Compression Therapy Device (Model: LGT-2201DVT) is intended to prevent DVT (Deep Vein Thrombosis) by improving the blood velocity of patients. LGT-2201DVT is Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, Edema - Acute, Edema - Chronic, Extremity Pain Incident to Trauma or Surgery, Leg Ulcers, Venous Stasis / Venous Insufficiency.
Device Story
Compression Therapy Device (LGT-2201DVT) consists of an intermittent pneumatic controller, connectable hoses, and limb sleeves. Device operates by sequentially inflating and deflating sleeves to apply circulating pressure to limbs; promotes blood and lymph circulation; prevents thrombus; reduces limb edema. Used in clinical or home settings under prescription. Healthcare providers use the device to manage venous/lymphatic conditions and prevent DVT. Benefits include improved blood velocity and reduced risk of venous stasis-related complications.
Clinical Evidence
Bench testing only. Evaluated for safety and performance per IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and IEC 62133-2 (battery safety). No clinical data provided.
Technological Characteristics
Intermittent pneumatic compression system. Class II, Type BF. Powered by 11.1V/6500mAh lithium battery or AC 100-240V. Pressure range 30-150mmHg. Biocompatibility per ISO 10993-5 and ISO 10993-10. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 62133-2.
Indications for Use
Indicated for patients requiring DVT prophylaxis, circulation enhancement, or treatment for chronic/acute edema, extremity pain from trauma/surgery, leg ulcers, and venous stasis/insufficiency.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 19, 2020
Guangzhou Longest Science & Technology Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F. No.1 TianTai road. Science City. LuoGang District Guangzhou. Cn
Re: K191856
Trade/Device Name: Compression Therapy Device (Model: LGT-2201DVT) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: February 12, 2020 Received: February 18, 2020
## Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support. Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Device Name
Compression Therapy Device (Model: LGT-2201DVT)
Indications for Use (Describe)
Compression Therapy Device (Model: LGT-2201DVT) is intended to prevent DVT (Deep Vein Thrombosis) by improving the blood velocity of patients.
LGT-2201DVT is Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, Edema - Chronic, Extremity Pain Incident to Trauma or Surgery, Leg Ulcers, Venous Stasis / Venous Insufficiency.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 870.1130.
#### Submitter Information 1.
Sponsor Name: Guangzhou Longest Science & Technology Co., Ltd.
Address: 5&6f, Building B4, No.11, Kaiyuan Avenue, Science City Guangzhou Hi-Tech Industrial Development Zone, Guangzhou Guangdong, CHINA 510530
Contact Person: Xiaobing Luo (Deputy general manager)
Phone: +86-020-66353999
E-mail: gzlongest@126.com
Application Correspondent:
Guangzhou KEDA Biological Tech Co., Ltd.
Contact person: Jet Li
Email: med-jl@foxmail.com
Tel: +86-20-22325619
Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China
#### 2. Subject Device Information
Type of 510(k) submission: Traditional
| Common Name: | Compressible limb sleeve |
|----------------------|-----------------------------------------------|
| Trade Name: | Compression Therapy Device Model: LGT-2201DVT |
| Classification Name: | sleeve, limb, compressible |
| Review Panel: | Cardiovascular |
| Product Code: | JOW |
| Regulation Number: | 870.5800 |
| Regulation Class: | II |
#### 3. Predicate Device Information
| Sponsor: | DAESUNG MAREF CO.,LTD |
|----------|-----------------------|
|----------|-----------------------|
1
{4}------------------------------------------------
| Common Name: | Intermittent Pneumatic Compression system |
|--------------------|-------------------------------------------|
| Trade Name: | DVT-4000 |
| 510(k) number: | K160180 |
| Review Panel: | Cardiovascular |
| Product Code: | JOW |
| Regulation Number: | 870.5800 |
| Regulation Class: | II |
# 4. Device Description
Compression Therapy Device LGT-2201DVT a compression therapy device comprised of intermittent pneumatic controller, sleeves and connectable hose. The working principle is the air inflating and deflating the sleeve sequentially to develop the circulating pressure on the human body. Squeezing the proximal and distal of the limbs to promote blood circulation lymphatic system and improve body microcirculation. Besides, prevent thrombus, reduce limbs drops and this kind disease which is related to blood and lymph circulation directly or indirectly .
## 5. Intended Use / Indication for Use
Compression Therapy Device (Model: LGT-2201DVT) is intended to prevent DVT (Deep Vein Thrombosis) by improving the blood velocity of patients.
LGT-2201DVT is Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, Edema -Acute, Edema - Chronic, Extremity Pain Incident to Trauma or Surgery, Leg Ulcers, Venous Stasis / Venous Insufficiency.
## 6. Test Summary
The Compression Therapy Device has been evaluated the safety and performance by lab bench testing according to the following standards:
- � IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988+A1:2012
- � IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, 2014
- ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro � cytotoxicity, 2009
- � ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
{5}------------------------------------------------
- � IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications- Part 2: Lithium systems
## 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Manufacturer | Guangzhou Longest Science &<br>Technology Co., Ltd | DAESUNG MAREF CO.,LTD | -- |
| Product Name | Compression Therapy Device | DVT-4000 | -- |
| Intended Use | Compression Therapy Device<br>(Model: LGT-2201DVT) is intended<br>to prevent DVT (Deep Vein<br>Thrombosis) by improving the<br>blood velocity of patients.<br>LGT-2201DVT is Circulation<br>Enhancement, Deep Vein<br>Thrombosis Prophylaxis, Edema -<br>Acute, Edema - Chronic, Extremity<br>Pain Incident to Trauma or Surgery,<br>Leg Ulcers, Venous Stasis /<br>Venous Insufficiency. | DVT-4000 is a system to prevent<br>DVT (Deep Vein Thrombosis) by<br>improving the blood velocity of<br>patients.<br>DVT-4000 is Circulation<br>Enhancement, Deep Vein<br>Thrombosis Prophylaxis, Edema -<br>Acute, Edema - Chronic, Extremity<br>Pain Incident to Trauma or Surgery,<br>Leg Ulcers, Venous Stasis /<br>Venous Insufficiency | SE |
| Type of use | Prescription Use | Prescription Use | SE |
| Size (L X W XH) | 27 x14.8 x 12.8 cm | 21.4 x18.1 x18.6 cm | Similar<br>Note 1 |
| Weight | 2.3kg | 2.0 kg | Similar<br>Note 1 |
| Input rating | AC 100-240 V 50/60Hz | AC 100-240 V 50/60Hz | SE |
| IEC<br>classification | Class II , Type BF | Class II , Type BF | SE |
| Input power | 90VA | 60-80VA | Similar<br>Note 2 |
| Battery Voltage | 11.1V /6500mAh, lithium<br>battery | 14.4V/2600mAh | Similar<br>Note 2 |
| Pressure range | 30 - 150mmHg | 20 - 140±10mmHg | Similar<br>Note 3 |
{6}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------|
| Setting Time | 1min~99hours or continuous | 1min~999hours | SE |
| Mode of<br>Operation | Continuous | Continuous | SE |
| Applied part | Compression sleeves | Compression sleeves | SE |
| Electrical,<br>Mechanical and<br>Thermal<br>Evaluation | IEC 60601-1<br>IEC 60601-1-2<br>IEC62133 | IEC 60601-1<br>IEC 60601-1-2<br>IEC62133 | SE |
| Biocompatibility<br>Evaluation | All the patient contracting materials<br>are evaluated by the<br>biocompatibility standard ISO<br>10993 -5, -10. | All the patient contracting materials<br>are evaluated by the<br>biocompatibility standard ISO<br>10993 -5, -10. | SE |
## Note 1
Although there is a bit difference of the size and weight between the predicate device and subject device, they are all complied with the IEC 60601-1. Thus the differences do not affect the safety and effectiveness.
# Note 2
The little difference of the input power does not affect the safety and effectiveness for the predicate device and subject device are complied with the IEC 60601-1. And although there is a bit difference of the battery, the LGT-2201DVT battery is verified in accordance with IEC62133. So the differences do not affect the safety and effectiveness.
## Note 3
Although there is a bit difference of the Pressure range between the predicate device and subject device, they are all complied with the IEC 60601-1. Thus the differences do not affect the safety and effectiveness.
## 8. Conclusion
The subject device Compression Therapy Device, model: LGT-2201DVT has same intended use, principle of operation and similar technological characteristics as the predicate device. Although there are some specification that are different between two device, the subject device has undergone safety and performance tests, and the results complied with the test requests. Therefore, the difference between the Compression Therapy Device, model: LGT-2201DVT and the predicate device did not raise any problem
{7}------------------------------------------------
of substantial equivalence. Thus the subject device is substantially equivalent to the predicate device.
# 9. Summary Prepared Date
12 February 2020