← Product Code [DZE](/productcode/DZE) · K191751

# EM Provisional (K191751)

_Hiossen, Inc. · DZE · Dec 16, 2019 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191751

## Device Facts

- **Applicant:** Hiossen, Inc.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Dec 16, 2019
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).

## Device Story

EM Provisional is a one-piece, screw-form endosseous dental implant; manufactured from titanium alloy (Ti-6Al-4V). Device serves as temporary support for provisional prosthetic devices during healing phase of permanent implants. Surgically placed in upper or lower jaw arches by dental professionals. Machined surface finish. Supplied sterile via radiation. Device provides temporary stabilization; allows immediate loading in edentulous sites. Clinical benefit includes maintaining prosthetic function during osseointegration of permanent implants.

## Clinical Evidence

No clinical data submitted. Substantial equivalence is supported by non-clinical performance data, including biocompatibility, sterilization validation, surface characterization, fatigue testing, compressive strength, and pull-out testing, all of which rely on the established performance of the predicate device.

## Technological Characteristics

Material: Titanium alloy Ti-6Al-4V (ASTM F 136). Form: One-piece screw-form fixture and abutment. Diameters: 1.8mm, 2.5mm. Lengths: 10.0mm, 13.0mm, 15.0mm. Surface: Machined finish. Sterilization: Radiation. Connectivity: None (mechanical device).

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- MS System (Provisional) ([K072871](/device/K072871.md))

## Reference Devices

- Hiossen Implant System ([K140934](/device/K140934.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 16, 2019

Hiossen, Inc. Peter Lee QA/RA Manager 85 Ben Fairless Drive Fairless Hills, Pennsylvania 19030

Re: K191751

Trade/Device Name: EM Provisional Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: September 16, 2019 Received: September 17, 2019

Dear Peter Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. There is a red swoosh above the "H". Below the word "Hiossen" is the word "Implant" in smaller, gray letters.

85 Ben Fairless Drive airless Hills, PA 19030 888-768-0001 www.hiossen.com

# Section 5 Indication for Use Statement - 1 PAGE

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name EM Provisional

Indications for Use (Describe)

The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).

Type of use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

[ Over-The-Counter Use (21 CFR 807 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Service (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters. The word "Implant" is written in gray letters below the word "Hiossen."

5 Ben Fairless Drive Hills, PA 19030 www.hiossen.com

# Section 6 510(k) Summary - 4 PAGES

| (1) Submitter Information: |  |
|----------------------------|--|
|----------------------------|--|

| Submitted by:                                                                                                            | Hiossen, Inc.<br>85 Ben Fairless Drive<br>Fairless Hills, PA 19030              |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Contact Person:<br>Telephone Number:<br>Fax Number:                                                                      | Peter Lee<br>267-759-7031<br>267-759-7031                                       |
| Date Prepared:                                                                                                           | November 8, 2019                                                                |
| (2) Device Name:<br>· Proprietary Name:<br>• Classification Name:<br>• CFR Number:<br>• Device Class:<br>• Product Code: | EM Provisional<br>Implant, Endosseous, Root-form<br>872.3640<br>Class II<br>DZE |

(3) Predicate Device:

| Device                  | 510(k)  | Manufacturer     |
|-------------------------|---------|------------------|
| MS System (Provisional) | K072871 | OSSTEM Co., Ltd. |
| Hiossen Implant System* | K140934 | Hiossen, Inc.    |

*Reference device

# (4) Description of Device:

The EM Provisional is a dental implant made of titanium alloy (Ti-6Al-4V) and is supplied sterile. The surface is machined finished and is intended to be surgically placed in the bone of the upper or lower jaw arches.

The EM Provisional is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the MS System (Provisional), OSSTEM Implant Co., Ltd, K072871

| Device         | Diameter(mm) | Thread Length(mm) |
|----------------|--------------|-------------------|
| EM Provisional | Ø1.8, Ø2.5   | 10.0, 13.0, 15.0  |

# (5) Indication For Use:

The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters. The "H" and the top part of the "i" are red, while the rest of the letters are gray. Below the word "Hiossen" is the word "Implant" in smaller, gray letters.

85 Ben Fairless Drive Fairless Hills, PA 19030 888-768-0001 www.hiossen.com

# (6) Substantial Equivalence:

The information and date provided in this submission established the EM Provisional is substantially equivalent to the primary predicate device, OSSTEM MS System (Provisional) in (510(k) number K072871).

|               | Proposed Device<br>EM Provisional                                                                                                                                                                                                                                                | Predicate Device<br>MS System (Provisional)                                                                                                                                                                                                                                             | Reference Device<br>HIOSSEN IMPLANT<br>SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device        | EM Provisional                                                                                                                                                                                                                                                                   | MS System (Provisional)                                                                                                                                                                                                                                                                 | HIOSSEN IMPLANT<br>SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Manufacturer  | Hiossen, Inc.                                                                                                                                                                                                                                                                    | Osstem Co., Ltd.                                                                                                                                                                                                                                                                        | Hiossen, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 510(K) No.    | New device                                                                                                                                                                                                                                                                       | K072871                                                                                                                                                                                                                                                                                 | K140934                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Design        | Image: EM Provisional Design                                                                                                                                                                                                                                                     | Image: MS System (Provisional) Design                                                                                                                                                                                                                                                   | Image: HIOSSEN IMPLANT SYSTEM Design                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended use  | The EM Provisional is<br>intended to be loaded<br>immediately in partially or<br>fully edentulous mandibles<br>and maxilla to serve as<br>temporary support for<br>provisional prosthetic<br>device during the healing<br>phase of permanent<br>endosseous dental<br>implant(s). | The MS system<br>(Provisional) is intended to<br>be loaded immediately in<br>partially or fully edentulous<br>mandibles and maxilla to<br>serve as temporary<br>support for provisional<br>prosthetic device during<br>the healing phase of<br>permanent endosseous<br>dental implants. | The HIOSSEN Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading. Ultra<br>wide Fixture System is<br>intended to be used in the<br>molar region. |
| Structure     | • Screw form<br>• One piece fixture and<br>abutment                                                                                                                                                                                                                              | • Screw form<br>• One-piece of Fixture and<br>Abutment                                                                                                                                                                                                                                  | • Internal Hex-connected<br>• Submerged Fixture<br>• Tapered body shape &<br>Straight body shape                                                                                                                                                                                                                                                                                                                                                               |
| Diameters     | 1.8, 2.5                                                                                                                                                                                                                                                                         | 1.8, 2.5                                                                                                                                                                                                                                                                                | 3.5~6.8                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Lengths       | 10.0, 13.0, 15.0                                                                                                                                                                                                                                                                 | 10.0, 13.0, 15.0                                                                                                                                                                                                                                                                        | 6.2~18.2                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Material      | Titanium alloy Ti-6Al-4V<br>(ASTM F 136)                                                                                                                                                                                                                                         | Titanium alloy Ti-6Al-4V<br>(ASTM F 136)                                                                                                                                                                                                                                                | Pure Titanium Grade 4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Surface       | Machined finished                                                                                                                                                                                                                                                                | Machined finished                                                                                                                                                                                                                                                                       | - SA(Sandblasted and<br>Acid etched)                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Sterilization | Radiation Sterile                                                                                                                                                                                                                                                                | Radiation Sterile                                                                                                                                                                                                                                                                       | Radiation Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Packaging     | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>tamper-evident outer<br>package.                                                                                                                                                            | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>outer box                                                                                                                                                                                          | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>tamper-evident outer<br>package.                                                                                                                                                                                                                                                                                                                                          |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Hiossen Implant. The logo is composed of the letters "HI" in red, followed by "OSSEN" in gray. The word "IMPLANT" is written in smaller gray letters below the word "OSSEN". There is a registered trademark symbol next to the word "OSSEN".

| S.E. | The proposed device is substantially equivalent with respect to the intended use, structure, material, surface, sterilization and design with the legally marketed predicate device as well as the sterilization and packaging of the legally marketed reference device. The proposed device does not pose any new or increased risks as compared to the legally marketed predicate and reference devices. |
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

### Substantial Equivalence Conclusion

In accordance with the Federal Food Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, HIOSSEN, INC. concludes since the EM Provisional has the same design, intended use, structure, diameters, lengths, material, surface, sterilization as the predicate device, MS System (Provisional) (K072871) and same sterilization and packaging to the reference Hiossen Implant System (K140934) it is substantially equivalent. The propose device does not pose any new or increased risk as compared to both the legally marketed predicate and reference devices.

#### Non-Clinical Performance Data

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include data from the following tests:

#### Biocompatibility Testing

The EM Provisional is manufactured using the same manufacturing process and same well known and well established material as the predicate device, the MS System (Provisional), K072871, therefore we reason it was not necessary to re-test biocompatibility in order to support the biological safety of the EM Provisional.

#### Sterilization Validation

The EM Provisional (manufactured using the same manufacturing process, material and utilizes the same sterilize barrier system) is gamma irradiated under the same conditions and process as the predicate device, the MS System (Provisional), K072871, therefore we reason it was not necessary to re-test validation in order to support the sterilization validity of the EM Provisional.

#### Surface Treatment Characterization Testing

The EM Provisional surface is manufactured using the same manufacturing process, material and is its surface is machined finished just as the predicate device, the MS System (Provisional), K072871, therefore no additional character testing was necessary to support the equivalency of the EM Provisional.

#### Fatique Testing

The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test fatigue on the proposed device since both are the same device.

#### Compressive Strength Testing

The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test compressive strength on the proposed device since both are the same device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters, with the "H" and the top part of the "i" in red, and the rest of the letters in gray. Below the word "Hiossen" is the word "IMPLANT" in gray, with a large amount of space between each letter.

85 Ben Fairless Drive Fairless Hills, PA 19030 888-768-0001 www.hiossen.com

Pull-out Testing The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test pull-out on the proposed device since both are the same device.

Clinical Performance Testing

No clinical performance report(s) is being submitted.

---

**Source:** [https://fda.innolitics.com/device/K191751](https://fda.innolitics.com/device/K191751)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
