← Product Code [MBI](/productcode/MBI) · K191783

# Parcus Synd-EZ Ti (K191783)

_Parcus Medical, LLC · MBI · Oct 9, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191783

## Device Facts

- **Applicant:** Parcus Medical, LLC
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Oct 9, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Parcus Synd-EZ Ti is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. The Parcus Synd-EZ Ti is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

## Device Story

Synd-EZ Ti is a bone fixation device consisting of two titanium buttons connected by UHMWPE suture; designed for stabilizing syndesmosis during healing. Device functions as an adjustable, knotless fixation system; replaces traditional screw fixation in specific trauma cases. Used by orthopedic surgeons in clinical/surgical settings. Device provides mechanical stabilization during the healing process; allows for adjustment of length; resists elongation. Benefits include secure fixation without knot-tying requirements. Output is physical stabilization of bone fragments/syndesmosis.

## Clinical Evidence

Bench testing only. Performance evaluated via mechanical testing on worst-case configurations in test fixtures. Metrics included strength and elongation under cyclic loading and ultimate failure conditions. Results demonstrated substantial equivalence to predicate device performance.

## Technological Characteristics

Materials: Medical-grade titanium and UHMWPE suture. Design: Two-button system with connecting suture; adjustable, knotless fixation. Sterilization: Provided sterile. No software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Parcus GFS Ultimate ([K162198](/device/K162198.md))
- Arthrex Tightrope™ Syndesmosis Device ([K043248](/device/K043248.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.

Parcus Medical, LLC Paul Vagts Director of Regulatory Affairs 6423 Parkland Drive Sarasota, Florida 34243

Re: K191783

Trade/Device Name: Parcus Synd-EZ Ti Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HTN Dated: September 9, 2019 Received: September 10, 2019

Dear Mr. Vagts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

#### K191783

Device Name

## Parcus Synd-EZ Ti

Indications for Use (Describe)

The Parcus Synd-EZ Ti is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

The Parcus Synd-EZ Ti is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for PARCUS MEDICAL, LLC. The word "PARCUS" is in large, blue, sans-serif font, with a red square to the left of the word. Below the word "PARCUS" is the phrase "MEDICAL, LLC." in a smaller, blue, sans-serif font.

Surgical Innovation | Customer Driven

| 510(k) Summary – K191783  |                                                                                                                 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------|
| 510(k) Owner & Submitter: | Parcus Medical, LLC                                                                                             |
|                           | 6423 Parkland Dr                                                                                                |
|                           | Sarasota, FL 34243                                                                                              |
| Company Contact:          | Paul Vagts                                                                                                      |
|                           | Phone: (941)755-7965                                                                                            |
|                           | Fax: (941)755-6543                                                                                              |
| Date Prepared:            | September 9th, 2019                                                                                             |
| Device Trade Name:        | Synd_EZ Ti                                                                                                      |
| Common Name:              | Fastener, Fixation, Non-Degradable, Soft Tissue                                                                 |
| Device Class:             | Class II                                                                                                        |
| Classification Name:      | 21 CFR 888.3040 - Product Code MBI                                                                              |
| Predicate Device:         | Parcus GFS Ultimate, K162198, cleared 11/18/16,<br>Arthrex Tightrope™ Syndesmosis Device, K043248,<br>2/16/2005 |

## Device Description:

The Parcus Synd-EZ Ti consists of two buttons with suture connecting them and are designed to be used for stabilizing the syndesmosis during healing. Building upon the technology of the Parcus GFS Ultimate, the Synd-EZ Ti is easily adjusted for length and resists elongation without requiring knots to secure the fixation. The device is made from medical grade titanium and UHMWPE suture and is provided sterile.

## Intended Use:

The Parcus Synd-EZ Ti is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

The Parcus Synd-EZ Ti is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

## Substantial Equivalence Summary:

The Synd-EZ Ti is very similar to the predicate Parcus Medical GFS Ultimate in that it is comprised of the same materials, intended for the same indications and utilizes similar designs. LAL testing was conducted on a representative device and it was concluded that the Synd-EZ Ti does not raise any additional concerns regarding pyrogenicity.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "PARCUS MEDICAL, LLC.". The word "PARCUS" is in large, bold, blue font. To the left of the word is a red square with a white line running through it. Below the word "PARCUS" is the phrase "MEDICAL, LLC." in a smaller, blue font.

Surgical Innovation | Customer Driven

## Summary Performance Data:

The Synd-EZ Ti was evaluated and testing was conducted on the worst case configurations. Devices were assembled to test blocks and placed in a test fixture. Devices were evaluated for strength and elongation under cycle loading and ultimate failure conditions. Results were compared with test data for the predicate device and demonstrated substantial equivalency.

---

**Source:** [https://fda.innolitics.com/device/K191783](https://fda.innolitics.com/device/K191783)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
