The Alpha Omega Disposable Sterile LeadConfirm for MER is intended to be used in neurosurgery for connecting a compatible DBS Lead to a compatible recording and stimulation device. The Alpha Omega Disposable Sterile LeadConfirm for MER is indicated for assisting Neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of a compatible DBS Lead. The Alpha Omega Disposable Sterile LeadConfirm should only be used to connect a compatible DBS Lead to a compatible recording and stimulation device.
Device Story
Sterile LeadConfirm cables and adaptors serve as hardware interfaces connecting deep brain stimulation (DBS) leads to Alpha Omega's Neuro Omega or NeuroSmart physiological navigation systems. Used in the operating room by neurosurgeons during functional neurosurgery, these devices facilitate intraoperative neurophysiological recording and stimulation. The system enables the transmission of electrical signals between the implanted DBS lead and the recording/stimulation console, aiding the surgeon in precise lead placement. The device supports stimulation parameters including square wave/Lili pulses, amplitudes up to 12mA, and durations up to 500uSec. By providing a reliable connection, the device assists in the clinical decision-making process regarding optimal lead positioning, potentially improving patient outcomes in functional neurosurgical procedures.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed design outputs meet design inputs. Sterilization validation performed using Gamma radiation (VDmax method) at 20kGy achieved a Sterility Assurance Level (SAL) of 10^-6.
Technological Characteristics
Sterile, disposable hardware cables and adaptors. Materials verified for stability post-irradiation. Sterilization via Gamma radiation (20kGy). Connectivity: physical interface between DBS leads (Medtronic, Boston Scientific, Abbott) and Alpha Omega navigation systems. No software or active electronic components; passive hardware.
Indications for Use
Indicated for assisting neurosurgeons in the operating room during functional neurosurgery to aid in the placement of a compatible DBS lead in patients requiring functional neurosurgery.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Alpha Omega Engineering Ltd. Maysana Mousa QA RA Manager Nazareth Industrial Park, Mount Precipice, St. 2015 Nazareth 1612102. Israel
Re: K191739
Trade/Device Name: Sterile LeadConfirm Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: November 25, 2019 Received: December 2, 2019
Dear Mrs. Maysana Mousa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191739
Device Name Sterile LeadConfirm
Indications for Use (Describe)
The Alpha Omega Disposable Sterile LeadConfirm for MER is intended to be used in neurosurgery for connecting a compatible DBS Lead to a compatible recording and stimulation device.
The Alpha Omega Disposable Sterile LeadConfirm for MER is indicated for assisting Neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of a compatible DBS Lead.
The Alpha Omega Disposable Sterile LeadConfirm should only be used to connect a compatible DBS Lead to a compatible recording and stimulation device.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of a stylized letter "A" in maroon color, with a gray bridge-like structure at the bottom. To the right of the "A" is the word "omega" in gray. Below the logo is the text "Defining Neuroscience Technology" in maroon color.
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92
## Submitter Information
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Nof Hagalil (Nazareth Elite) 1789062, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075
#### Submission contact person:
Mrs. Maysana Mousa Tel:+972-4-6563-327 Fax:+972-4-6574-075 Mobile: +972-54-9043303
#### Device Classification
| Proprietary Device Name: | Sterile LeadConfirm |
|----------------------------|--------------------------------------------------------------------|
| Common name: | Intraoperative neurophysiological recording and stimulating device |
| Product Code: | GZL |
| Classification Name: | Depth Electrode |
| Classification Regulation: | 21 CFR 882.1330 |
| Regulatory Class: | II |
| Predicate Devices | |
Sterile Disposable Recording Cables K120098
Page 1 of 10
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Alpha Omega. The logo features the Greek letters alpha and omega, with the word "ALPHA" in red and the word "omega" in gray. The words "Defining Neuroscience Technology" are written in smaller font below the logo. The alpha and omega symbols are stylized, with the alpha symbol resembling a triangle and the omega symbol resembling a semicircle.
### Device Description
Alpha Omega's Sterile LeadConfirm cable and adaptor connects the DBS lead to Alpha Omega's approved physiological navigation systems for recording and stimulation system: Neuro Omega (K171581) and NeuroSmart (K172042).
The purpose of the submission is to introduce the Sterile LeadConfirm Cable and Adaptor, the subject devices are sterile Hardware only.
There are three types of the subject device, Alpha Omega's Sterile LeadConfirm Cables:
- 1) Sterile LeadConfirm M Cable that connects to Medtronic Lead P960009 (3389, 3387)
- 2) Sterile LeadConfirm B Cable that connects to Boston Scientific Lead P150031 (BSC-DB-2201, BSC-DB-2202)
- 3) Sterile LeadConfirm A Cable that connects to Abbott Lead P140009 (6170/ 6171/ 6172/ 6173/ 6178/ 6170/ 6180/ 6181)
In addition, there are two types of Sterile LeadConfirm Adaptors:
- 1) Sterile LeadConfirm M Adaptor
- 2) Sterile LeadConfirm A Adaptor
AOE (Alpha Omega Engineering) stimulation with LeadConfirm cable and adaptor is identical to IPG/ test stimulator, with the following parameters:
- a) Pulse shape: Square wave, Lili pulse
- b) Amplitude up to 12mA
- c) Duration up to 500uSec
#### Indications for Use Statement
The Alpha Omega Disposable Sterile LeadConfirm for MER is intended to be used in neurosurgery for connecting a compatible DBS Lead to a compatible recording and stimulation device.
The Alpha Omega Disposable Sterile LeadConfirm for MER is indicated for assisting Neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of a compatible DBS Lead.
Page 2 of 10
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Nof Hagalil 1789062, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com www.alphaomega-eng.com
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Alpha Omega. The logo has the letters Alpha and Omega in it. The word Alpha is in a dark red color, and the word Omega is in a gray color.
#### -Defining Neuroscience Technology
The Alpha Omega Disposable Sterile LeadConfirm should only be used to connect a compatible DBS Lead to a compatible recording and stimulation device
Page 3 of 10
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Alpha Omega, a company that defines neuroscience technology. The logo features the letters "A" and "O" in a stylized font, with the "A" in maroon and the "O" in gray. Above the letters "ALPHA" is a maroon line that resembles a heartbeat. Below the logo is the tagline "Defining Neuroscience Technology" in a simple, sans-serif font.
#### Safety & Effectiveness
The Sterile LeadConfirm devices have been compared to the predicate device, Sterile Disposable Recording Cables (under K12008), in terms of intended use, indications for use, components, principles of operation, technological characteristics and safety features.
#### Intended Use Comparison .
| # | Comparison<br>parameter | Subject device:<br>Sterile LeadConfirm | Predicate device:<br>Sterile Disposable Recording<br>Cables | Substantial Equivalent discussion |
|---|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| 1 | Legally<br>distribution<br>clearance No. | Subject device | K120098 | |
| 2 | Owner | Alpha Omega Engineering<br>Ltd. | Alpha Omega Engineering<br>Ltd. | |
| 3 | Intended use and<br>indications for<br>use. | The Alpha Omega Disposable<br>Sterile LeadConfirm for MER<br>is intended to be used in<br>neurosurgery for connecting a<br>compatible DBS Lead to a<br>compatible recording and<br>stimulation device.<br><br>The Alpha Omega Disposable<br>Sterile LeadConfirm for MER<br>is indicated for assisting | Alpha Omega's Sterile<br>Disposable Recording Cables<br>for MER are intended to be<br>used in neurosurgery for<br>connecting NeuroProbes to<br>recording and stimulation<br>device. | Similarities:<br>Similar<br>Differences:<br>None |
| # | Comparison<br>parameter | Subject device:<br>Sterile LeadConfirm | Predicate device:<br>Sterile Disposable Recording | Substantial Equivalent discussion |
| | | | Cables | |
| | | Neurosurgeons, in the<br>operating room during<br>functional neurosurgery, to aid<br>in the placement of a<br>compatible DBS Lead.<br>The Alpha Omega Disposable<br>Sterile LeadConfirm should<br>only be used to connect a<br>compatible DBS Lead to a<br>compatible recording and<br>stimulation device | | |
| 4 | indications for<br>use and<br>environment | Alpha Omega's Disposable<br>Sterile LeadConfirm for MER<br>is indicated for assisting<br>Neurosurgeons, in the<br>operation room during<br>functional neurosurgery, to aid<br>in placement of Lead. | Alpha Omega's Sterile<br>Disposable Recording Cables<br>for MER are indicated for<br>assisting Neurosurgeons, in the<br>operation room during<br>functional neurosurgery, to aid<br>in placement of depth<br>electrodes. | Similarities:<br>Similar<br>Differences:<br>None |
| 5 | Device code and<br>regulation | Product Code:<br>GZL<br>Regulation #: 21CFR882.1330 | Product Code:<br>GZL<br>Regulation #: 21CFR882.1330 | Similarities:<br>Identical<br>Differences:<br>None |
| # | Comparison<br>parameter | Subject device:<br>Sterile LeadConfirm | Predicate device:<br>Sterile Disposable Recording<br>Cables | Substantial Equivalent discussion |
| 1 | Legally<br>distribution<br>clearance No. | Subject device | K120098 | |
| 2 | Owner | Alpha Omega Engineering<br>Ltd. | Alpha Omega Engineering<br>Ltd. | |
| 3 | Target treatment<br>Population | All patients that need<br>functional neurosurgery | All patients that need<br>functional neurosurgery | Similarities:<br>Identical<br>Differences:<br>None |
| 4 | Human Factors | Used by professional<br>Neurosurgeons | Used by professional<br>Neurosurgeons | Similarities:<br>Identical<br>Differences:<br>None |
| 5 | Use environment | Operating Room | Operating Room | Similarities:<br>Identical<br>Differences:<br>None |
Page 4 of 10
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Nof Hagalil 1789062, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com www.alphaomega-eng.com
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the words "ALPHA" and "omega". The word "ALPHA" is in a dark red color and has a line going through it. The word "omega" is in gray and is located below the word "ALPHA".
Defining Neuroscience Technology
Page 5 of 10
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Alpha Omega, a company that defines neuroscience technology. The logo features the letters "Alpha" and "Omega" in a stylized font. The word "Alpha" is in red, and the word "Omega" is in gray. The tagline "Defining Neuroscience Technology" is below the logo.
#### Performance Comparison .
Page 6 of 10
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Nof Hagalil 1789062, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com www.alphaomega-eng.com
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the Greek letters alpha and omega, with the word "ALPHA" written in a stylized font above the word "omega". The alpha symbol is red, while the omega symbol is gray. The word "ALPHA" is also red, and the word "omega" is gray.
Defining Neuroscience Technology
| # | Comparison<br>parameter | Subject device:<br>Sterile LeadConfirm | Predicate device:<br>Sterile Disposable Recording<br>Cables | Substantial Equivalent discussion |
|---|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| 6 | Integrity of<br>materials and<br>functionality<br>after sterilization | Functionality and materials<br>stability, including packaging<br>verified, after irradiation<br>sterilization. | Functionality and materials<br>stability, including packaging<br>verified, after irradiation<br>sterilization. | Similarities:<br>Identical<br>Differences:<br>None |
| 7 | Sterility<br>achievement<br>method | Sterilized by Alpha omega<br>using irradiation sterilization<br>and delivered sterile, validated<br>process | Sterilized by Alpha omega<br>using irradiation sterilization<br>and delivered sterile. | Similarities:<br>Identical<br>Differences:<br>None |
Page 7 of 10
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Nof Hagalil 1789062, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com www.alphaomega-eng.com
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Alpha Omega. The logo features the letters "Alpha" and "Omega" in a stylized font. The word "Alpha" is in a dark red color, while the word "Omega" is in gray. Below the words is the phrase "Defining Neuroscience Technology" in a smaller font and dark red color.
Based on the performance results provided in this submission (including test results) and the analysis of similarities and differences presented above, Alpha Omega Technologies Ltd. believes that the proposed device is substantially equivalent to the predicate device without raising new safety and/or effectiveness issues.
### Rational for Substantial Equivalency
The proposed Sterile LeadConfirm devices have been compared to the predicate and reference device in terms of intended use, indications for use, components, principles of operation, technological characteristics and safety features.
Based on the performance results provided in this submission (including test results and clinical data) and the analysis of similarities and differences presented above, Alpha Omega believes that the proposed device is substantially equivalent to the predicate device without raising new safety and/or effectiveness issues.
#### Substantial Equivalence Statement
Based on the above, it is Alpha Omega's opinion that the proposed Sterile LeadConfirm devices are substantially equivalent in terms of design principles, performance features and of safety & effectiveness to the predicate legally cleared devices referred to in section 4 of this document.
#### Non Clinical and Clinical validation data
Verification and validation bench tests were performed, to demonstrate the safety and effectiveness of the proposed Sterile LeadConfirm devices.
The purposes of the design verification & validation processes were to verify the Device specifications and to verify the proper function of all device components and options.
After comparing the predicate device to the subject device, results show that with the above intended use, the device is equivalent in safety and effectiveness.
Therefore, the subject devices of this 510(k) notification, the Sterile LeadConfirm devices, did not clinical studies to to support safety and effectiveness of of the require device.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the Greek letters alpha and omega, with the word "ALPHA" in red and the word "omega" in gray. The alpha symbol is stylized with a red arch above a gray bridge-like structure. A red line extends from the top of the "A" in ALPHA, creating a wave-like pattern above the word.
-Defining Neuroscience Technology Performance Tests
| Test | Test Method Summary | Results |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verification and<br>Validation | This verification performed on<br>complete Sterile LeadConfirm<br>devices, the subject devices, and<br>checked that the design output<br>meets the design input | Sterile LeadConfirm devices have<br>been verified under a complete<br>verification plan traceable to Sterile<br>LeadConfirm design input. All<br>samples passed the acceptance<br>criteria which determines the<br>effectiveness of Sterile<br>LeadConfirm devices. |
| Sterilization<br>Validation<br>LeadConfirm<br>Adaptors final<br>Report | This Sterilization validation<br>performed on complete Sterile<br>LeadConfirm Adaptors | The purpose of the study was to<br>validate the effectiveness of the<br>Gamma Radiation Sterilization for<br>Alpha Omega Engineering Ltd.<br>LeadConfirm adaptors, according<br>to VDmax Method. According to<br>the test results (section 13), the<br>gamma radiation sterilization<br>process of LeadConfirm Adaptors<br>Cat. No. STR-000071-00, STR-<br>000072-00 and STR-000073-00, at<br>a sterilization dose of 20kGy, was<br>sustained. This dose gave a SAL of<br>6 magnitudes as requested by the<br>standards |
| Sterilization<br>Validation<br>LeadConfirm B<br>Cable final Report | This Sterilization validation<br>performed on complete Sterile<br>LeadConfirm B Cable | The purpose of the study was to<br>validate the effectiveness of the<br>Gamma Radiation Sterilization for<br>LeadConfirm B Cable, according to<br>Vdmax Method. According to the<br>test results (Section 13), the gamma<br>radiation sterilization process of<br>Electrodes lead cable, at a<br>sterilization dose of 20kGy, was<br>sustained. This dose gave a SAL of<br>6 magnitudes as requested by the<br>standards |
| Sterilization<br>Validation | This Sterilization validation<br>performed on complete Sterile | The purpose of the study was to<br>validate the effectiveness of the |
| -Defining Neuroscience Technology | | |
| LeadConfirm M &<br>A Cables final<br>Report | LeadConfirm M & A Cables | Gamma Radiation Sterilization for<br>LeadConfirm M & A Cables,<br>according to Vdmax Method.<br>According to the test results<br>(Section 13), the gamma radiation<br>sterilization process of Electrodes<br>lead cable, at a sterilization dose of<br>20kGy, was sustained. This dose<br>gave a SAL of 6 magnitudes as<br>requested by the standards |
Page 9 of 10
Alpha Omega Engineering Ltd. Hamerkava St. 6 Tsiporit Industrial Zone Not Hagali 1789062, Israel
Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image shows a logo with the words "ALPHA omega" in a stylized font. The word "ALPHA" is in a dark red color, and the word "omega" is in gray. The "A" in "ALPHA" is stylized with a dark red arch over a gray bridge-like structure. A red line extends from the "A" in "ALPHA" to the "A" in "omega", creating a visual connection between the two words.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.