AMICUS Separator System

K192150 · Fresenius Kabi AG · LKN · Nov 13, 2019 · GU

Device Facts

Record IDK192150
Device NameAMICUS Separator System
ApplicantFresenius Kabi AG
Product CodeLKN · GU
Decision DateNov 13, 2019
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Pediatric

Indications for Use

The AMICUS Separator System is an automated blood component separator intended for use in the collection of blood components and mononuclear cells. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform Therapeutic Plasma Exchange (TPE). The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures.

Device Story

AMICUS Separator System is a continuous-flow, centrifugal blood cell separator. Device draws whole blood from donor/patient; separates components via centrifugal density differences; collects specific components; returns remainder to donor/patient. System utilizes pumps, clamps, and valves to move blood through single-use, sterile fluid path disposable kits. Operator monitors procedure via touch screen interface with audible/visual alarms. Software version 6.0 introduces protocol enhancements for platelet, mononuclear cell, TPE, and RBCX procedures, including blood warmer priming, configurable mini-cycles, and adjusted citrate concentration parameters. Device used in clinical settings for therapeutic apheresis and component collection. Output allows healthcare providers to perform therapeutic exchanges or collect blood products for transfusion. Benefits include automated, precise component separation and management of sickle cell disease.

Clinical Evidence

Bench testing only. Design verification and validation were performed in accordance with IEC 62304, ISO 14971, and FDA software validation guidance. No clinical data was required or provided for this software-only modification.

Technological Characteristics

Centrifugal blood separator; continuous-flow. Hardware includes pumps, clamps, valves, and touch screen interface. Disposable fluid path kits. Connectivity: data management capabilities. Standards: IEC 62304 (software), ISO 14971 (risk management), ES 60601-1 (electrical safety). Software version 6.0.

Indications for Use

Indicated for therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX), including exchange and depletion/exchange for sickle cell disease in adults and children. Also indicated for collection of blood components (platelets, red blood cells, plasma, mononuclear cells). Contraindications include donor post-platelet count <100,000/microliter or extracorporeal volume >10.5ml/kg.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products. November 13, 2019 Fresenius Kabi AG Aunica Jones Senior Regulatory Affairs Specialist Regulatory Affairs Transfusion and Cell Technologies Division Fenwal, Inc. Three Corporate Drive Lake Zurich, IL 60047 Re: K192150 Trade/Device Name: AMICUS Separator System Regulatory Class: Unclassified Regulation Name: None Regulation Number: None Product Code: LKN, GKT Dated: August 14, 2019 Received: August 15, 2019 Dear Aunica Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192150 Device Name AMICUS Separator System #### Indications for Use (Describe) Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following: The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702) The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702) The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615) The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615) The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615) The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615) The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells. Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect: - · Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) (BK960005) (BK990009) - · Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double or triple units) (BK090065) - · Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039) - Mononuclear Cells (BK000047) - · Plasma (BK960005, BK120041) - o Fresh Frozen Plasma · Must be prepared and placed in a freezer at -18° C or colder within 8 hours after phlebotomy. - o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) - Must be stored at 1-6℃ within 8 hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy. · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma. o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24) - · Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at {3}------------------------------------------------ -18° C or colder within 24 hours after phlebotomy. - · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma. o Source Plasma The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter. (BK080018) Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Additive Solution (InterSol) (single, double, or triple units). (BK090065) The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to 7 days. Additionally, for platelet units stored past 5 days and through 7 days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure." (BK040059, BK150242) Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ #### Fresenius Kabi Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us ## 510(k) SUMMARY #### Date Prepared: October 17, 2019 ## Owner/Operator: #9027285 Fresenius Kabi AG 61346 Bad Homburg Germany #### Contact Person: Aunica Jones Senior Regulatory Affairs Specialist Fresenius Kabi USA, LLC Three Corporate Drive Lake Zurich, IL 60047 USA T: 847-550-2387 F: 847-550-2960 aunica.jones@fresenius-kabi.com ## Device Trade Name: AMICUS Separator System ## Common Name/Usual Name: Automated Separator, Blood Cell and Plasma, Therapeutic Automated Blood Cell Separator (Centrifugal Separation Principle) ## Classification Name: Automated separators, used for separation of blood cells and plasma for therapeutic purposes, have not been classified under a regulation by the Center for Devices and Radiological Health due to preamendment status. ## 21 CFR 864.9245 Automated Blood Cell Separator Automated blood cell separators which are based on centrifugation type technology have been classified by the Center for Biologics Evaluation and Research as Class II devices with Special Controls (Docket 2005N-0017, Final Rule, 30-Nov-07, updated March 28, 2011 OMB Control No: 0910-0594). ## Product Code and Classification Panel: LKN (Gastroenterology/Urology panel) - Unclassified (due to pre-amendment status) {5}------------------------------------------------ # GKT (Hematology panel) - Separator, Automated, Apheresis #### Legally Marketed Device Under Which Substantial Equivalence is Being Claimed: Fresenius Kabi is claiming substantial equivalence with the currently cleared version of the AMICUS Separator System. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications including Therapeutic Plasma Exchange (TPE), Red Blood Cell Exchange and Red Blood Cell Depletion (RBCX procedures). The AMICUS Separator System was most recently cleared by CDRH under K180615 on Dec 4, 2018. This includes all operating protocols and changes previously cleared for the AMICUS Separator System. #### Device Description: The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a single-use, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences. The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS separator during the procedure. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alarms. Once the cell separation is complete, the operator disconnects the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray. ## Modification to the Existing Device: Software version 6.0 has been developed for use with the AMICUS Separator System. The AMICUS software 6.0 release included enhancements for Single Needle Platelet (SNP), Double Needle Platelet (DNP), Mononuclear Cell (MNC), Therapeutic Plasma Exchange (TPE), and Red Blood Cell Exchange (RBCX) protocols. These functional changes include: - Added Blood Warmer Primer Option MNC Procedure ● - Added Prime and Go Option MNC Procedure ● - Ability to Configure Mini-Cycle as On/Off MNC Procedure ● - Ability to Decrease Custom AC Citrate Concentration – TPE and RBCX Procedures - Change to MNC Reinfusion Rate - MNC Procedure - Update Parameters for ACD Ratio - MNC and ECP Procedures - Additional Output Values – MNC, TPE and RBCX Procedures - Data Management Enhancements - Additional Minor Enhancements/Updates ● Modifications have also been made to the operating instructions, both in support of the software changes and as a result of review and update. The cumulative changes in the software and operator's manual are such that a 510(k) should be submitted for the changes. {6}------------------------------------------------ The AMICUS instrument hardware remains the same as the currently cleared device. The AMICUS apheresis Kits remain the same as the currently cleared kits. # Statement of Intended Use: The AMICUS Separator System is an automated blood component separator intended for use in the collection of blood components and mononuclear cells. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform Therapeutic Plasma Exchange (TPE). The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures. ## Indications for Use: Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following: The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702) The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702) The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615) The AMICUS Exchange Kit - Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K11702, K180615) The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615) The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615) The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells. Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect: {7}------------------------------------------------ - . Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) (BK960005) (BK990009) - . Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double or triple units) (BK090065) - Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039) ● - Mononuclear Cells (BK000047) - Plasma (BK960005, BK120041) - o Fresh Frozen Plasma - . Must be prepared and placed in a freezer at -18° C or colder within 8 hours after phlebotomy. - o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) - . Must be stored at 1-6℃ within 8 hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy. - Indicated for replacement of non-labile clotting factors. This product ● is not equivalent to Fresh Frozen Plasma. - o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24) - . Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy. - Indicated for replacement of non-labile clotting factors. This product ● is not equivalent to Fresh Frozen Plasma. - o Source Plasma The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter. (BK080018) Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Pheresis, Platelet Additive Solution (InterSol) (single, double, or triple units). (BK090065) The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to 7 days. Additionally, for platelet units stored past 5 days and through 7 days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure." (BK040059, BK150242) # Technological Comparison as Compared to the Predicate Device The technological characteristics of the AMICUS separator remain the same as the predicate AMICUS device. This includes the centrifuge system, fluid control system, safety management system (including safety sensors and alarms), and anticoagulant management system. The physical design of the AMICUS separator instrument is identical to the marketed AMICUS device. The AMICUS apheresis kits remain the same as the currently cleared kits, including design, materials and manufacturing methods. The data management capabilities remain the same as the cleared AMICUS device. {8}------------------------------------------------ # Performance Data: The Design Verification and Design Validation processes are designed to meet the requirements for software development and testing as described in the following standards and guidance documents: - IEC 62304 Ed. 1.0, "Medical device software Software life cycle processes" . - ISO 14971:2012 Medical Devices – Applications of Risk Management to Medical Devices - FDA/CDRH "General Principles of Software Validation; Final Guidance for Industry and ● FDA Staff." Issued January 11, 2002. - ES 60601-1:2005/(R)2012 and A1:2012, C1: 2009/(R)2012 and A2:2010/(R)2012 . Medical electrical equipment - Part 1: General requirements for basic safety and essential performance Software and systems verifications were performed in support of this submission. The results of the testing were acceptable. # Conclusion: Based on the verification activities performed, the AMICUS Separator System modified with software 6.0 provides a device system that is substantially equivalent to the currently marketed AMICUS Separator System.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...