← Product Code [JAA](/productcode/JAA) · K191713

# CS-series-FP Radiographic / Fluoroscopic Systems with Optional CA-100S / FluoroShield™ (ROI Assembly) (K191713)

_Omega Medical Imaging, LLC · JAA · Oct 4, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191713

## Device Facts

- **Applicant:** Omega Medical Imaging, LLC
- **Product Code:** [JAA](/productcode/JAA.md)
- **Decision Date:** Oct 4, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML

## Indications for Use

The Omega Medical Imaging, LLC CS-series-FP (SSXI) systems with optional accessory device CA-100S as a modification device to provide an automated Region of interest that reduces exposure to the patient and operator. The system is intended for use in Radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging for General Populations. At no time will the CA-100S be considered as a replacement for the primary collimator shall always be used, in accordance with good medical practice, to define a Region of Interest

## Device Story

The CS-series-FP is a radiographic/fluoroscopic X-ray system featuring the CA-100S/FluoroShield™ secondary collimator. The device takes X-ray image inputs from a Varex 2020 flat panel detector. The CA-100S uses a shutter assembly (four lead blades driven by magnetic coils) to define a Region of Interest (ROI) within the primary collimator's field of view. An Auto-ROI Processor analyzes images to automatically track devices (e.g., catheters, endoscopes) and adjust shutter positions. The system produces a 'blended' image, combining high-frequency live ROI data with lower-frequency peripheral background data. Used in interventional suites by physicians and technicians, the device reduces X-ray scatter and patient/staff radiation exposure by shielding anatomy outside the ROI. The output is displayed on a monitor, allowing clinicians to maintain peripheral visualization and orientation without compromising workflow. The system is permanently floor-mounted with motorized C-arm positioning.

## Clinical Evidence

Prospective clinical study of 100 patients undergoing fluoroscopy-guided endoscopic procedures compared conventional fluoroscopy to the FluoroShield system. Results demonstrated up to 61.8% reduction in patient radiation exposure (DAP) and 59.4% reduction in staff scatter radiation, with no significant differences in demographics, BMI, or procedural times.

## Technological Characteristics

System includes X-ray generator, Varex 2020 flat panel detector, and CA-100S secondary collimator. Shutter assembly uses lead blades, magnetic coils, and spring retraction. Enclosure made of plated steel and aluminum. Connectivity via Gigabit Ethernet (CAT5E/CAT6). Complies with 21 CFR 1020.30-32, EN 60601-1, IEC 60601-1-2, and IEC 60601-2-54. Software performs image blending and automated ROI tracking.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- CS-series-FP Fluoroscopic / Radiographic Systems with Optional CA-100S (ROI Assembly) ([K182834](/device/K182834.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Omega Medical Imaging, LLC c/o John Newman Manager of Regulatory Affairs 3400 St. Johns Parkway, Suite 1020 SANFORD, FL 32771

October 4, 2018

# Re: K191713

Trade/Device Name: CS-series-FP Radiographic / Fluoroscopic Systems with Optional CA-100S / FluoroShield™ (ROI Assembly) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: JAA, OWB Dated: May 2, 2019 Received: June 26, 2019

# Dear John Newman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### Device Name

CS-series-FP Radiographic / Fluoroscopic Systems with Optional CA-100S / FluoroShield (ROI Assembly)

#### Indications for Use (Describe)

The Omega Medical Imaging, LLC CS-series-FP (SSXI) systems with optional accessory device CA-100S as a modification device to provide an automated Region of interest that reduces exposure to the patient and operator. The system is intended for use in Radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging for General Populations. At no time will the CA-100S be considered as a replacement for the primary collimator shall always be used, in accordance with good medical practice, to define a Region of Interest

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Omega Medical Imaging. The logo consists of a blue stylized omega symbol on the left, followed by the words "OMEGA MEDICAL" in a gray sans-serif font. Below "OMEGA MEDICAL" is the word "IMAGING" in a smaller, lighter gray font, with hyphens separating the letters.

|                          | Traditional 510(k) SUMMARY K191713                                                                        |
|--------------------------|-----------------------------------------------------------------------------------------------------------|
| Company Name:            | Omega Medical Imaging, LLC                                                                                |
| Address:                 | 3400 St. Johns Parkway, Suite 1020, Sanford, FI 32771                                                     |
| Telephone No:            | 407-323-9400                                                                                              |
| Registration No:         | 1052701                                                                                                   |
| Contact person:          | John Newman (Manager of Regulatory Affairs)                                                               |
| Date Prepared:           | 08/28/2019                                                                                                |
| Device (trade) name:     | CS-series-FP Radiographic / Fluoroscopic Systems with Optional CA-<br>100S / FluoroShield™ (ROI Assembly) |
| Common/usual name:       | Fluoroscopic/Radiographic X-ray system                                                                    |
| Classification Name:     | Image-intensified Fluoroscopic x-ray system                                                               |
| Classification Panel:    | Radiology                                                                                                 |
| CFR section:             | 892.1650                                                                                                  |
| Device Class:            | Class II                                                                                                  |
| Primary Product code:    | JAA                                                                                                       |
| Secondary product code:  | OWB                                                                                                       |
| Predicate Device K182834 |                                                                                                           |
| Company Name:            | Omega Medical Imaging, LLC                                                                                |
| Address:                 | 675 Hickman Circle, Sanford, Florida 32771                                                                |
| Telephone No:            | 407-323-9400                                                                                              |
| Registration No:         | 1052701                                                                                                   |
| Contact person:          | John Newman                                                                                               |
| Date Prepared:           | 02/13/2019                                                                                                |
| Device (trade) name:     | CS-series-FP Fluoroscopic / Radiographic Systems with Optional CA-<br>100S (ROI Assembly) K182834         |
| Common/usual name:       | Fluoroscopic/Radiographic X-ray system                                                                    |
| Classification Name:     | Image-intensified fluoroscopic x-ray system                                                               |
| Classification Panel:    | Radiology                                                                                                 |
| CFR section:             | 892.1650                                                                                                  |
| Device Class:            | Class II                                                                                                  |
| Primary Product code:    | JAA                                                                                                       |
| Secondary Product code:  | OWB                                                                                                       |

Traditional 510(k) Summary for Optional CA-100S / FluoroShield™ Region of Interest Device, Rev. 4 page-1

{4}------------------------------------------------

## Device description:

This 510(k) submission is for the addition of an optional secondary fast collimator system (Model: CA-100S / FluoroShield™ K182834 cleared 02/18/19). The CA-100S/ FluoroShield™ will be used with the following FDA cleared device: Omega's CS-series-FP ( K100102) which utilizes a Varex 2020 Flat Panel Detector. (Note: the only difference between this submission and the predicate is the size of the FPD. The predicate device utilizes a Varex 3030 Flat Panel Detector)

The-CA-100S/ FluoroShield™ system is composed of a Shutter, a ROI Electronics Unit, an Auto-ROI Processor, two ROI Control Panels and a Monitor (See Figure 1) .

Conventional Collimation

Image /page/4/Picture/5 description: The image is a black and white medical image, possibly an X-ray or fluoroscopy, showing internal structures. A thin, dark line, likely a catheter or wire, is visible, curving through the image. The image also shows a section of the body, possibly the heart or a major blood vessel, with varying shades of gray indicating different densities of tissue or fluid.

CA-100S/ FluoroShield™ Auto ROI

Image /page/4/Picture/7 description: This image is an x-ray of a human chest. There are medical wires and devices visible inside the chest cavity. A white box highlights a specific area of the x-ray, showing a closer view of one of the medical devices. The image is in grayscale, with the bones and medical devices appearing as lighter shades against the darker background of the soft tissues.

Optimized Dose Reduction: Advanced Auto Collimation

Omega's (CA-100S/ FluoroShield™) allows for auto collimation while maintaining a perspective of surrounding anatomy. The blended image incorporates a lower frequency refresh of the peripheral image area. This combined image (live fluoroscopy of ROI + background refreshed at a rate of once or twice per second) increases the quality of information presented during interventional procedures.

- . Image quality is improved via auto-collimation resulting in a reduced FOV and subsequently less X-ray scatter.
- . Anatomical landmarks and devices visible outside the Region of Interest (ROI) provide important clinical information which are viewed at a reduced exposure level / Rate.
- The CA-100S can be sized and positioned in manual mode, as opposed to conventional collimation, which is generally limited to positioning about the center of the image.
- Auto ROI automatically follows the movement of devices i.e. endoscopes, catheters, etc., minimizing . distraction and input requirements for the operator.

{5}------------------------------------------------

Image /page/5/Picture/0 description: This image is a diagram of an X-ray tube system. The diagram shows the secondary collimator CA-100S mounted below the primary collimator. The diagram also shows the ROI control panels, the ROI electronics unit (1), the auto ROI processor (2), and the monitor (5). The auto ROI processor has ports labeled A and B, and a port labeled 3, and the monitor has a port labeled 4.

During a fluoroscopy procedure, the physician's attention is focused on a narrow region of interest (ROI) or a small region within the full field of view after collimating down to an area that will best suit the procedure with the primary collimator. (Figure 2)

Image /page/5/Picture/2 description: The image is a black and white X-ray of a human chest. There are medical wires visible inside the chest cavity. The wires appear to be connected to a device that is implanted in the chest. The image is likely taken during a medical procedure to implant or adjust the device.

Image /page/5/Picture/3 description: The image shows a medical image, possibly an X-ray or fluoroscopy, of a device implanted within a body. The device appears to be a catheter or lead with visible markers or electrodes along its length. The image is grayscale and shows the device's position within the body, likely in relation to internal organs or structures. The image is set against a black background.

Image /page/5/Picture/4 description: This image contains the text "Figure 2, Primary / Conventional Collimator". The text is written in a clear, sans-serif font and is left-aligned. The phrase "Figure 2" indicates that this is likely a label or title for a diagram or illustration within a larger document.

Figure 3, Secondary Collimator, CA-100S/FluoroShield™, Shutter

Image /page/5/Picture/6 description: This image shows a medical x-ray of a patient's chest. There are medical wires that are inserted into the patient's heart. The wires are used to monitor the patient's heart activity. The image is in black and white.

Figure 4, Blended image from both collimators.

This smaller region of interest (ROI) captured by the secondary collimator (CA-100S) is a small area within the area defined by the pre-existing primary collimator. (See Figure 3) The area outside the ROI receives much less attention by the physician but is exposed. The CA-100S secondary collimator may be used to additionally shield anatomy that is not the primary focus of the physician but is required to maintain peripheral imaging. The CA-100S ROI image processing combines the live ROI with the legacy image of the full field of view (defined by the primary collimator). This allows for the collimation of the ROI without impacting the doctors normal work flow, visualisation, orientation and navigation. (See Figure 4)

The ROI always contains the central beam for accurate calculations of Air Kerma values.

The CA-100S / FluoroShield™ offers additional collimation by shielding parts of the FOV outside the ROI without loss of visual information available to the physician or operator. the X-Ray Generator and Generator Interface are not changed. The system is configured with Omega's permanently floor mounted C-Arm that provides motorized positioning of the x-ray source and image receptor, along with a patient table with options for elevating and tilting.

{6}------------------------------------------------

- System components integrated for this option are: ●
	- CA-100S / FluoroShield™ secondary collimator Components detailed in Yellow. a.
	- Image Processor / Workstation b.
	- Associated electrical / electronic / hardware components (See illustration below) C.

Typical system configuration consisting of integrated components such as the X-Ray generator, Image Processor, collimator, X-ray Tube, Positioner and patient table.

Image /page/6/Figure/5 description: This image shows a diagram of the CA-100S components and FDA-approved system. The diagram includes components such as a detector, ion chamber, primary collimator, secondary collimator, and X-ray tube. It also shows the ROI electronics unit, image processor, auto ROI processor, and generator interface/X-ray generator. The diagram illustrates the flow of image data and status between the different components.

Figure 5

{7}------------------------------------------------

# OPERATION

The CA-100S / FluoroShield™ collimator is installed between the X-ray source and the primary collimator (See Figure 5). The opening of the secondary collimator shutters is adjustable in position. When the secondary collimator shutter function is enabled, only the ROI is continuously exposed and always includes the central beam from the X-ray tube this does not affect Air Kerma rate, Accumulated Air Kerma or Accumulated Fluoro Time as calculations are derived from kV. mA. exposure time and SOD. The secondary collimator shutter periodically opens fully to allow the acquisition of a FOV frame (defined by the primary collimator) that provides the background peripheral area. The image processing software then provides a combined, or "blended" image comprising the ROI frame and the most recently acquired FOV frame. The Product is permanently incorporated in the host fluoroscopy system (Omega's CS-series-FP) and its clinical environment. The system is intended to be installed by trained technicians and operated by professionals trained in its use and the associated medical interventional procedures. The CA-100S/ FluoroShield™ Product functions in ON Mode, OFF Mode or Bypass Mode. In ON Mode, the secondary collimator shutter is activated, and the images are blended. In OFF Mode, the secondary collimator shutter is not driven, and the shutter blades are retracted fully while the images pass through both collimators without blending images (i.e. all image frames are FOV image frames defined by the primary collimator). In BYPASS Mode, the secondary collimator shutter driver is powered down (all shutter blades are retracted) and all received images bypass the CA 100-S/ FluoroShield™ image processing path.

The Product consists of a Shutter, a ROI Electronics Unit, an Auto-ROI Processor with Status / Reduction Monitor and two ROI Control Panels. Their operations are described below.

Image /page/7/Picture/3 description: The image shows a close-up of a robotic arm with a circular platform mounted on top. The platform appears to be made of metal and has a square opening in the center, revealing some internal components. An arrow points to the platform, drawing attention to its structure and placement on the robotic arm.

# Shutter

The shutter consists of four lead blades that are moved into positions defined by the operator or automatically by the ROI Processor to determine and select the Region of Interest. Each shutter blade is moved by the application of a current which drives its magnetic coil. The electromotive force is counterbalanced by the retracting force of the spring. This current is limited to 1 Amp. The driving voltage is limited to ±30V peak.

The Shutter provides position feedback via low voltage (10V peak) signaling. Power for the feedback circuit is provided externally as 12V, limited to approximately 200mA. When not driven, the shutter is mechanically retracted by strong springs. In this state, it is essentially transparent to the optical path of the Host System.

Neither operators nor patients will come into contact with the shutter.

The enclosure is made of plated steel and aluminum. All moving parts and electrical components are completely enclosed. External electrical connections are made with shielded cables. Electrical components are galvanically isolated from the housing. The shutter and associated cables are enclosed by the housing of the Host System.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a gray electronic device with a blue arrow pointing to the front panel. The device has several ports and vents on the back panel. The device has metal brackets on either side with holes for mounting.

# ROI Electronics Unit ("REU")

The REU controls the overall operations of the Product. It receives images from the receptor panel and transmits images to the Host System's image processor. If the received image is an ROI frame, it combines the frame with the background from the most recent FOV frame.

The REU drives the four shutter blades through independent channels and receives position feedback through independent channels. The REU holds the blade positions constant while X-ray emission is in progress. The shutter blades are only moved when there is no X-rav emission.

The REU receives and responds to user inputs via the ROI Control Panels. (See section below for details.)

The REU receives and responds to signals representing the state of a fluoro sequence, the state of X-ray emission, the source-to-image distance, pending or in-progress table movement, flip and rotation offset, and magnification mode.

The REU sends positions of the boundaries of the ROI and the overall operating status to the Host System for display on its main screen.

The REU sends blended images to the Auto-ROI Processor and receives the calculated (recommended) locations for the shutter blades. In auto-ROI mode the REU drives the blades to these locations.

*On detecting conditions that may compromise operations or longevity of the Product, the REU puts the Product in OFF Mode. On detecting conditions that may compromise the image, a loss of communication with, or control over, other modules, the REU puts the Product in Bypass Mode.

The REU is installed in an electronics cabinet. Neither operators nor patients will come into contact with the REU.

The REU's electrical and electronic components are housed in a metallic enclosure. All panels of the enclosure are bonded to each other. A protective earth terminal is provided.

The REU is intended to operate on a 120/240 Vac, 50/60Hz, single-phase output from an uninterruptible power supply. The component power supply unit ("PSU") is

recognized to UL60950-1 and ES60601-1. The Earth terminal of the PSU is bonded to the enclosure.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains two identical panels with a screen and several buttons. The buttons are labeled "Video Bypass", "ROI ON/OFF", "Auto/Manual ROI", and "High/Low Reduction". The screen is black and rectangular. The panels are gray and have rounded corners.

# ROI Control Panels ("RCP")

A bedside RCP and a control room/booth RCP provide facilities for the user to control the operation of the Product. Four buttons and a trackpad are provided on each RCP to control ROI functionality.

The RCP communicates with the REU over a shielded cable which also carries 12V to power the RCP. The 12V power source provides 2A of current to both RCPs and other circuits drawing on this rail. Each RCP consumes about 100mA.

The RCP's electrical and electronics components are fully enclosed in a metallic enclosure that provides ingress protection aqainst fluids (jets or strong sprays). The trackpad is exposed to touch. The touch surface is tied to the enclosure and isolated from the rest of the circuits. A shielded cable connects the RCP to the REU. The shield of the cable is tied to the enclosure.

Image /page/9/Picture/5 description: The image shows a computer setup with a desktop tower, a monitor, and a network adapter. The desktop tower is labeled with the number 2, and it has a sleek design with red accents. The monitor is labeled with the number 5, and it is connected to the network adapter, which is labeled with the number 4.

# Auto-ROI Processor ("ARP")

A dedicated computing device receives and analyzes the processed images from the REU. From these images, the ARP calculates the ROI settings that provide the most useful image to the physician or operator. The ARP also generates display data that indicates the operating status and/or status messages. The ARP communicates with the REU via a single Gigabit Ethernet ("GigE") connection over CAT5E or CAT6 copper cables with magnetic isolation at the connectors.

The ARP is enclosed in a metallic housing and is intended to be installed in a control room or electronics room. It is intended to operate on a 120Vac, 60Hz, single-phase output from an uninterruptible power supply.

{10}------------------------------------------------

# No Mechanical Differences other than the sized of the Flat Panel Detector between Predicate device and this submission:

- No mechanical changes to mount the CA-100S Sutter under the Primary Collimator as illustrated in Figure 6. Both the predicate device and this submission utilize the same mechanical mounting for the CA-100S / FluoroShield™ Shutter assembly.
Image /page/10/Figure/2 description: This image shows a device with several labeled components. The components labeled are the primary collimator, the CA-100S/FluoroShield shutter (secondary collimator), and the X-Ray tube. The image is labeled as Figure 6.

{11}------------------------------------------------

# New and adapted Features for implementation of CA-100S into this submission:

- . This release provides both REU and ARP support for the Varex 2020 Panel based CA-100S system. There are no changes to the existing software that supports the Varex 3030 Panel Based CA-100S system.
- Image resolution: image width and height are set to 1024 to match the 2020 system resolution.
- . Frame rate: image processing and display frame rates are set to 15 frames per second to match the 2020 system's effective frame rate.
- . Frame duplication: system will replay the last good known frame when a dark frame (image with no matching X-RAY exposure) is input
- . Blades' indices switch: The blade location index is updated to match the panel orientation of the 2020 system.
- . Mag modes: Mag mode 1 and Mag mode 2 is set to 80% and 66.7% of normal mode (Mag mode 0) image size, respectively.

### Auto ROI Processor Software

- Image resolution: image width and height are set to 1024 to match the 2020 system resolution. ●
- Frame rate: image processing and display frame rates are set to 15 frames per second to match the . 2020 system's effective frame rate.
- . lgnore duplicate frames in image receiving: a received frame with the same ID as the previous frame is skipped for only processing valid frames (such ID duplication occurs due to the valid frame duplication in the REU to discard the 2020 system's alternation between valid and dark frames).
- . Shutter's ROI size and path in calibration: the ROI's size used in calibration is reduced and its path is trimmed to match the narrower field of view of the 2020 system.
- . Blades' indices switch in calibration: the blades' indices are re-assigned to match the panel orientation of the 2020 system.
- . Segmentation model: updated model based on trained images from the 2020 system.
- Motion detection module and its incorporation in tracking: a module that computes motion detection ● between consecutive frames, based on which a motion map is generated to improve tracking of the ROI in combination with the segmentation map.
- . Initial ROI size in Mag modes: the initial ROI in Mag mode 1 and Mag mode 2 is set to 80% and 66.7% of normal mode (Mag mode 0) image size, respectively.
- . Prior probability in tracking and updated parameters: to improve ROI tracking performance.

{12}------------------------------------------------

## Reason for Submission:

- Modification of a cleared device.
### Comparison with Predicate Devices:

| Indications for use comparison:           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number<br>and<br>Device<br>Name | K182834 (Predicate Device)<br>CS-series-FP with optional CA-100S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | (This Submission)<br>CS-series-FP with optional CA-100S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended<br>Use                           | The Omega Medical Imaging, LLC CS-series-FP<br>(SSXI) systems with optional accessory device CA-<br>100S as a modification device to provide an<br>automated Region of interest that manages<br>exposure to the patient and operator. The System<br>is intended for use in Radiographic/fluoroscopic<br>applications including cardiac, vascular, general<br>radiographic/fluoroscopic diagnostic, and<br>interventional x-ray imaging for General<br>Populations.<br>At no time will the CA-100S be considered as a<br>replacement for the primary collimator. The primary<br>collimator shall always be used, in accordance with<br>good medical practice, to define a Region of<br>Interest | The Omega Medical Imaging, LLC CS-series-FP<br>(SSXI) systems with optional accessory device CA-<br>100S as a modification device to provide an<br>automated Region of interest that reduces<br>exposure to the patient and operator. The System<br>is intended for use in Radiographic/fluoroscopic<br>applications including cardiac, vascular, general<br>radiographic/fluoroscopic diagnostic, and<br>interventional x-ray imaging for General<br>Populations.<br>At no time will the CA-100S be considered as a<br>replacement for the primary collimator. The primary<br>collimator shall always be used, in accordance with<br>good medical practice, to define a Region of<br>Interest |
| Classification Name:                      | Image-intensified Fluoroscopic X-ray system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Image-intensified Fluoroscopic X-ray system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| CFR<br>Regulation<br>#:                   | 892.1650                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 892.1650                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device<br>Class                           | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification<br>on<br>Product<br>Code:  | JAA, OWB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | JAA, OWB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

# PRODUCT OVERVIEW

The CA-100S / FluoroShield™ is a secondary collimator that can only be used in conjunction with the primary collimator. When using the CA-100S / FluoroShield™, the primary collimator shall be used to manually define a region of interest. Once this ROI has been established by the primary collimator, the CA-100S / FluoroShield™ can be used to further reduce the size of the ROI beyond the ROI initially established with the primary collimator. Due to the reduction in size of the ROI, the CA-100S/ FluoroShield™ can offer a reduction in Dose Area Product to patients and medical staff by reducing the area of exposure. At no time will the CA-100S/ FluoroShield™ be considered as a replacement for the primary collimator. The primary collimator shall always be used, in accordance with good medical practice, to define a ROI.

Additional testing for this submission was performed utilizing a DAP meter to demonstrate the dose reduction results utilizing the CA-100S / FluoroShield™ accessory. A clinical study was conducted sampling 100 patients as well to demonstrate the dose reduction to patients and staff.

- · Predicate device Omega CS-series-FP utilizes the CA-100S.

{13}------------------------------------------------

- · The only difference for this submission is use of the CA-100S with a smaller Varex 2020 FPD that is 20cm x 20cm verses the predicate device that utilizes a Varex 3030 FPD that is 30cm x 30cm.
# Substantial Equivalence:

- SE was determined on Bench performance testing. Included in this report is detailed data comparing performance with the existing Omega Medical Imaging CS-series-FP system utilizing the CA-100S integrated with the Varex 3030 FPD verses the Varex 2020 FPD. The tests that were performed utilized commercially available Phantoms such as the Phillips Phantom, and including a fabricated Moving Catheter to exercise the auto ROI functionality.
- Additional Testing utilizing a DAP meter to further demonstrate the reduction of Patient dose was conducted and demonstrated in house bench testing and a 3rd party Clinical Study. This substantiates our position in changing the IFU statement from "Manages exposure" to "Reduces exposure"

This device is substantially equivalent to Omega Medical Imaging CS-series-FP Radiographic Fluoroscopy System with Optional Accessory Device CA-100S (ROI Assembly) K182834 marketed by Omega Medical Imaging LLC. The Omega CS-series-FP with Optional accessory device Model CA-100S. The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. The only difference is the Flat Panel detector, Predicate device utilizes a Varex 3030 FPD, this submission utilizes a Varex 2020 FPD.

# Safety information:

- · The Omega CS-series-FP with the CA-100S option systems comply with the applicable requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
- · The Omega CS-series-FP with the CA-100S option systems comply with the international safety standards EN 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-2-54, EN ISO 15223-1 and EN ISO 14971.
- · The Omega CS-series-FP with the CA-100S option systems comply with UL 60601-1 and CAN/USA C22.2 No.601.1-M90
- The device is designed and manufactured under the Quality System Requlations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with applicable parts of the IEC60601-1 standards and its collateral standards. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020, that apply to this device will be met and reported in this initial report.
- This device conforms to applicable Performance Standards for lonizing Radiation Emitting . Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

Safety is assured through a risk management process and manufacturing is in compliance with the Quality System Regulations

# Conclusion:

- The technological characteristics of Omega CS-series-FP with Optional accessory device ● Model CA-100S is substantially equivalent to the Omega Medical Imaging CS-series-FP

{14}------------------------------------------------

K182834. This determination was made by comparing intended uses, designs, materials and performance. Any differences between the devices do not raise any issues of safety and effectiveness. It is the contention of Omega Medical Imaging that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.

- It is believed that the clinical studies presented in this Submission, taken together with the . extensive internal bench testing with FluoriSheild™ are sufficient to support the safety and effectiveness of FluoroShield to reduce radiation during fluoroscopic imaging procedures. A prospective study of 100 consecutive patients who underwent fluoroscopy-guided endoscopic procedure. Patients underwent interventions utilizing either conventional fluoroscopy or FluoroSheild™ fluoroscopy system to limit radiation exposure to the region of interest. There was no significant difference in demographics, body mass index, procedural type, procedural or fluoroscopy time between the conventional and the FluoroShield fluoroscopy systems.
- There were two main Objectives of this clinical study
	- 1. To compare and measure radiation exposure to patients by using dose area product (DAP).
	- 2. To compare and measure scatter radiation to endoscopy personnel measured using Landauer Luxel personal dosimetry badges.
	- . Clinical studies conducted showed When using FluoroShield, unnecessary radiation can be reduced up to 61.8% in patients and 59.4% to the staff. FluoroShield technology ensures radiation safety while maintaining image quality so physician and staff can continue to perform interventional procedures on patients effectively.
	- The use of the CA-100S / FluoroShield ™ is consistent with a best medical practice for . interventional fluoroscopy with Floroshield, which in real-time optimizes image quality in the region of interest while reducing unnecessary radiation in the periphery.
	- Taken together, the clinical trials and in-house testing reports presented in this Submission are believed to be sufficient to support the safety and effectiveness of FluoroShield in reducing unnecessary radiation exposure to both patient and staff. This also substantiates are decision to modify the Indications for Use statement by changing the phrase: "Manages exposure" to "Reduces Exposure"
	- . The FluoroShield™ was fully integrated with the Varex 2020 FPD, Image Processor ARP processor, ROI electronics and all supporting equipment to establish full functionality and integration into Omega Medical Imaging host system including Verification and Validation of Software and operational functions.
	- . Product verification plan and the Test Reports are written in accordance to applicable regulations Historically, OMI includes the purpose (test objectives) scope and lists methods (test instructions) within one design verification protocol and captures the results in one report. The full validation/Verification and test report for this project were included.

{15}------------------------------------------------

The functional test also serves as the validation of the software modules whose outputs are observable in system behavior and effects on, or interaction with, the fluoroShield™

- OMI does not create separate test reports and protocols for each test. A table listing all . the testing performed to ensure full integration of the system is indicated below in Table 1.
- . OMI has established Test Instructions that were utilized to perform and record all of the Design Verification tests identified in Protocol after a full integration of the device into a host system. These controlled test instructions were explicitly created for the CA-100S / FluoroShield™ with a Varex 2020 PFD but have been based on OMI's experience with previous CA-100S with a Varex 3030 FPD device project. Within each Testing Instruction, the Purpose, Contents (testing material and equipment) and Data Recording are defined. These Test Instructions were referenced in the Design Verification the Testing Instructions are listed in Table 1 below:

| Test Description:                    | Reference:    | Section: |
|--------------------------------------|---------------|----------|
| Calibration Process                  | Doc#IK08-0118 | 7.1      |
| Power- On                            | Doc#IK08-0118 | 7.2      |
| ROI Control Panel                    | Doc#IK08-0118 | 7.3      |
| Host System Communication via Relays | Doc#IK08-0118 | 7.4      |
| Mag Modes                            | Doc#IK08-0118 | 7.5      |
| Responses to MX- Ikomed Serial Link  | Doc#IK08-0118 | 7.6      |
| Responses to Abnormal Condition      | Doc#IK08-0118 | 7.7      |
| Responses to Abnormal Signals        | Doc#IK08-0118 | 7.8      |
| Video Bypass Button                  | Doc#IK08-0118 | 7.9      |
| Housekeeping Task                    | Doc#IK08-0118 | 7.10     |
| Auto ROI Processor Access            | Doc#IK08-0118 | 7.11     |
| Video Quality                        | Doc#IK08-0118 | 7.12     |
| Shutter Blade Position Consistency   | Doc#IK08-0118 | 7.13     |
| Shutter Vibration                    | Doc#IK08-0118 | 7.14     |

Table 1

{16}------------------------------------------------

### Referenced Guidance Documents:

- · Guidance for this submission of 510(k) for Indications of use as provided in: Pediatric Information for X-ray Imaging Device Premarket Notifications (Document issued on November 28th, 2017) Guidance for Industry and Food and Drug Administration Staff.
- · Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. (Document issued on October 2, 2014)
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Document issued on May 11, 2005)
- · Guidance for Industry and FDA Staff: Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-powered Medical Devices. (Document issued on July 11, 2016)
- Guidance for this submission of 510(k) for (SSXID) Solid State X-ray Imaging Devices issued on: September 1, 2016 was used to establish substantial equivalence
- Guidance for industry and FDA Staff User Fees and Refunds for Premarket Notification Submissions 510(k)s, (Document issued on October 2, 2017)
- Guidance for industry and FDA Staff Refuse to Accept Policy for 510(k) (Document issued on January 30, 2018)
- Guidance for industry and FDA Staff -Format for Traditional and Abbreviated 510(k)s (Document issued on August 12, 2005)
- Guidance for industry and FDA Staff Deciding when to submit a 510(k) for a change to an existing device. (Document issued on October 25, 2017)
- · Guidance for industry and FDA Staff The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (Document Issued on July 28, 2014)
- Guidance for industry and FDA Staff Guidance for Off-The-Shelf Software Use in Medical Devices (Document issued on September 9, 1999)
- Guidance for industry and FDA Staff Guidance for the Content of Premarket Submission for Software in Medical Devices. (Document issued February 3, 2016)
- Guidance for industry and Food and Drug Administration Staff Policy Clarification for Certain Fluoroscopic Equipment Requirements (Document issued on May 8, 2019)
- Guidance for Industry and FDA Staff Medical X-Ray Imaging Devices Conformance with IEC Standards. (Document issued on May 8,2019)
- Guidance for Industry and FDA Staff Clarification of Radiation Control Requlations for Manufacturers of Diagnostic X-Ray Equipment. (Document issued on December 17, 2018)

---

**Source:** [https://fda.innolitics.com/device/K191713](https://fda.innolitics.com/device/K191713)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
