← Product Code [DXN](/productcode/DXN) · K191673

# Upper Arm Blood Pressure Monitor (K191673)

_Famidoc Technology Company Limited · DXN · Nov 27, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191673

## Device Facts

- **Applicant:** Famidoc Technology Company Limited
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Nov 27, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

FDBP A series Upper Arm Blood Pressure Monitor is intended to measure the blood pressure and pulse and children at least 12 years of age, at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). with the cuff around the left upper arm according to the instruction in the user's guide manual.

## Device Story

Upper arm blood pressure monitor; uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Device consists of pneumatic cuff, flow control module, pressure sensor, and single-chip microcomputer. Operation: user places cuff on left upper arm; device automatically inflates cuff to exceed arterial pressure; pneumatic module releases pressure while sensor collects pulse pressure signals; signals are filtered and amplified; microcomputer processes signals via proprietary algorithm to calculate blood pressure and pulse; results displayed on LCD/LED and stored in memory. Used in household or medical center settings by patients or clinicians. Provides objective measurements to assist in monitoring cardiovascular health.

## Clinical Evidence

Clinical validation conducted per ISO 81060-2:2013. Non-clinical evidence includes biocompatibility testing (ISO 10993-5, -10), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and home-use performance (IEC 60601-1-11, ISO 80601-2-30).

## Technological Characteristics

Oscillometric measurement; pressure range 0-280 mmHg; pulse range 40-200 bpm. Powered by 4xAA/AAA batteries or DC adapter. Patient-contacting materials evaluated per ISO 10993-5 and -10. Software-controlled via single-chip microcomputer. Connectivity: standalone device with internal memory storage. Sterilization: N/A.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- GE Automatic Upper Arm Blood Pressure Monitor with Models BA-815, BA-816, BA-818 and BA-819 ([K172895](/device/K172895.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue.

November 27, 2019

Famidoc Technology Company Limited Leon Cao General Manager No. 212 Yilong Road, Hexi Industrial Zone Jinxia. Changan Town Dongguan, Guangdong 523853 CHINA

Re: K191673

Trade/Device Name: FDBP A Series Upper Arm Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 28, 2019 Received: November 6, 2019

Dear Leon Cao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191673

Device Name

FDBP A Series Upper Arm Blood Pressure Monitor

### Indications for Use (Describe)

FDBP A series Upper Arm Blood Pressure Monitor is intended to measure the blood pressure and pulse and children at least 12 years of age, at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). with the cuff around the left upper arm according to the instruction in the user's guide manual.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for FAMIDOC, a company that appears to be based in China. The logo consists of the word "FAMIDOC" in a bold, sans-serif font, with a blue plus sign inside of a circle in the middle of the word. To the right of the English word is the Chinese translation of the company name, "费米博士", and the number "K191673" is in the upper right corner of the image.

Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section

# 510(K) SUMMARY

This summary of 510(K) safety and effective information is being submitted in accordance with the requirement SMDA and 21 CFR 807.92.

#### Submitter of 510(K): 1.

| Date of Prepared: | 11/27/2019                                                                                     |
|-------------------|------------------------------------------------------------------------------------------------|
| Submitter's Name: | Famidoc Technology Company Limited                                                             |
| Address:          | No.212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan<br>Town Dongguan Guangdong 523853 CN |
| Contact person:   | Mr. Leon Cao                                                                                   |
| TEL:              | +(86) -769-8927-2488                                                                           |
| FAX:              | +(86) -769-8927-2498                                                                           |
| E-mail:           | leon@famidoc.com                                                                               |

#### 2. Proposed Device and code:

| Device Trade Name:           | FDBP A series Upper Arm Blood Pressure Monitor<br>(Model:FDBP-A8,FDBP-A11,FDBP-A12,FDBP-A14 ) |
|------------------------------|-----------------------------------------------------------------------------------------------|
| Regulation Medical Specialty | Noninvasive blood pressure measurement system.                                                |
| Product Code:                | DXN                                                                                           |
| Regulation number            | 21 CRF 870.1130                                                                               |
| Device Class                 | 2                                                                                             |

{4}------------------------------------------------

3. **Predicate Device:**

001 510 (k) Summary Version: 1.1

| 510(K) | Trade or Proprietary or Model Name                                                       | Manufacturer |
|--------|------------------------------------------------------------------------------------------|--------------|
|        | K172895   GE Automatic Upper Arm Blood Pressure Monitor   Ageless Health Industrial Co., |              |
|        | with Models BA-815, BA-816, BA-818 and BA-819                                            | Ltd.         |
|        |                                                                                          |              |

### Description of Proposed Device: 4.

FDBP A series Upper Arm Blood Pressure Monitor (Model:FDBP-A8,FDBP-A11,FDBP-A12,FDBP-A14 )includes utilize modular design method, It consists of nine main modules:

- power-on self-test module, system initialization module,sampling data processing and pressure, pulse rate calculation module, display processing module, power detection processing module, data storage module, key scanning processing module, sampling processing module, voice broadcast processing module, and each module communicates through a message queue.
- The blood pressure monitor controls the pneumatic flow control module through singlechipped microcomputer to pressurize the cuff module in order to exceed the lower pressure of patients, the blood being pushed against the artery walls;
- Pneumatic Flow Control Module being directed to release the pressure, while the pressure detection module collect pulse pressure signal and amplify filter;
- amplified filtersignal being read by single-chipped microcomputer for pressure and pulse signal,through unique algorithm to obtain the systolic and diastolic pressure with pulse;
- Single-chipped microcomputer will control the inflation module to release the pressure after receive measurements;
- in the meanwhile, display the measurements results then stored the values with memory module.

#### 5. Intended for Use

FDBP A series Upper Arm Blood Pressure Monitor is intended to measure the blood pressure and pulse rate of adults and children at least 12 years of age, at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). with the cuff around the left upper arm according to the instruction in the user's guide manual.

#### 6. Technical and Performance

The following table compares the device to the predicate device with basic technological characteristics.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for FAMIDOC, a healthcare company. The logo consists of the word "FAMIDOC" in a sans-serif font, with the "+" symbol in a circle. To the right of the English name is the Chinese name "费米博士" in a similar sans-serif font. The English name is in gray and blue, while the Chinese name is in gray.

| Elements of<br>Comparison            | Subject Devices                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                                                                         | Verdoc |  |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|--|
| Device Name                          | Upper Arm Blood Pressure Monitor                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   | AGE Automatic Upper<br>Arm Blood Pressure                                                                                                                                                                                                                                                                                                | /      |  |
| Device Model                         | FDBP-A8                                                                                                                                                                                                                                                                                                                                           | FDBP-A12                                                                                                                                                                                                                                                                                                                                          | FDBP-A14                                                                                                                                                                                                                                                                                                                                       | FDBP-A11                                                                                                                                                                                                                                                                                                                                          | BA-815, BA-816, BA-818,<br>BA-819                                                                                                                                                                                                                                                                                                        | /      |  |
| 510 (k) Number                       | N/A                                                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                                                                               | K172895                                                                                                                                                                                                                                                                                                                                  | /      |  |
| Product Code                         | DXN                                                                                                                                                                                                                                                                                                                                               | DXN                                                                                                                                                                                                                                                                                                                                               | DXN                                                                                                                                                                                                                                                                                                                                            | DXN                                                                                                                                                                                                                                                                                                                                               | DXN                                                                                                                                                                                                                                                                                                                                      | SE     |  |
| Regulation No.                       | 870.1130                                                                                                                                                                                                                                                                                                                                          | 870.1130                                                                                                                                                                                                                                                                                                                                          | 870.1130                                                                                                                                                                                                                                                                                                                                       | 870.1130                                                                                                                                                                                                                                                                                                                                          | 870.1130                                                                                                                                                                                                                                                                                                                                 | SE     |  |
| Classification                       | II                                                                                                                                                                                                                                                                                                                                                | II                                                                                                                                                                                                                                                                                                                                                | II                                                                                                                                                                                                                                                                                                                                             | II                                                                                                                                                                                                                                                                                                                                                | II                                                                                                                                                                                                                                                                                                                                       | SE     |  |
| Intended Use and indications for Use |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |
|                                      |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |
| Intended Use                         | FDBP A series Upper<br>Arm Blood Pressure<br>Monitor is intended to<br>measure the blood<br>pressure and pulse rate<br>of adult at household or<br>medical center. (Not<br>suitable for neonate,<br>pregnancy or pre-<br>eclampsia). with the<br>cuff around the left<br>upper arm according to<br>the instruction in the<br>user's guide manual. | FDBP A series Upper<br>Arm Blood Pressure<br>Monitor is intended to<br>measure the blood<br>pressure and pulse rate<br>of adult at household or<br>medical center. (Not<br>suitable for neonate,<br>pregnancy or pre-<br>eclampsia). with the<br>cuff around the left<br>upper arm according to<br>the instruction in the<br>user's guide manual. | FDBP A series Upper Arm<br>Blood Pressure Monitor is<br>intended to measure the<br>blood pressure and pulse<br>rate of adult at household<br>or medical center. (Not<br>suitable for neonate,<br>pregnancy or pre-<br>eclampsia). with the cuff<br>around the left upper arm<br>according to the<br>instruction in the user's<br>guide manual. | FDBP A series Upper<br>Arm Blood Pressure<br>Monitor is intended to<br>measure the blood<br>pressure and pulse rate<br>of adult at household or<br>medical center. (Not<br>suitable for neonate,<br>pregnancy or pre-<br>eclampsia). with the<br>cuff around the left<br>upper arm according to<br>the instruction in the<br>user's guide manual. | AGE Automatic Upper<br>Arm Blood Pressure<br>Monitor is intended for<br>use by medical<br>professionals or at home<br>to monitor and display<br>diastolic, systolic blood<br>pressure and pulse rate<br>on adult each time, with<br>the cuff around the left<br>upper arm according to<br>the instruction in the<br>user's guide manual. | SE     |  |
|                                      |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |
| Performance Specification            |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |
|                                      |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |
| Measuring<br>Method                  | Oscillometric<br>Method                                                                                                                                                                                                                                                                                                                           | Oscillometric<br>Method                                                                                                                                                                                                                                                                                                                           | Oscillometric<br>Method                                                                                                                                                                                                                                                                                                                        | Oscillometric<br>Method                                                                                                                                                                                                                                                                                                                           | Oscillometric<br>Method                                                                                                                                                                                                                                                                                                                  | SE     |  |
| Measuring<br>Range                   | Pressure: 0~280 mmHg<br>Pulse: 40~200 beats/minute<br>SYS(systolic pressure ): 60-<br>255mmHg<br>DIA (diastolic pressure): 40-<br>200mmHg                                                                                                                                                                                                         | Pressure: 0~280 mmHg<br>Pulse: 40~200 beats/minute<br>SYS(systolic pressure ): 60-<br>255mmHg<br>DIA (diastolic pressure): 40-<br>200mmHg                                                                                                                                                                                                         | Pressure: 0~280 mmHg<br>Pulse: 40~200 beats/minute<br>SYS(systolic pressure ): 60-<br>255mmHg<br>DIA (diastolic pressure): 40-<br>200mmHg                                                                                                                                                                                                      | Pressure: 0~280 mmHg<br>Pulse: 40~200 beats/minute<br>SYS(systolic pressure ): 60-<br>255mmHg<br>DIA (diastolic pressure): 40-<br>200mmHg                                                                                                                                                                                                         | Pressure: 0~280 mmHg<br>Pulse: 40~199 beats/minute<br>SYS(systolic pressure ): 60-<br>255mmHg<br>DIA (diastolic pressure): 40-<br>200mmHg                                                                                                                                                                                                | SE     |  |
| Pressure<br>resolution               | 1 mmHg or 0.1 kPa                                                                                                                                                                                                                                                                                                                                 | 1 mmHg or 0.1 kPa                                                                                                                                                                                                                                                                                                                                 | 1 mmHg or 0.1 kPa                                                                                                                                                                                                                                                                                                                              | 1 mmHg or 0.1 kPa                                                                                                                                                                                                                                                                                                                                 | 1 mmHg or 0.1 kPa                                                                                                                                                                                                                                                                                                                        | SE     |  |
| Accuracy                             | Pressure:<br>$±3$ mmHg ( $±0.4$ kPa)<br>Pulse: $±5$ %                                                                                                                                                                                                                                                                                             | Pressure:<br>$±3$ mmHg ( $±0.4$ kPa)<br>Pulse: $±5$ %                                                                                                                                                                                                                                                                                             | Pressure:<br>$±3$ mmHg( $±0.4$ kPa)<br>Pulse: $±5$ %                                                                                                                                                                                                                                                                                           | Pressure:<br>$±3$ mmHg( $±0.4$ kPa)<br>Pulse: $±5$ %                                                                                                                                                                                                                                                                                              | Pressure:<br>$±3$ mmHg<br>Pulse: $±5$ %                                                                                                                                                                                                                                                                                                  | SE     |  |
|                                      |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                          |        |  |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for FAMIDOC 菱米博士. The logo is composed of the word "FAMIDOC" in gray, with a blue plus sign in the middle of the "O". To the right of the English word is the Chinese characters "菱米博士" in gray. The logo is simple and modern, and the use of blue and gray gives it a professional look.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for FAMIDOC, a company that appears to be based in China. The logo consists of the word "FAMIDOC" in a dark gray sans-serif font, with a blue plus sign inside of a circle. To the right of the English text is the company name in Chinese characters, which translates to "Fei Mi Bo Shi".

| Patient<br>Population       | Adult                                                                                                                                                                  | Adult                                                                                                                                                                  | Adult                                                                                                                                                                  | Adult                                                                                                                                                                  | Adult                                                                                                                                                                  | SE                |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Measurement<br>Site of Body | Upper Arm                                                                                                                                                              | Upper Arm                                                                                                                                                              | Upper Arm                                                                                                                                                              | Upper Arm                                                                                                                                                              | Upper Arm                                                                                                                                                              | SE                |
| Inflation and<br>Deflation  | Automatic                                                                                                                                                              | Automatic                                                                                                                                                              | Automatic                                                                                                                                                              | Automatic                                                                                                                                                              | Automatic                                                                                                                                                              | SE                |
| Memory Size                 | 2x60 sets record                                                                                                                                                       | 2x60 sets record                                                                                                                                                       | 2x60 sets record                                                                                                                                                       | 2x60 sets record                                                                                                                                                       | 2x90 sets record                                                                                                                                                       | Similar<br>Note 1 |
| Indicators                  | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse, Date,<br>Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | Blood Pressure (Systolic<br>and Diastolic), Pulse,<br>Date, Time, WHO BP<br>Classification Indicating<br>Bar, Low Battery Icon,<br>Heart Icon, Memory<br>Record Number | SE                |

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for FAMIDOC, a company that appears to be based in China. The logo consists of the word "FAMIDOC" in a combination of gray and blue colors, with a plus sign inside a blue circle. To the right of the English name is the Chinese translation of the company name, "费米博士", which is also in gray.

| Cuff<br>Circumference        | 220mm~320mm                                                                          | 220mm~320mm                                                                           | 220mm~320mm                                                                          | 220mm~320mm                                                                          | For BA-815: 22-34 cm;<br>For BA-816: 28-42 cm;<br>For BA-818 and BA-819, there are 6 size:<br>size A: 17cm--22cm (SMALL ADULT CUFF)<br>size B: 22cm--30cm (ADULT CUFF-1)<br>size C: 24cm--34cm (ADULT CUFF-2)<br>size D: 22cm--42cm (L-LARGE ADULT CUFF)<br>size E: 30cm--42cm (LARGE ADULT CUFF)<br>size F: 42cm--50cm (EXTRA LARGE ADULT CUFF) | SE                |
|------------------------------|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Power Battery                | DC 6V (4xAA 1.5V<br>alkaline batteries or DC<br>Adaputer)                            | DC 6V (4xAA 1.5V<br>alkaline batteries or DC<br>Adaputer)                             | DC 6V (4xAAA 1.5V alkaline<br>batteries or DC Adaputer)                              | DC 6V (4xAAA 1.5V<br>alkaline batteries or DC<br>Adaputer)                           | DC 6V (4xAA 1.5V alkaline<br>batteries )                                                                                                                                                                                                                                                                                                         | Similar<br>Note 2 |
| Display                      | LCD Digital Display                                                                  | LCD Digital Display                                                                   | LCD Digital Display                                                                  | LED Digital Display                                                                  | LCD Digital Display                                                                                                                                                                                                                                                                                                                              | Similar<br>Note 3 |
| OPERATING&STORAGE CONDITIONS |                                                                                      |                                                                                       |                                                                                      |                                                                                      |                                                                                                                                                                                                                                                                                                                                                  |                   |
| Operating<br>Environment     | Temperature:<br>5 °C-40 °C                                                           | Temperature:<br>5 °C -40 °C                                                           | Temperature:<br>5 °C -40 °C                                                          | Temperature:<br>5 °C -40 °C                                                          | Temperature:<br>5°C ~ 40°C<br>Humidity: 15~90%RH<br>Atmospheric Pressure:<br>86 kPa~106 kPa                                                                                                                                                                                                                                                      | Similar<br>Note 4 |
|                              | Humidity:<br>15%RH-90%RH,<br>No condensation<br>Atmospheric pressure<br>70kPa~106kPa | Humidity :<br>15%RH-90%RH,<br>No condensation<br>Atmospheric pressure<br>70kPa~106kPa | Humidity:<br>15%RH-90%RH,<br>No condensation<br>Atmospheric pressure<br>70kPa~106kPa | Humidity:<br>15%RH-90%RH,<br>No condensation<br>Atmospheric pressure<br>70kPa~106kPa |                                                                                                                                                                                                                                                                                                                                                  |                   |

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for FAMIDOC, a company that appears to be based in China. The logo consists of the company name in English, "FAMIDOC," followed by the Chinese characters "费米博士". The English portion of the logo is in a dark gray font, while the "+" symbol in the middle of the name is in blue. The Chinese characters are also in dark gray.

| Storage<br>Environment | Temperature:<br>-25 °C-55 °C<br>Humidity:<br>15%RH-95%RH,<br>No condensation<br>Atmospheric pressure<br>700hPa~1060hPa | Temperature:<br>-25 °C-55 °C<br>Humidity:<br>15%RH-95%RH,<br>No condensation<br>Atmospheric pressure<br>700hPa~1060hPa | Temperature:<br>-25 °C-55 °C<br>Humidity:<br>15%RH-95%RH,<br>No condensation<br>Atmospheric pressure<br>700hPa~1060hPa | Temperature:<br>-25 °C-55 °C<br>Humidity:<br>15%RH-95%RH,<br>No condensation<br>Atmospheric pressure<br>700hPa~1060hPa | Temperature:<br>-20°C ~ +65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:<br>86 kPa~106 kPa                         | Similar<br>Note 5 |
|------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------|
| COMPLIANCE STANDARDS   |                                                                                                                        |                                                                                                                        |                                                                                                                        |                                                                                                                        |                                                                                                                        |                   |
| Performance            | ANSI/AAMI/ISO<br>81060-2                                                                                               | ANSI/AAMI/ISO<br>81060-2                                                                                               | ANSI/AAMI/ISO<br>81060-2                                                                                               | ANSI/AAMI/ISO<br>81060-2                                                                                               | ANSI/AAMI/ISO<br>81060-2                                                                                               | SE                |
| Electrical Safety      | IEC<br>60601-1                                                                                                         | IEC<br>60601-1                                                                                                         | IEC<br>60601-1                                                                                                         | IEC<br>60601-1                                                                                                         | IEC<br>60601-1                                                                                                         | SE                |
| EMC                    | IEC<br>60601-1-2                                                                                                       | IEC<br>60601-1-2                                                                                                       | IEC<br>60601-1-2                                                                                                       | IEC<br>60601-1-2                                                                                                       | IEC<br>60601-1-2                                                                                                       | SE                |
| Home Use               | IEC 60601-1-11                                                                                                         | IEC 60601-1-11                                                                                                         | IEC 60601-1-11                                                                                                         | IEC 60601-1-11                                                                                                         | IEC 60601-1-11                                                                                                         | SE                |
| Performance            | IEC<br>80601-2-30                                                                                                      | IEC<br>80601-2-30                                                                                                      | IEC<br>80601-2-30                                                                                                      | IEC<br>80601-2-30                                                                                                      | IEC<br>80601-2-30                                                                                                      | SE                |
| Biocompatibility       | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | SE                |

### Note 1 and 2

The Memory Size, Power Battery of proposed device are different. But the difference is very slight,it will not affect the main function and the intended use of the device. and Memory Size,Power Battery of proposed device is clearly indicated in user manual and outer carton. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

Note 3

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for FAMIDOC, a company that appears to be based in China. The logo consists of the word "FAMIDOC" in gray and blue letters, followed by the trademark symbol. To the right of the English name is the Chinese translation of the company name, "费米博士", which is also in gray.

The Display of proposed device and predicate device are difference is very slight,it will not affect the main function and the intended use of the device. And they are both compliance with IEC 60601-1-2 Standard. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

### Note 4 and 5:

The Temperature, Relative Humidity and Atmospheric pressure of Operation and storage environment of subject devices is difference with predicate device,and they are both compliance with IEC 60601-1-11 standard,it will not raise any safety or effectiveness issue.

### Conclusion:

The subject device AGE Automatic Upper Arm Blood Pressure Monitor has all features of the differences between them do not affect the safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device.

{11}------------------------------------------------

### 7. Performance

### Testing:

Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as

below.

### 7.1 Non-Clinical Data:

The following performance data were provided in support of the substantial equivalence determination.

### 7.2 Biocompatibility testing

The biocompatibility evaluation for the FDBP A series Upper arm Blood Pressure Monitor and

the NIBP CUFF were conducted in accordance with the International Standard ISO 10993-1 "Biological

Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as

recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less

than 24 hours. And the testing included the following tests:

- Cytotoxicity
- Skin Sensitization
- Skin Irritation

## 7.3 Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the FDBP A series Upper arm Blood Pressure Monitor, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance - Collateral Standard:Electromagnetic disturbances -Requirements and tests standard for EMC.

### 7.4 Bench Testing

Bench testing was conducted on the FDBP A series Upper arm Blood Pressure Monitor,consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2015 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, ISO 80601-2-30: 2009 Medical electrical equipment — Part 2-56: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers standards for performance effectiveness.

### 7.5 Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was

{12}------------------------------------------------

provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.

### 7.6 Usability Testing

Usability testing according to following FDA Guidance 1757, Applying Human Factors and Usability

Engineering to Optimize Medical Device Design, was conducted.

### 7.7 Clinical data:

Clinical testing is conducted per ISO 81060-2: 2013 Non-invasive sphygmomanometers -Part 2: Clinical

validation of automated measurement type.

### 7.7 Summary

Based on the non-clinical and clinical performance as documented in the device development, the subject

devices were found to have a safety and effectiveness profile that is similar to the predicate device.

### 8. Conclusions:

The proposed device has the same intended use and similar characteristics as the predicate device, AGE Automatic Upper Arm Blood Pressure Monitor with Models BA-815, BA-816, BA-818 and BA-819(K172895) Meanwhile, performance testing, bench testing, and safety report documentation supplied in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Based on performance testing, the proposed device is Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K191673](https://fda.innolitics.com/device/K191673)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
