← Product Code MQC · K191836

# Ceramill A-Splint (K191836)

_Amann Girrbach AG · MQC · Dec 20, 2019 · DE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191836

## Device Facts

- **Applicant:** Amann Girrbach AG
- **Product Code:** MQC
- **Decision Date:** Dec 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE
- **Attributes:** Therapeutic

## Intended Use

Ceramill A-Splint is indicated for the fabrication of removable bite splints.

## Device Story

Ceramill A-Splint consists of ready-to-use polymethyl methacrylate (PMMA) blanks designed for CAD/CAM milling of removable bite splints. Available in round disk or arch-form shapes, the material is processed by dental laboratories or clinicians using CAD/CAM milling systems to create custom splints. The device is non-sterile, single-use, and intended for prescription use. By providing a biocompatible, durable PMMA material, it enables the fabrication of custom-fit splints to assist in dental management. The output is a finished, milled bite splint used by the patient to address dental conditions as determined by the prescribing clinician.

## Clinical Evidence

Bench testing only. Performance was validated against ISO 20795-2:2013, including residual monomer content, 3-point bending strength, water solubility, water absorption, flexural modulus, and fracture toughness. Biocompatibility was assessed per ISO 10993-1, with cytotoxicity testing showing no adverse effects. Additional testing included a chewing simulator (wearing period up to 3 years) and shelf-life testing (4 years).

## Technological Characteristics

Material: Polymethyl methacrylate (PMMA) without pigments. Form factor: CAD/CAM milling blanks (round or arch-shaped). Standards: ISO 20795-2:2013 (dentistry base polymers), ISO 10993-1 (biocompatibility). Processing: CAD/CAM milling. Sterilization: Non-sterile. Single-use.

## Predicate Devices

- M-PM-Disc (Clear) ([K134015](/device/K134015.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue box, and the full name is in a sans-serif font.

December 20, 2019

Amann Girrbach AG % Rachel Paul Senior Consultant QA & RA Emergo Europe Consulting Prinsessegracht 20 The Hague, 2514AP N1

Re: K191836

Trade/Device Name: Ceramill A-Splint Regulatory Class: Unclassified Product Code: MQC Dated: October 28, 2019 Received: October 31, 2019

Dear Rachel Paul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K191836

Device Name

Ceramill A-Splint

Indications for Use (Describe)

Ceramill A-Splint is indicated for the fabrication of removable bite splints.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) K191836 Summary

## Ceramill A-Splint

#### 1. Submission Sponsor

Amann Girrbach AG

Herrschaftswiesen 1

6842 Koblach

AUSTRIA

Contact: Debora Engel

Title: Regulatory Affairs Manager

Email: Debora.engel@amanngirrbach.com

Office number: + 49 (7231) 957-260

#### 2. Submission Correspondent

Emergo Europe Consulting

Prinsessegracht 20

The Hague, 2514 AP

THE NETHERLANDS

Office Phone: +31 (0) 70 345 8570 / Direct: +31 (0) 70 850 8249

Contact: Rachel Paul

Title: Senior Consultant, QA & RA

Email: LST.AUS.ProjectManagement@ul.com

Cell Phone : 00 6 89 83 16 09

3. Date Prepared

28 October 2019

- 4. Device Identification

{4}------------------------------------------------

Trade/Proprietary Name: Ceramill A-Splint Common/Usual Name: Occlusal splints Classification Name: Mouthguard, Prescription N/A - Unclassified Regulation Number: Product Code: MQC Device Class: N/A - Unclassified Classification Panel: Dental

#### 5. Legally Marketed Predicate Device

K134015, M-PM-Disc (Clear), Merz Dental GmbH

#### 6. Indication for Use Statement

Ceramill A-Splint is indicated for the fabrication of removable bite splints.

#### 7. Device Description

Ceramill A-Splint are ready to use, polymethyl methacrylate (PMMA) based computer-aided design/computer-aided manufacturing (CAD-CAM) blanks, for milling splints. Ceramill A-Splint is available as a disk (round form) or as an arch-form (or "U"-shape).

#### 8. Comparison of Technology

The following table compares the Ceramill A-Splint to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. There are no differences in technology.

| Attribute                                                                                     | Ceramill A-Splint                                                                                                            | M-PM-Disc                                                                           | Comparison                                                                                                                                                                                                            |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                                                  | Amann Girrbach AG                                                                                                            | Merz Dental GmbH                                                                    | N/A                                                                                                                                                                                                                   |
| 510(k) Number                                                                                 | To be determined (TBD)                                                                                                       | K134015                                                                             | N/A                                                                                                                                                                                                                   |
| Product Code                                                                                  | MQC                                                                                                                          | MQC                                                                                 | Same                                                                                                                                                                                                                  |
| Regulation<br>Reference                                                                       | Unclassified                                                                                                                 | Unclassified                                                                        | Same                                                                                                                                                                                                                  |
| Indications for Use                                                                           | Ceramill A-Splint is<br>indicated for the                                                                                    | The Merz Dental GmbH<br>M-PM-Disc (Clear) is<br>indicated for the                   | Similar; both devices are<br>indicated for the<br>production of removable                                                                                                                                             |
| Attribute                                                                                     | Ceramill A-Splint                                                                                                            | M-PM-Disc                                                                           | Comparison                                                                                                                                                                                                            |
|                                                                                               | fabrication of removable<br>bite splints.                                                                                    | fabrication of removable<br>bite splints such as<br>mouthguards and<br>nightguards. | bite splints. The<br>indications for use of<br>Ceramill A-Splint fall<br>within the intended use of<br>unclassified mouthguard<br>devices. Both devices are<br>for prescription only.                                 |
| Material                                                                                      | Solid<br>Polymethylmethacrylate<br>(PMMA)<br>No pigments                                                                     | Solid<br>Polymethylmethacrylate<br>(PMMA)<br>No pigments                            | Same                                                                                                                                                                                                                  |
| Residual monomer                                                                              | Less than 1% residual<br>monomer                                                                                             | Less than 1% residual<br>monomer                                                    | Same; both devices meet<br>requirements from ISO<br>20795-2:2013 (<5).                                                                                                                                                |
| Shape, sizes                                                                                  | Available in round and<br>arch shapes respectively<br>of 98 and 71 mm with<br>three (3) different height<br>(14, 16, 20 mm). | Available only in round<br>shape from 90 x 15 to<br>114.8 x 25 mm.                  | Similar; the arch shape<br>does not introduce any<br>additional safety or<br>efficacy concerns.<br>Dimensions are in the<br>same range.                                                                               |
| 3-point bending<br>strength                                                                   | >100 MPa                                                                                                                     | 91.5 MPa - 96.6 MPa                                                                 | Similar; the specifications<br>are in the same range.<br>This minor variance does<br>not introduce additional<br>safety or efficacy<br>concerns. Both devices<br>meet requirements from<br>ISO 20795-2:20131 (≥50).   |
| Water solubility                                                                              | <0.7 µg/mm3                                                                                                                  | 0.2 µg/mm³                                                                          | Similar; the specifications<br>are in the same range.<br>This minor variance does<br>not introduce additional<br>safety or efficacy<br>concerns.                                                                      |
| Attribute                                                                                     | Ceramill A-Splint                                                                                                            | M-PM-Disc                                                                           | Comparison                                                                                                                                                                                                            |
|                                                                                               |                                                                                                                              |                                                                                     | Both devices meet<br>requirements from ISO<br>20795-2:2013 (≤5).                                                                                                                                                      |
| Water absorption                                                                              | <25 µg/mm3                                                                                                                   | 26.5 µg/mm³                                                                         | Similar; the specifications<br>are in the same range.<br>This minor variance does<br>not introduce additional<br>safety or efficacy<br>concerns.<br>Both devices meet<br>requirements from ISO<br>20795-2:2013 (≤32). |
| Flexural modulus                                                                              | ≥ 2000 MPa                                                                                                                   | Unknown                                                                             | The submission device<br>meet requirements from<br>ISO 20795-2:2013 (≥<br>1500).                                                                                                                                      |
| Fracture toughness<br>(maximum stress<br>intensity (Kmax) and<br>total fracture work<br>(Wf)) | Kmax≥ 1.1 MPa m1/2<br>Wf ≥ 250 J/m²                                                                                          | Unknown                                                                             | The submission device<br>meet requirements from<br>ISO 20795-2:2013 (Kmax ≥<br>1.1 and Wf≥ 250).                                                                                                                      |
| Sterile                                                                                       | Delivered and use non-<br>sterile                                                                                            | Delivered and use non-<br>sterile                                                   | Same                                                                                                                                                                                                                  |
| Single-Use                                                                                    | Yes                                                                                                                          | Yes                                                                                 | Same                                                                                                                                                                                                                  |
| Complies with ISO<br>10993-1                                                                  | Yes                                                                                                                          | Yes                                                                                 | Same                                                                                                                                                                                                                  |

#### Table 5A - Comparison of Technology

{5}------------------------------------------------

¹ ISO 20795-2 :2013 Dentistry – Base polymers : Part 2 : Orthodontic base polymers (FDA Recognition Number: 4-233).

{6}------------------------------------------------

#### 9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of Ceramill A-Splint and to show substantial equivalence to the predicate device, Amann Girrbach AG completed the following non-clinical tests.

The Ceramill A-Splint passed the testing in accordance with internal requirements, national standards, and international standards shown below:

- . Biocompatibility assessment was conducted according to ISO 10993-1. Considering the chemically defined composition, the existing dental history with PMMA polymers and the inert material properties substantiated by highly sensitive chemical analysis, it was determined that no other

{7}------------------------------------------------

biocompatibility testing than cytotoxicity testing was required. No cytotoxic effects were observed in presence of the test material extracts.

- . Bench performance testing conducted in conformity with ISO 20795-2:2013 met all of the requirements set forth in the standard for residual monomer testing, 3-point bending strength, water solubility, water absorption, flexural modulus, fracture toughness, polishability, and porosity.
- Wearing testing based on an adapted Ivoclar method in a chewing simulator – the wearing period has been determined to be of a maximum of three years.
- . Shelf Life Testing – it was demonstrated that key characteristics do not degrade after four years of storage.

#### 10. Conclusion

Based on similarities in intended use, indications for use statements, technology, and non-clinical performance testing, Amann Girrbach believes that Ceramill A-Splint is substantially equivalent to the predicate device (K134015).

---

**Source:** [https://fda.innolitics.com/device/K191836](https://fda.innolitics.com/device/K191836)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
