← Product Code [GEX](/productcode/GEX) · K191611

# Diode Laser Therapy Systems (K191611)

_Wingderm Electro-Optics , Ltd. · GEX · Jul 31, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191611

## Device Facts

- **Applicant:** Wingderm Electro-Optics , Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 31, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

## Device Story

Diode Laser Therapy System (Model WLA-01) is a laser surgical instrument for hair removal. It utilizes selective photothermolysis via an 808nm diode laser to target melanin in hair follicles, causing thermal damage to follicles and nutrient vessels while sparing surrounding tissue through epidermal cooling. The system comprises a main console with an LCD touch screen, a treatment handpiece, and a foot switch. It is intended for use in healthcare facilities or hospitals by trained personnel. The operator uses the touch screen to control parameters; the handpiece is applied directly to the skin. The device provides effective hair removal and prevents hair regeneration. It is designed for professional use only.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence was supported by non-clinical performance data, including biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993), electrical safety, and electromagnetic compatibility testing.

## Technological Characteristics

Class IV 808nm diode laser source; selective photothermolysis principle; epidermal cooling system; console with LCD touch screen; foot switch control. Complies with ANSI AAMI ES60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-2. Biocompatibility per ISO 10993.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Diode Laser Therapy Systems ([K152898](/device/K152898.md))

## Reference Devices

- GME LinScan System ([K141063](/device/K141063.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Wingderm Electro-Optics Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd 8-9th Floor, R&D Building, No.26 Qinglan Street Panyu District July 31, 2019 Guangzhou, 510006 CN

Re: K191611

Trade/Device Name: Diode Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 14, 2019 Received: June 17, 2019

Dear Mike Gu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191611

Device Name Diode Laser Therapy Systems

Indications for Use (Describe)

The Diode Laser Therapy Systems are intended for hair reduction on all skin types (Fitzpatrick skin type I-VI).

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span style="font-size: 20px;">☐</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Wingderm. The logo consists of a blue abstract symbol on the left and the word "WINGDERM" in black on the right. The symbol is made up of three connected circles, with a gradient from light blue to dark blue. The text is in a sans-serif font.

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

### SUBMITTER l.

Wingderm Electro-Optics Ltd.

Room 312, Building D-3, Dongsheng Science Park, No.66 of Xixiaokou Road, Haidian

District, Beijing 100192, China

Phone: 086-010-57734966

Fax: 086-010-82290038

| Primary Contact Person:   | Mike Gu                                            |
|---------------------------|----------------------------------------------------|
|                           | Regulatory Affairs Manager                         |
|                           | Guangzhou Osmunda Medical Device Technical Service |
|                           | Co., Ltd                                           |
|                           | Tel: (+86) 20-6231 6262                            |
|                           | Fax: (+86) 20-8633 0253                            |
| Secondary Contact Person: | Juan Zhou                                          |
|                           | Quality Manager                                    |
|                           | Wingderm Electro-Optics Ltd.                       |
|                           | Tel: 86-010-62910812                               |
| Date Prepared:            | June 13, 2019                                      |

II. DEVICE

| Name of Device:       | Diode Laser Therapy Systems, Model WLA-01           |
|-----------------------|-----------------------------------------------------|
| Common or Usual Name: | Diode Laser Therapy Systems                         |
| Classification Names: | Powered Laser Surgical Instrument (21 CFR 878.4810) |
| Regulation Class:     | II                                                  |
| Product Code:         | GEX                                                 |

lll. PREDICATE DEVICE

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for "WINGDERM". To the left of the company name is a blue graphic that looks like a molecule. The company name is in all caps and bolded.

| Predicate device | Diode Laser Therapy Systems: K152898 |
|------------------|--------------------------------------|
|------------------|--------------------------------------|

- Reference device GME LinScan System: K141063
These predicates have not been subject to a design-related recall.

### IV. DEVICE DESCRIPTION

The Diode Laser Therapy Systems has one model WLA-01, which mainly consist of the main console, treatment hand piece and a foot switch. The treat hand piece consist of the laser aperture, laser emission indicator, laser emission and hand piece display screen. The main console consist of power switch, hand piece holder, LCD touch screen, emergency shut off, hand-grip grab bar, connector hand piece, access door, observation Window, power input, inlet, air switch, gate interlock connector, foot switch connector, vent, drain and fans. The device has an embedded software named Diode Laser Therapy Systems Control Software.

The principle of laser hair removal is selective photothermolysis, 808nm wavelength effectively penetrates deep into and absorbed by the target chromophore. The adequate pulse duration, energy density and epidermal cooling ensure an adequate thermal damage to the target tissue without damaging the surrounding tissue to achieve effective hair removal.

The laser window of the hand piece will contact the skin directly and the laser output will contact the skin through the window. The Diode Laser Therapy System is suitable for use in healthcare facility/hospital. Only trained personnel who have been trained in the use of laser products and clinical training should be able to use this product for treatment.

### V. INDICATION FOR USE

The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI).

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE

The Diode Laser Therapy Systems are substantially equivalent to the cleared predicate device (K152898) and the reference device (K141063).

The Diode Laser Therapy Systems have same indications for use, patient population and material with the predicates.

{5}------------------------------------------------

Selective photothermolysis is the principle of laser hair removal for both the subject and predicate devices. Melanin selectively absorbs laser energy with wavelength of 808nm, and immediately forms local high temperature. When the temperature in hair follicles rises to some certain degree, thermal expansion occurs in hair follicle. Fractures melanin cells are pushed out of hair pores by steam. Meanwhile, dermal papilla nutrient vessels are damaged because of hemoglobin solidification. Under above dual functions, effective hair removal will be achieved and hair regeneration will be prevented. Treatment time for each pulse is no longer than target tissue s thermal relaxation time, which enables that photothermy effect limits in target tissues and prevent thermal energy damaging surrounding tissues. Epidermal cooling is also designed for the protection of surrounding tissues.

At a high level, the subject and predicate devices are based on the following same technological elements:

- Use the same Class IV Diode Laser as the laser source.
- Use the same 808nm laser wavelength as selective absorption wavelength.
- Use the same energy density
- Use the same pulse width
- Use the same frequency
- Use the same cooling system
- Footswitch or Laser Emission button can control the system.
- Use of a touch Screen Control Panel to display.
- Hand piece is cleaned between patients.
- Anatomical Sites include axilla, facial, bikini, upper limb and lower limb.
- The proposed device is similar with the predicate in Spot Size.

### PERFORMANCE DATA VII.

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility testing

The biocompatibility evaluation for the Diode Laser Therapy Systems was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Wingderm. The logo consists of a blue, abstract, interconnected network of circles on the left, resembling a molecule or a cluster of cells. To the right of the graphic is the word "WINGDERM" in bold, black, sans-serif font.

of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

- Cytotoxicity
- Sensitization
- Irritation

The laser aperture is considered tissue contacting for a duration of less than 24 hours.

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Diode Laser Therapy Systems, consisting of the main console, treatment hand piece and foot switch. The system complies with the ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text), IEC 60601-2-22 Edition 3.1 2012-10 and IEC 60825-1: 2014 standards for safety and the IEC 60601-1-2:2014 standard for EMC.

## Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure of latent design flaw could directly result in minor injury to the patient or operator.

## Clinical Testing

Based on the similarities of the device specifications, intended use, indications for use between the Diode Laser Therapy Systems and its predicate devices, no clinical studies were needed to support this 510(k) Premarket Notification.

### VIII. CONCLUSION

The Diode Laser Therapy Systems and its application comply with standards as detailed in section 9, 11 and 17 of this premarket notification. Non-clinical tests determined that the Diode Laser Therapy Systems to be as safe, as effective and performance is substantially equivalent to the predicates.

---

**Source:** [https://fda.innolitics.com/device/K191611](https://fda.innolitics.com/device/K191611)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
