← Product Code [DIO](/productcode/DIO) · K191638

# Pointe Scientific Cocaine Metabolite Enzyme Immunoassay (K191638)

_Medtest DX · DIO · Mar 12, 2020 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191638

## Device Facts

- **Applicant:** Medtest DX
- **Product Code:** [DIO](/productcode/DIO.md)
- **Decision Date:** Mar 12, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3250
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay is intended for the qualitative determination of benzoylecgonine (a cocaine metabolite) in human urine at a cutoff value of 150 ng/mL. Rx only. This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatograph/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

## Device Story

Pointe Scientific Cocaine Metabolite Enzyme Immunoassay is an in vitro diagnostic reagent kit for qualitative detection of benzoylecgonine in human urine. Device utilizes competitive enzyme immunoassay principle; drug in sample competes with enzyme-labeled drug (G6PDH) for fixed antibody binding sites. In absence of drug, enzyme-labeled conjugate binds antibody, inhibiting enzyme activity. Presence of drug displaces conjugate, allowing enzyme to convert NAD to NADH, producing absorbance change measured spectrophotometrically at 340 nm. Assay is performed on BS-480 or BA-800 Chemistry Analyzers in clinical laboratory settings by trained personnel. Results provide preliminary screening; positive results require confirmation via GC/MS or LC/MS. Clinical decision-making must incorporate professional judgment and confirmatory testing. Benefits include rapid, automated screening for cocaine metabolite presence in urine samples.

## Clinical Evidence

Bench testing only. Precision evaluated over 20 days on BS-480 and BA-800M analyzers using spiked urine samples. Method comparison performed on 114 clinical urine samples against LC/MS reference method. No clinical studies (sensitivity/specificity in patient populations) were required or performed.

## Technological Characteristics

Competitive enzyme immunoassay; Reagent 1: mouse monoclonal anti-benzoylecgonine antibody, G6P, NAD, stabilizers, 0.09% sodium azide; Reagent 2: G6PDH-labeled benzoylecgonine, buffer, 0.09% sodium azide; Detection: spectrophotometric (340 nm); Storage: 2-8°C; Analyzers: Mindray BS-480, BA-800M; Traceability: NIST-traceable Cerilliant standards.

## Regulatory Identification

A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.

## Special Controls

*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- Cocaine Metabolite Enzyme Immunoassay (k113139)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

k191638

B Applicant

MedTest Dx

C Proprietary and Established Names

Pointe scientific cocaine metabolite enzyme immunoassay

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DIO | Class II | 21 CFR 862.3250 - Cocaine And Cocaine Metabolite Test System | TX - Clinical Toxicology  |

## II Submission/Device Overview:

A Purpose for Submission:

New Device

B Measurand:

Benzoylecgonine

C Type of Test:

Immunoassay

k191638 - Page 1 of 9

{1}

k191638 - Page 2 of 9

## III Intended Use/Indications for Use:

### A Intended Use(s):

See Indications for Use below.

### B Indication(s) for Use:

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay is intended for the qualitative determination of benzoylecgonine (a cocaine metabolite) in human urine at a cutoff value of 150 ng/mL. Rx only.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatograph/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

### C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

### D Special Instrument Requirements:

- BS-480 Chemistry Analyzer
- BA-800 Chemistry Analyzer

## IV Device/System Characteristics:

### A Device Description:

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay consists of ready-to-use liquid reagents:

- Reagent 1 contains a mouse monoclonal anti-benzoylecgonine antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers and sodium azide (0.09%) as a preservative.
- Reagent 2 contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with benzoylecgonine in buffer with sodium azide (0.09%) as a preservative.

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay has a cutoff of 150 ng/mL benzoylecgonine.

### B Principle of Operation:

{2}

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, benzoylecgonine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody binds to the free drug; the unbound benzoylecgonine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at a 340 nm primary wavelength.

k191638 - Page 3 of 9

V Substantial Equivalence Information:

A Predicate Device Name(s):
Cocaine Metabolite Enzyme Immunoassay

B Predicate 510(k) Number(s):
k113139

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | k191638 | k113139  |
| --- | --- | --- |
|  Device Trade Name | Pointe Scientific Cocaine Metabolite Enzyme Immunoassay | Cocaine Metabolite Enzyme Immunoassay  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | To be used for the determination of benzoylecgonine in human urine.
This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatograph/Mass Spectrometry (GC/MS or LC/MS) is the preferred confirmatory method. | Same  |
|  Contents | The assay consists of ready-to-use liquid reagents. Reagent 1 contains a mouse | Same  |

{3}

k191638 - Page 4 of 9

|  Device & Predicate Device(s): | k191638 | k113139  |
| --- | --- | --- |
|   | monoclonal anti-benzoylecgonine antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers and sodium azide (0.09%) as a preservative. Reagent 2 contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with benzoylecgonine in buffer with sodium azide (0.09%) as a preservative |   |
|  Analyte | Benzoylecgonine | Same  |
|  Cutoff | 150 ng/mL | Same  |
|  Matrix | Urine | Same  |
|  Controls | 2 levels (112.5 ng/mL, 187.5 ng/mL) | Same  |
|  Storage | 2-8°C until expiration date | Same  |
|  General Device Characteristic Differences |  |   |
|  Calibrators | 2 Levels (0 ng/mL, 150 ng/mL) | 5 Levels (0, 75, 150, 300, 1000 ng/mL)  |

VI Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Precision was determined by spiking benzoylecgonine into drug free urine at various concentrations (zero, -75%, -50%, -25%, at the cutoff, 125%, 150%, 175% and 200% of the cutoff). Concentrations were confirmed by LC/MS. Testing for both the with-in run and between-run studies were performed by testing each sample in 2 replicates, with two runs per day, for 20 days, on both the BS-480 Chemistry Analyzer and the BS-800M Chemistry Analyzer. The qualitative results are presented below as Positive/Negative.

BS-480 Chemistry Analyzer:

|   | Within Run |   | Between Run  |   |
| --- | --- | --- | --- | --- |
|  Sample concentration (ng/mL) | No. Observations | # Neg/#Pos | No. Observations | # Neg/#Pos  |

{4}

k191638 - Page 5 of 9

|  0 (negative) | 20 | 20/0 | 80 | 80/0  |
| --- | --- | --- | --- | --- |
|  37.5 (-75% c/o) | 20 | 20/0 | 80 | 80/0  |
|  75 (-50% c/o) | 20 | 20/0 | 80 | 80/0  |
|  112.5 (-25% c/o) | 20 | 20/0 | 80 | 80/0  |
|  150 (cutoff) | 20 | 19/1 | 80 | 68/12  |
|  187.5 (+25% c/o) | 20 | 0/20 | 80 | 0/80  |
|  225 (+50% c/o) | 20 | 0/20 | 80 | 0/80  |
|  262.5 (+75% c/o) | 20 | 0/20 | 80 | 0/80  |
|  300 (+100% c/o) | 20 | 0/20 | 80 | 0/80  |

BA-800M Chemistry Analyzer:

|   | Within Run |   | Between Run  |   |
| --- | --- | --- | --- | --- |
|  Sample concentration (ng/mL) | No. Observations | # Neg/#Pos | No. Observations | # Neg/#Pos  |
|  0 (negative) | 20 | 20/0 | 80 | 80/0  |
|  37.5 (-75% c/o) | 20 | 20/0 | 80 | 80/0  |
|  75 (-50% c/o) | 20 | 20/0 | 80 | 80/0  |
|  112.5 (-25% c/o) | 20 | 20/0 | 80 | 80/0  |
|  150 (cutoff) | 20 | 7/13 | 80 | 41/39  |
|  187.5 (+25% c/o) | 20 | 0/20 | 80 | 0/80  |
|  225 (+50% c/o) | 20 | 0/20 | 80 | 0/80  |
|  262.5 (+75% c/o) | 20 | 0/20 | 80 | 0/80  |
|  300 (+100% c/o) | 20 | 0/20 | 80 | 0/80  |

2. Linearity:

Not applicable.

3. Analytical Specificity/Interference:

The following endogenous compounds were added into drug-free urine, urine sample spiked with benzoylecgonine to 75% of cutoff value and urine spiked with benzoylecgonine to 125% of cutoff value. The substances listed in the table below were determined not to interfere at the concentration shown.

Results were identical for both the BS-480 Chemistry Analyzer and the BS-800M Chemistry Analyzer.

|  Interfering Substances | Spiked [ ] (mg/dL) | Qualitative Result  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  0 ng/mL (Pos/Neg) | 112.5 ng/mL Control (Pos/Neg) | 187.5 ng/mL Control (Pos/Neg)  |
|  Acetaminophen | 10 | Neg | Neg | Pos  |

{5}

To test for possible positive and/or negative interference from pH, urine samples having pH from 3, 4, 5, 6, 7, 8, 9, 10 and 11 were used. Each of these samples were divided into two aliquots for each drug and spiked with benzoylecgonine to 75% of the cutoff and 125% of the cutoff. No positive or negative interference due to pH was observed on either the BS-480 Chemistry Analyzer or the BS-800M Chemistry Analyzer.

To test for possible positive and/or negative interference from specific gravity, urine samples having specific gravity ranging from 1.009 g/mL to 1.031 g/mL were used. Each of these samples were divided into three aliquots and two were spiked with benzoylecgonine to 75% of the cutoff and 125% of the cutoff, the third was not spiked. No positive or negative interference due to specific gravity was observed on either the BS-480 Chemistry Analyzer or the BS-800M Chemistry Analyzer.

## Cross Reactivity

Cross-reactivity was established by spiking various concentrations of structurally related and unrelated compounds into drug-free urine.

The following tables summarize approximate the quantity of each compound that is equivalent in assay reactivity to the 150 ng/mL benzoylecgonine cutoff. Results are expressed as a minimum concentration of metabolite or compound required to produce a response approximately equivalent to the cutoff concentration of the assay or the maximal concentration of the compound tested that gave a response with cross-reactivity below the cutoff calibrator. The percent cross-reactivity of those compounds are presented below.

Structurally Related Cocaine Compound:

|  Cross-reactant | Target Concentration (ng/mL) | % Cross Reactivity  |
| --- | --- | --- |
|  Benzoylecgonine | 150 | 100 %  |
|  Cocaethylene | 4,000 | 4.58 %  |
|  Cocaine (BS-480) | 25,000 | 0.62%  |

k191638 - Page 6 of 9

{6}

Structurally Unrelated Pharmacological Compounds

|  Cross-reactant | Target Concentration (ng/mL) | % Cross Reactivity  |
| --- | --- | --- |
|  Acetaminophen | 500,000 | 0.00 %  |
|  Acetylsalicylic Acid | 500,000 | 0.00 %  |
|  Amobarbital | 500,000 | 0.00 %  |
|  Amoxicillin | 500,000 | 0.00 %  |
|  Amphetamine | 500,000 | 0.00 %  |
|  Bupropion | 500,000 | 0.00 %  |
|  Captopril | 500,000 | 0.00 %  |
|  Caffeine | 500,000 | 0.00 %  |
|  Chlordiazepoxide | 500,000 | 0.00 %  |
|  Chlorpheniramine | 500,000 | 0.00 %  |
|  Chlorpomazine | 500,000 | 0.00 %  |
|  Codeine | 500,000 | 0.00 %  |
|  Dextromethorphan | 500,000 | 0.00 %  |
|  Diazepam | 500,000 | 0.00 %  |
|  Digoxin | 500,000 | 0.00 %  |
|  Enalapril | 500,000 | 0.00 %  |
|  Fluoxetine | 100,000 | 0.01 %  |
|  Glyburide | 500,000 | 0.00 %  |
|  Ibuprofen | 500,000 | 0.00 %  |
|  Lidocaine | 500,000 | 0.00 %  |
|  Meperidine | 500,000 | 0.00 %  |
|  Methadone | 100,000 | 0.01 %  |
|  Methamphetamine | 500,000 | 0.00 %  |
|  Methaqualone | 500,000 | 0.00 %  |
|  Morphine | 500,000 | 0.00 %  |
|  Nicodine | 500,000 | 0.00 %  |
|  Nifedipine | 100,000 | 0.00 %  |
|  Oxazepam | 100,000 | 0.00 %  |
|  Phencyclidine | 500,000 | 0.00 %  |
|  Phenobarbital | 500,000 | 0.00 %  |
|  Propoxyphene | 100,000 | 0.00 %  |
|  Ranitidine | 500,000 | 0.00 %  |
|  Salicyluric acid | 500,000 | 0.00 %  |
|  Secobarbital | 500,000 | 0.00 %  |
|  11-nor- THC-COOH | 500,000 | 0.00 %  |
|  Valproic Acid | 500,000 | 0.00 %  |
|  Verapamil | 500,000 | 0.00 %  |

k191638 - Page 7 of 9

{7}

4. Assay Reportable Range:

Not applicable.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Two levels of calibrators (0 and 150 ng/mL) are available for use with the Pointe Scientific Cocaine Metabolite Enzyme Immunoassay. A commercially available benzoylecgonine standard solution from Cerilliant Analytical Reference Standards is used that is traceable to NIST standard.

6. Detection Limit:

Not applicable.

7. Assay Cut-Off:

The Pointe Scientific Cocaine Metabolite Enzyme Immunoassay has a cutoff of 150 ng/mL benzoylecgonine.

## B Comparison Studies:

1. Method Comparison with Predicate Device:

Two methods comparison studies were conducted, one on the BS-480 Chemistry Analyzer and one on the BS-800M Chemistry Analyzer. 114 unaltered clinical urine remnant samples were evaluated over 3 nonconsecutive days by the Pointe Scientific Cocaine Metabolite Enzyme Immunoassay and compared to LC/MS. Results. The data is summarized below for each analyzer.

BS-480 Chemistry Analyzer:

|  Candidate Device Results | Negative | < 75 ng/mL by LC/MS | 75-150 ng/mL by LC/MS | 150-225 ng/mL by LC/MS | >225 ng/mL by LC/MS  |
| --- | --- | --- | --- | --- | --- |
|  Positive | 0 | 0 | 1 | 4 | 36  |
|  Negative | 56 | 8 | 4 | 5 | 0  |

BA-800M Chemistry Analyzer:

|  Candidate Device Results | Negative | < 75 ng/mL by LC/MS | 75-150 ng/mL by LC/MS | 150-225 ng/mL by LC/MS | >225 ng/mL by LC/MS  |
| --- | --- | --- | --- | --- | --- |
|  Positive | 0 | 0 | 1 | 6 | 36  |
|  Negative | 56 | 8 | 4 | 3 | 0  |

2. Matrix Comparison:

k191638 - Page 8 of 9

{8}

Not applicable.

## C Clinical Studies:

1. **Clinical Sensitivity:**
Not applicable.

2. **Clinical Specificity:**
Not applicable.

3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not applicable.

## D Clinical Cut-Off:

Not applicable.

## E Expected Values/Reference Range:

Not applicable.

## VII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## VIII Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

k191638 - Page 9 of 9

---

**Source:** [https://fda.innolitics.com/device/K191638](https://fda.innolitics.com/device/K191638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
