The NeuroVention Cranial Fixation System is intended for use as a burr hole cover and/or skull bone fixation following craniotomy, cranioplasty, or craniectomy surgery.
Device Story
NeuroVention Cranial Fixation System comprises titanium, titanium alloy, and PEEK plates, burr hole covers, and screws used for cranial bone fixation. System includes manual surgical instruments (screwdriver, drivers, forceps, drill bit) for implantation. Provided non-sterile; requires steam sterilization by user facility. Used in OR by neurosurgeons to stabilize bone flaps or cover burr holes post-craniotomy/craniectomy. Surgeon selects appropriate implant configuration based on patient anatomy and surgical needs. Device provides rigid fixation; mechanical performance verified via static compression, skull conformity, and screw pushout/torque testing. MR conditional status established for 1.5T and 3.0T environments.
Clinical Evidence
Bench testing only. No clinical data was required. Mechanical testing included static compression, skull conformity, screw axial pushout, and screw torque to failure (per ASTM F543-13). Usability testing was performed for the drill bit. All testing met or exceeded established protocols and acceptance criteria.
Indicated for burr hole coverage and skull bone fixation in patients undergoing craniotomy, cranioplasty, or craniectomy surgery.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
{0}------------------------------------------------
March 13, 2020
NeuroVention, LLC Rohit Khanna President 645 S. Beach St. Daytona Beach, Florida 32114
### Re: K192162
Trade/Device Name: NeuroVention Cranial Fixation System. NuCrani Plates, KTC Burr Hole Cover, DC Plate Regulation Number: 21 CFR 882.5250 Regulation Name: Burr Hole Cover Regulatory Class: Class II Product Code: GXR, GWO, HBW Dated: February 11, 2020 Received: February 12, 2020
Dear Rohit Khanna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192162
Device Name
NeuroVention Cranial Fixation System
#### Indications for Use (Describe)
The NeuroVention Cranial Fixation System is intended for use as a burr hole cover and/or skull bone fixation following craniotomy, cranioplasty, or craniectomy surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
(as required by 21 CFR 807.92)
| Submitter | NeuroVention LLC |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 645 S Beach Street<br>Daytona Beach, FL 32114 |
| Telephone | 386-238-9638 |
| Fax | 386-253-8174 |
| Contact Person | Rohit Khanna, M.D. |
| email | khanna@neurovention.com |
| Date Prepared | January 27, 2020 |
| Trade Name | Neurovention Cranial Fixation System<br>Includes: NuCrani Plates, DC Plate, and KTC burr hole cove |
| Common Name | Neuro plating system |
| Panel Code | Neurological Devices |
| Classification Name | Burr Hole Cover (GXR)<br>Preformed Alterable Cranioplasty Plate (GWO)<br>Cranioplasty Plate Fastener (HBW)<br>Drills, Burrs, Trephines & Accessories- Manual (HBG) |
| Class | Class II |
| Regulation Number | 21 CFR 888.5250 21 CFR 888.5320<br>21 CFR 888.5360<br>21 CFR 882.4300 |
| Product Code | GXR, GWO, HBW, HBG |
| Predicate Device Name | 510(k) Number | Manufacturer |
|----------------------------------|---------------|--------------|
| Stryker Universal Neuro 3 System | K112557 | Stryker |
{4}------------------------------------------------
### Description
The NeuroVention Cranial Fixation System is a series of burr hole covers and plates with various configurations to facilitate surgeon selection of the implant he/she determines to be most appropriate for the patient and the surgical circumstances. Each is provided non-sterile single use and is made of titanium as per ASTM F67, titanium alloy (Ti-6AI4V ELI) implantable components that comply with ASTM F136 or PEEK per ASTM F2026. Class I exempt instrumentation is available for delivery and removal: Screwdriver Adapter (handle), Torx Drivers, Forceps. Additionally, a Class II Drill bit is included to create pilot holes for the screws.
#### Indications and Intended Use
The NeuroVention Cranial Fixation System is intended for use as a burr hole cover and/or skull bone fixation following craniotomy, cranioplasty, or craniectomy surgery.
#### Technological Characteristics and Substantial Equivalence
Documentation was provided to demonstrate that the NeuroVention Cranial Fixation System is substantially equivalent to the legally marketed predicates. It consists of a storage module that contains the various shapes of implants, screws and implantation instruments. The devices and accessories included in the Subject device and the predicate devices are both burr hole covers, cranioplasty plates and screws. The NeuroVention Cranial Fixation System is substantially equivalent to the predicate devices in intended use, site of application, patient populations of use, mechanical performances, basic design, and operating principles. The NeuroVention Cranial Fixation System is comparable to its predicate in size and materials. Mechanical testing shows the mechanical strength of the Subject device to be equivalent or better than the predicate devices.
| | NeuroVention Cranial<br>Fixation System | Stryker Universal Neuro 3 System | | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Manufacturer | NeuroVention | Stryker | | |
| 510(k)<br>number | This application | K112557 | | |
| Product<br>Code | GXR, GWO, HBW | GXR, GWO, HBW | | |
| Classification<br>Regulation | 21 CFR 882.5250, 882.5320,<br>882.5360 | 21 CFR 882.5250, 882.5320,<br>882.5360 | | |
| Common<br>Name | Burr Hole Cover, Preformed<br>Alterable Cranioplasty Plate,<br>Cranioplasty Plate Fastener | Burr Hole Cover, Preformed<br>Alterable Cranioplasty Plate,<br>Cranioplasty Plate Fastener | | |
| Class | II | II | | |
| Implantable<br>components | Yes | Yes | | |
| Anatomic<br>Locations | Skull | Skull | | |
| Intended<br>Use | Cover burr holes and skull<br>bone fixation following<br>craniotomy, cranioplasty, or<br>craniectomy surgery. | Cover burr holes and skull bone<br>fixation following craniotomy,<br>cranioplasty, or craniectomy<br>surgery. | | |
| Indications<br>for Use<br>Statement | The NeuroVention Cranial<br>Fixation System is intended<br>for use as a burr hole cover<br>and/or skull bone fixation<br>following craniotomy,<br>cranioplasty, or | The Stryker Universal Neuro 3<br>System is intended for<br>reconstruction, stabilization, and/or<br>rigid fixation of non-loadbearing<br>areas subsequent to craniotomy,<br>craniectomy, and cranial fractures | | |
| | craniectomy surgery. | In adults and adolescents (age 12 and higher). | | |
| Implant Material | Titanium, Titanium alloy (Ti6AL4V ELI)<br>PEEK (VESTAKEEP) per ASTM F2026 | Titanium, Titanium alloy (Ti6AL4V ELI) | | |
| Instrument materials | Stainless Steel | Stainless Steel | | |
| Sterilization | Provided non-sterile to be steam sterilized by user facility | Provided non-sterile to be steam sterilized by user facility | | |
| MRI compatibility | Non-clinical testing has demonstrated the NeuroVention Cranial Plate System is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:<br>• Static magnetic field of 1.5 T and 3.0 T;<br>• Maximum spatial field gradient of 6,200 G/cm (62 T/m);<br>• Quadrature driven, Circular Polarized (CP) Mode Only<br>• Maximum MR system-reported, whole-body averaged specific absorption rate (SAR) of 2 W/kg (Normal Control Mode).<br>• Maximum Head SAR of 3.2 W/kg.<br><br>Under the scan conditions defined above, the NeuroVention Cranial Plate System is expected to produce a maximum temperature rise of less than 2.5 °C after 15 minutes of continuous scanning at 1.5 T and less than 3.5 °C after 15 minutes of continuous scanning at 3.0 T.<br>In non-clinical testing, the image artifact caused by the device extends approximately 13 mm from the NeuroVention Cranial Plate System when imaged with a gradient echo pulse sequence and a 3 T MRI system. | "has not been evaluated for safety and compatibility in the MR environment" and that it "has not been tested for heating or migration in the MR environment." | | |
{5}------------------------------------------------
{6}------------------------------------------------
| Name of<br>Device | NeuroVention Cranial Fixation System | Stryker Universal<br>Neuro 3 System |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implant Size<br>Ranges | NuCrani Small Rectangular 4 hole plate, 12 x 20 mm<br>· NuCrani Medium Circular 8 hole plate<br>· NuCrani Large Circular 7 hole Plate, 20 x 27 mm<br>• KTC Burr Hole Cover 26.5 mm<br>• DC plate, 15 x 24 mm<br>· 2.0mm Self Tapping Screws, 6-10 mm<br>· 2.3 mm Self Tapping Rescue Screws, 6-10 mm | • 2 hole dogbone<br>plate<br>· Straight plate 2<br>hole<br>· Straight plate 8<br>hole<br>· Straight plate 16<br>hole<br>· Double Y plate<br>• Box plate 2x2 hole<br>· Rectangle plate<br>· X plate<br>· Gap plate, small<br>· Gap plate, large<br>• Burr hole cover,<br>7mm<br>• Burr hole cover,<br>10mm<br>• Burr hole cover<br>14mm<br>• Burr hole cover,<br>20mm<br>• Burr hole cover,<br>24mm<br>• Round malleable<br>plate, 42-58 mm<br>· Suboccipital<br>malleable plate,<br>62 x 30 mm<br>· Subocciptial<br>malleable plate,<br>65 x 40 mm<br>• Temporal<br>malleable plate,<br>47 x 29 mm<br>• Translabyrithine<br>malleable plate, 54<br>x 52 mm<br>· Translabyrinthine<br>malleable plate, 60<br>x 60 mm<br>· 1.5 mm Self Drilling<br>Screws, 3-5 mm<br>• 1.5 mm Self<br>Tapping screws, 4-<br>6 mm<br>· 1.7 mm Emergency<br>Screws, 4mm |
| Manual Drill Bit | 1.5 mm diameter | 1.2 mm diameter |
{7}------------------------------------------------
Performance Data The potential hazards have been evaluated and controlled through Risk Management.
> All testing met or exceeded the requirements as established by the test protocols and applicable standards. A review of the mechanical data indicates that the components of the Subject device are capable of withstanding expected loads without failure. The Subject device was therefore found to be substantially equivalent to the Predicates. Clinical data was not needed to support the safety and effectiveness of the Subject Device.
The following mechanical testing was performed:
- Static Compression
- . Skull Conformity
- Screw Axial Pushout •
- . Screw Torque to Failure per ASTM F543-13
- . Usability testing of Drill Bit
All results passed acceptance criteria and were equivalent or better when compared to the predicates.
Biocompatibility The components of the predicate device for the implantable components, are manufactured from the same titanium and titanium alloy for implantable components in conformity with ASTM F67-13 and ASTM F136. The Subject device also has a PEEK configuration made from VESTAKEEP by Evonik Industries per ASTM F2026. The instruments are made of the same stainless steel in conformance with ASTM F899. These materials are well known and well characterized have been used in other commercially available cranial implants. The predicate devices are used for the same duration and come into contact with the patient in the same manner and in the same body parts. The materials are considered to be biocompatible and have a long history of safe use as part of medical devices marketed in the United States and around the world. The Subject and predicate devices as well as a large number of implantable devices are manufactured in the same way with the same materials.
- Conclusion Based on design, materials, intended use, technological characteristics, and comparison to predicate devices, the Subject Neurovention Cranial Fixation System has been shown to be substantially equivalent to legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.