Stat Profile Prime ES Comp Plus Analyzer System

K191626 · Nova Biomedical Corporation · JFP · Aug 16, 2019 · Clinical Chemistry

Device Facts

Record IDK191626
Device NameStat Profile Prime ES Comp Plus Analyzer System
ApplicantNova Biomedical Corporation
Product CodeJFP · Clinical Chemistry
Decision DateAug 16, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1145
Device ClassClass 2

Intended Use

The Stat Profile Prime ES Comp Plus Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, Hematocrit, Ionized Calcium and Ionized Magnesium in heparinized venous whole blood, and pH. Ionized Calcium and Ionized Magnesium in plasma and serum.

Device Story

Stat Profile Prime ES Comp Plus Analyzer is a microprocessor-based, benchtop blood electrolyte analyzer for clinical laboratory use. It accepts lithium heparinized whole blood (syringes, open tubes, cups), serum, or plasma. Device utilizes ion-selective electrode (ISE) technology for pH, ionized calcium, and ionized magnesium, and impedance sensing for hematocrit. System features automated sampling, calibration, and quality management. User interface includes a 5.7" VGA touch screen, thermal printer, and integrated barcode scanner. Output provides quantitative analyte concentrations and calculated parameters (e.g., temperature-corrected pH, normalized calcium/magnesium). Healthcare professionals use results to diagnose and manage metabolic, renal, and acid-base conditions. Device benefits include rapid, automated point-of-care or lab-based testing to guide clinical decision-making.

Clinical Evidence

No clinical trials performed. Analytical performance established via bench testing: precision (within-run and run-to-run), linearity, interference, and method comparison against the predicate. Method comparison (N=150-191 whole blood, 100-121 plasma, 100-115 serum) showed high correlation (r² > 0.98) across all analytes. LoB, LoD, and LoQ determined per CLSI EP17-A2.

Technological Characteristics

Microprocessor-based analyzer; ion-selective electrodes (ISE) for iCa, iMg, pH; impedance sensor for Hct. Dimensions/form factor: benchtop analyzer with sensor card housing. Connectivity: barcode reader, onboard printer, QMS. Calibration: automated via sealed pouches. Sterilization: N/A (in vitro diagnostic).

Indications for Use

Indicated for use by healthcare professionals in clinical laboratories for quantitative measurement of pH, Hct, iCa, and iMg in heparinized venous whole blood, plasma, and serum to assist in diagnosing/treating parathyroid disease, bone disease, renal disease, tetany, magnesium imbalances, acid-base disturbances, anemia, and erythrocytosis. Not for point-of-care use.

Regulatory Classification

Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom. August 16, 2019 Nova Biomedical Corporation Cesidio Tempesta Regulatory Affairs Manager 200 Prospect St. Waltham, MA 02454 Re: K191626 Trade/Device Name: Stat Profile Prime ES Comp Plus Analyzer System Regulation Number: 21 CFR 862.1145 Regulation Name: Calcium test system Regulatory Class: Class II Product Code: JFP, CHL, CFA, GKF Dated: June 17, 2019 Received: June 18, 2019 Dear Cesidio Tempesta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k191626 #### Device Name Stat Profile Prime ES Comp Plus Analyzer System Indications for Use (Describe) The Stat Profile Prime ES Comp Plus Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, Hematocrit, Ionized Calcium and Ionized Magnesium in heparinized venous whole blood, and pH. Ionized Calcium and Ionized Magnesium in plasma and serum. lonized Calcium (iCa) measurements are used in the diagnosis and treatment of parathyroid disease, a variety ofbone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). lonized Magnesium (iMg) measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). pH measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances. Hematocrit measurements of the packed red blood cell volume are used to distinguish normal from abnormal states, such as anemia and erythrocytosis. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | 510(K) Owner: | Nova Biomedical Corporation | |----------------------|----------------------------------------------| | Registration Number: | 1219029 | | Address: | 200 Prospect St.<br>Waltham, MA 02454 | | Phone: | 781-894-0800 | | Fax Number: | 784-891-4806 | | Contact Person: | Cesidio Tempesta, Regulatory Affairs Manager | | Date Prepared: | June 14, 2019 | | 510k Number: | k191626 | Proprietary Name: Stat Profile Prime ES Comp Plus Analyzer System Common or Usual Name: Blood Gas/Electrolyte Analyzer # Classification Name: Multiple | Classification Names: | Reg. No. | Class | Product Code | Panel | |---------------------------------|----------|-------------|--------------|-----------------| | Calcium Test System | 862.1145 | II | JFP | Chemistry (75) | | Magnesium Test System | 862.1495 | I, reserved | CFA | | | Blood Gases and Blood pH system | 862.1120 | II | CHL | | | Hematocrit Test System | 864.6400 | II | GKF | Hematology (81) | # Product Codes: CHL, JFP, CFA, GKF # Predicate Devices: K110648 - Nova Stat Profile pHOx Ultra Blood Gas Analyzer # Device Description: The Stat Profile Prime ES Comp Plus Analyzer is a small, low cost blood electrolyte analyzer. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets). The Stat Profile Prime ES Comp Plus Analyzer has an enhanced test menu and multiple quality control options. External Control Solutions (ampules) shall be offered, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly. The Stat Profile Prime ES Comp Plus Analyzer can accommodate either of two sensor cards in the sensor card housing. The analyzer will determine the test configuration of the system by detecting which sensor card is installed. The two options for the sensor card are: - Sensor Card 1 (Basic Electrolyte Panel plus Hct) shall enable and report the following listed analytes: Hct, Na, K, Cl - Sensor Card 2(Full Electrolyte Panel plus pH & Hct) shall enable and report the following listed analytes: pH, Hct, Na, K, Cl, iCa, iMg {4}------------------------------------------------ As with the predicates, the Stat Profile Prime ES Comp Plus Analyzer is microprocessor-based and incorporates ion selective electrode technology to measure pH, ionized calcium, ionized magnesium. The Prime ES Comp Plus can be configured with an optional sample tray, which allows the user to run up to 10 consecutive samples may be any combination of Serum/Plasma or control solutions. Whole Blood samples may only be run in STAT Mode (not tray mode). Calibration standards are provided in sealed pouches within a calibrator pack. Liquid quality control materials are available as external ampules. Sampling and calibration are fully automated. The Stat Profile Prime ES Comp Plus Analyzer accepts lithium heparinized whole blood sample from syringes, open tubes, and small cups. The minimum sample sizes for analysis is 100 µL. # Measured Parameters: The Stat Profile Prime ES Comp Plus Analyzer measures iCa, iMg, pH, and Hct,. # Calculated Parameters: From the directly measured results, the following calculated results are available: - Normalized Calcium, nCa ● - Normalized Magnesium, nMq - Ratio of nCa/nMg ● - . Hb - . pH (patient temperature corrected) ### Intended Use: The Stat Profile Prime ES Comp Plus Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, Hematocrit. Ionized Calcium and Ionized Magnesium in heparinized venous whole blood, and pH. Ionized Calcium and Ionized Magnesium in plasma and serum. #### Indication for Use: Ionized Calcium measurements are used in the diagnosis and treatment of parathyroid disease. a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). Ionized Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). pH measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances. Hematocrit measurements of the packed red blood cell volume are used to distinguish normal from abnormal states, such as anemia and erythrocytosis. #### Summary of the Technological Characteristics: The Stat Profile Prime ES Comp Plus Analyzer is substantially equivalent to the previously cleared for market Nova Stat Profile pHOx Ultra Blood Gas Analyzer System in intended use. It uses the same sensor technology and measurement algorithms, and the formulations of the external controls and the calibration cartridge are the same for the tested parameters. The linearity standards and External Control solutions for use with the Stat Profile Prime ES Comp {5}------------------------------------------------ Plus Analyzer are substantially equivalent to those cleared for use with the predicate Nova Stat Profile pHOx Ultra Blood Gas Analyzer System (K110648). | Characteristic | Predicate Stat Profile pHOx Ultra Blood<br>Gas Analyzer (K110648) | Proposed Stat Profile Prime ES Comp<br>Plus Analyzer | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication For Use | The STP pHOx Ultra Analyzer without<br>CO-Oximeter is intended for in vitro<br>diagnostic use by health care<br>professionals and/or point-of-care usage<br>in the quantitative<br>determination of pH, PCO2, PO2, SO2%,<br>Hematocrit (Hct), Hemoglobin (Hb) in<br>heparinized whole blood; Na+, K+, Cl-,<br>Ca++, Mg++, Glucose (Glu), Lactate<br>(Lac), BUN (Urea), and Creatinine (Creat)<br>in heparinized whole blood, serum, or<br>plasma. | The Stat Profile Prime ES Comp Plus<br>Analyzer System is intended for in vitro<br>diagnostic use by health care<br>professionals in clinical laboratory<br>settings for the quantitative determination<br>of pH, Hematocrit, Ionized Calcium and<br>Ionized Magnesium in heparinized<br>venous whole blood, and pH, Ionized<br>Calcium and Ionized Magnesium in<br>plasma and serum. | | Acceptable Samples | Sodium or lithium heparinized whole<br>blood, serum, or plasma samples from<br>syringes, open tubes, small cups, and<br>capillary tubes can be used on the STP<br>pHOx Ultra Analyzer. | Lithium heparin whole blood, serum and<br>plasma samples from syringes, open<br>tubes, and small cups | | Sample Volumes | 60-210µL dependent on selected test<br>panel | 100µL | | Measurement Range | | | | iCa (Ca++) | 0.10 - 2.70 mmol/L | 0.20 - 2.70 mmol/L | | iMg | 0.1 - 1.5 mmol/L | Same | | pH | 6.500 - 8.000 | Same | | Hematocrit | 12% - 70% | Same | | Principles of Measurement | | | | iCa (Ca++) | Calcium ion-selective sensor | Same | | iMg | Magnesium ion-selective sensor | Same | | pH | Hydrogen ion-selective glass sensor | Same | | Hematocrit | Impedance Sensor | Same | | Touch Screen | 12.1" LCD, 1024x768 pixel, Resistive<br>Touch | 5.7" VGA full color display with LED<br>backlight and integrated touch panel | | Menu | Fully configurable test menu based on<br>above sensors | Same | | Bar Code Scanner | External (optional) 1D | Internal Integrated 1D/2D | | Printer | 2" Roll, Thermal Transfer | Same | | Pump | Peristaltic Pump w/ Pressure Plate, TPE<br>Tubing (Pharmed BPT) | Same | | Analog Board | Precision low level analog front end w/<br>amperometric and potentiometric<br>amplifiers, air detector circuitry and<br>temperature control circuitry | Same | Table 2-1: Comparison of Predicate and Proposed devices {6}------------------------------------------------ # Summary of Performance Testing: Performance testing was completed to demonstrate that the Stat Profile Prime ES Comp Plus Analyzer is substantially equivalent in performance, safety and efficacy to the predicate Nova Stat Profile pHOx Ultra Blood Gas Analyzer (K110648). The performance testing included: - Method Comparison Studies . - Precision/Reproducibility Studies . - . Run to Run Precision - Linearity Testing . - . Specificity / Interference Testing - . Detection Limit - . Shelf Life Stability Testing The results of the testing confirmed that the performance of the Stat Profile Prime ES Comp Plus Analyzer System is substantially equivalent to that of the Nova Stat Profile pHOx Ultra Blood Gas Analyzer (predicate device). # Conclusion: The results of performance verification testing confirmed that the Stat Profile Prime ES Comp Plus Analyzer is safe and effective for its intended purpose and that the Stat Profile Prime ES Comp Plus Analyzer System is substantially equivalent to that of K110648 - Nova Stat Profile pHOx Ultra Blood Gas Analyzer (predicate device).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...