K192211 · Soundbite Medical Solution, Inc. · PDU · Jan 17, 2020 · Cardiovascular
Device Facts
Record ID
K192211
Device Name
SoundBite Console, SoundBite Active Wire 18
Applicant
Soundbite Medical Solution, Inc.
Product Code
PDU · Cardiovascular
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SoundBite™ Crossing System - Peripheral is indicated to facilitate the intraluminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions. The SoundBite™ Crossing System – Peripheral is not intended for use in the carotid arteries.
Device Story
System consists of reusable SoundBite Console and single-use sterile SoundBite Active Wire 18. Console generates controlled mechanical pulses (shock waves) transmitted to wire; causes distal tip to accelerate axially in reciprocating motion (micro-jackhammer effect). Used in peripheral vasculature to cross chronic total occlusions (CTO). Operated by physicians in clinical settings. Provides mechanical recanalization to allow subsequent placement of conventional guidewires or treatment devices. Benefits patient by facilitating crossing of occluded vessels, potentially avoiding more invasive procedures.
Clinical Evidence
Single-arm clinical study of 52 patients with symptomatic infrainguinal CTOs. Primary endpoint: technical success (ability to allow crossing/treatment of target lesions) achieved in 92.3% (48/52). Secondary endpoints: 88.5% post-procedural patency; 98.1% crossed ≥0.5 cm and ≥1.0 cm of CTO; 59% full traversal without additional devices. No device-related adverse events reported.
Technological Characteristics
System includes AC-powered console and Titanium alloy Active Wire 18 (0.46 mm OD, 300 cm length). Principle: axial reciprocating mechanical pulses. No irrigation required. Sterilization: Ethylene Oxide. Software verified per IEC 62304. Electrical safety per ANSI AAMI ES60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients with peripheral artery chronic total occlusions (CTO) to facilitate intraluminal placement of guidewires or treatment devices. Not for use in carotid arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2020
SoundBite Medical Solution, Inc. Mr. Marc-Andre Cote Director. Regulatory Affairs 2300 Blvd Alfred Nobel Montreal, QC Canada H4S 2A4
Re: K192211
Trade/Device Name: SoundBite Crossing System - Peripheral Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: December 13, 2019 Received: December 16, 2019
Dear Mr. Cote:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192211
#### Device Name
SoundBite Crossing System - Peripheral
#### Indications for Use (Describe)
SoundBite™ Crossing System - Peripheral is indicated to facilitate the intra-luminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions. The SoundBite™ Crossing System -Peripheral is not intended for use in the carotid arteries.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K192211 p. 1 of 4
# 510(K) Summary
As required by 21 CFR 807.92
| Date Prepared: | 17 January 2020 | |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitted by: | SoundBite Medical Solutions, Inc.<br>2300 Alfred Nobel, Suite 230<br>Montreal, Quebec<br>Canada H4S 2A4 | |
| Contact: | Marc-André Côté<br>Director, Regulatory Affairs<br>Tel: (514) 956-2525 x 3352<br>Fax: (514) 956-2529<br>E-mail : marc-andre.cote@soundbitemedical.com | |
| Trade/Device Name: | SoundBite™ Crossing System ">– Peripheral | |
| Common Name: | System for Crossing Total Occlusions | |
| Regulation: | 21 CFR 870.1250, Percutaneous Catheter | |
| Device Class: | Class II | |
| Product Code: | PDU (Catheter for Crossing Total Occlusions) | |
| Predicate Device: | The CROSSER® S6 System by FlowCardia, Inc. (K092175) | |
| Reference Devices: | CROSSER® 14S RX Catheter – The CROSSER® System (K072776)<br>TruePath™ CTO Device by Boston Scientific Corp. (K140288)<br>V-18™ ControlWire™ Guidewire by Boston Scientific Corp. (K033742) | |
### Device Description:
The SoundBite™ Crossing System - Peripheral is a recanalization tool, designed to help physicians place conventional guidewires or treatment devices in the intraluminal space beyond chronic total occlusions in the peripheral vasculature. The SoundBite™ Crossing System - Peripheral consists of the reusable mobile SoundBite™ Console, the single-use sterile SoundBite™ Active Wire 18, and their accessories.
The SoundBite™ Console generates controlled mechanical pulses (i.e., shock waves) which are transmitted to the SoundBite™ Active Wire 18 and cause the distal tip of the wire to accelerate axially in a reciprocating motion, acting like a micro-jackhammer.
The SoundBite™ Active Wire 18 is similar in construction to other commercially available CTO crossing wires, with friction reducing PTFE coating (except for the distal tip), a radiopaque marker near the tip, and enhanced flexibility at the distal end. It has an outer diameter of 0.46 mm (0.018") and it is 300 cm long, with a working length of 150 cm; the proximal end flares up to a larger diameter for connection to the console. The single-use SoundBite™ Active Wire 18 is supplied sterile with a shelf life of 24 months.
{4}------------------------------------------------
### Intended Use / Indications for Use:
The SoundBite™ Crossing System - Peripheral is indicated to facilitate the intraluminal placement of conventional guidewires or treatment devices beyond peripheral artery chronic total occlusions.
The SoundBite™ Crossing System – Peripheral is not intended for use in the carotid arteries.
### Substantial Equivalence Comparison:
The SoundBite™ Crossing System - Peripheral is substantially equivalent to the predicate device - The CROSSER® S6 System (K092175) - in intended use and indications for use, fundamental technologies, principles of operation, and labeling. Non-clinical bench testing data, and performance data from animal and clinical studies have been submitted to demonstrate that the differences in technological characteristics do not raise different questions of safety and effectiveness. A side-by-side comparison of key device characteristics is presented in the following table:
| Device<br>Characteristic | The CROSSER® S6 System<br>(K092175) | SoundBite™ Crossing System – Peripheral |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System<br>components | - AC-powered, mobile CROSSER®<br>Generator with footswitch;<br>- High-frequency transducer;<br>- Single-use CROSSER® S6 Catheter;<br>- FLOWMATE® Injector (optional) | - AC-powered, mobile SoundBite™ Console<br>generator and footswitch;<br>- Single-use SoundBite™ Active Wire 18 |
| Catheter<br>Irrigation | Irrigation port and lumen on catheter.<br>Requires refrigerated saline to control<br>heat generation. | Not required |
| Mechanism of<br>action | High-frequency mechanical vibrations<br>are propagated through a Nitinol core<br>wire to the Stainless Steel tip of the<br>CROSSER® S6 Catheter. | Mechanical pulses sent along the length of the<br>Titanium alloy SoundBite™ Active Wire 18<br>cause the distal tip to accelerate axially in a<br>reciprocating (back-and-forth) motion, acting<br>like a micro-jackhammer. |
| Guidewire<br>Compatibility | No guidewire required | No guidewire required |
| Catheter/Wire<br>Connection | Connection hub on catheter with<br>transducer coupling and irrigation port | Titanium alloy section reducer for connection<br>to the SoundBite™ Console |
| Working<br>Length | 154 cm or 106 cm | 150 cm |
| Radiopaque<br>Marker | Stainless Steel catheter tip | 10 mm radiopaque coil starting at 1.5 mm<br>from the distal tip |
| Wire/Catheter<br>Shaft Diameter | 1.3 mm (0.051") | 0.46 mm (0.018") |
{5}------------------------------------------------
| Device<br>Characteristic | The CROSSER® S6 System<br>(K092175) | SoundBiteTM Crossing System - Peripheral |
|-------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Wire/Catheter<br>Tip Diameter | 0.6 mm (0.025") | 0.29 mm (0.0115") |
| Wire/Catheter<br>Material | Core: Nitinol<br>Main body: Pebax<br>Tip: Stainless Steel | Titanium alloy |
| Wire/Catheter<br>Coating | Main body and catheter tip:<br>Hydrophilic coating<br>(material not known) | PTFE coating on main body of the wire only;<br>no coating on distal end |
| Wire/Catheter<br>Sold Sterile | Yes | Yes |
| Sterilization<br>Method | Gamma Irradiation | Ethylene Oxide |
| Shelf Life | Not known | 24 months |
# Summary of Non-Clinical Testing:
Design verification and validation testing was conducted following systematic risk assessment in accordance with the FDA-recognized consensus standard ISO 14971 and FDA guidance.
Basic safety and performance of the device, both at the system and component levels (i.e., sterility assurance to the point of care, packaging integrity and shelf-life, biocompatibility, electrical/mechanical safety and electromagnetic compatibility) have been verified and/or validated in accordance with current FDArecognized consensus standards and regulatory requirements. The medical device software system has been designed, developed and verified in compliance with the FDA-recognized consensus standard IEC 62304.
For device characteristics where objective performance standards and acceptance criteria have not been established, and where direct comparison with the predicate device was not possible, the SoundBite™ Crossing System – Peripheral was compared to commercially available reference devices having substantially similar technological, physiological and anatomical site of use characteristics. The overall acute and chronic safety profile of the proposed system were compared to a reference device in an animal study.
SoundBite™ Crossing System - Peripheral was assessed in the following areas:
- . Visual and dimensional inspection
- Simulated use
- Coating integrity
- . Tip pull
- Flexibility resistance ●
- Fracture resistance
- . Torque strength
- Tensile strength ●
- Corrosion resistance ●
- . Torqueability
- Distal temperature
{6}------------------------------------------------
- Tip flexibility ●
- Catheter qualification ●
- Lubricity assessment ●
- Particulate testing
- Shelf life testing ●
SoundBite™ Console was assessed in the following areas:
- Console Output Stability .
- . Life-Cycle Testing
- . Console and Shipping Container Labels Verification
- Electronics and Software Verification
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 .
- IEC 60601-1-2 Edition 4.0 2014-02
Biocompatibility: A full panel of biocompatibility tests was successfully performed in accordance with product classification, under GLP, demonstrating that all utilized materials and methods of construction/processing passed biocompatibility rigors.
Usability: Representative users were included in a summative evaluation. The study confirms that the SoundBite™ Crossing System - Peripheral can be used without serious use errors or problems, for the intended uses under the expected use conditions.
The results from bench and animal testing indicate that the performance characteristics of the SoundBite™ Crossing System - Peripheral are comparable to the reference devices and do not raise different questions of safety and effectiveness.
# Summary of Clinical Testing:
The performance of the SoundBite™ Crossing System – Peripheral was also evaluated under actual clinical conditions in a single-arm clinical study involving 52 patients with documented symptomatic infrainguinal chronic total occlusions (CTO). All CTOs were confirmed angiographically to be 100% occluded.
The SoundBite™ Crossing System - Peripheral has met the primary performance endpoint of technical success in 92.3% of cases (48 of 52 study subjects). Technical success was defined as the abilitate treatment of the target lesions by allowing additional crossing and/or treatment devices to cross the CTO.
In assessing the secondary endpoints, 88.5% (46/52) of subjects had post-procedural patency following successful crossing of the CTO. In 98.1% (51/52) of cases, ≥0.5 cm of any segment of the CTO was crossed. Similarly, in 98.1% (51/52) of cases, ≥1.0 cm of any segment of the CTO was crossed. Furthermore, in 59% (31/52) of cases, the system fully traversed the CTO with entry into the distal lumen without the need for additional guidewires or re-entry devices.
There were no adverse events attributed to the device, as per independent physician adjudication. The study results demonstrate the favorable clinical safety and performance profile of the proposed device.
# Conclusion
The data submitted with this 510(k) premarket notification demonstrate that the SoundBite™ Crossing System - Peripheral is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.