← Product Code [LPN](/productcode/LPN) · K191872

# Menicon 3% Hydrogen Peroxide Cleaning and Disinfecting Solution (K191872)

_Menicon Co, Ltd. · LPN · Mar 2, 2020 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191872

## Device Facts

- **Applicant:** Menicon Co, Ltd.
- **Product Code:** [LPN](/productcode/LPN.md)
- **Decision Date:** Mar 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5928
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfection, and storing of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) and rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses, as recommended by your eye care professional.

## Device Story

System comprises 3% hydrogen peroxide solution and specialized lens case with neutralizer catalyst disk. User places lenses in holder; adds solution to cup; catalyst disk neutralizes peroxide over 6-hour minimum period. Used by contact lens wearers for daily cleaning, protein removal, and disinfection. Preservative-free aqueous solution contains hydrogen peroxide, phosphonic acid stabilizer, citric acid, PEG, sodium chloride, sodium hydroxide, and phosphate buffer. System must be used together to ensure safe neutralization of peroxide before lens re-insertion. Benefits include effective antimicrobial activity and removal of lens deposits, maintaining lens hygiene and patient comfort.

## Clinical Evidence

No clinical data. Evidence based on non-clinical bench testing: lens compatibility (6 hydrogel/silicone hydrogel and 2 RGP lens types), neutralization profile (residual peroxide levels), disinfecting/preservative efficacy (ISO level reduction at 6/24 hours and 7 days), cleaning efficacy (protein/lipid removal), and biocompatibility (GLP-compliant in-vitro/in-vivo testing).

## Technological Characteristics

Aqueous solution: 3% hydrogen peroxide, phosphonic acid, citric acid, PEG, NaCl, NaOH, phosphate buffer. Lens case: transparent cup with lens holder/cap assembly containing neutralizer catalyst disk. Disinfection principle: oxidative disinfection followed by catalytic neutralization. Minimum disinfection time: 6 hours. Preservative-free.

## Regulatory Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

## Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

## Predicate Devices

- Clear Care Plus Cleaning & Disinfecting Solution ([K142284](/device/K142284.md))
- Clear Care Cleaning & Disinfecting Solution ([K030522](/device/K030522.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym with the full name of the agency on the right. The FDA part is in blue, with the acronym in a solid blue square and the full name written in a smaller font next to it.

March 2, 2020

Menicon Co., Ltd. Li Haosheng, Ph.D. International Regulatory Affairs Menicon Co., Ltd. 3-21-19, Aoi, Naka-ku, Nagoya, 460-0006 JAPAN

Re: K191872

Trade/Device Name: Menicon 3% Hydrogen Peroxide Cleaning and Disinfecting Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) contact lens care products Regulatory Class: Class II Product Code: LPN, MRC Dated: January 29, 2020 Received: February 3, 2020

Dear Dr. Li Haosheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191872

Device Name

Menicon 3 % Hydrogen Peroxide Cleaning & Disinfecting Solution

#### Indications for Use (Describe)

Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfection, and storing of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) and rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses, as recommended by your eye care professional.

Type of Use (Select one or both, as applicable)

|  | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
|  | Over-The-Counter Use (21 CFR 801 Subpart C)  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows a circular logo that is divided into two halves. The top half is red, and the bottom half is green. A white, stylized shape resembling a sine wave or a connected 'f' and 'a' is overlaid across the center of the circle, spanning both the red and green sections. The logo has a clean and modern design.

# 510(k) Summary

# Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution

### 1. Applicant Information

Menicon Co., Ltd. 21-19, Aoi 3, Naka-ku, Nagoya, Aichi 460-0006 JAPAN

| Contact Person: | Tetsuji Kawai         |
|-----------------|-----------------------|
| Telephone No.:  | +81-52-935-1676       |
| Fax No.:        | +81-52-935-1633       |
| E-mail:         | t-kawai@menicon.co.jp |
| Date Prepared:  | July 10, 2019         |

### 2. Device Information

| Device classification: | Class II                                                                                                                            |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Classification name:   | 21 CFR 886.5928<br>Soft (hydrophilic) Contact Lens Care Product<br>21 CFR 886.5918<br>Rigid Gas Permeable Contact Lens Care Product |
| Product code:          | LPN, MRC                                                                                                                            |
| Proprietary name:      | Menicon 3% Hydrogen Peroxide<br>Cleaning & Disinfecting Solution                                                                    |

## 3. Predicate Device

Menicon claims substantial equivalence to K142284 Clear Care Plus Cleaning & Disinfecting Solution and K030522 Clear Care Cleaning & Disinfecting Solution.

{4}------------------------------------------------

### 4. Description of Device

The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution system consists of: Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case. The lens case consists of a transparent cup and a lens holder-lens case cap assembly with a neutralizer catalyst disk. Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution and the special lens case must always be used together.

The preservative free, aqueous Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution contains: hydrogen peroxide 3%, phosphonic acid compound (stabilizer), citric acid, polyethylene glycol, sodium chloride, sodium hydroxide, and phosphate (buffer system).

### 5. Indications for Use

Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfection, and storing of soft (hydrophilic) contact lenses (including silicone hydrogel lenses) and rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses, as recommended by your eve care professional.

### 6. Performance Data

### Non-Clinical Data

A series of preclinical tests were performed to demonstrate the substantial equivalence of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution to the predicate devices. All tests were conducted in accordance with the Premarket Notification 510(k) Guidance Document for Contact Lens Care Products (1997) or according to valid scientific protocols.

#### . Lens Compatibility

Compatibility testing was conducted with six commercially available hydrogel and silicone hydrogel soft contact lenses: Acuvue2 (etafilcon A), Acuvue Oasys (senofilcon A), Biofinity (comfilcon A), Air Optix Aqua (lotrafilcon B), Definitive74 (efrofilcon A) and PureVision2 (balafilcon A), and 2 commercially available RGP lenses: Menicon Z (tisilfocon A) and Boston IV (itafocon B). Results demonstrated that the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution was compatible with the test contact lenses.

{5}------------------------------------------------

#### . Neutralization Profile

Tests were conducted to evaluate the effectiveness of the neutralization lens case of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution in terms of residual hydrogen peroxide level. The Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution showed a similar neutralization profile to the predicate device. In addition, stability of the neutralizer catalyst disk was confirmed for 90-cycles treatments.

#### . Disinfecting and Preservative Efficacy

Disinfecting efficacy test was conducted to evaluate the antimicrobial activity of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution as a contact lens disinfection product. Results demonstrated that harmful microorganisms were effectively reduced to an ISO level at 6 and 24 hours and 7 days after neutralization. In addition, Preservative Efficacy test was conducted to ensure 3-month open bottle discard date of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution.

#### ● Cleaning Efficacy

Studies were conducted in an in-vitro system to quantify the ability of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution to remove proteins and lipids from commercially available hydrogel and silicone hydrogel soft contact lenses and RGP lenses, and to compare its performance to the predicate device. Results suggested that both solutions performed similarly in the removal of lens deposition.

#### ● Toxicology

A series of in-vitro and in-vivo biocompatibility tests were conducted in accordance with the GLP regulation (21 CFR Part 58) to ensure the safety of the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution. Results demonstrated that the neutralized solution, the lens case materials and the neutralizer catalyst were non-toxic and nonirritating under the experimental conditions.

### Clinical Data

Clinical studies were unnecessary for this application.

### Conclusion

Based upon the data presented, the Menicon 3% Hydrogen Peroxide Cleaning & Disinfecting Solution is as safe, as effective and performs as well as the predicate devices.

{6}------------------------------------------------

## 7. Substantial equivalence

The claim of substantial equivalence to the previously cleared devices is supported by the following Comparison of Characteristics in Table 1.

|                              | Menicon 3% Hydrogen Peroxide<br>Cleaning & Disinfecting Solution                                                                                                                                                                                                                                                         | Clear Care Plus                                                                                                                                                                                                                                                                                                                          | Clear Care                                                                                                                                                                                                                                                  |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                | New                                                                                                                                                                                                                                                                                                                      | K142284                                                                                                                                                                                                                                                                                                                                  | K030522                                                                                                                                                                                                                                                     |
| Product Codes                | LPN, MRC                                                                                                                                                                                                                                                                                                                 | LPN, MRC                                                                                                                                                                                                                                                                                                                                 | LPN                                                                                                                                                                                                                                                         |
| Indications for Use          | Indicated for use in cleaning, daily<br>protein removal, disinfection, and<br>storing of soft (hydrophilic) contact<br>lenses (including silicone hydrogel<br>lenses) and rigid gas permeable<br>(fluoro silicone acrylate and<br>silicone acrylate) contact lenses, as<br>recommended by your eye care<br>professional. | Indicated for use in simultaneous<br>cleaning, daily protein removal,<br>disinfection, and storing of soft<br>(hydrophilic) contact lenses<br>(including silicone hydrogel<br>lenses) and rigid gas permeable<br>(fluoro silicone acrylate and<br>silicone acrylate) contact lenses,<br>as recommended by your eye<br>care professional. | Indicated for use in<br>simultaneous cleaning,<br>daily protein removal,<br>disinfecting, and storing of<br>soft (hydrophilic) contact<br>lenses (including silicone<br>hydrogel lenses) contact<br>lenses as recommended by<br>your eye care practitioner. |
| Lens Case Design             | specially designed lens<br>case with neutralizing disk                                                                                                                                                                                                                                                                   | specially designed lens<br>case with neutralizing disk                                                                                                                                                                                                                                                                                   | specially designed lens<br>case with neutralizing disk                                                                                                                                                                                                      |
| Minimum<br>Disinfection Time | 6 hours                                                                                                                                                                                                                                                                                                                  | 6 hours                                                                                                                                                                                                                                                                                                                                  | 6 hours                                                                                                                                                                                                                                                     |
| Color of Bottle Cap          | red                                                                                                                                                                                                                                                                                                                      | red                                                                                                                                                                                                                                                                                                                                      | red                                                                                                                                                                                                                                                         |
| Tamper Resistant             | yes                                                                                                                                                                                                                                                                                                                      | yes                                                                                                                                                                                                                                                                                                                                      | yes                                                                                                                                                                                                                                                         |

#### Table 1 Comparison of Characteristics

---

**Source:** [https://fda.innolitics.com/device/K191872](https://fda.innolitics.com/device/K191872)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
