K192065 · Siemens Medical Solutions USA, Inc. · MUJ · Sep 18, 2019 · Radiology
Device Facts
Record ID
K192065
Device Name
Syngo.Via RT Image Suite
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MUJ · Radiology
Decision Date
Sep 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ-at-risk (OAR) segmentation
Deep Image-to-Image Network
—
100% detection rate for Brain, Liver, Kidney Left, and Kidney Right; mean ASSD 0.64 mm to 3.04 mm; mean DICE 0.85 to 0.97
32 datasets
—
32 datasets
—
Indications for Use
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to efficiently view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create and modify simple treatment plans. Contours, images and treatmently be exported to a Treatment Planning System. The software combines following digital image processing and visualization tools: · Multi-modality viewing and contouring of anatomical, functional, and multi-parametric images such as but not limited to CT. PET. PET/CT. MRI. Linac Cone Beam CT (CBCT) images and dose distributions · Multiplanar reconstruction (MPR) thick, minimum intensity projection (MIP), volume rendering technique (VRT) - · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique - · Creation of contours on any type of images without prior assignment of a planning CT · Manual and semi-automatic registration using rigid and deformable registration · Supports the user in comparing, and adapting contours based on datasets acquired with different imaging modalities and at different time points · Supports the user in comparing images and contours of different patients - · Supports multi-modality image fusion - · Visualization and contouring of moving tumors and organs - · Management of points of interest including but not limited to the isocenter - · Management of simple treatment plans - · Generation of a synthetic CT based on multiple pre-define MR acquisitions
Device Story
Software tool for radiation therapy planning; inputs include multi-modality medical images (CT, PET, PET/CT, MRI, CBCT) and dose distributions. Transforms inputs via digital image processing, visualization (MPR, MIP, VRT), and deep learning-based segmentation to produce contours, structure sets, and simple treatment plans. Used in clinical settings by medical professionals (e.g., radiation oncologists, physicists). Features include manual/semi-automatic registration (rigid/deformable), contouring of organs-at-risk/targets, and synthetic CT generation. Output exported to external Treatment Planning Systems. Assists clinicians in defining treatment volumes and isocenters; improves workflow efficiency and segmentation accuracy for radiation therapy preparation.
Clinical Evidence
Clinical evaluation compared subject (SOMARIS/8 VB40) to predicate (SOMARIS/8 VB30) using 32 datasets. Evaluated organ-at-risk (OAR) segmentation performance. Detection rate for Brain, Liver, and Kidneys reached 100%. Segmentation quality measured via Average Symmetric Surface Distance (ASSD) and DICE coefficient. Mean ASSD ranged 0.64 mm to 3.04 mm; mean DICE ranged 0.85 to 0.97. Results demonstrated improved segmentation quality for the subject device.
Technological Characteristics
Software-based image analysis tool. Features multi-modality viewing, MPR, MIP, VRT, and manual/semi-automatic registration. Implements deep learning-based segmentation using an adversarial network. Conforms to DICOM (PS 3.1-3.20), IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366-1 (usability). Operates on standard clinical hardware platforms.
Indications for Use
Indicated for 3D/4D image visualization, multimodality manipulation, and contouring to assist in preparation and response assessment for radiation therapy (e.g., Brachytherapy, Particle Therapy, External Beam Radiation Therapy). Applicable to patients requiring anatomical/functional image analysis for treatment planning.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
September 18, 2019
Siemens Medical Solutions USA, Inc. % Ms. Veronica Padharia Regulatory Affairs Specialist 2501 N Barrington Road HOFFMAN ESTATES IL 60192
# Re: K192065
Trade/Device Name: syngo.via RT Image Suite Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: July 31, 2019 Received: August 1, 2019
# Dear Ms. Padharia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
# K192065
Device Name syngo.via RT Image Suite
#### Indications for Use (Describe)
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).
It provides tools to efficiently view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create and modify simple treatment plans. Contours, images and treatmently be exported to a Treatment Planning System.
The software combines following digital image processing and visualization tools:
· Multi-modality viewing and contouring of anatomical, functional, and multi-parametric images such as but not limited to CT. PET. PET/CT. MRI. Linac Cone Beam CT (CBCT) images and dose distributions
· Multiplanar reconstruction (MPR) thick, minimum intensity projection (MIP), volume rendering technique (VRT)
- · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- · Creation of contours on any type of images without prior assignment of a planning CT
· Manual and semi-automatic registration using rigid and deformable registration
· Supports the user in comparing, and adapting contours based on datasets acquired with different imaging modalities and at different time points
· Supports the user in comparing images and contours of different patients
- · Supports multi-modality image fusion
- · Visualization and contouring of moving tumors and organs
- · Management of points of interest including but not limited to the isocenter
- · Management of simple treatment plans
- · Generation of a synthetic CT based on multiple pre-define MR acquisitions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
K192065
# 510(K) SUMMARY FOR
# SYNGO.VIA RT IMAGE SUITE
Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: June 25, 2019
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
# I. Submitter
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Establishment Registration Number 2240869
#### Manufacturing Site
Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany
Establishment Registration Number 3004977335
## Contact Person
Veronica Padharia Regulatory Affairs Specialist, CNMT Siemens Medical Solutions, Inc. USA 2501 N. Barrington Road Hoffman Estates, IL 60192 Phone: (630) 877-5761 Fax: (847) 304-6023 email: veronica.padharia(@siemens-healthineers.com
# II. Device Name and Classification
Product Name: Propriety Trade Name: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
syngo.via RT Image Suite syngo.via RT Image Suite System, Planning, Radiation Therapy Treatment Radiology 21 CFR §892.5050 Class II MUJ
# III. Predicate Device
- Predicate Device: Trade Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section:
syngo.via RT Image Suite K173635 01/12/2018 System, Planning, Radiation Therapy Treatment Radiology 21 CFR §892.5050
{4}------------------------------------------------
Device Class: Class II Product Code: MUJ Recall Information:
There have been no recalls for this device
# IV. Device Description
The subject device with the current software version SOMARIS/8 VB40 is an image analysis software for viewing, manipulation, 3D and 4D visualization, comparison of medical images from multiple imaging modalities and for the segmentation of tumors and organs-at-risk, prior to dosimetric planning and response assessment in radiation therapy. syngo.via RT Image Suite combines routine and advanced digital image processing and visualization tools for easy manual and software assisted contouring of volumes of interest, identification of points of interest, sending isocenter points to an external laser system, registering images and exporting final results. syngo.via RT Image Suite supports the medical professional with tools to use during different steps in radiation therapy case preparation. At a high-level, the following features have been modified:
- 1. Beam Placement
- Reference Point Management 2.
- 3. Patient Marking
- 4. Contouring
- 5. Structure Set Management
# V. Indications for Use
syngo.via RT Image Suite is a 3D and 4D image visualization, multi-modality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).
It provides tools to efficiently view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. It also provides functionalities to create and modify simple treatment plans. Contours, images and treatment plans can subsequently be exported to a Treatment Planning System.
The software combines following digital image processing and visualization tools:
- . Multi-modality viewing and contouring of anatomical, functional, and multi-parametric images such as but not limited to CT, PET, PET/CT, MRI, Linac Cone Beam CT (CBCT) images and dose distributions
- Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
- Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- Creation of contours on any type of images without prior assignment of a planning CT .
- Manual and semi-automatic registration using rigid and deformable registration ●
- Supports the user in comparing, contouring, and adapting contours based on datasets acquired ● with different imaging modalities and at different time points
- . Supports the user in comparing images and contours of different patients
- Supports multi-modality image fusion ●
- Visualization and contouring of moving tumors and organs ●
- Management of points of interest including but not limited to the isocenter
- Management of simple treatment plans
- Generation of a synthetic CT based on multiple pre-define MR acquisitions ●
{5}------------------------------------------------
# SIEMENS
# VI. Comparison of Technological Characteristics with the Predicate Device
As with the predicate device, the subject device supports viewing, manipulation, 3D and 4D visualization, comparison of medical images from multiple imaging modalities and the segmentation of tumors and organs-at-risk, prior to dosimetric planning and response assessment in radiation therapy.
At a high-level a tabular summary of the subject and predicate device's technological differences is provided as Table 4 below for the software version SOMARIS/8 VB40:
| Feature | Comparison of the Subject Device to the Predicate Device |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Beam<br>Placement | Creation of new geometric treatment plans for photon radiotherapy. The syngo.via workflow<br>supports a more user intuitive workflow (i.e. a more modern look and feel).<br><br>Modification: Beam placement auto shape now also allows in creating rectangular block<br>fields. |
| Reference<br>Point<br>Management | Reference Point Management allows the user to edit the position, color, and type of Points of<br>Interest (POI).<br><br>Modification: The workflow for reference point creation has been simplified and a<br>streamlined workflow for breast iso-centering has been implemented. |
| Patient<br>Marking | Sending of reference points with offset details to a laser system.<br><br>Modification: Support of simplified workflow for breast iso-centering. Enabled by combining<br>improvements in iso-centering contouring the breast. It identifies the breast boundaries and<br>suggests a proper iso-center for breast treatment. |
| Routine<br>Contouring | Routine Contouring tools (e. g. freehand drawing tools, creation of margins etc.)<br><br>Modifications include capability of interpolating contours, 3D Pan/Scale options, and option<br>to add rectangular and ellipse as contour shapes, improvement in nudge tool to enable auto<br>filling of the inner part of a closed contour, and restricting the margin extension to external<br>contour. |
| Advanced<br>Contouring | Advanced Contouring tools (automatic contouring of different structures, nudge 3D tool,<br>contour interpolation etc.). Automatic contouring of pelvic and head/neck regions, support of<br>Rapid Results Technology.<br><br>Modification: This subject device provides the following extensions:<br>• Automatic Contouring can be applied on further structures (thoracic and abdominal<br>regions)<br>• A new deep learning-based approach that uses an adversarial network has been<br>implemented |
| Structure Set<br>Management | The Structure Set Management provides the following functionality:<br>• Loading and storing of DICOM RT structure sets, creating, editing and deletion of<br>structures and POIs.<br>• Creating, editing and deletion of structure templates.<br>• Customize predefined structure database with mapping to international nomenclature<br>schemes.<br><br>Modification includes simplifying the user interface. |
Table 1 Differences in Technical Characteristics
# VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Software Verification and Validation
Software Documentation for a Major Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claim of substantial equivalence.
{6}------------------------------------------------
# SIEMEN
## Non-Clinical Testing Summary
Non-clinical tests (integration and functional) were conducted for syngo.via RT Image Suite during product development. Performance tests were conducted to test the functionality of the syngo.via RT Image Suite. The modifications described in this Premarket Notification were supported with verification/validation testing. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.
### Clinical Testing Summary
The results of the evaluation show improved performance of the subject device syngo.via RT Image Suite (software version SOMARIS/8 VB40), compared to the predicate device syngo.via RT Image Suite (software version SOMARIS/8 VB30). The detailed report is provided in support of the device modifications.
The organ-at-risk (OAR) segmentation of the subject device consists of a region of interest detection based on anatomical landmarks, followed by a Deep Image-to-Image Network performing organ segmentation. Those both parts of the segmentation algorithm were validated separately using a testing set of 32 datasets.
The detection rate of the subject device improved for the OAR Brain, Liver, Kidney Left, and Kidney Right compared to the predicate device, achieving 100% for all evaluated organs.
The segmentation quality was assessed by comparing a manually annotated ground truth with the algorithm result using the distance measure Average Symmetric Surface (ASSD) and the overlap measure DICE coefficient. The mean ASSD computed for the subject device ranged between 0.64 mm (for Femur Head Right) and 3.04 mm (for Heart) and improved for all evaluated OAR. The mean DICE coefficient ranged between 0.85 (for Prostate and Rectum) and 0.97 (for Heart). The results for mean ASSD and mean DICE coefficient both demonstrate the improved segmentation quality provided by the subject device compared to the predicate device.
#### Standards
Siemens claims conformance to the following performance standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Publication<br>Date | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|------------------------------------------------------------------------------------------------------------|---------------------|------------------------------------------|
| 12-300 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set; PS 3.1 - 3.20 | 06/27/2016 | NEMA |
| 13-32 | Software | Medical Device Software -Software Life<br>Cycle Processes; 62304:2006 (1st Edition) | 08/20/2012 | AAMI, ANSI,<br>IEC |
| 5-40 | Software/<br>Informatics | Medical devices – Application of risk<br>management to medical devices; 14971<br>Second Edition 2007-03-01 | 08/20/2012 | ISO |
| 5-114 | General I<br>(QS/RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices<br>IEC 62366-1:2015 | 06/27/2016 | IEC |
## Summary
This subject device provides tools designed help the medical professional in contouring and evaluating volumes of interest, for example gross target volumes, or organs-at-risk. The fundamental software technology which is provided within the scope of the subject device is already cleared and remains unchanged in comparison to the predicate device. The Indications for Use for the subject device remains unchanged. The modifications described in this Premarket Notification were supported with verification and validation testing and clinical performance evaluation. The Risk analysis was completed and risk control implemented to mitigate identified hazards.
{7}------------------------------------------------
# SIEMENS
# General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. syngo.via RT Image Suite is designed to fulfill the requirements of the applicable safety and performance standards as listed above.
# VIII. Conclusions
The predicate device was cleared based on non-clinical testing including verification and validation, phantom tests, and supportive literature. The results of these tests demonstrate that the predicate device is adequate for the intended use. The subject device is also tested using the same methods as used for the predicate devices. The comparison of technological characteristics, non-clinical performance data, and software validation included in this submission demonstrates that the subject device is as safe and effective when compared to the predicate devices that are currently marketed for the same intended use. Since both devices were tested using the same methods, Siemens believes that the data generated from the syngo.via RT Image Suite testing supports a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.