K191659 · Instylla, Inc. · FMF · Oct 10, 2019 · General Hospital
Device Facts
Record ID
K191659
Device Name
Instylla Delivery Kit
Applicant
Instylla, Inc.
Product Code
FMF · General Hospital
Decision Date
Oct 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.
Device Story
The Instylla Delivery Kit is a manual piston syringe system used for the intra-arterial or intra-venous administration of radiographic contrast media, saline, and other aqueous solutions. The kit consists of two 1mL delivery syringes, a syringe holder, and a plunger clip. The device is operated manually by a clinician who advances the dual syringe plungers simultaneously using the plunger clip to deliver fluids. It is used in clinical settings to facilitate controlled fluid administration. The device provides a mechanical means for delivery, aiding the clinician in precise fluid management during procedures. It does not involve automated processing or software.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence was demonstrated through bench testing, functional verification (syringe holder/clip force, glide/break force, delivery integrity), biocompatibility testing (ISO 10993-1), and sterility validation (ISO 11135).
Technological Characteristics
Materials: Polycarbonate, silicone, ABS. Principle: Manual piston syringe. Volume: 1 mL. Connectivity: None. Sterilization: Ethylene oxide (EO) to SAL 10^-6. Compliance: ISO 7886-1 (graduation), ISO 594-2 (luer connectors), ISO 11135 (sterilization).
Indications for Use
Indicated for intra-arterial or intra-venous administration of radiographic contrast media, saline, and other aqueous solutions in patients requiring such procedures.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
October 11, 2019
Instylla, Inc. Amita Smith Vice President Regulatory and Quality Affairs 204 Second Avenue Waltham, Massachusetts 02451
Re: K191659
Trade/Device Name: Instylla Delivery Kit Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: September 12, 2019 Received: September 13, 2019
Dear Amita Smith:
This letter corrects our substantially equivalent letter of October 10. 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image contains two words, "Rumi" and "Young", stacked on top of each other. The word "Rumi" is on the top line, and the word "Young" is on the second line. The text is in a simple, sans-serif font and is black. There is a light blue graphic in the background.
Rumi Young -S c=US, o=U.S. Government, ou=HHS, ou=FDA. ou=People. cn=Rumi Young -0.9.2342.19200300.100.1.1=200246791
2019.10.11 08:34:51 -04'00'
For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191659
Device Name Instylla Delivery Kit
Indications for Use (Describe)
The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "instylla" in black font. There is a red dot above the "i" and a red curved line that goes from the dot, under the word, and ends at the end of the word. The font is simple and easy to read.
### 510(k) Summary
#### Submitter Information:
Instylla, Inc. 204 2nd Avenue Waltham, MA 02451
#### Contact Person:
Amita Shah VP Regulatory and Quality Affairs Phone: 781-790-4857 Email: amitas@instylla.com
#### Date Prepared:
October 9, 2019
#### Subject Device:
| Proprietary Name: | Instylla Delivery Kit |
|------------------------|-------------------------------------------------------|
| Common Name: | Syringe, Piston |
| Classification Name: | Piston syringe<br>(21 CFR 880.5860, Product Code FMF) |
| Device Classification: | Class II |
| Classification Panel: | General Hospital |
#### Predicate Device:
| Proprietary Name: | NAMIC RCS |
|-------------------|------------------------|
| Manufacturer: | Navilyst Medical, Inc. |
| 510(k) Number: | K140194 |
The FibriJet Aerosol Applicator cleared under 510(k) Premarket Notification K012868 was used as a secondary predicate device in this submission.
#### Device Description:
The Instylla Delivery Kit is comprised of two 1mL delivery syringes, a syringe holder, and a plunger clip (link).
{4}------------------------------------------------
### Indications for Use:
The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.
The indications for use statement is similar to the predicate device, NAMIC RCS.
## Comparison of Technological Characteristics to the Predicate Device:
The Instylla Delivery Kit is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate devices. The below table summarizes the similarities in design and configuration of the Instylla Delivery Kit compared with the predicate devices.
| Attribute | Subject Device:<br>Instylla Delivery Kit | Primary Predicate<br>Device:<br>NAMIC RCS (K140194) | Secondary Predicate<br>Device:<br>FibriJet Aerosol Applicator<br>(K012868) |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Instylla Delivery Kit is<br>intended to be used for the<br>intra-arterial or intra-<br>venous administration of<br>radiographic contrast<br>media, saline and other<br>aqueous solutions. | The NAMIC RCS is<br>intended to be used for<br>the intra-arterial or intra-<br>venous administration of<br>radiographic contrast<br>media and saline<br>solutions during an<br>angiographic procedure. | The Micromedics FibriJet<br>Aerosol Applicator is<br>intended for use in<br>applying two non-<br>homogenous fluids to a<br>treatment site. |
| Basic Design | Two standard delivery<br>piston syringes | One standard delivery<br>piston syringe | Two standard delivery<br>piston syringes |
| Components | Two syringes, syringe<br>holder, plunger clip | One syringe | Two syringes, syringe<br>holder, plunger link |
| Material | Polycarbonate, silicone,<br>ABS | Polycarbonate | Polypropylene, silicone,<br>ABS |
| Principle of<br>Operation: | Manually operated by<br>advancing dual syringe<br>plungers simultaneously<br>with the aid of a plunger<br>clip. | Manually operated by<br>advancing the single<br>syringe plunger. | Manually operated by<br>advancing dual syringe<br>plungers simultaneously<br>with the aid of a plunger<br>link. |
| Operation<br>Volume | 1 mL | 7, 8, 10, 12 and 20 mL | 1-12 mL |
| Gradation | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1. | Printed with graduation<br>lines. | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1. |
The intended use and principal of operation are the same between the subject and both predicate devices. Both devices are standard piston syringes that are manually operated by advancing a plunger.
{5}------------------------------------------------
## Performance Data
Performance testing of the final sterilized Instylla Delivery Kit included bench testing and functional testing to verify specifications fundamental to the device. Testing included the following:
- . Visual Inspection
- . Dimensional Verification
- . Functional Testing
- o Syringe Holder and Plunger Clip Connection Force
- Syringe Holder and Plunger Clip Disconnection Force o
- Syringe Compatibility o
- Syringe Glide Force and Break Force о
- Fluids Compatibility о
- o Delivery Integrity
- Particulate Matter Testing
Further, the syringes of the subject device are complaint against FDA recognized standards ISO 7886-1 and ISO 594-2.
### Biocompatibility Testing
A biocompatibility evaluation was conducted in accordance with the FDA Guidance for Industry, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," consistent with a device indirectly contacting blood for a limited duration (≤24 hours). The following biocompatibility tests were successfully completed on the final, sterilized Instylla Delivery Kit:
- . Cytotoxicity
- Sensitization
- Irritation or Intracutaneous toxicity ●
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity
- Hemolysis
### Sterility
The Instylla Delivery Kit is sterilized via a validated ethylene oxide (EO) process to a Sterility Assurance Level (SAL) of 10-6. The sterilization process was validated per ISO 11135 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices. A bacterial endotoxin test (BET), also known as the Limulus amebocyte lysate (LAL) test, was also validated to establish that the delivery kit endotoxin level will be <20 endotoxin units (EU)/device.
{6}------------------------------------------------
### Shelf Life
The Instylla Delivery Kit has a shelf life of 6-months. Shelf life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life.
#### Clinical Performance Data
The technological characteristics, indications for use, material, manufacturing and sterilization processes are the same as the predicate device and therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.
#### Conclusion
Instylla has demonstrated that the Instylla Delivery Kit is substantially equivalent in fundamental design, function, device materials, packaging, sterilization, operating principle, intended use/ indication for use and fundamental technology as the legally marketed predicate device, NAMIC RCS, which was cleared under 510(k) Premarket Notification K140194 on April 14, 2014.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.