The Thermo Scientific Sensititre ARIS HiQ System is part of the Sensititre AST system and is an automated plate management device containing an incubator and embedded OptiRead module. The Thermo Scientific Sensititre ARIS HiQ System is designed for use with the Thermo Scientific Sensititre SWIN Software System. The ARIS HiQ and the SWIN system work together to read Sensititre (18-24 hr) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results for non-fastidious microorganisms.
Device Story
The Thermo Scientific Sensititre ARIS HiQ System is an automated plate management device for antimicrobial susceptibility testing (AST). It functions as an incubator and reader for Sensititre susceptibility plates. The device integrates with the Sensititre SWIN Software System to process plates containing non-fastidious microorganisms. The system reads plates after 18-24 hours of incubation using an embedded OptiRead module. The resulting data is processed by the SWIN software to generate MIC values and interpret breakpoint results. Used in clinical laboratory settings by trained laboratory personnel. The output assists healthcare providers in determining appropriate antimicrobial therapy for patients by identifying microbial susceptibility profiles.
Clinical Evidence
No clinical data provided. The submission relies on bench testing and performance validation of the automated incubation and reading system to demonstrate substantial equivalence.
Technological Characteristics
Automated plate management system; incubator; embedded OptiRead fluorescence reader. Connectivity: USB to external PC. Dimensions: 100-plate capacity. Software: Sensititre SWIN Software v3.4. Sterilization: Not applicable (instrument). Sensing: Fluorescence-based detection of microbial growth in microtiter wells.
Indications for Use
Indicated for use with Sensititre (18-24 hr) susceptibility plates to generate MIC and interpret BP results for non-fastidious Gram-positive and Gram-negative microorganisms.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Submission Summary (Full Text)
{1}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K191914
B Applicant
Thermo Fisher Scientific
C Proprietary and Established Names
Thermo Scientific Sensititre ARIS HiQ System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LRG | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To demonstrate substantial equivalence of the Thermo Scientific Sensititre ARIS HiQ System to the predicate Sensititre ARIS Module, used in conjunction with the OptiRead, to read Sensititre (18-24 hour) susceptibility plates for non-fastidious Gram-positive and Gram-negative microorganisms.
### B Measurand:
Used in conjunction with the embedded OptiRead module and Sensititre SWIN Software System to automatically read Sensititre MIC and Breakpoint (BP) Susceptibility test panels and interpret the antimicrobial susceptibility test results for non-fastidious Gram-positive and Gram-negative microorganisms.
K191914 - Page 1 of 20
{2}
C Type of Test:
Automated plate management system containing an incubator and embedded OptiRead module, which is an automated fluorescence-based detection instrument used to read Sensititre Susceptibility plates.
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Thermo Scientific Sensititre ARIS HiQ System is part of the Sensititre AST system and is an automated plate management device containing an incubator and embedded OptiRead module. The Thermo Scientific Sensititre ARIS HiQ System is designed for use with the Thermo Scientific Sensititre SWIN Software System. The ARIS HiQ and the SWIN system work together to read Sensititre (18-24 hr) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results for non-fastidious microorganisms.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
The Thermo Scientific Sensititre ARIS HiQ System can only be used with version 3.4 of the Thermo Scientific Sensititre SWIN Software System.
IV Device/System Characteristics:
A Device Description:
The Thermo Scientific Sensititre ARIS HiQ System is an automated plate management device containing an incubator and an embedded OptiRead reader. In its current version, it is designed for use with Sensititre SWIN Software System (version 3.4). The ARIS HiQ system handles plate incubation and management. The ARIS HiQ and SWIN Software System work together to read and interpret up to 100 Sensititre plates, generating minimum inhibitory concentration (MIC) results and corresponding categorical interpretation based on breakpoints (BP) information within the SWIN software for non-fastidious Gram-negative and non-fastidious Gram-positive microorganisms.
B Principle of Operation:
The Thermo Scientific ARIS HiQ System is for use with Sensititre (18-24 hr) Susceptibility plates which are micro-versions of the classic broth dilution method for providing qualitative and quantitative susceptibility results in a dried plate format. Each plate (panel) contains a fluorogenic substrate to monitor growth and antimicrobial agents at specific two-fold dilutions to achieve drug concentrations over a specified range. When microbial metabolism and growth occurs, fluorescence is released from plate wells in amounts directly related to the activity of bacterial
K191914 - Page 2 of 20
{3}
growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism.
To use the ARIS HiQ, plates are inoculated with a standardized suspension of microorganism using the Sensititre AutoInoculator/AIM instrument, covered, and then loaded onto the ARIS HiQ for automated incubation and reading. After the incubation time indicated for each susceptibility panel, the microtiter plates are automatically moved by the robotic arm component of the ARIS HiQ to the embedded OptiRead module where the fluorescence count values are measured from each plate well. The fluorescence count values are transmitted to the SWIN software on an external PC for processing and interpretation of MIC and BP results. No result processing is performed within the ARIS HiQ.
C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name:
Thermo Scientific Sensititre ARIS HiQ System
2. Specimen Identification:
Specimen (Sensititre plate) identification information should be manually entered into the SWIN Software, followed by scanning the plate barcode. If plates have not yet been loaded in the ARIS HiQ, plate identification information can be sent from the SWIN to the ARIS HiQ. If plates have been loaded in the ARIS HiQ, the SWIN Software will automatically send plate identification information to the ARIS HiQ.
3. Specimen Sampling and Handling:
Sensititre plates inoculated with pure cultures can be loaded and unloaded from the ARIS HiQ in one of two ways: (1) individually, interacting with a rack mounted to the turntable at the front of the ARIS HiQ, or (2) in batches, using transportable microplate racks to simultaneously load, unload and handle up to ten microplates at a time.
4. Calibration:
All calibrations and adjustments to the ARIS HiQ System are performed by Sensititre system-trained service engineers during installation and regular servicing. No adjustments or calibrations are required by the end user during day-to-day operations.
5. Quality Control:
K191914 - Page 3 of 20
{4}
QC organisms were tested as appropriate for each drug and in accordance with CLSI recommendations. For results, see Section VII.A.5.
V Substantial Equivalence Information:
A Predicate Device Name(s): Sensititre Aris Module
B Predicate 510(k) Number(s): K911419
C Comparison with Predicate(s):
| Similarities | | |
| --- | --- | --- |
| Device & Predicate Device(s): | ARIS HiQ K191914 (Device) | Sensititre ARIS Module K911419 (Predicate) |
| Intended Use | Automated plate handling and incubator instrument, used in conjunction with a plate reader and external software to read Sensititre (18-24 hour) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results | Same |
| Plate Compatibility | FDA cleared Sensititre susceptibility plates designed for automated reading | Same |
| Barcode Reading | • Linear barcode reading capability on microtiter plates
• The entire plate inventory is read when plates are loaded/ unloaded | Same |
| Reader Unit | OptiRead | Same |
| Incubation Temperature | 35°C ±1°C | Same |
| System Control & Software | Embedded processor running firmware | Same |
| Results Processing | Handled by established algorithms in SWIN system software | Same |
| Differences | | |
| Device & Predicate Device(s): | ARIS HiQ K191914 (Device) | Sensititre ARIS Module K911419 (Predicate) |
K191914 - Page 4 of 20
{5}
VI Standards/Guidance Documents Referenced:
- Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA, August 2009.
- CLSI M100-28th ed., Performance Standard for Antimicrobial Susceptibility Testing; Approved Standard, Twenty-eighth Edition, January 2018.
- CLSI M07, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard, Tenth Edition, January 2015.
| Plate Loading, Unloading & Handling | Plates are loaded, unloaded and handled in one of two ways:
(1) Individually, on a plate rack mounted to a turntable;
(2) In batches, using transportable plate racks to simultaneously load, unload and handle up to ten plates at a time | Plates are loaded, unloaded and handled individually, on a fixed plate carousel |
| --- | --- | --- |
| Incubation Control | Five heater chimneys with a fan for each chimney to provide heated air circulation to the microplate racks | Heated microplate carousel and a single heater chimney (no fan) |
| Reader Unit Implementation | OptiRead is fully integrated into the ARIS HiQ | OptiRead is a separate instrument which plugs into the Sensititre ARIS module beneath the instrument |
| User Interface | Touch screen display with icon driven graphical user interface | Seven segment display with keypad |
| SWIN interface/communication | USB | RS-232 |
| Plate Capacity | 100 | 64 |
| Automated Plate Organization | Shuffle function to automatically relocate plates within the instrument during incubation and to consolidate plates for removal | No automated plate relocation function |
K191914 - Page 5 of 20
{6}
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
# 1. Precision/Reproducibility:
A reproducibility study was conducted using a Gram-positive non-fastidious panel that included 25 isolates and a Gram-negative non-fastidious panel consisting of 25 isolates. The isolates tested are listed in Table 1.
Table 1. Non-Fastidious Isolates Evaluated in the Reproducibility Study
| Organisms (# Tested) | |
| --- | --- |
| Gram-Positive Isolates | Gram-Negative Isolates |
| Staphylococcus aureus (9) | Escherichia coli (6) |
| Enterococcus faecalis (6) | Klebsiella pneumoniae (5) |
| Staphylococcus epidermidis (1) | Enterobacter cloacae (3) |
| Staphylococcus lugdunesis (1) | Serratia marcescens (3) |
| Staphylococcus haemolyticus (1) | Acinetobacter spp. (3) |
| Staphylococcus saprophyticus (1) | Pseudomonas aeruginosa (4) |
| Staphylococcus hominis (1) | Aeromonas hydrophila (1) |
| Enterococcus raffinosus (1) | |
| Enterococcus gallinarum (1) | |
| Enterococcus faecium (3) | |
The Gram-positive isolates listed above were tested against the following antimicrobial agents: Chloramphenicol, Oxacillin, Penicillin, Vancomycin, Piperacillin/Tazobactam, Ciprofloxacin, Clindamycin, Tetracycline, Dalbavancin, Telavancin, Daptomycin, Linezolid, Tedizolid, Gentamicin, Erythromycin, Ceftaroline and Quinupristin/Dalfopristin. The antimicrobial agents were selected as representatives of the major antimicrobial classes.
The Gram-negative isolates listed above were tested against the following antimicrobial agents: Ampicillin, Ticarcillin, Piperacillin/Tazobactam, Chloramphenicol, Cefazolin, Cefepime, Ceftazidime, Ertapenem, Tetracycline, Levofloxacin, Gentamicin, and Trimethoprim/Sulfamethoxazole (TMP/SMX). The antimicrobial agents were selected as representatives of the major antimicrobial classes.
Testing was performed at one site over three days by three operators using Sensititre susceptibility plates on three separate ARIS HiQ instruments. Percent reproducibility was calculated as the number of isolates with MIC values falling within $+/-$ one doubling dilution from the mode MIC value.
Review of the reproducibility study data revealed two antibiotic/Gram-positive organism combinations that lacked a sufficient number of on-scale MIC results. The antibiotics included the following:
- Gram-positive reproducibility study - Clindamycin and Quinupristin/Dalfopristin
K191914 - Page 6 of 20
{7}
To resolve this issue, an additional modified reproducibility study was conducted which included the testing of 10 isolates (with known on-scale results), in triplicate with three different operators; performed on three different days; using three separate ARIS HiQ instruments.
In summary, despite a number of isolates with off-scale MIC results, the percent reproducibility for all antibiotic/organism combinations tested met the acceptance criteria of >95% based on the "best-case" percent calculation as described in the AST Special Controls Guidance Document.
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality Control (QC) Testing. The FDA and CLSI recommended QC strains for Gram-negative (E. coli ATCC 25922, E. coli ATCC 35218, and P. aeruginosa ATCC 27853) and Gram-positive (S. aureus ATCC 29213, S. aureus ATCC BAA-977, E. faecalis ATCC 29212, and E. faecalis ATCC 51299) organisms were tested at two sites using both the ARIS HiQ and legacy ARIS (ARIS 2X), used with the OptiRead instrument. Twenty replicates were conducted per site for each QC strain/antimicrobial combination. For all antimicrobial/QC organisms evaluated, at least one QC organism exhibited acceptable results (≥95% in-range results), except for Cefepime tested with E. coli ATCC 25922 and P. aeruginosa ATCC 27853. This was attributed to not including the full CLSI/FDA recommended dilution range for QC testing with these two organisms.
Results are summarized in Tables 2 – 7 by QC strain, while QC results for macrolide-inducible resistance to clindamycin (D-test) and the Cefoxitin Screen are illustrated in Table 8 and Table 9, respectively. Shading in the tables indicate the expected concentration range or expected result.
| Table 2. Quality Control
E. coli ATCC 25922 | Concentration
(μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Ampicillin
Expected MIC Range:
2-8 μg/mL | 1 | | |
| | 2 | | |
| | 4 | 39 | 40 |
| | 8 | | |
| | 16 | 1 | |
| Ceftazidime^{a} | ≤0.25 | 34 | 34 |
K191914 - Page 7 of 20
{8}
K191914 - Page 8 of 20
| Table 2. Quality Control
E. coli ATCC 25922 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Expected MIC Range
0.06-0.5 μg/mL | 0.5 | 5 | 6 |
| | 1 | 1 | |
| Chloramphenicol
Expected MIC Range
2-8 μg/mL | 1 | | |
| | 2 | | |
| | 4 | 38 | 40 |
| | 8 | 2 | |
| | 16 | | |
| Ertapenem
Expected MIC Range
0.004-0.016 μg/mL | 0.002 | | |
| | 0.004 | | |
| | 0.008 | 38 | 40 |
| | 0.016 | 1 | |
| | >0.03 | 1 | |
| Cefazolin
Expected MIC Range
1-4 μg/mL | ≤1 | | |
| | 2 | 39 | 40 |
| | 4 | 1 | |
| | 8 | | |
| Cefepime^{b}
Expected MIC Range
0.016-0.12 μg/mL | ≤0.5 | 40 | 40 |
| | 1 | | |
| Gentamicin
Expected MIC Range
0.25-1 μg/mL | 0.12 | | |
| | 0.25 | | |
| | 0.5 | 32 | 32 |
| | 1 | 8 | 8 |
| | 2 | | |
| Levofloxacin
Expected MIC Range
0.008-0.06 μg/mL | 0.004 | | |
| | 0.008 | | |
| | 0.015 | 33 | 32 |
| | 0.03 | 7 | 8 |
| | 0.06 | | |
| | 0.12 | | |
| Piperacillin/Tazobactam^{c}
Expected MIC Range
1/4-4/4 μg/mL | 0.5 | | |
| | 1 | | |
| | 2 | 39 | 40 |
| | 4 | | |
| | 8 | 1 | |
| Trimethoprim/Sulfamethoxazole
Expected MIC Range
≤0.5/9.5 μg/mL | ≤0.25/4.75 | 40 | 40 |
| | 0.5/9.5 | | |
| | 1/19 | | |
| Tetracycline
Expected MIC Range
0.5-2 μg/mL | 0.25 | | |
| | 0.5 | | |
| | 1 | 40 | 40 |
| | 2 | | |
| | 4 | | |
| Ticarcillin
Expected MIC Range
4-16 μg/mL | 2 | | |
| | 4 | | |
| | 8 | 36 | 38 |
| | 16 | 3 | 1 |
{9}
K191914 - Page 9 of 20
| Table 2. Quality Control
E. coli ATCC 25922 | Concentration
(μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| | ≥32 | 1 | 1 |
a The Gram-negative panel did not include the full CLSI/FDA recommended dilution range for QC testing of this drug with this organism. However, at least one other QC organism was tested using the full CLSI/FDA recommended dilution range for this drug and exhibited acceptable results (≥95% in-range results)
b The Gram-negative panel did not include the full CLSI/FDA recommended dilution range for QC testing of this drug with this organism
c Tazobactam concentration was fixed at 4 ug/mL
| Table 3. Quality Control
E. coli ATCC 35218 | Concentration
(μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Ampicillin
Expected MIC Range
>32 μg/mL | 32 | | |
| | >32 | 40 | 40 |
| Piperacillin/Tazobactam^{a, b}
Expected MIC Range
0.5/4-2/4 μg/mL | ≤0.5 | 24 | 25 |
| | 1 | 16 | 14 |
| | 2 | | 1 |
| | 4 | | |
a The Gram-negative panel did not include the full CLSI/FDA recommended dilution range for QC testing of this drug with this organism. However, at least one other QC organism was tested using the full CLSI/FDA recommended dilution range for this drug and exhibited acceptable results (≥95% in-range results)
b Tazobactam concentration was fixed at 4 ug/mL
| Table 4. Quality Control
P. aeruginosa ATCC 27853 | Concentration
(μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Ceftazidime
Expected MIC Range
1-4 μg/mL | 0.5 | | |
| | 1 | 37 | 34 |
| | 2 | 2 | 6 |
| | 4 | 1 | |
| | ≥8 | | |
| Ertapenem
Expected MIC Range
2-8 μg/mL | ≤1 | | |
| | 2 | 12 | 14 |
| | 4 | 28 | 26 |
| | 8 | | |
| Cefepime^{a}
Expected MIC Range
0.5-4 μg/mL | ≤0.5 | 9 | 8 |
| | 1 | 30 | 29 |
| | 2 | 1 | 3 |
| | 4 | | |
| | 8 | | |
| | 16 | | |
| Gentamicin
Expected MIC Range
0.5-2 μg/mL | 0.25 | | |
| | 0.5 | | 1 |
| | 1 | 40 | 39 |
| | 2 | | |
{10}
K191914 - Page 10 of 20
| Table 4. Quality Control P. aeruginosa ATCC 27853 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| | ≥4 | | |
| Levofloxacin Expected MIC Range 0.5-4 μg/mL | 0.25 | 1 | 1 |
| | 0.5 | 21 | 24 |
| | 1 | 18 | 15 |
| | 2 | | |
| | 4 | | |
| | 8 | | |
| Piperacillin/Tazobactam Expected MIC Range 1/4-8/4 μg/mL | 0.5 | | |
| | 1 | | 3 |
| | 2 | 32 | 30 |
| | 4 | 8 | 7 |
| | 8 | | |
| | 16 | | |
| Trimethoprim/Sulfamethoxazole Expected MIC Range 8/152-32/608 μg/mL | ≤4/76 | | |
| | 8/152 | 11 | 10 |
| | 16/304 | 28 | 29 |
| | 32/608 | 1 | 1 |
| | >32/608 | | |
| Tetracycline Expected MIC Range 8-32 μg/mL | 4 | | |
| | 8 | 36 | 34 |
| | 16 | 4 | 6 |
| | 32 | | |
| | >32 | | |
| Ticarcillin Expected MIC Range 8-32 μg/mL | 4 | | |
| | 8 | | |
| | 16 | 36 | 37 |
| | 32 | 4 | 3 |
| | 64 | | |
a The Gram-negative panel did not include the full CLSI/FDA recommended dilution range for QC testing of this drug with this organism
b Tazobactam concentration was fixed at 4 ug/mL
| Table 5. Quality Control S. aureus ATCC 29213 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Chloramphenicol Expected MIC Range 2-16 μg/mL | 1 | | |
| | 2 | | |
| | 4 | 4 | 1 |
| | 8 | 36 | 39 |
| | 16 | | |
| | >16 | | |
| Ciprofloxacin Expected MIC Range 0.12-0.5 μg/mL | ≤0.12 | | |
| | 0.25 | 28 | 23 |
| | 0.5 | 12 | 17 |
| | 1 | | |
| Clindamycin Expected MIC Range 0.06-0.25 μg/mL | ≤0.12 | 40 | 40 |
| | 0.25 | | |
| | 0.5 | | |
{11}
K191914 - Page 11 of 20
| Table 5. Quality Control
S. aureus ATCC 29213 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Daptomycin
Expected MIC Range
0.12-1 μg/mL | ≤0.12 | 2 | |
| | 0.25 | 18 | 21 |
| | 0.5 | 20 | 19 |
| | 1 | | |
| | 2 | | |
| Erythromycin
Expected MIC Range
0.25-1 μg/mL | ≤0.12 | 1 | |
| | 0.25 | 5 | 8 |
| | 0.5 | 34 | 32 |
| | 1 | | |
| | 2 | | |
| Gentamicin
Expected MIC Range
0.12-1 μg/mL | ≤0.12 | | |
| | 0.25 | 28 | 30 |
| | 0.5 | 12 | 10 |
| | 1 | | |
| | 2 | | |
| Linezolid
Expected MIC Range
1-4 μg/mL | 0.5 | | |
| | 1 | 2 | 2 |
| | 2 | 29 | 38 |
| | 4 | 9 | |
| | >4 | | |
| Oxacillin^{a}
Expected MIC Range
0.12-0.5 μg/mL | ≤0.25 | 4 | 4 |
| | 0.5 | 36 | 36 |
| | 1 | | |
| Penicillin
Expected MIC Range
0.25-2 μg/mL | 0.12 | | |
| | 0.25 | | |
| | 0.5 | 26 | 24 |
| | 1 | 13 | 14 |
| | 2 | 1 | 2 |
| Streptomycin 1000
Expected MIC Range
≤1000 μg/mL | ≤1000 | 40 | 40 |
| | >1000 | | |
| Quinupristin/Dalfopristin^{a}
Expected MIC Range
0.25-1 μg/mL | ≤1 | 40 | 40 |
| | 2 | | |
| Tetracycline^{a}
Expected MIC Range
0.12-1 μg/mL | ≤0.12 | 7 | 2 |
| | 0.25 | 34 | 38 |
| | 0.5 | | |
| | 1 | | |
| | 2 | | |
| Vancomycin
Expected MIC Range
0.5-2 μg/mL | ≤0.5 | | |
| | 1 | 39 | 39 |
| | 2 | 1 | 1 |
| | 4 | | |
| Ceftaroline
Expected MIC Range
0.12-0.5μg/mL | 0.06 | | |
| | 0.12 | | |
| | 0.25 | 40 | 40 |
| | 0.5 | | |
| | 1 | | |
{12}
K191914 - Page 12 of 20
| Table 5. Quality Control S. aureus ATCC 29213 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Dalbavancin Expected MIC Range 0.03-0.12 μg/mL | 0.015 | | |
| | 0.03 | 11 | 5 |
| | 0.06 | 29 | 35 |
| | 0.12 | | |
| | 0.25 | | |
| Tedizolid Expected MIC Range 0.12-1 μg/mL | 0.12 | | |
| | 0.25 | 37 | 30 |
| | 0.5 | 3 | 10 |
| | 1 | | |
| | 2 | | |
a The Gram-positive panel did not include the full CLSI/FDA recommended dilution range for QC testing of this drug with this organism. However, at least one other QC organism was tested using the full CLSI/FDA recommended dilution range for this drug and exhibited acceptable results ( $\geq 95\%$ in-range results)
| Table 6. Quality Control E. faecalis ATCC 29212 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Chloramphenicol Expected MIC Range 4-16 μg/mL | 2 | | |
| | 4 | 40 | 37 |
| | 8 | | 3 |
| | 16 | | |
| | >16 | | |
| Ciprofloxacin Expected MIC Range 0.25-2 μg/mL | 0.12 | | |
| | 0.25 | | |
| | 0.5 | 40 | 40 |
| | 1 | | |
| | 2 | | |
| | 4 | | |
| Clindamycin Expected MIC Range 4-16 μg/mL | 2 | | |
| | 4 | 1 | |
| | 8 | 27 | 31 |
| | 16 | 11 | 9 |
| | >16 | | |
| Daptomycin Expected MIC Range 1-4 μg/mL | 0.5 | | |
| | 1 | 3 | 1 |
| | 2 | 33 | 37 |
| | 4 | 3 | 2 |
| | ≥8 | 1 | |
| Erythromycin Expected MIC Range 1-4 μg/mL | 0.5 | | |
| | 1 | 14 | 12 |
| | 2 | 26 | 28 |
| | 4 | | |
| | 8 | | |
| Gentamicin Expected MIC Range 4-16 μg/mL | 2 | | |
| | 4 | 3 | |
| | 8 | 31 | 34 |
| | 16 | 6 | 6 |
{13}
K191914 - Page 13 of 20
| Table 6. Quality Control
E. faecalis ATCC 29212 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| | >16 | | |
| Gentamicin 500
Expected MIC Range
≤ 500 μg/mL | ≤ 500 | 40 | 40 |
| | >500 | | |
| Oxacillin
Expected MIC Range
8-32 μg/mL | 4 | | |
| | 8 | | 2 |
| | 16 | 40 | 36 |
| | 32 | | 2 |
| | >32 | | |
| Penicillin
Expected MIC Range
1-4 μg/mL | 0.5 | | |
| | 1 | | |
| | 2 | 39 | 39 |
| | 4 | 1 | 1 |
| | 8 | | |
| Streptomycin 1000
Expected MIC Range
≤1000 μg/mL | ≤1000 | 40 | 40 |
| | >1000 | | |
| Quinupristin/Dalfopristin
Expected MIC Range
2-8 μg/mL | 1 | | |
| | 2 | | |
| | 4 | 40 | 40 |
| | 8 | | |
| | >8 | | |
| Tetracycline
Expected MIC Range
8-32 μg/mL | 2 | 1 | 1 |
| | 4 | | |
| | 8 | 6 | 7 |
| | 16 | 33 | 32 |
| | 32 | | |
| | >32 | | |
| Vancomycin
Expected MIC Range
1-4 μg/mL | 0.5 | | |
| | 1 | | |
| | 2 | 38 | 34 |
| | 4 | 2 | 6 |
| | 8 | | |
| Ceftaroline
Expected MIC Range
0.25-2 μg/mL | 0.12 | | |
| | 0.25 | | |
| | 0.5 | 8 | 2 |
| | 1 | 32 | 38 |
| | 2 | | |
| | 4 | | |
| Dalbavancin
Expected MIC Range
0.03-0.12 μg/mL | 0.015 | | |
| | 0.03 | 7 | 9 |
| | 0.06 | 33 | 31 |
| | 0.12 | | |
| | 0.25 | | |
| Tedizolid
Expected MIC Range
0.25-1 μg/mL | 0.12 | 1 | |
| | 0.25 | 36 | 28 |
| | 0.5 | 3 | 12 |
| | 1 | | |
{14}
| Table 6. Quality Control
E. faecalis ATCC 29212 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| | 2 | | |
| Table 7. Quality Control
E. faecalis ATCC 51299 | Concentration (μg/mL) | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| Gentamicin 500
Expected MIC Range
>500 μg/mL | ≤ 500 | 1 | 1 |
| | >500 | 39 | 39 |
| Streptomycin 1000
Expected MIC Range
>1000 μg/mL | ≤1000 | | |
| | >1000 | 40 | 40 |
| Table 8. D-Testa | Test Version | Result | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- | --- |
| S. aureus ATCC BAA-977
Expected Result: Growth | D-Test 1b | No Growth | | |
| | | Growth | 40 | 40 |
| S. aureus ATCC 29213
Expected Result: No Growth | D-Test 1b | No Growth | 40 | 40 |
| | | Growth | | |
| S. aureus ATCC BAA-977
Expected Result: Growth | D-Test 2c | No Growth | | 1 |
| | | Growth | 40 | 39 |
| S. aureus ATCC 29213
Expected Result: No Growth | D-Test 2c | No Growth | 40 | 40 |
| | | Growth | | |
a D-Test is conducted on Staphylococcus spp. that are resistant to Erythromycin (MIC's ≥ 8 ug/mL) and susceptible or intermediate to Clindamycin (MIC's ≤ 2 ug/mL) to determine if they exhibit inducible Clindamycin resistance
b D-Test 1: 4 ug/mL Erythromycin and 0.5 ug/mL Clindamycin are included in a single well
c D-Test 2: 8 ug/mL Erythromycin and 1.5 ug/mL Clindamycin are included in a single well
| Table 9. Cefoxitin Screena | Result | Legacy ARIS | ARIS HiQ |
| --- | --- | --- | --- |
| S. aureus ATCC 29213 | No Growth | 40 | 40 |
| Expected Result: No Growth | Growth | | |
a Cefoxitin Screen predicts the presence of mecA-mediated resistance in S. aureus. Cefoxitin Screen was performed at 6 ug/mL
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Not applicable
8. Accuracy (Instrument):
K191914 - Page 14 of 20
{15}
Not applicable
9. Carry-Over:
Not applicable
## B Comparison Studies:
1. Method Comparison with Predicate Device:
The performance of the ARIS HiQ was evaluated by means of a method comparison study conducted at one internal site. The study was designed to evaluate the performance of the ARIS HiQ compared to the legacy ARIS and OptiRead instruments using Sensititre (18-24 hour) susceptibility plates. Specially prepared susceptibility plates containing serial dilutions of select antibiotics representing a range of antimicrobial agent classes were used in the study. Each microorganism suspension was prepared using the Sensititre Nephelometer. Each plate was inoculated with the prepared microorganism suspension using the Sensititre Autoinoculator, incubated for 18-24 hours at 35°C on either the ARIS HiQ or legacy ARIS, and then read using the OptiRead. The MIC results obtained by the ARIS HiQ and legacy ARIS were then compared. The performance criteria described in the "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems" were used to evaluate the performance of the ARIS HiQ.
Method comparison testing was performed of a total of 75 Gram-negative non-fastidious challenge isolates to include (56) Enterobacteriaceae [E. cloacae (4), E. coli (8), K. aerogenes (5), K. oxytoca (4), K. ozaenae (1), K. pneumoniae (10), M. morganni (1), P. mirabilis (3), P. vulgaris (3), P. rettgeri (3), P. stuartii (2), S. marsescens (10), P. agglomerans (1), and S. odorifera (1)], (2) A. hydrophila, (6) Acinetobacter spp., and (11) P. aeruginosa isolates.
In addition, method comparison testing was performed on a total of 75 Gram-positive non-fastidious challenge isolates including (29) Enterococcus spp. [E. faecalis (15), E. faecium (10), E. durans (2), E. gallinarum (1) and E. raffinosus (1)], (27) S. aureus and (19) coagulase-negative Staphylococci (CoNS) [S. epidermidis (12), S. haemolyticus (3), S. warneri (2), S. saprophyticus (1) and S. capitis (1)] isolates. The full Gram-positive challenge panel was evaluated with each antimicrobial/screen performed, except for the Cefoxitin Screen, which was only conducted with S. aureus isolates.
Performance was evaluated using current FDA-recognized susceptibility interpretative criteria (STIC) on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/antibacterial-susceptibility-test-interpretive-criteria). As previously noted, performance of the ARIS HiQ System was being evaluated in a comparative study against the legacy ARIS using Sensititre MIC panels processed by both systems and there was no reference method testing required. Therefore, many representative drugs and organisms were used to encompass the multiple test combinations. The FDA STIC webpage information was used to assess categorical agreement. For some given organism/antimicrobial agent combinations, CLSI M100 (29th edition) was also used for this evaluation.
K191914 - Page 15 of 20
{16}
A single $P$ . aeruginosa isolate did not grow in the Sensititre susceptibility plate, so no results were obtained by either the ARIS HiQ or the legacy ARIS. In addition, the only $E$ . gallinarum isolate included in the study did not grow on either the ARIS HiQ or the legacy ARIS. Accounting for these growth failures, the growth rate for the Gram-negative and Gram-positive challenge isolate panels were $98.7\%$ (74/75) and $98.7\%$ (74/75), respectively.
Performance evaluations are shown in Table 10 and Table 11 for the Gram-negative and Gram-positive challenge isolates.
| Table 10. Non-Fastidious Gram-Negative Organisms (ARIS HiQ vs. Legacy ARIS/OptiRead) | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Drug | Organism Group | Total | EA N | EA % | Eval. Total | Eval. EA N | Eval. EA % | CA N | CA % | #R or NS | min | maj | vmj |
| Ampicillin | Aeromonas hydrophila | 2 | 2 | 100 | 0 | 0 | 0 | 2 | 100 | 2 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 7 | 7 | 100 | 53 | 94.6 | 51 | 3 | 0 | 0 |
| | | | | | | | | | | | | | |
| Cefazolin | Aeromonas hydrophila | 2 | 2 | 100 | 0 | 0 | 0 | 2 | 100 | 2 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 9 | 9 | 100 | 59 | 100 | 55 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Cefepime | Acinetobacter spp. | 6 | 6 | 100 | 2 | 2 | 100 | 6 | 100 | 4 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 1 | 1 | 100 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 18 | 18 | 100 | 51 | 91.1 | 11 | 5 | 0 | 0 |
| | | | | | | | | | | | | | |
| Ceftazidime | Acinetobacter spp. | 6 | 6 | 100 | 1 | 1 | 100 | 6 | 100 | 5 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 1 | 50 | 1 | 1 | 100 | 1 | 50 | 1 | 1 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 11 | 11 | 100 | 56 | 100 | 26 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Chloramphenicol | Acinetobacter spp. | 6 | 6 | 100 | 3 | 3 | 100 | 6 | 100 | 3 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 1 | 1 | 100 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 40 | 40 | 100 | 54 | 96.4 | 23 | 2 | 0 | 0 |
| | | | | | | | | | | | | | |
| Ertapenem | Aeromonas hydrophila | 2 | 2 | 100 | 1 | 1 | 100 | 2 | 100 | 1 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 54 | 96.4 | 48 | 46 | 95.8 | 55 | 98.2 | 22 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| Gentamicin | Acinetobacter spp. | 6 | 6 | 100 | 2 | 2 | 100 | 6 | 100 | 3 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 2 | 2 | 100 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 55 | 98.2 | 40 | 39 | 97.5 | 55 | 98.2 | 15 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| Levofloxacin | Acinetobacter spp. | 6 | 6 | 100 | 5 | 5 | 100 | 6 | 100 | 1 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 2 | 2 | 100 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 40 | 40 | 100 | 54 | 96.4 | 24 | 2 | 0 | 0 |
| | | | | | | | | | | | | | |
| Piperacillin/Tazobactam | Acinetobacter spp. | 6 | 6 | 100 | 2 | 2 | 100 | 6 | 100 | 3 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 1 | 1 | 100 | 2 | 100 | 1 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 30 | 30 | 100 | 54 | 96.4 | 19 | 2 | 0 | 0 |
K191914 - Page 16 of 20
{17}
| Table 10. Non-Fastidious Gram-Negative Organisms (ARIS HiQ vs. Legacy ARIS/OptiRead) | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Drug | Organism Group | Total | EA N | EA % | Eval. Total | Eval. EA N | Eval. EA % | CA N | CA % | #R or NS | min | maj | vmj |
| | P. aeruginosa | 10 | 10 | 100 | 9 | 9 | 100 | 10 | 100 | 1 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| TMP/SMX | Acinetobacter spp. | 6 | 6 | 100 | 3 | 3 | 100 | 6 | 100 | 2 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 0 | 0 | 0 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 9 | 9 | 100 | 56 | 100 | 25 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Tetracycline | Acinetobacter spp. | 6 | 6 | 100 | 5 | 5 | 100 | 6 | 100 | 1 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 2 | 2 | 100 | 2 | 100 | 0 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 36 | 36 | 100 | 52 | 92.9 | 26 | 4 | 0 | 0 |
| | | | | | | | | | | | | | |
| Ticaracillin | Acinetobacter spp. | 6 | 6 | 100 | 1 | 1 | 100 | 6 | 100 | 5 | 0 | 0 | 0 |
| | Aeromonas hydrophila | 2 | 2 | 100 | 0 | 0 | 0 | 2 | 100 | 2 | 0 | 0 | 0 |
| | Enterobacteriaceae | 56 | 56 | 100 | 13 | 13 | 100 | 55 | 98.2 | 43 | 1 | 0 | 0 |
| | P. aeruginosa | 10 | 10 | 100 | 4 | 4 | 100 | 10 | 100 | 9 | 0 | 0 | 0 |
| Table 11. Non-Fastidious Gram-Positive Organisms (ARIS HiQ vs. Legacy ARIS/OptiRead) | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Drug | Organism Group | Total | EA N | EA % | Eval. Total | Eval. EA N | Eval. EA % | CA N | CA % | #R or NS | min | maj | vmj |
| Ceftaroline | S. aureus | 27 | 27 | 100 | 27 | 27 | 100 | 26 | 96.3 | 0 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| Chloramphenicol | Enterococcus spp. | 28 | 28 | 100 | 22 | 22 | 100 | 28 | 100 | 5 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 22 | 22 | 100 | 26 | 96.3 | 5 | 1 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 17 | 17 | 100 | 19 | 100 | 2 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Ciprofloxacin | Enterococcus spp. | 28 | 28 | 100 | 19 | 19 | 100 | 27 | 96.4 | 9 | 1 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 18 | 18 | 100 | 27 | 100 | 9 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 15 | 15 | 100 | 19 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Clindamycin | S. aureus | 27 | 27 | 100 | 0 | 0 | 0 | 27 | 100 | 3 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 1 | 1 | 100 | 19 | 100 | 5 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Dalbavancin | S. aureus | 27 | 26 | 96.3 | 27 | 26 | 96.3 | 27 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Daptomycin | E. faecalis | 15 | 15 | 100 | 15 | 15 | 100 | 15 | 100 | 0 | 0 | 0 | 0 |
| | S. aureus | 27 | 26 | 96.3 | 25 | 24 | 96.0 | 27 | 100 | 0 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 17 | 17 | 100 | 19 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Erythromycin | Enterococcus spp. | 28 | 28 | 100 | 8 | 8 | 100 | 27 | 96.4 | 17 | 1 | 0 | 0 |
| | S. aureus | 27 | 26 | 96.3 | 10 | 9 | 90 | 25 | 92.6 | 17 | 2 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 1 | 1 | 100 | 19 | 100 | 7 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Gentamicin | S. aureus | 27 | 26 | 96.3 | 23 | 22 | 95.7 | 27 | 100 | 2 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 2 | 2 | 100 | 18 | 94.7 | 5 | 1 | 0 | 0 |
K191914 - Page 17 of 20
{18}
K191914 - Page 18 of 20
| Table 11. Non-Fastidious Gram-Positive Organisms (ARIS HiQ vs. Legacy ARIS/OptiRead) | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Drug | Organism Group | Total | EA N | EA % | Eval. Total | Eval. EA N | Eval. EA % | CA N | CA % | #R or NS | min | maj | vmj |
| | | | | | | | | | | | | | |
| Linezolid | Enterococcus spp. | 28 | 28 | 100 | 25 | 25 | 100 | 28 | 100 | 0 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 27 | 27 | 100 | 27 | 100 | 0 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 4 | 4 | 100 | 19 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Oxacillin | S. aureus | 27 | 27 | 100 | 15 | 15 | 100 | 27 | 100 | 11 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 8 | 8 | 100 | 19 | 100 | 13 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Penicillin | Enterococcus spp. | 28 | 28 | 100 | 17 | 17 | 100 | 28 | 100 | 10 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 4 | 4 | 100 | 27 | 100 | 21 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 8 | 8 | 100 | 18 | 94.7 | 16 | 0 | 0 | 1e |
| | | | | | | | | | | | | | |
| Quinupristin / Dalfopristin | Enterococcus spp. | 28 | 28 | 100 | 17 | 17 | 100 | 26 | 92.9 | 18 | 2 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 0 | 0 | 0 | 27 | 100 | 0 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 0 | 0 | 0 | 19 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Tedizolid | E. faecalis | 15 | 15 | 100 | 15 | 15 | 100 | 15 | 100 | 0 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 27 | 27 | 100 | 27 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Telavancin | E. faecalis | 15 | 15 | 100 | 12 | 12 | 100 | 15 | 100 | 3 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 27 | 27 | 100 | 27 | 100 | 0 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Tetracycline | Enterococcus spp. | 28 | 28 | 100 | 11 | 11 | 100 | 28 | 100 | 22 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 7 | 7 | 100 | 27 | 100 | 5 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 11 | 11 | 100 | 19 | 100 | 2 | 0 | 0 | 0 |
| | | | | | | | | | | | | | |
| Vancomycin | Enterococcus spp. | 28 | 28 | 100 | 14 | 14 | 100 | 28 | 100 | 11 | 0 | 0 | 0 |
| | S. aureus | 27 | 27 | 100 | 26 | 26 | 100 | 27 | 100 | 0 | 0 | 0 | 0 |
| | CoNS | 19 | 19 | 100 | 16 | 16 | 100 | 18 | 94.7 | 0 | 1 | 0 | 0 |
| | | | | | | | | | | | | | |
| D-Test 1a | Enterococcus spp. | 28 | NA | 28 | 100 | 19 | NA | 0 | 0 | | | | |
| | S. aureus | 27 | | | | | | 16 | 96.3 | 16 | 1f | 0 | |
| | CoNS | 19 | | | | | | 19 | 100 | 6 | 0 | 0 | |
| | | | | | | | | | | | | | |
| D-Test 2b | Enterococcus spp. | 28 | NA | 28 | 100 | 18 | NA | 0 | 0 | | | | |
| | S. aureus | 27 | | | | | | 27 | 100 | 17 | 0 | 0 | |
| | CoNS | 19 | | | | | | 19 | 100 | 6 | 0 | 0 | |
| | | | | | | | | | | | | | |
| Cefoxitin Screenc | S. aureus | 27 | NA | | | | | 27 | 100 | 11 | NA | 0 | 0 |
| | | | | | | | | | | | | | |
| HLAR Screen: Gentamycin 500 ug/mL | Enterococcus spp. | 28 | NA | 28 | 100 | 12 | NA | 0 | 0 | | | | |
| | S. aureus | 27 | | | | | | 27 | 100 | 0 | 0 | 0 | |
| | CoNS | 19 | | | | | | 19 | 100 | 1 | 0 | 0 | |
{19}
K191914 - Page 19 of 20
| Table 11. Non-Fastidious Gram-Positive Organisms (ARIS HiQ vs. Legacy ARIS/OptiRead) | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Drug | Organism Group | Total | EA N | EA % | Eval. Total | Eval. EA N | Eval. EA % | CA N | CA % | #R or NS | min | maj |
| | | | | | | | | | | | | |
| HLAR Screen: Streptomycin 1000 ug/mLd | Enterococcus spp. | 28 | NA | | | | | 28 | 100 | 12 | NA | 0 |
a. Erythromycin – 4 ug/mL; Clindamycin – 0.5 ug/mL
b. Erythromycin – 8 ug/mL; Clindamycin – 1.5 ug/mL
c. Cefoxitin – 6 ug/mL
d. 27 S. aureus isolates and 19 coagulase-negative Staphylococci (CoNS) challenge isolates were also tested but performance was not included because this test is only cleared for use with Enterococcus spp.
e. The categorical very major error rate for penicillin when testing coagulase-negative Staphylococci (CoNS) challenge isolates is 6.3% (1/16). Based on the essential agreement and lack of an intermediate breakpoint for penicillin, the overall adjusted very major error rate for coagulase-negative staphylococci challenge isolates is 0% (0/16)
f. A single major error was obtained for D-test 1 with S. aureus. This result was considered a random error occurrence and is acceptable
NA – Not applicable, as these are qualitative tests
EA – Essential Agreement
CA – Category Agreement
EVAL – Evaluable isolates
R – Resistant isolates
NS – Non-Susceptible isolates
min – minor errors
maj – major errors
vmj – very major errors
Essential Agreement (EA) is when the ARIS HiQ result agrees exactly or is within one doubling dilution of the legacy ARIS result. Category Agreement (CA) is when the ARIS HiQ result interpretation agrees exactly with the legacy ARIS result interpretation. Evaluable EA is when the MIC result for the ARIS HiQ and the legacy ARIS are on-scale.
The %EA is acceptable when compared to the reference method as described in the FDA guidance document, “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) System; Guidance for Industry and FDA”.
Overall, the MIC results of the ARIS HiQ and the legacy ARIS were very similar, for all antimicrobials tested. The %EA and %CA for all microorganism group/antimicrobial combinations were greater than 90%, except for A. hydrophila challenged with Ceftazidime, for which only two isolates were evaluated.
## Resistance Markers
Resistance markers for indicated Gram-negative non-fastidious isolates were provided in the submission. They consisted of the Extended Spectrum β-lactamase (ESBL) gene CTX-M-1 and the carbapenemase genes KPC and NDM.
## 2. Matrix Comparison:
Not applicable
{20}
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
See comparison studies in Section VII.B above.
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Not applicable
F Other Supportive Instrument Performance Characteristics Data:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K191914 - Page 20 of 20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.