Panthera Dental Milled Bars

K192108 · Panthera Dental, Inc. · NHA · Apr 8, 2020 · Dental

Device Facts

Record IDK192108
Device NamePanthera Dental Milled Bars
ApplicantPanthera Dental, Inc.
Product CodeNHA · Dental
Decision DateApr 8, 2020
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Panthera Dental Milled Bar is indicated for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient for purpose of restoring chewing function. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained. The Panthera Dental Milled Bars are indicated for compatibility with: NobelActive: NP Ø3.5; RP Ø4.3 / Ø5.0; WP Ø5.5 NobelParallel CC: NP Ø3.75; RP Ø4.3 / Ø5.0; WP Ø5.5 NobelReplace: NP Ø3.5; RP Ø4.0 / Ø4.3 / Ø5.0; WP Ø5.0; 6 Ø6.0 NobelSpeedy: RP Ø4.0 / Ø5.0; WP Ø5.0 / Ø6.0 Brånemark: NP Ø3.3; RP Ø3.75 / Ø4.0; WP Ø5.0

Device Story

Metallic dental restorative device (milled bar) used as an accessory to endosseous dental implants; supports fixed or removable dentures. Input: 3D optical/digital scan of patient impression. Operation: CAD/CAM software designs bar based on scan; bar milled from titanium (Ti-6Al-4V grade 5) by manufacturer. Output: Custom-milled titanium bar. Used in dental laboratory/clinic; operated by dental professionals. Healthcare provider matches bar to denture for final placement. Provides retention/support for prosthetic devices (e.g., resin composite). Benefits patient by restoring chewing function and esthetics.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing: fatigue testing per ISO 14801, biocompatibility testing (ISO 10993-1, -5, -12, ANSI/AAMI ST72, USP <85>), sterilization validation (ANSI/AAMI-ST8), and reverse engineering analysis of connection interfaces.

Technological Characteristics

Material: Titanium alloy Ti-6Al-4V. Design: CAD/CAM milled custom bars. Connection: Internal conical, Internal Hex, External Hex, Internal Tri-channel. Sterilization: Steam autoclave (132°C, 4 min, 30 min dry). Connectivity: N/A (mechanical device).

Indications for Use

Indicated for partially or edentulous patients requiring support for multiple tooth prostheses (screw-retained) in the mandible or maxilla to restore chewing function. Compatible with specific Nobel Biocare implant systems (NobelActive, NobelParallel CC, NobelReplace, NobelSpeedy, Brånemark).

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 8, 2020 Panthera Dental Inc. Martine Fortin Director Regulatory Affairs and Ouality Assurance 9105 rue John-Simons Quebec City, Québec, G2B 0S6 CANADA Re: K192108 Trade/Device Name: Panthera Dental Milled Bars Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 6, 2020 Received: March 11, 2020 Dear Martine Fortin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K192108 Device Name Panthera Dental Milled Bars ### Indications for Use (Describe) The Panthera Dental Milled Bar is indicated for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient for purpose of restoring function. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained. The Panthera Dental Milled Bars are indicated for compatibility with: - · NobelActive: NP Ø3.5; RP Ø4.3 / Ø5.0; WP Ø5.5 - NobelParallel CC: NP Ø3.75: RP Ø4.3 / Ø5.0: WP Ø5.5 - NobelReplace: NP Ø3.5; RP Ø4.0 / Ø4.3 / Ø5.0; WP Ø5.0; 6 Ø6.0 - NobelSpeedy: RP Ø4.0 / Ø5.0; WP Ø5.0 / Ø6.0 - Brånemark: NP Ø3.3; RP Ø3.75 / Ø4.0; WP Ø5.0 | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black silhouette of a panther in a crouching position, ready to pounce. Below the panther is the word "PANTHERA" in a bold, sans-serif font, with the word "DENTAL" in a smaller font underneath. A purple line underlines the word "PANTHERA". ### 510 (k) Summary 5 | Date Prepared: | April 7, 2020 | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Panthera Dental Inc.<br>9105 rue John-Simons<br>Quebec City (QC) G2B 0S6<br>Canada | | Official Contact: | Martine Fortin<br>Director Regulatory Affairs and Quality Assurance<br>regulatory@pantheradental.com<br>Tel: +1 (418) 527-0388<br>Fax: +1 (418) 431-9942 | | Proprietary Name: | Panthera Dental Milled Bars | | Device Common Name: | Overdenture Bar | | Classification: | 21 CFR 872.3630 (Class II) | | Product Code: | NHA | | Primary Predicate: | K173466 Panthera Dental Milled Bars | | Reference Devices: | K071370, K133731 NobelActive<br>K173418 NobelParallel Conical Connection<br>K062566, K073142, K023113 NobelReplace<br>K050406, K160119 NobelSpeedy | ## 5.1 Description The Panthera Dental Milled Bar is a metallic dental restorative device that is intended for attaching by screw retention to dental implants to aid in the treatment of partial and totally edentulous patients for the purpose of restoring chewing function. The Panthera Dental milled bars for which clearance is requested, are included in one of the following bar types, which have distinct design specifications. The Type I bars are specific for removable overdenture and include: - . Panthera Dental Dolder Bar, Hader Bar, Milled Bar, REBourke Bar and Paris Bar. 9105 rue John-Simons, Québec (Qc), G2B 0S6, Canada, Tel: +1 418-527-0388 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a purple sans-serif font, with the word "DENTAL" written in a smaller font below it. The Type II bars are specific for fixed prostheses and include: - Panthera Dental Wrap-around Bar, Montreal Bar, Montreal Bar with metallic lingual, Pin Lingual Bar and Pin Wrap-Around Bar. Table 5.1 presents the design specifications per bar types. | | Type I | | Type II | | |-----------------------------------------|---------|---------|---------|---------| | Description | Minimum | Maximum | Minimum | Maximum | | Platform Seating Diameter (mm) | 2 | 8 | 2 | 8 | | Total Cylinders | 2 | 10 | 2 | 10 | | Bar Span Between Cylinders (mm) | 0 | 30 | 0 | 30 | | Bar Height (mm) | 2.51 | 8 | 3.52 | 22 | | Bar Width (mm) | 1.53 | 124 | 2.55 | 106 | | Distal Extension (mm) | 0 | 307 | 0 | 30 | | Cylinder Height (mm) | 0 | 10 | | | | Cylinder Diameter (mm) | 3 | 8 | | | | Maximum Divergence Between<br>Cylinders | 0° | 30° | | | Table 5.1: Design Specifications of Panthera Dental Bars The Panthera Dental Milled Bar is designed to match an individual patient. Panthera Dental designs the bar from a three-dimensional optical and/or digital scanner system that scans the patient's impression; the dental professional prepares the model cast beforehand. The designed bar is then machined using a computer-aided design/ computer-aided manufacturing (CAD/CAM) software system. The bar is milled from titanium (Ti-6AI-4V grade 5). CAD/CAM fabrication is only performed by Panthera Dental, within our manufacturing control and not by the dental laboratory. The Panthera Dental Milled Bar is packaged as non-sterile, and delivered to a dental laboratory for completion. Once received at the laboratory, the Panthera Dental Milled Bar is matched to a denture for final placement. The Panthera Dental Milled Bar provides <sup>1</sup> Except for the Paris Bar that is 3.5 mm. <sup>2</sup> Except for the Wrap-Around Bar that is 2.5 mm and the Pin Wrap-Around that is 3 mm. <sup>3</sup> Except for the Dolder Bar that is 1.5 mm; the Hader and Milled Bars that are 1.8 mm and the Paris Bar that is 4 mm. <sup>4</sup> Except for the Dolder Bar and the Hader Bar that are 5 mm; the Milled Bar that is 10 mm. <sup>5</sup> Except for the Montreal Bar and the Montreal with Metallic Lingual that are 4 mm. <sup>6</sup> Except for the Montreal Bar and the Montreal with Metallic Lingual that are 12 mm. <sup>7</sup> Except for the Dolder Bar and the Hader bar that are 20 mm. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, ready to pounce. Below the panther, the word "PANTHERA" is written in a bold, purple font, with the word "DENTAL" written in a smaller font below it. retention and support for a removable or fixed denture made of standard laboratory dental materials such as resin composite. The Panthera Dental Milled Bars are indicated for compatibility with the following Nobel Biocare implant system platforms: NobelActive/ Parallel CC/ Speedy/ Brånemark: NP/RP/WP; NobelReplace: NP/RP/WP/6; cleared under K071370, K133731, K173418, K062566, K073142, K023113, K050406, K160119 and K022562. # 5.2 Indications for Use The Panthera Dental Milled Bar is indicated for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient for purpose of restoring chewing function. It is intended for use to support multiple tooth prostheses in the mandible or maxilla. The prostheses can be screw retained. The Panthera Dental Milled Bars are indicated for compatibility with: NobelActive: NP Ø3.5; RP Ø4.3 / Ø5.0; WP Ø5.5 NobelParallel CC: NP Ø3.75; RP Ø4.3 / Ø5.0; WP Ø5.5 NobelReplace: NP Ø3.5; RP Ø4.0 / Ø4.3 / Ø5.0; WP Ø5.0; 6 Ø6.0 NobelSpeedy: RP Ø4.0 / Ø5.0; WP Ø5.0 / Ø6.0 Brånemark: NP Ø3.3; RP Ø3.75 / Ø4.0; WP Ø5.0 ### Technological 5.3 Aside from incorporating features to facilitate connection to Nobel Biocare implant systems, the materials, design, operating principles, manufacturing and sterilization method are identical to those of the primary predicate intended for use with Zimmer implant systems. No new surgical instruments or secondary components are being introduced as a result of this submission. The specific features from the proposed device allow a perfect connection between the Panthera bar cylinder and the Nobel Biocare implant necessary to meet the intended use that is to support a prosthetic device in the mandible or maxilla, in a partially or edentulous patient. The additional Nobel Biocare implant systems compatibility do not add new bar designs, in fact, the same bar designs are available for both OEM (Zimmer and Nobel Biocare). Therefore, we believe that the new device is substantially equivalent to the primary predicate. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black silhouette of a panther in a stalking pose, with its tail curved upwards. Below the panther, the word "PANTHERA" is written in a stylized, sans-serif font, with the letters in a light purple color. Underneath "PANTHERA", the word "DENTAL" is written in a smaller, simpler font, also in light purple. K192108 Table 5.2: Comparison chart between the proposed device and the reference devices | Feature | The Panthera Dental<br>Milled Bars<br>(proposed device) | The Panthera Dental<br>Milled Bars<br>(Primary Predicate<br>K173466) | NobelActive<br>(K071370, K133731),<br>NobelParallel CC<br>(K173418) | NobelReplace<br>(K062566,<br>K073142, K023113) | NobelSpeedy<br>(K160119 incl.<br>K050406) | Brånemark<br>(K022562) | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation<br>description | Endosseous dental<br>implant abutment | Endosseous dental<br>implant abutment | Endosseous dental<br>implant | Endosseous dental<br>implant | Endosseous dental<br>implant | Endosseous dental<br>implant | | Indications<br>for Use | As an accessory to an<br>endosseous dental<br>implant to support a<br>prosthetic device in a<br>partially or edentulous<br>patient. It is intended for<br>use to support multiple<br>tooth prostheses in the<br>mandible or maxilla.<br>The prostheses can be<br>screw retained.<br>Compatibility with the<br>Nobel Biocare systems. | As an accessory to an<br>endosseous dental<br>implant to support a<br>prosthetic device in a<br>partially or edentulous<br>patient. It is intended for<br>use to support multiple<br>tooth prostheses in the<br>mandible or maxilla.<br>The prostheses can be<br>screw retained.<br>Compatibility with the<br>Zimmer tapered screw-<br>vent system | NobelActive: As<br>endosseous implant<br>intended to be surgically<br>placed in the bone of the<br>upper or lower jaw<br>arches to provide<br>support for prosthetic<br>devices, such as an<br>artificial tooth, in order<br>to restore patient<br>esthetics and chewing<br>function. Indicated for<br>single or multiple unit<br>restorations in splinted<br>or non-splinted<br>applications. May be<br>placed immediately and<br>put into immediate<br>function provided that<br>initial stability<br>requirements detailed in<br>the manual are satisfied. | As endosseous dental<br>implants intended to<br>be surgically placed in<br>the bone of the upper<br>or lower jaw arches to<br>provide support for<br>prosthetic devices,<br>such as an artificial<br>tooth, in order to<br>restore patient<br>esthetics and chewing<br>function. Indicated for<br>single or multiple unit<br>restorations. Can be<br>used in splinted or<br>nonsplinted<br>applications. May be<br>placed immediately<br>and put into immediate<br>function provided that<br>initial stability<br>requirements detailed<br>in the manual are<br>satisfied. | K050406: Intended to<br>be surgically placed in<br>the bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such<br>as an artificial tooth, in<br>order to restore patient<br>esthetics and chewing<br>function. Indicated for<br>single or multiple unit<br>restorations in splinted<br>or non-splinted<br>applications. May be<br>placed immediately and<br>put into immediate<br>function providing that<br>the initial stability<br>requirements detailed in<br>the surgical manuals are<br>satisfied. Indicated for<br>use in soft bone or<br>whenever immediate or<br>early loading is applied.<br>Incorporate a groove on<br>the implant thread and | For single-stage or<br>two-stage surgical<br>procedures and cement<br>or screw retained<br>restorations. Intended<br>for immediate<br>placement and<br>function on single<br>tooth and or multiple<br>tooth applications<br>recognizing sufficient<br>bone stability (type I<br>or I1 bone) and<br>appropriate occlusal<br>loading, to restore<br>chewing function.<br>Multiple tooth<br>applications may be<br>splinted with a bar. | | Feature | The Panthera Dental<br>Milled Bars<br>(proposed device) | The Panthera Dental<br>Milled Bars<br>(Primary Predicate<br>K173466) | NobelActive<br>(K071370, K133731),<br>NobelParallel CC<br>(K173418) | NobelReplace<br>(K062566,<br>K073142, K023113) | NobelSpeedy<br>(K160119 incl.<br>K050406) | Brånemark<br>(K022562) | | | | | NobelParallel CC: As<br>endosseous implants<br>intended to be<br>surgically placed in the<br>upper or lower jaw<br>bone for anchoring or<br>supporting<br>replacements to restore<br>patient esthetics and<br>chewing function.<br>Indicated for single or<br>multiple restorations in<br>splinted or non-splinted<br>applications. This can<br>be achieved by a 2-<br>stage or 1-stage<br>surgical techniques in<br>combination with<br>immediate, early or<br>delayed loading<br>protocols, recognizing<br>sufficient primary<br>stability and<br>appropriate occlusal<br>loading for the selected<br>technique. Implants<br>with <7mm length are<br>for delayed loading<br>only when appropriate | Addition for<br>K050258: Groovy<br>implants are indicated<br>for use in soft bone in<br>posterior regions or<br>whenever immediate<br>or early loading is<br>applied. The Groovy<br>implants incorporate a<br>groove on the implant<br>thread and are<br>preferred over models<br>without the groove in<br>these soft bone<br>indications because<br>bone forms more<br>rapidly in the<br>groove than on other<br>parts of the implant<br>resulting in increased<br>stability when<br>compared to non-<br>grooved implants.<br>Addition for<br>K023113: If the single<br>stage procedure is<br>used, these implants<br>may be loaded<br>immediately following<br>insertion provided at | are preferred over<br>models without the<br>groove in these soft bone<br>indications because bone<br>forms more rapidly in<br>the groove than on other<br>parts of the implant<br>resulting in increased<br>stability when compared<br>to non-grooved<br>implants. In addition,<br>preferred in these soft<br>bone indications because<br>bone formation on the<br>Ti~lnite surface is more<br>rapid and greater than on<br>machined surface<br>implants resulting in<br>better maintenance of<br>initial implant stability,<br>faster and stronger<br>osseointegration, and<br>higher success rates.<br>Implants may be tilted<br>up to 45°. When used<br>with angulations<br>between 30° and 45° a<br>minimum of four<br>implants must be used<br>and splinted. | | | Feature | The Panthera Dental<br>Milled Bars<br>(proposed device) | The Panthera Dental<br>Milled Bars<br>(Primary Predicate<br>K173466) | NobelActive<br>(K071370, K133731),<br>NobelParallel CC<br>(K173418) | NobelReplace<br>(K062566,<br>K073142, K023113) | NobelSpeedy<br>(K160119 incl.<br>K050406) | Brånemark<br>(K022562) | | | | | stability has been<br>achieved. | least four implants are<br>placed, and are<br>splinted with a bar.<br>These implants must<br>be placed<br>predominantly in the<br>anterior mandible<br>(between the mental<br>foramina) where good<br>initial stability of the<br>implants, with or<br>without bi-cortical<br>anchorage, can most<br>often be obtained. | K160119: Intended to be<br>surgically placed in the<br>upper or lower jaw bone<br>for anchoring or<br>supporting tooth<br>replacements to restore<br>patient esthetics and<br>chewing function.<br>Indicated for single or<br>multiple unit restorations<br>in splinted or non-<br>splinted applications.<br>This can be achieved by<br>a 2-stage or 1-stage<br>surgical technique in<br>combination with<br>immediate, early or<br>delayed loading<br>protocols, recognizing<br>sufficient primary<br>stability and appropriate<br>occlusal loading for the<br>selected technique.<br>Implants allow also for<br>bicortical anchorage in<br>cases of reduced bone<br>density. Implants 20, 22,<br>25 mm when placed in<br>the maxilla are only ind-<br>icated for multiple unit<br>restorations in splin-ted | | | Feature | The Panthera Dental<br>Milled Bars<br>(proposed device) | The Panthera Dental<br>Milled Bars<br>(Primary Predicate<br>K173466) | NobelActive<br>(K071370, K133731),<br>NobelParallel CC<br>(K173418) | NobelReplace<br>(K062566,<br>K073142, K023113) | NobelSpeedy<br>(K160119 incl.<br>K050406) | Brånemark<br>(K022562) | | | | | | | applications that utilize<br>at least two implants. | | | Device<br>Material | Titanium alloy Ti-6Al-<br>4V | Titanium alloy Ti-6Al-<br>4V | CP Titanium | CP Titanium | CP Titanium | CP Titanium | | Design/<br>Technology | CAD/CAM milling from<br>single milling blanks. | CAD/CAM milling from<br>single milling blanks. | N/A | N/A | N/A | N/A | | Connection<br>Type | Internal conical,<br>Internal Hex, External<br>Hex, Internal Tri-<br>channel | N/A | Internal conical<br>(previously called<br>Internal Hex) | Internal Hex<br>Internal Tri-channel | External Hex | External Hex | | Implant<br>Platform /<br>diameter<br>(mm) | NobelActive:<br>NP Ø3.5; RP Ø4.3 /<br>Ø5.0; WP Ø5.5<br>NobelParallel CC:<br>NP Ø3.75; RP Ø4.3 /<br>Ø5.0; WP Ø5.5<br>NobelReplace<br>NP Ø3.5; RP Ø4.0 /<br>Ø4.3 / Ø5.0; WP Ø5.0; 6<br>Ø6.0<br>NobelSpeedy:<br>RP Ø4.0 / Ø5.0; WP<br>Ø5.0 / Ø6.0<br>Brånemark:<br>NP Ø3.3; RP Ø3.75 /<br>Ø4.0; WP Ø5.0 | Zimmer implant systems | NobelActive:<br>NP Ø3.5; RP Ø4.3 /<br>Ø5.0; WP Ø5.5<br>NobelParallel CC:<br>NP Ø3.75; RP Ø4.3 /<br>Ø5.0; WP Ø5.5 | NobelReplace<br>Tapered:<br>NP Ø3.5; RP Ø4.3 /<br>Ø5.0; WP Ø5.0;<br>6 Ø6.0<br>NobelReplace<br>NP Ø3.5; RP Ø4.0 | NobelSpeedy:<br>NP Ø3.5; RP Ø4.0 /<br>Ø5.0; WP Ø5.0 / Ø6.0 | Brånemark:<br>NP Ø3.3; RP Ø3.75 /<br>Ø4.0; WP Ø5.0 | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a stylized font, with the word "DENTAL" written in a smaller font below it. The text is purple. K192108 {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black silhouette of a panther in a crouching position, with its tail curved upwards. Below the panther, the word "PANTHERA" is written in a bold, sans-serif font, with the word "DENTAL" written in a smaller font size below it. The text is purple. K192108 Confidential {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position above the words "PANTHERA DENTAL" in purple. The word "PANTHERA" is in a larger font than the word "DENTAL". K192108 {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows a logo for "Panthera Dental". The logo features a black silhouette of a panther in a crouching or stalking pose. Below the panther image, the word "PANTHERA" is written in a stylized, purple font. Underneath "PANTHERA", the word "DENTAL" is written in a smaller, simpler font, also in purple. # 5.4 Risks Panthera Dental Inc. did not performed clinical testing. However, a FMEA risk analysis, and evaluation of the materials of construction and the design were performed. The function of the bars requires that the prescribing dentist is cognizant of the potential for soreness, soft tissue soreness, and dentition complications by the bars. Management of these risks is achieved by advising the patient and dentist in the directions for use that the prescribing dentist must perform early and repeated examination of the performance of the device, and its fit, in the dental office. The precautions, storage directions and prescription preparation instructions are written to avoid potential problems from arising or persisting with the dentition or tissue, caused by the bars. The proprietary manufacturing of the Panthera Dental bar includes materials for the milling, the polishing and the cleaning that are biocompatible and from standard manufacturing in dental practices. This manufacturing process does not alter the chemical or physical properties of the Panthera Dental bar. No new risks are introduced with the new device that are not present in the primary predicate device. ## 5.5 Non-Clinical Testing The non-clinical testing includes assessment of the physical properties of the bars and its ability to achieve its intended use. The bars meet the same specifications as set for the primary predicate device. #### 5.5.1 Sterilization A sterilization validation was conducted for the primary predicate (K173466). The complete validation of the following steam autoclave sterilization process has been conducted: Pre-vacuum steam sterilization cycle for wrapped instruments for 4 minutes, at 132°C followed by a drying period of 30 minutes. The first part of the validation consisted in the validation of the sterility test used to test the sterility of the dental bars after the sterilization cycle; validation parameters where based on USP<71> (Sterility Tests). In the second part, the sterility test has demonstrated that the Panthera Dental bars were sterile once the above cycle was completed. The sterilization test was performed in an FDA cleared sterilizer (# K111223), Amsco Chimeron Small Prevacuum Steam Sterilizer. Based on the effectiveness statement of this sterilizer, the Prevac cycle was validated using full load instrument trays, described in 5.5.4.1 of {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for Panthera Dental. The logo features a black panther in a crouching position, ready to pounce. Below the panther, the word "PANTHERA" is written in a stylized, purple font, with the word "DENTAL" written in a smaller font below it. ANSI/AAMI-ST8, and was qualified according to Section 5.5.4 of ANSI/AAMI-ST8. This cycle demonstrated a sterility assurance level of at least 10° using half-cycle analysis. The sterilization process is a typical steam autoclave sterilization process, it does not alter the chemical or physical properties of the Panthera Dental bar. The complete sterility testing conducted for the primary predicate (K173466) used the same materials and the same sterilization cycle as for the proposed device. Therefore, no additional sterility testing was required for the proposed device. #### 5.5.2 Biocompatibility The biocompatibility analysis conducted for the primary predicate (K173466) supports the substantial equivalence in the safety and effectiveness of the predicate device. The raw material used in the manufacturing of the Panthera Dental bar is the same as for the predicate device: Titanium Ti-6Al-4V. Both Panthera Dental bar and the predicate bar are categorized as external communicating devices that contact tissue/bone/dentin for more than 30 days. The cytotoxicity, the extractable/leachable chemical analysis, the toxicological risk assessment and the bacterial endotoxins testing performed with the Panthera Dental bars support substantial equivalence to the predicate, following the standards of ISO 10993-1, ISO 10993-5, ISO 10993-12, ANSI/AAMI ST72, and USP <85>. Therefore, we believe that the new device is as safe as the legally marketed device. #### 5.5.3 FEA Finite Element Analysis (FEA) was done on the bar cylinders prior to starting any testing. #### 5.5.4 Fatigue Testing Fatigue testing was performed on the Panthera Dental bar. The first test was performed on the bar itself as part of the primary predicate submission (cleared K173466). The second one was performed on the bar cylinders. The testing was performed in accordance with the ISO 14801standard "Dentistry - Implants-Dynamic fatigue test for endosseous dental implants" and the FDA guidance "Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Implant Abutments; guidance for industry and FDA." The results have demonstrated that the Panthera Dental Milled Bar as well as the cylinders constituting the bar have the mechanical strength for its intended clinical application. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a stylized, purple font, and below that, the word "DENTAL" is written in a smaller, sans-serif font. #### 5.5.5 Reverse engineering analysis Panthera Dental has used the reverse engineering (RE) method to determine the critical parameters of the design of each interface connection including the tolerance limits, between each components per size and type, for the Nobel Biocare implant systems, including OEM Implant bodies, OEM abutments, and OEM abutment fixation screws. With those in hand, it was possible to reproduce an exact copy of the Nobel Biocare cylinder's interface connection and to ensure a perfect fit between the proposed cylinder and its Nobel Biocare RE counterparts (implant and screw). The RE of the Nobel Biocare components gives the assurance that the RE values of the connection interfaces of our proposed device fall within the measured sizes of the Nobel Biocare counterparts and confirms that the Pantal assemblies will always be functional with no possible interference. To complete the RE, the different gaps were verified in the final Panthera assemblies and confirmed proper seating of the cylinder on the implant as well as the insertion and the seating of the screw in the cylinder. The RE analysis performed for the proposed device is identical to the one performed for the primary predicate. #### 5.5.6 Process capability study In order to confirm the reliability of the manufacturing process of the Panthera Dental bar. a process capability test was conducted for the primary predicate (K173466). This test is valid for the proposed device because the manufacturing process is identical for both the proposed and the primary predicate device. ### 5.6 Clinical Testing Human clinical study was not deemed necessary to support substantial equivalence. The Panthera Dental Milled Bar do not: use designs dissimilar from the primary predicate device; do not use new technologies different from legally marketed milled bars; and do not deviate from the indications for use identified in the primary predicate device: Panthera Dental Milled Bar K173466. {13}------------------------------------------------ ### Substantial Equivalence Conclusion 5.7 The new device, the Panthera Dental Milled Bar, is considered substantially equivalent to the primary predicate device based on the following. Both devices (proposed and primary predicate) have the same indications for use and are indicated for the same user population; use the same operating principle; incorporate the same basic design; use the same biocompatible material and the same manufacturing process; and have the same technological characteristics.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%