K192053 · Tav Medical , Ltd. · DZE · Oct 29, 2019 · Dental
Device Facts
Record ID
K192053
Device Name
W Zirconia Implants
Applicant
Tav Medical , Ltd.
Product Code
DZE · Dental
Decision Date
Oct 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Indications for Use
TAV Medical's W Zirconia Implants are intended for surgical placement in the patient's upper and lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. The ø3.6mm reduced diameter implants are recommended for central and lateral incisors only.
Device Story
W Zirconia Implants are endosseous dental implants (one-piece or two-piece) designed for surgical placement in the jaw to support prosthetic restorations. The device consists of a Y-TZP zirconia body that osseo-integrates with alveolar bone, with various accessories including titanium/PEEK healing caps, cover screws, and titanium abutments. Used in dental clinics by clinicians; the device provides a foundation for artificial teeth. The system supports immediate loading protocols when primary stability is achieved. The device benefits patients by restoring chewing function and aesthetics. The two-piece system uses screw-retained abutments to connect the implant to the crown. The device is non-active and mechanical in nature.
Clinical Evidence
Bench testing only. No clinical data provided. Performance established via ISO 14801 dynamic fatigue testing, biocompatibility (ISO 10993-1, 10993-5), and mechanical wear testing of implant-to-abutment connections. Sterilization validated per ANSI/AAMI/ISO 11137-2 and 17665-1.
Technological Characteristics
Materials: Yttria-stabilized tetragonal zirconia (Y-TZP) per ISO 13356; Titanium Ti 6Al 4V ELI (ASTM F136-13) for abutments/screws; PEEK for healing/temporary caps. Manufacturing: Ceramic Injection Molding (CIM). Design: Tissue-level, macro/micro-rough surface topography. Sterilization: Gamma irradiation (implants), steam heat (abutments). Connectivity: Mechanical screw-retained connection.
Indications for Use
Indicated for patients requiring dental implants in the upper or lower jaw to support prosthetic devices (artificial teeth) and restore chewing function. Includes immediate loading if primary stability and occlusal conditions are met. ø3.6mm implants restricted to central and lateral incisors.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 29, 2019
TAV Medical Ltd. Revital Shabtai General Manager - Dental Division Dora Industrial Park, P.O. Box 88 Shlomi, 2283202 ISRAEL
Re: K192053
Trade/Device Name: W Zirconia Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 31, 2019 Received: July 31, 2019
Dear Revital Shabtai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar for Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K192053
Device Name W Zirconia Implants
#### Indications for Use (Describe)
TAV Medical's W Zirconia Implants are intended for surgical placement in the patient's upper and lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the wing function. The implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
The ø3.6mm reduced diameter implants are recommended for central and lateral incisors only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K192053
# 510(k) Summary TAV Medical Ltd. — W Zirconia Implants
# 1. Submitter Information
| Submitter's Name: | TAV Medical Ltd.<br>Dora Industrial Park<br>P.O.B 88, Shlomi 2283202<br>Israel |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Phone Number: +972 (4) 9808615/503 | |
| Fax Number: +972 (4) 9808356 | |
| Contact Person: Mrs. Revital Shabtai<br>E-mail: revitals@tavmedical.com | |
| Date: October 29, 2019 | |
# 2. Device
| Trade Name: | W Zirconia Implants |
|-------------------------|------------------------------------|
| Common Name: | Endosseous Dental Implant |
| | Endosseous Dental Implant Abutment |
| Classification Name: | 21 CFR 872.3640 |
| Device Class: | Class II |
| Primary Product Code: | DZE |
| Secondary Product Code: | NHA |
# 3. Predicate Devices
| Primary Predicate: | K172668 W Zirconia Implants |
|--------------------|-----------------------------------------|
| Reference Devices: | K151328 Pure Ceramic Implants |
| | K180477 Straumann Pure Ceramic Implants |
{4}------------------------------------------------
# 4. Device Description
TAV Medical's W Zirconia Implants are dental implants, composed of the following implant models:
W One Piece - monotype implant with integrated abutment. W Two Piece – implant for screw retained Abutment.
The implants are tissue level designed and includes a body portion and a neck (1.8mm). The implant body portion is configured to extend into the bone and osseo-integrate with the alveolar bone. The neck should be positioned 1.8mm above the bone level.
TAV Medical's W Zirconia Implants are made of Yttria stabilized tetragonal zirconia (Y-TZP). This material conforms with ISO 13356:2015 standard for Implants for surgery - Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
The Titanium abutments are going through anodizing process using an electrolytic process that adjusts the oxide level of the metal surface. This adjustment changes the spectrum of light, resulting in perceived color. By controlling the surface oxide level, an entire range of colors can be achieved.
# The Subject Device includes the following dimensions:
W One-Piece implants: Diameter of 3.6mm for lengths of 8mm, 10mm, 12mm and 14mm and; Diameters of 4.1mm and 4.8mm for length of 14mm. W Two-Piece implants: Diameters of 4.1mm and 4.8mm for lengths of 8mm, 10mm, 12mm and 14mm.
Cover Screw are screwed into the implant to protect the inner configuration of the implant during the healing phase in cases of submucosal healing protocol. Cover screws are made of Titanium alloy Ti 6Al 4V ELI & Anodize and are available in diameters of 4.1mm and 4.8mm.
Titanium Healing Caps are intended to protect the 2-piece implants during the healing phase. The healing caps also support the emergence profile and keep the implant shoulder ideal for the impression phase. The healing caps are available in different geometrical features such as height and diameter. The Titanium Healing Caps manufactured from Ti 6Al 4V ELI & Anodize and are available in diameters
{5}------------------------------------------------
of 4.1mm and 4.8mm. Each diameter is provided with height dimensions of 1mm, 2mm, 3mm and 4mm.
PEEK Healing Caps are designed to protect the two-piece implant during the healing phase. The healing caps also support the emergence profile and keep the implant shoulder ideal for the impression phase. Peek healing caps are available in different geometrical features such as height and diameter and available in 2 different designs, one design for the one-piece implant and second design for the two-piece implant. Both designs are manufactured from PEEK material and are indicated to be placed in patients' mouth for a maximum duration of 180 days. The PEEK healing caps are available in diameters of 4.1mm and 4.8mm and height dimensions of 1mm, 2mm, 3mm and 4mm.
PEEK Temporary Caps serves as a basis for temporary restoration. TAV Medical temporary restoration caps available in two configurations, temporary Cap for Crown and Temporary Cap for Bridge. The caps are manufactured from PEEK material and are indicated to be placed in patients' mouth for a maximum duration of 180 days. The PEEK temporary caps are available in diameter of 3.6mm.
Titanium Abutments for the W Two Piece Zirconia Implants are intended for use as an adapter between the implant and the crown. The abutments are characterized by distinct geometrical features such as length and angulation. The titanium abutments are manufactured from ASTM F136-13 compatible Titanium Ti 6Al 4V ELI & Anodize. TAV Medical abutments are intended to be connected to 2-piece W Zirconia implants with titanium screw, to provide support for prosthetic reconstructions. The abutments are available in the following dimensions: Length of 5.0mm and 6.0mm and Angles of 0°, 5°, 10° and 15°. The 0º serves also for bridges.
#### 5. Indications for Use
TAV Medical's W Zirconia Implants are intended for surgical placement in the patient's upper and lower jaw to provide support for prosthetic devices, such as artificial teeth and in order to restore the patient chewing function. The implants are indicated for immediate loading when qood primary stability is achieved and with appropriate occlusal loading.
The ø3.6mm reduced diameter implants are recommended for central and lateral incisors only.
{6}------------------------------------------------
## 6. Technological Characteristics
The proposed W Zirconia Implants have similar indications for use, technological characteristics, mode of operation and performance specifications as its predicate devices listed.
The following abutments interacts with the two-piece implants as follows:
Cover Screw protect the inner configuration of the implant during the healing phase in cases of submucosal healing protocol.
Titanium Healing Caps Protect the implant during the healing phase, and maintain, stabilize and form the soft tissue during the healing process.
PEEK Healing Caps Protect the implant during the healing phase up to 180 days, and maintain, stabilize and form the soft tissue during the healing process.
PEEK Temporary Caps Serves as a basis for temporary restoration for crown or bridge. Up to 180 days.
Titanium Abutments for the W Two Piece Zirconia Implants are intended for use as an adapter between the implant and the crown.
The following tables summarize the equivalence comparison between TAV Medical's W Zirconia Implants systems and its primary predicate and reference devices:
{7}------------------------------------------------
# Implants:
| Owner: | TAV Medical Ltd. | TAV Medical Ltd. | Institute<br>Straumann AG | Institute Straumann<br>AG |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| Trade Name | W One Piece<br>Zirconia implant | W One Piece<br>Zirconia implant | Pure Ceramic Implants | -- |
| | W Two Piece<br>Zirconia Implant | W Two Piece<br>Zirconia Implant | -- | Straumann Pure<br>Ceramic Implant<br>systems |
| 510(k) Number | K192053 | Cleared under:<br>K172668 | Cleared under:<br>K151328 | Cleared under:<br>K180477 |
| Product Code | Primary: DZE<br>Secondary: NHA | Primary: DZE<br>Secondary: NHA | Primary: DZE<br>Secondary: NHA | Primary: DZE<br>Secondary: NHA |
| Manufacturer | TAV Medical Ltd. | TAV Medical Ltd. | Institute Straumann AG | Institute Straumann<br>AG |
| Intended Use /<br>Indications for<br>Use | TAV Medical's W<br>Zirconia<br>Implants are<br>intended for<br>surgical<br>placement in the<br>patient's upper<br>and lower jaw<br>to provide<br>support for<br>prosthetic<br>devices, such as<br>artificial teeth<br>and in order to<br>restore the<br>patient chewing<br>function. The<br>implants are<br>indicated for<br>immediate<br>loading when<br>good primary<br>stability is<br>achieved and<br>with appropriate<br>occlusal loading.<br>The ø3.6 mm<br>reduced<br>diameter<br>implants are<br>recommended<br>for central and<br>lateral incisors<br>only. | TAV Medical's W<br>Zirconia Implants<br>are intended for<br>surgical placement<br>in the patient's<br>upper and lower<br>jaw to provide<br>support for<br>prosthetic devices,<br>such as artificial<br>teeth and in order<br>to restore the<br>patient chewing<br>function. The<br>implants are<br>indicated for<br>immediate loading<br>when good primary<br>stability is achieved<br>and with<br>appropriate<br>occlusal loading. | The Straumann® PURE<br>Ceramic Implant<br>(Monotype) is indicated<br>for restoration in single<br>tooth gaps and in an<br>edentulous or partially<br>edentulous jaw. The<br>prosthetic restorations<br>used are single crowns,<br>fixed partial or full<br>dentures, which are<br>connected to the<br>implants through the<br>corresponding<br>components.<br>The ø3.3 mm reduced<br>diameter implants are<br>recommended for<br>central and lateral<br>incisors only.<br>The Straumann® PURE<br>Ceramic Implant<br>Protective Cap is<br>intended to protect the<br>Straumann® PURE<br>Ceramic Implant<br>(Monotype) during the<br>healing phase after<br>implant placement for<br>up to 6 months.<br>Temporary copings are<br>intended to serve as a<br>base for temporary<br>crown or bridge<br>restoration for the<br>Straumann® PURE<br>Ceramic Implant<br>(Monotype) for up to 30<br>days. | Straumann PURE<br>Ceramic Implant:<br>The Straumann<br>PURE Ceramic<br>Implant is indicated<br>for the restoration of<br>single-tooth gaps<br>and in edentulous or<br>partially edentulous<br>jaws. The prosthetic<br>restorations used are<br>single crowns, fixed<br>partial or full<br>dentures, which are<br>connected to the<br>implants through the<br>corresponding<br>compnents.<br>Closure and healing<br>caps:<br>Closure and Healing<br>caps are intended for<br>use with the<br>Straumann Dental<br>Implant System<br>(SDIS) to protect the<br>inner<br>configuration of the<br>implant and<br>maintain, stabilize<br>and form the soft<br>tissue during the<br>healing process.<br>Closure and<br>Healing caps should<br>be used only with<br>suitable implant<br>connections. Do not<br>use healing<br>components for<br>longer than 6<br>months.<br>Temporary<br>Abutments:<br>The provisional<br>components are<br>intended to serve as<br>a base for temporary<br>crown or bridge<br>restoration out of<br>occlusion for<br>the Straumann®<br>PURE Ceramic<br>Implant System. The<br>Straumann®<br>Temporary |
| | | | | |
| Owner: | TAV Medical Ltd. | TAV Medical Ltd. | Institute<br>Straumann AG | Institute Straumann<br>AG |
| Feature | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| | | | | Abutment VITA CAD-<br>Temp® for the<br>Straumann® PURE<br>Ceramic Implant is<br>indicated for<br>temporary usage of<br>up to 180 days.<br>CI RD Straumann<br>PUREbase<br>Abutments:<br>CI RD Straumann<br>PUREbase abutment<br>is a titanium base<br>placed onto<br>Straumann ceramic<br>dental implants to<br>provide support for<br>customized<br>prosthetic<br>restorations and is<br>indicated for screw-<br>retained single tooth<br>or cement-retained<br>single tooth and<br>bridge restorations.<br>All digitally designed<br>copings and/or<br>crowns for use with<br>the Straumann®<br>Variobase Abutment<br>system are intended<br>to be sent to<br>Straumann for<br>manufacture at a<br>validated milling<br>center. |
| Material | yttria-stabilized<br>zirconia<br>(Y-TZP) | yttria-stabilized<br>zirconia<br>(Y-TZP) | yttria-stabilized<br>zirconia<br>(Y-TZP) | yttria-stabilized<br>zirconia<br>(Y-TZP) |
| Manufacturing<br>Technology | CIM: Ceramic<br>injection<br>molding | CIM: Ceramic<br>injection molding | Turning | Turning |
| Diameter (mm) | W two Piece<br>Zirconia implant<br>Ø 4.1, 4.8 | W two Piece<br>Zirconia implant<br>Ø 4.1, 4.8 | -- | Ø 4.1 |
| | W one piece<br>Zirconia implant<br>Ø 3.6 | W One Piece<br>Zirconia implant<br>Ø 4.1, 4.8 | Ø 3.3, 4.1 | -- |
| Length (mm) | W two piece<br>Zirconia implant | W two piece<br>Zirconia implant | -- | 8, 10, 12, 14 |
| Owner: | TAV Medical Ltd. | TAV Medical Ltd. | Institute<br>Straumann AG | Institute Straumann<br>AG |
| Feature | Subject Device | Primary Predicate<br>Device | Reference Device | Reference Device |
| | W One Piece<br>Zirconia implant<br>8, 10, 12, 14 | W One Piece<br>Zirconia implant<br>8, 10 ,12 | One Piece<br>8, 10, 12, 14 | -- |
| Design | W Two Piece<br>Zirconia Implant | W Two Piece<br>Zirconia Implant | -- | Two-piece |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.