The F&P Optiflow™ 3S nasal cannula is a 14 day single use nasal cannula interface for use with specified respiratory gas humidifiers to deliver Nasal High Flow (NHF) therapy to spontaneously breathing adult patients. This product is designed to be used by appropriately qualified healthcare professionals in a hospital / institutional environment.
Device Story
F&P Optiflow 3S Nasal Cannula is a patient interface for delivering humidified respiratory gases; connects to F&P 850 and AIRVO systems via dual 22 mm male ISO taper or AIRVO-specific connector. Used in hospital/institutional environments by qualified healthcare professionals. Features include headgear release mechanism and user-side swapping functionality. Device is single-patient use for up to 14 days. Provides humidified gas to patient; supports clinical decision-making by enabling NHF therapy delivery. Benefits include respiratory support for spontaneously breathing adults.
Clinical Evidence
Bench testing only. Performance data included shelf-life simulation (ASTM F1980-07, ISO 11607-1), transportation simulation (ISTA 2A), humidification output, thermal overshoot, connection strength, gas path leak testing, and condensation testing. Biocompatibility testing performed per ISO 10993 (parts 1, 3, 5, 10, 11, 12, 17, 18) and ISO 18562 (parts 1, 2, 3) for breathing gas pathways.
Technological Characteristics
Nasal cannula interface; 3 sizes (S, M, L). Connectors: 22 mm male ISO taper and AIRVO-specific. Materials evaluated per ISO 10993 and ISO 18562. Non-sterile. Shelf life: 3 years. Operating temperature: 18–28 °C. Compatible with F&P 850 and AIRVO/AIRVO 2 systems. Features: headgear release mechanism, user-side swapping.
Indications for Use
Indicated for spontaneously breathing adult patients requiring Nasal High Flow (NHF) therapy via respiratory gas humidifiers in hospital or institutional settings.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 19, 2020
Fisher & Paykel Healthcare Ltd Reena Daken Senior Regulatory Affairs Specialist 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand
Re: K191818
Trade/Device Name: F&P Optiflow 3S Nasal Cannula Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: February 14, 2020 Received: February18 , 2020
Dear Reena Daken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191818
Device Name F&P Optiflow 3S Nasal Cannula
#### Indications for Use (Describe)
The F&P Optiflow™™ 3S nasal cannula is a 14 day single use nasal cannula interface for use with specified respiratory gas humidifiers to deliver Nasal High Flow (NHF) therapy to spontaneously breathing adult patients. This product is designed to be used by appropriately qualified healthcare professionals in a hospital / institutional environment.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date prepared | 18 March 2020 |
|--------------------------|--------------------------------------------------------------------------------------------------------|
| Company Name and Address | Fisher and Paykel Healthcare<br>15 Maurice Paykel Place, East Tamaki,<br>Auckland 2013,<br>New Zealand |
| Submitter | Masar Mohammad |
| Contact Person | Reena Daken |
| | Senior Regulatory Affairs Specialist |
| | Telephone: +64 9 574 0100 |
| | Fax: +64 9 574 0158 |
| | Reena.Daken@fphcare.co.nz |
| Trade name | F&P OptiflowTM 3S Nasal Cannula |
| Common name | Nasal Cannula |
| Classification name | Respiratory Gas Humidifier (accessory to) |
| | Class II (21 CFR §868.5450 |
| | Product Code BTT (Anaesthesiology) |
| Predicate Device | F&P Optiflow+ Nasal Cannula – K162553 |
## Device Description
The F&P Optiflow™ 3S Nasal Cannula is intended to be used in a hospital / institutional environment. A dual 22 mm male ISO taper and AIRVO™ specific connector allows connection to breathing circuits to receive humidified breathing gases from F&P 850™ and AIRVO™ systems. The F&P Optiflow™ 3S Nasal Cannula is offered in three sizes being small (S), medium (M) and large (L). The F&P Optiflow™ 3S Nasal Cannula is a prescription only device, provided in a nonsterile state.
{4}------------------------------------------------
### Indications for Use Statement
The F&P Optiflow™ 3S nasal cannula is a 14 day single use nasal cannula interface for use with specified respiratory gas humidifiers to deliver Nasal High Flow (NHF) therapy to spontaneously breathing adult patients. This product is designed to be used by appropriately qualified healthcare professionals in a hospital / institutional environment.
## Non-Clinical Performance Data
### Performance Testing
Performance testing of the F&P Optiflow™ 3S Nasal Cannula was completed to demonstrate substantial equivalence of the F&P Optiflow™ 3S Nasal Cannula to the predicate device.
- Shelf life simulation was based on ASTM F1980-07, ISO 11607-1:2006/A1 2014 and ISO . 291:2008(E).
- . Transportation simulation was based on ISTA 2A Packaged-Products weighing 150lb (68kg) or less.
- Relevant Humidification Output and Thermal Overshoot testing. .
- Gas path and headgear joint and connection strength testing following selected pre-conditioning . intended to simulate worst case life scenarios.
- . Gas path leak testing following selected pre-conditioning intended to simulate worst case life scenarios.
- . 24-hour maximum accumulated and emitted condensation testing intended to simulate worst case.
The F&P Optiflow™ 3S Nasal Cannula has been tested to applicable requirements of the following standards:
- ISO 5356-1:2015 Anaesthetic and respiratory equipment- Conical connectors: Part 1: Cones . and sockets
### Biocompatibilitv
- ISO 10993-1:2009, Biological evaluation of medical devices Part 1: Evaluation and testing . within a risk management process.
- ISO 10993-3:2014, Biological evaluation of medical devices Part 3: Tests for Genotoxicity, ● Carcinogenicity, and reproductive toxicity.
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
- . ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic Toxicity
- ISO 10993-12:2012, Biological evaluation of medical devices Part 12: Sample preparation and ● reference material
{5}------------------------------------------------
- ISO 10993-17:2002, Biological evaluation of medical devices Part 17: Establishment of . allowable limits for leachable substances
- . ISO 10993-18:2005, Biological evaluation of medical devices - Part 18: chemical characterization of materials
- ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare . applications, Part 1: Evaluation and testing within a risk management process
- . ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications, Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare ● applications, Part 3: Tests for emissions of volatile organic compounds (VOCs)
### Technological Characteristics Comparison
A comparison of the F&P Optiflow™ 3S Nasal Cannula and the predicate device is provided in table below.
{6}------------------------------------------------
| | Subject Device | Predicate Device | Comments | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| | F&P Optiflow 3S Nasal Cannula | F&P Optiflow+ Nasal Cannula | | |
| Intended use / Indications for use | Intended Use /<br>Indications for Use | The F&P Optiflow™ 3S nasal cannula is a<br>14 day single use nasal cannula interface<br>for use with specified respiratory gas<br>humidifiers to deliver Nasal High Flow<br>(NHF) therapy to spontaneously breathing<br>adult patients. This product is designed to<br>be used by appropriately qualified<br>healthcare professionals in a hospital /<br>institutional environment. | Nasal cannula patient interface for delivery<br>of humidified respiratory gases. | The Intended Use / Indications for Use<br>statement for the subject device is more<br>specific than the predicate device. |
| Availability | | Prescription use.<br>(Part 21 CFR 801 Subpart D) | Prescription use.<br>(Part 21 CFR 801 Subpart D) | Identical. |
| Patient Population | | Adult patients. | Adult patients. | Identical. |
| Operating<br>Environment | | Hospital environment. | Home or hospital environment. | Similar.<br>The Operating Environment for the subject<br>device is a subset of the predicate device |
| Reusability | | Single patient use only,<br>< 14 days hospital | Single patient use only,<br>< 14 days hospital | Identical. |
| Specifications | | | | |
| Ambient Operating<br>Temperature | | 18 – 28 °C | 18 – 28 °C | Identical. |
| AIRVO System<br>Specifications | | AIRVO / AIRVO 2 Humidifier with<br>900PT56x-series tube; or tube and<br>chamber kit (e.g. 900PT561)<br><br>Flow Range:<br>OPT1042 (Small) 10 - 50L/min<br>OPT1044 (Medium) 10 - 60 L/min<br>OPT1046 (Large) 10 - 60 L/min | AIRVO / AIRVO 2 Humidifier with<br>900PT56x-series tube; or tube and<br>chamber kit (e.g. 900PT561)<br><br>Flow Range:<br>OPT942 (Small) 10 - 50L/min<br>OPT944 (Medium) 10 - 60 L/min<br>OPT946 (Large) 10 - 60 L/min | Identical. |
| | Subject Device | Predicate Device | Comments | |
| | F&P Optiflow 3S Nasal Cannula | F&P Optiflow+ Nasal Cannula | | |
| F&P 850 System<br>Specifications | MR850 Humidifier in invasive mode, RT<br>series kit with 22 mm heated inspiratory<br>tube and chamber. | MR850 Humidifier in invasive mode, RT<br>series kit with 22 mm heated inspiratory<br>tube and chamber. | Identical. | |
| | Flow Range:<br>OPT1042 (Small) 10 - 60L/min<br>OPT1044 (Medium) 10 - 60 L/min<br>OPT1046 (Large) 10 - 60 L/min | Flow Range:<br>OPT942 (Small) 10 - 60L/min<br>OPT944 (Medium) 10 - 60 L/min<br>OPT946 (Large) 10 - 60 L/min | | |
| Shelf-Life | 3 years. | Shelf-life not claimed on labelling | The subject device claims a 3 year shelf life. | |
| Storage Temperature | -10°C to +50°C | -10°C to +50°C | Identical. | |
| Sterility | Device not provided sterile. | Device not provided sterile. | Identical. | |
| Disposal | Dispose of according to hospital protocol. | Not defined. | Disposal information added in the User<br>Instruction. | |
| Cannula Function & Design | | | | |
| Headgear Release<br>Mechanism | Included. | Not present. | New feature. | |
| User Side Swapping<br>functionality | Included. | Not present. | New feature. | |
| 22 mm Male ISO<br>Taper Connector | Included. | Included. | Identical. | |
| AIRVO Specific<br>Connector | Included. | Included. | Identical. | |
### Comparison of Technological Characteristics with the Predicate Device
{7}------------------------------------------------
#### Comparison of Technological Characteristics with the Predicate Device
### Conclusion
The comparison of features, performance, and intended use demonstrate that the F&P Optifiow™ 3S Nasal Cannula is substantially equivalent to the predicate F&P Optiflow+ Nasal Cannula (K162553).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.