linda Braddon, Ph.D. 7828 Hickory Flat Highway Suite 120 Woodstock, GA 30188 770-837-2681 Requlatory@SecureBME.com

#### Name of Device and Classification Name

Device Name: MedCAD® AccuPlan® System Regulation Name: Bone Cutting Instruments and Accessories Requlation Number: 872.4120 Product Code: Primary - DZJ; Secondary - LLZ Classification Panel: Dental

#### Predicate Device

VSP® System, Medical Modeling, Inc. (K120956)

#### Indication for Use:

The MedCAD® AccuPlan® System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the MedCAD® AccuPlan® System and the result is an output data file that may then be provided as digital models or used as input to a rapid prototyping portion of the system that produces physical outputs including anatomical models, surgical guides and dental splints for use in maxillofacial surgery. The surgical guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery. The MedCAD® AccuPlan® System is also intended as a pre-operative software tool for simulating /evaluating surgical treatment options.

#### Device Description

The MedCAD® AccuPlan® System is a collection of software and associated additive manufacturing equipment intended to provide a variety of outputs to support orthognathic or reconstructive surgery. The system uses electronic medical images of the patient's anatomy or stone castings made from the patient anatomy with input from the physician, to manipulate

{4}------------------------------------------------

# Med C

original patient images for planning and executing surgery. The patient specific outputs from the system includes anatomical models, surgical quides, dental splints, and patient-specific case reports.

Following the MedCAD® Quality System and specific Work Instructions, trained employees utilize Commercial Off-The-Shelf (COTS) software to manipulate 3-D medical scan images which can include Computed Tomography (CT), Cone Beam CT (CBCT), and/or 3-D scan images from patient physical models of the patient's teeth) to create patientspecific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. While the process and dataflow vary somewhat based on the requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how thev interact to produce the defined system outputs. It should be noted that the system is operated only by trained MedCAD employees, and the physician does not directly input information. The physician provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.

The AccuPlan® System is made up of 4 individual pieces of software for the design and various manufacturing equipment integrated to provide a range of anatomical models (physical and digital), dental splints, surgical guides, and patient-specific planning reports for reconstructive surgery in the maxillofacial region.

The AccuPlan® System requires an input 3-D image file from medical imaging systems (i.e. -CT) and/or implant file. This input is then used, with support from the prescribing physician to provide the following potential outputs to support reconstructive surgery. Each system output is designed with physician input and reviewed by the physician prior to finalization. All outputs are used only with direct physician involvement to reduce the criticality of the outputs.

System outputs include:

- Anatomical Models .
- Surgical Guides .
- . Dental Splints
- Patient-Specific Case Reports ●

# Performance Data

The performance data indicates that the verification and validation testing performed on the MedCAD® AccuPlan® system successfully demonstrates that design outputs meet design inputs.

# Device Performance Validation

The performance testing for the device includes processes validation methods such as IQ, OQ. and PQ to ensure that the manufacturing process can effectively produce patient matched devices. Equipment used for production purposes have been qualified to ensure the equipment used for manufacturing of surgical guides, dental splints, and anatomical models meet production needs. Dental splints and surgical quides are manufactured based on

{5}------------------------------------------------

recommendations outlined in the FDA Guidance Document "Technical Considerations for Additive Manufactured Medical Devices."

Additionally, a dimensional analysis was performed to ensure the proper fit of the final output. Mechanical performance evaluations assessing dynamic compressive strength and ligature wire pullout testing were conducted on final, finished devices demonstrating they are equivalent to the predicate device.

### Software System Validation

In order to ensure the off-the-shelf software packages are operating correctly and any necessary file conversions will not neqatively impact the final output, software verification and validation activities include:

- · Verification of each independent software subsystem against defined requirements
- Verification of compatibility between software subsystems aqainst defined . requirements
- Validation of fully integrated system including all subsystems against overall . system requirements

# Sterilization Validation

Sterilization validation was conducted in accordance with international standard ISO 17665 and FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling." to a sterility Assurance Level (SAL) of 1x10-9. All test method acceptance criteria were met.

#### Biocompatibility Validation

Biocompatibility validation was conducted in accordance with international standard ISO 10993-1 and FDA quidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The results of the testing adequately address biocompatibility for the output devices and their intended use.

#### Substantial Equivalence

MedCAD® AccuPlan® system is substantially equivalent to the identified predicate based on indications for use, principles of operation, technological characteristics, inputs, Minor differences in the surgical planning and manufacturing processes are verified and validated in the performance data in accordance with the intended use.

The input is images from medical scanners, dental models, and/or implants files. Physical outputs include dental splints and surgical guides. Biocompatible materials are used in the creation of the subject devices.

All devices are intended to aid in maxillofacial surgeries. These systems are intended to be utilized by trained employees with the approval by the physician.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MedCAD. The text "Med" is in a large, bold, black font. The text "CAD" is in a white, bold font and is set against a teal background. There is a registered trademark symbol in the upper right corner of the teal background.

| COMPARISON OF TECHNOLOGICAL CHARACTERISITICS             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                          | MedCAD® AccuPlan® System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | VSP® System Medical<br>Modeling Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 510(k) number                                            | K182282                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K120956                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use                                      | The MedCAD® AccuPlan® System is<br>intended for use as a software<br>system and image segmentation<br>system for the transfer of imaging<br>information from a medical scanner<br>such as a CT based system. The<br>input data file is processed by the<br>MedCAD® AccuPlan® System and<br>the result is an output data file that<br>may then be provided as digital<br>models or used as input to a rapid<br>prototyping portion of the system that<br>produces physical outputs including<br>anatomical models, surgical guides,<br>and dental splints for use in<br>maxillofacial surgery. The surgical<br>guides and dental splints are<br>intended to be used for the<br>maxillofacial bone in maxillofacial<br>surgery. The MedCAD® AccuPlan®<br>System is also intended as a pre-<br>operative software tool for simulating<br>/ evaluating surgical treatment<br>options. | The Medical Modeling VSP®<br>System is intended for use as a<br>software system and image<br>segmentation system for the<br>transfer of imaging information<br>from a medical scanner such as a<br>CT based system. The input data<br>file is processed by the VSP®<br>System and the result is an output<br>data file that may then be provided<br>as digital models or used as input<br>to a rapid prototyping portion of the<br>system that produces physical<br>outputs including anatomical<br>models, templates, and surgical<br>guides for use in maxillofacial<br>surgery. The VSP® System is also<br>intended as a pre-operative<br>software tool for simulating /<br>evaluating surgical treatment<br>options. |
| Preoperative software                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Additive manufacturing of<br>splints, guides, and models | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Data inputs                                              | Images from medical scanners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Images from medical scanners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Data outputs                                             | Output for dental splints, surgical<br>guides, and anatomical models                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Output for anatomical models,<br>templates, and surgical guides                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Physical outputs                                         | Dental splints, surgical guide,<br>anatomical models, and patient-specific<br>case reports                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Templates, surgical guides, and<br>anatomical models                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Materials                                                | Biocompatible polymers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Biocompatible polymers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Sterilization                                            | Subject device is provided non-sterile<br>and is steam sterilized by the end-user                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate device is provided non-<br>sterile and is steam sterilized by<br>the end-user                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Manufacturing Method                                     | Additive Manufacturing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Additive Manufacturing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Patient Contact                                          | Guides: Surface Contacting - Tissue<br>/ Bone / Mucosal Membrane Limited<br>(< 24 hours)<br><br>Splints: Surface Contacting -<br>Mucosal Membrane Limited (< 24<br>hours)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Limited (< 24 hours)<br>Prolonged (< 30 days)<br>Surface Contacting - Mucosal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

# Comparison of Technological Characteristics with the Predicate Device

MedCAD® AccuPlan® system is substantially equivalent to and it is safe and effective as its predicate device (VSP® System, Medical Modeling, Inc. – K120956).

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MedCAD. The word "Med" is in a large, bold, black serif font. The word "CAD" is in a smaller, white sans-serif font inside of a teal box. There is a small registered trademark symbol to the right of the teal box.

# Similarities to Predicate

The MedCAD® AccuPlan® System has the same intended use and similar technological characteristics as the identified predicate device. The system employs similar fundamental technologies as the identified predicates including: Software image transfer, manipulation, and surgical planning. The principals of operation and technological characteristics are either identical or substantially equivalent to the predicate. The system has similar technological characteristics including:

- . System inputs: Images from medical scanners (ex: CT), dental models, and/or implant files (.STL)
- . System outputs: Physical and/ or Digital outputs such as patient-specific anatomical models, guides, and splints
- Materials: Biocompatible polymers
- . Sterility assurance level of 1x10-6

The intended use of the subject device and the predicate both provide tools and accessories (software for image manipulation, anatomical models, splints and quides) for use in reconstructive surgery. Additionally, both the subject and the predicate device are intended to be used by trained personnel with active support from the surgeon.

#### Differences to Predicate

The following differences exist between the subject and predicate device:

- The predicate specifically refers to support of reconstructive surgery. The subject ● device will support both orthognathic (correction of jaw conditions) and reconstructive surgeries.
- . The predicate device contains a set of stainless-steel cutting and drill inserts which are intended to be used by the physician to quide cutting and drilling activities during the surgical procedure. The subject device is intended to be used only for positioning and marking as determined by the physician. There is no use of stainless-steel cutting and drill inserts with the subject device.
- . The predicate system includes the use of surgical guides for graft bone harvesting which can include the fibula and hip. The subject device is not intended to be used for bone harvesting and is to be used only in the maxillofacial region of the body.
- . The Indications for Use statements differ slightly in that the predicate device lists the physical outputs to include "anatomical models, templates, and surgical quides" whereas the subject device includes "anatomical models, surgical quides, and dental splints". Additionally, the subject device contains the statement "The surgical guides and dental splints are intended to be used for the maxillofacial bone in maxillofacial surgery."

# Conclusion

MedCAD® AccuPlan® is substantially equivalent to and is as safe and effective as its predicate device. Both devices incorporate similar inputs, operations, and outputs.

---

**Source:** [https://fda.innolitics.com/device/K192282](https://fda.innolitics.com/device/K192282)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
