← Product Code [ODP](/productcode/ODP) · K191621

# ChoiceSpine Octane A/T/P Spinal Implant System, ChoiceSpine Cervical Interbody Spacer System Ascendant PC, ChoiceSpine Cervical Interbody Spacer System Ascendant, ChoiceSpine Cervical Spacer System Blackhawk, ChoiceSpine Intervertebral Fusion Device Octane Straight, ChoiceSpine Intervertebral Fusion Device Octane Straight PC, ChoiceSpine VEO Laterial Access & Interbody Fusion System (K191621)

_Choice Spine, LLC · ODP · Aug 15, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191621

## Device Facts

- **Applicant:** Choice Spine, LLC
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Aug 15, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a vertebral body replacement: The ChoiceSpine Octane-A/T/P Spinal Implant is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture), to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Octane device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column, even in the absence of fusion for a prolonged period. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and supplemental fixation to facilitate fusion. When used as an intervertebral body fusion device: The Octane-A/T/P Spinal Implant is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD), with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

## Device Story

Spinal implant systems (cervical spacers and lumbar interbody fusion/vertebral body replacement devices) designed to restore spinal height and facilitate fusion. Implants feature internal chambers for autogenous/allogenic bone graft; some models include internal locking anchors or titanium plasma coatings. Devices are implanted by surgeons via open anterior or lateral approaches. Output is structural support for the spinal column; used in conjunction with supplemental fixation (e.g., anterior cervical plates or lumbar fixation systems). Benefits include restoration of biomechanical integrity and stabilization of the spinal segment in patients with DDD, tumor, or trauma.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, and performance similarities to previously cleared predicate devices. Bench testing was not required as the subject components were previously cleared.

## Technological Characteristics

Implants constructed of PEEK-OPTIMA LT1 (ASTM F2026) or titanium alloy (Ti-6Al-4V ELI, ASTM F3001/F1472). Some models feature CP titanium plasma coating (ASTM F1580) and tantalum radiographic markers (ASTM F560). Cervical systems may include nitinol locking components (ASTM F2063). Devices are provided sterile for single use. No software or electronic components.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Nexxt Spine, LLC MatrixxTM System ([K171140](/device/K171140.md))
- ChoiceSpine TOMCAT™ Cervical Spacer System ([K170953](/device/K170953.md))
- ChoiceSpine Blackhawk™ Cervical Spacer System ([K171489](/device/K171489.md))
- ChoiceSpine Ascendant Cervical Spacer System ([K141129](/device/K141129.md))
- ChoiceSpine Cervical Spacer System Ascendant PC ([K150130](/device/K150130.md))
- SeaSpine Spacer System – Hollywood, Hollywood VI, Ventura, Pacifica ([K173260](/device/K173260.md))
- ChoiceSpine Intervertebral Fusion Device Octane StraightPC ([K150152](/device/K150152.md))
- ChoiceSpine Intervertebral Fusion Device Octane Straight ([K130434](/device/K130434.md))
- ChoiceSpine VEO™ Lateral Access interbody Fusion Device ([K123997](/device/K123997.md))
- ChoiceSpine Octane A/T/P™ Vertebral Body Implant System ([K082270](/device/K082270.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 15, 2019

ChoiceSpine, LLC Kim Finch Director of Regulatory Affairs 400 Erin Drive Knoxville, TN 37919

Re: K191621

Trade/Device Name: ChoiceSpine Cervical Spacer System "Blackhawk", ChoiceSpine Cervical Interbody Spacer System "Ascendant", ChoiceSpine Cervical Interbody Spacer System "Ascendant PC", ChoiceSpine Intervertebral Fusion Device "Octane Straight", ChoiceSpine Intervertebral Fusion Device "Octane Straight PC", ChoiceSpine VEO Lateral Access & Interbody Fusion System, ChoiceSpine Octane A/T/P Spinal Implant System

Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP, OVE, MQP Dated: July 15, 2019 Received: July 16, 2019

Dear Kim Finch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine Octane A/T/P Spinal Implant System

## Indications for Use (Describe)

When used as a vertebral body replacement:

The ChoiceSpine Octane-A/T/P Spinal Implant is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture), to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Octane device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column, even in the absence of fusion for a prolonged period. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and supplemental fixation to facilitate fusion.

When used as an intervertebral body fusion device:

The Octane-A/T/P Spinal Implant is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD), with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of non-operative treatment. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine Cervical Interbody Spacer System "Ascendant PC"

#### Indications for Use (Describe)

The Ascendant PC Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Ascendant PC Cervical Spacer System is to be used autogenous and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine Cervical Interbody Spacer System "Ascendant"

#### Indications for Use (Describe)

The Ascendant Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one-disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment.The Ascendant Cervical Spacer System is to be used with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

### Device Name

ChoiceSpine Cervical Spacer System "Blackhawk'

## Indications for Use (Describe)

The BLACKHAWK™ Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The BLACKHAWK™ Cervical Spacer System is to be used autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plated via an open, anterior approach.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine Intervertebral Fusion Device "Octane Straight"

## Indications for Use (Describe)

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be sketally mature and have had at least 6 months of non-operative treatment. The device system is designed for use with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{7}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine Intervertebral Fusion Device "Octane Straight PC"

#### Indications for Use (Describe)

The Octane Straight PC Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be sketally mature and have had at least 6 months of non-operative treatment. The device system is designed for use with autogenous bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{8}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191621

## Device Name

ChoiceSpine VEO Laterial Access & Interbody Fusion System

## Indications for Use (Describe)

The VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous and/or allogenic bone graft composed of cancellous and / or corticocancellous bone graft, and supplemental spinal fixation that is cleared for use in the lumbar spine.

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of infonmation unless it displays a currently valid OMB number."

{9}------------------------------------------------

| 510(k) Summary<br>Date                | August 12, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor                               | ChoiceSpine, LLC<br>400 Erin Drive<br>Knoxville, TN 37919                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Phone<br>Fax                          | 865-246-3333<br>865-246-3334                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Contact Person                        | Kim Finch, Director of Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Proposed<br>Proprietary Trade<br>Name | ChoiceSpine Cervical Spacer System "Blackhawk"<br>ChoiceSpine Cervical Interbody Spacer System "Ascendant"<br>ChoiceSpine Cervical Interbody Spacer System "Ascendant PC"<br>ChoiceSpine Intervertebral Fusion Device "Octane Straight"<br>ChoiceSpine Intervertebral Fusion Device "Octane Straight PC"<br>ChoiceSpine VEO Lateral Access & Interbody Fusion System<br>ChoiceSpine Octane A/T/P Spinal Implant System                                                                     |
| Product Class                         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name                   | 21 CFR 888.3080 - Intervertebral Body Fusion Device<br>21 CFR 888.3060 - Intervertebral Body Fixation Orthosis                                                                                                                                                                                                                                                                                                                                                                             |
| Device<br>Product Code                | ChoiceSpine Cervical Spacer System "Blackhawk" (OVE)<br>ChoiceSpine Cervical Interbody Spacer System "Ascendant" (ODP)<br>ChoiceSpine Cervical Interbody Spacer System "Ascendant PC" (ODP)<br>ChoiceSpine Intervertebral Body Fusion Device "Octane Straight" (MAX)<br>ChoiceSpine Intervertebral Body Fusion Device "Octane Straight PC" (MAX)<br>ChoiceSpine Lateral Access Interbody Fusion Device "VEO" (MAX)<br>ChoiceSine Vertebral Body Implant System "Octane A/T/P" (MAX), (MQP) |
| Purpose of<br>Submission              | The purpose of this submission is for ChoiceSpine Intervertebral Body Fusion<br>Device Systems to gain clearance for the change in indications for use to allow the<br>use of autogenous bone and/or allogenic bone graft comprised of cancellous and /<br>or corticocanellous bone graft material.                                                                                                                                                                                        |

{10}------------------------------------------------

Device Description This submission includes 7 subject devices. The device descriptions for each are listed below:

> The ChoiceSpine Octane Straight Intervertebral Fusion Device is an implant constructed of medical grade Polyetheretherketone, (PEEK-OPTIMA® LT1) as described by ASTM F2026. The implant incorporates ridges on the superior and inferior surfaces to resist expulsion. The device is open in the transverse plane to allow insertion of bone graft prior to placement, and fenestrated along the sides. The radiolucent PEEK-OPTIMA® material allows visualization of the defect site on radiography to assess bone growth, and incorporates tantalum markers conforming to ASTM F560 to permit verification of position. The Octane Straight Intervertebral Fusion Device is provided sterile for single use.

> The ChoiceSpine Octane Straight PC Intervertebral Fusion Device is an implant constructed of medical grade Polyetheretherketone, (PEEK-OPTIMA® LT1) as described by ASTM F2026. The implant incorporates ridges on the superior and inferior surfaces to resist expulsion. The device is open in the transverse plane to allow insertion of bone graft prior to placement, and fenestrated along the sides. The radiolucent PEEK-OPTIMA® material allows visualization of the defect site on radiography to assess bone growth and incorporates tantalum markers conforming to ASTM F560 to permit verification of position. The device is plasma coated with commercial pure titanium (CPTi) per ASTM F1580. The Octane Straight PC Intervertebral Fusion Device is provided sterile for single use.

> The BLACKHAWK™ Cervical Spacer System is an anterior cervical interbody device consisting of a PEEK implant cage per ASTM F2026 with tantalum radiographic markers per ASTM F560, nitinol internal locking components per ASTM F2063 and two internal titanium alloy anchors (Ti-6Al-4V ELI) per ASTM F136. They are intended for use as interbody fusion devices and offered in a variety of heights, footprints, and lordotic angles to accommodate varying anatomical conditions. The devices feature a chamber intended to be filled with autogenous bone and or/allogenic bone graft material. They are used with two internal anchors that lock on deployment and provide additional fixation. The BLACKHAWK™ Cervical Spacer System is to be used with autogenous bone and or/allogenic bone graft and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach.

> The Ascendant® Cervical Spacer System is an anterior cervical interbody device consisting of a PEEK Optima® LT1 (polyetheretherkeytone) implant cage with tantalum radiographic markers. It is intended for use as an interbody fusion device and is offered in a variety of heights, footprints and lordotic angles to accommodate varying anatomical conditions. The device features an enclosed chamber intended to be filled with autogenous and or/allogenic bone graft material. The Ascendant® Cervical Spacer System is intended to be used with supplemental fixation (i.e., an anterior cervical plate).

{11}------------------------------------------------

The Ascendant™ PC Cervical Spacer System is an anterior cervical interbody device consisting of a PEEK Optima® LT1 (polyetheretherkeytone) implant cage with CP titanium coating and tantalum radiographic markers. It is intended for use as an interbody fusion device and is offered in a variety of heights, footprints and lordotic angles to accommodate varying anatomical conditions. The device features an enclosed chamber intended to be filled with autogenous and or/allogenic bone graft material.

The VEO® Lateral Access & Interbody Fusion System is a multi-component system including instrumentation made of biocompatible materials such as Stainless Steel, Aluminum, and Radel R and implants made of Tantalum (ASTM F560) and PEEK (ASTM F2026) or Ti-6AI-4V ELI (ASTM F136).

The ChoiceSpine Octane®-A/T/P Spinal Implant is an implant constructed of medical grade Polyetheretherketone, (PEEK-OPTIMA® LT1) as described by ASTM F2026. The implant incorporates ridges on the superior and inferior surfaces to resist expulsion. The device is open in the transverse plain to allow insertion of bone graft prior to placement, and fenestrated along the sides. The radiolucent PEEK-OPTIMA® material allows visualization of the defect site on radiography to assess bone growth, and incorporates tantalum markers conforming to ASTM F560 to permit verification of position. The ChoiceSpine Octane Spinal Implant is provided sterile in three (3) styles:

- . The Octane-T Spinal Implant, available heights of 7mm to 17mm, in 2mm increments;
- . The Octane-P Spinal Implant, available in heights of 7mm to 17mm, in 2mm increments; and
- . The Octane-A Spinal Implant, available in heights of 9mm to 19mm, in 2mm increments, and with a convex superior and inferior surface, angled at either 6° or 12° and in Small, Medium, or Large transverse profile.

Proposed The Ascendant PC Cervical Spacer System is indicated for anterior cervical interbody Indications for Use fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Ascendant PC Cervical Spacer System is to be used with autogenous and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

> The Ascendant® Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at onedisc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment.

{12}------------------------------------------------

The Ascendant Cervical Spacer System is to be used with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

The BLACKHAWK™ Cervical Spacer System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease at one disc level from C2-T1. Degenerative Disc Disease (DDD) is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The BLACKHAWK™ Cervical Spacer System is to be used with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and supplemental fixation (i.e., an anterior cervical plate), and is implanted via an open, anterior approach.

The Octane Straight Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have had at least 6 months of nonoperative treatment. The device system is designed for use with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

The Octane Straight PC Intervertebral Fusion Device is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have had at least 6 months of nonoperative treatment. The device system is designed for use with autogenous bone and/or allogenic bone graft composed of cancellous and /or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

The VEO® Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous and/or allogenic bone graft composed of cancellous and / or corticocancellous bone graft, and supplemental spinal fixation that is cleared for use in the lumbar spine.

{13}------------------------------------------------

The ChoiceSpine Octane-A/T/P Spinal Implant System:

## When used as a vertebral body replacement:

The ChoiceSpine Octane®-A/T/P Spinal Implant is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture), to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Octane device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column, even in the absence of fusion for a prolonged period. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and / or corticocancellous bone graft, and supplemental fixation to facilitate fusion.

The ChoiceSpine Octane-A/T/P Spinal Implant System:

When used as an intervertebral body fusion device: The Octane®-A/T/P Spinal Implant is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with Degenerative Disc Disease (DDD,) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature, and have had at least 6 months of nonoperative treatment. The device may be used with autogenous bone and/or allogenic bone graft composed of cancellous and / or corticocancellous bone graft, and with supplemental fixation systems cleared for use in the lumbosacral spine.

- Materials The implants are made from either titanium alloy (Ti-6AI-4V ELI per ASTM F3001 or F1472) or polyetheretherketone (PEEK-Optima) per ASTM F2026 with tantalum markers per ASTM F560.
- Non-Clinical All subject components have been cleared and therefore do not present a new Testing mechanical worst-case and no non-clinical testing is needed.

Predicate Device Primary Predicate: Nexxt Spine, LLC MatrixxTM System (K171140)

> Additional Predicate: ChoiceSpine TOMCAT™ Cervical Spacer System (K170953) Additional Predicate: ChoiceSpine Blackhawk™ Cervical Spacer System (K171489)

Additional Predicates: ChoiceSpine Ascendant Cervical Spacer System "Ascendant" (K141129), ChoiceSpine Cervical Spacer System "Ascendant PC" (K150130)

Additional Predicate: SeaSpine Spacer System – Hollywood, Hollywood VI, Ventura, Pacifica (K173260),

Additional Predicates: ChoiceSpine Intervertebral Fusion Device "Octane StraightPC" (K150152), ChoiceSpine Intervertebral Fusion Device "Octane Straight" (K130434), ChoiceSpine VEO™ Lateral Access interbody Fusion Device (K123997), and ChoiceSpine Octane A/T/P™ Vertebral Body Implant System (K082270)

{14}------------------------------------------------

Substantial Equivalence Nexxt Spine, LLC Matrixx TM System (K171140).

Substantial Equivalence Conclusion

The implants proposed in this submission are similar to the predicates in principle of operation, materials, indications for use, biocompatibility, manufacturing and post processing steps, stabilization methods, sterilizations method, anatomic location and approach, product classification and product codes. The only differences are the indications for use for the subject devices indicate the use of autograft bone and the primary predicates allow with autograft bone and/or allogenic bone graft composed of cancellous and / or corticocancellous bone graft. Both the subject devices and predicate devices require supplemental fixation.

After considering all similarities and differences to the predicate devices, the subject device has shown to be equivalent when compared to the predicate devices in safety, effectiveness, and performance.

---

**Source:** [https://fda.innolitics.com/device/K191621](https://fda.innolitics.com/device/K191621)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
