← Product Code [NQW](/productcode/NQW) · K191927

# Hinged Laminoplasty System (K191927)

_Life Spine, Inc. · NQW · Sep 20, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191927

## Device Facts

- **Applicant:** Life Spine, Inc.
- **Product Code:** [NQW](/productcode/NQW.md)
- **Decision Date:** Sep 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Life Spine Hinged Laminoplasty System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Hinged Laminoplasty System is used to hold or buttresses the allograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

## Device Story

System consists of titanium alloy plates and screws; used in lower cervical/upper thoracic spine (C3-T3) following laminoplasty. Plates feature screw holes for bone attachment and a center hole for affixing allograft or autograft material. Device functions as a mechanical buttress to stabilize graft material, preventing expulsion or spinal cord impingement. Used by surgeons in clinical settings. Implants are single-use only.

## Clinical Evidence

Bench testing only. Performance demonstrated via screw pull-out testing per ASTM F543 to establish substantial equivalence to the predicate device (K181717).

## Technological Characteristics

Titanium alloy 6AL-4V-ELI per ASTM F-136. System comprises plates and screws. Mechanical fixation via screw-to-bone and plate-to-graft interface. Single-use. No software or electronic components.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Life Spine Laminoplasty System ([K181717](/device/K181717.md))
- Synthes Arch ([K032534](/device/K032534.md))
- Choice Spine Laminoplasty Fixation System ([K173215](/device/K173215.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

September 20, 2019

Life Spine, Inc. Angela Batker RA/QA Specialist 13951 South Quality Drive Huntley, Illinois 60142

Re: K191927

Trade/Device Name: Hinged Laminoplasty System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: NQW Dated: July 16, 2019 Received: July 19, 2019

### Dear Ms. Batker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191927

Device Name Hinged Laminoplasty System

#### Indications for Use (Describe)

The Life Spine Hinged Laminoplasty System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Hinged Laminoplasty System is used to hold or butresses the allograft or autograft material in place to prevent the graft material from expulsion or impinging the spinal cord.

| Type of Use (Select one or both, as applicable)                                  |                                                                      |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary Life Spine Hinged Laminoplasty System

| Submitted By:         | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118                  |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:       | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 |
| Date Prepared:        | September 12, 2019                                                                                                               |
| Trade Name:           | Life Spine Hinged Laminoplasty System                                                                                            |
| Common Name:          | Spinal Interlaminal Fixation Orthosis                                                                                            |
| Classification:       | NQW, CFR 888.3050, Class II                                                                                                      |
| Primary Predicate:    | Life Spine Laminoplasty System (K181717)                                                                                         |
| Additional Predicate: | Synthes Arch (K032534)<br>Choice Spine Laminoplasty Fixation System (K173215)                                                    |

### Device Description:

The Life Spine Hinged Laminoplasty System consists of various plates and screw configurations to address surgeon and patient needs. The plates have screw holes at the ends that allow attachment to bone. The center hole of the plate allows it to be affixed to allograft or autograft material. All components are fabricated and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136.

All implants are intended for single use only and should not be reused under any circumstances. Do not use any of the Life Spine Hinged Laminoplasty System components with components from any other system or manufacturer. The Life Spine Laminoplasty System components should never be reused under any circumstances.

Intended Use of the Device:

{4}------------------------------------------------

The Life Spine Hinged Laminoplasty System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Life Spine Hinged Laminoplasty System is used to hold or buttresses the allograft material in place in order to prevent the graft material from expulsion or impinging the spinal cord.

## Technological Characteristics:

The Life Spine Hinged Laminoplasty System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.

# Material:

This submission seeks clearance of a device made from implant grade (Ti-6AL-4V) Titanium alloy according to F136. This this is the same material used in the predicate devices.

# Performance Data:

Testing according to Screw Pull-Out testing to ASTM F543 was presented to demonstrate the substantial equivalency of the Life Spine Laminoplasty (K181717).

# Substantial Equivalence:

The Life Spine Hinged Laminoplasty System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.

# Conclusion:

The information presented demonstrates the substantial equivalency of The Life Spine Hinged Laminoplasty System.

---

**Source:** [https://fda.innolitics.com/device/K191927](https://fda.innolitics.com/device/K191927)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).