The aqua-tools FILT'RAY 2G Disposable Water Filter is intended to filter EPA (Environmental Protection Agency in USA) quality drinking water. By retaining bacteria, the filters may aid in infection control. The filters produce water that is suitable for washing and drinking, superficial wound cleansing (minor cuts, scrapes, or abrasions), cleaning of equipment used in medical procedures and washing of surgeon's hands. The filters are not intended to provide water that can be used as a substitute for USP grade sterile water.
Device Story
Disposable water filter; utilizes hollow fiber microfiltration technology (0.1 µm nominal / 0.2 µm absolute pore size) to retain waterborne pathogens (e.g., Legionella spp., Pseudomonas aeruginosa). Device connects to water distribution networks (showers, faucets, in-line). Operates via physical filtration; housing contains silver ion-based bacteriostatic agent to reduce external surface contamination. Used by healthcare providers or patients in clinical environments for washing, drinking, and equipment cleaning. Provides bacteriologically controlled water; aids in infection control. No electronic components or software.
Clinical Evidence
Bench testing only. Microbial retention validated per ASTM F838-15a (>7 log reduction of Brevundimonas diminuta). Bacteriostatic effectiveness of silver ion-based housing tested per ISO 22196 (>99.9% reduction of P. aeruginosa, E. coli, S. aureus, MRSA). Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity) per ISO 10993-1 passed. Sterility validated per ISO 11137-1/2 (VDMax method, SAL 10^-6). Shelf life and packaging integrity confirmed via accelerated and real-time aging.
Technological Characteristics
Hollow fiber microfiltration membrane (0.1 µm nominal/0.2 µm absolute pore size). Housing incorporates silver ion-based bacteriostatic agent. Mechanical filtration principle. No energy source. Sterile, disposable. Biocompatibility per ISO 10993-1. Sterilization per ISO 11137-1/2. Shelf life up to 3 years.
Indications for Use
Indicated for filtering EPA-quality drinking water for infection control in clinical settings. Suitable for washing, drinking, superficial wound cleansing, cleaning medical equipment, and washing surgeon's hands. Not for use as a substitute for USP grade sterile water.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an eagle and text around it. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the text "U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue.
October 25, 2019
aqua-tools Virginie Grondin Medical Devices Quality and Regulatory Affairs Manager 26 Rue Charles Edouard Jeanneret 78300 POISSY FRANCE
Re: K191563
Trade/Device Name: FILT'RAY 2G Disposable Water Filters Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: NHV Dated: September 17, 2019 Received: September 23, 2019
Dear Virginie Grondin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191563
Device Name FILT'RAY2G Diposable Water Filter
### Indications for Use (Describe)
The aqua-tools FILT'RAY 2G Disposable Water Filter is intended to filter EPA (Environmental Protection Agency in USA) quality drinking water. By retaining bacteria, the filters may aid in infection control. The filters produce water that is suitable for washing and drinking, superficial wound cleansing (minor cuts, scrapes, or abrasions), cleaning of equipment used in medical procedures and washing of surgeon's hands. The filters are not intended to provide water that can be used as a substitute for USP grade sterile water.
Type of Use (Select one or both, as applicable):
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: 2;"> <span style="text-align: left;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: 2;"> <span style="word-spacing: 0px;"> <span style="display: inline !important;"> <input checked="true" type="checkbox"/> </span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: 2;"> <span style="text-align: left;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: 2;"> <span style="word-spacing: 0px;"> <span style="display: inline !important;"> <input type="checkbox"/> </span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Aqua-Tools. The logo features a series of blue and green circles that fade from dark blue to light green. Below the circles is the text "aqua-tools" in a sans-serif font, with the word "aqua" in lowercase and "tools" in uppercase. Below that is the text "INFECTION CONTROL SOLUTIONS" with "INFECTION CONTROL" in a larger font than "SOLUTIONS".
# 510(K) SUMMARY
#### I. SUBMITTER
| Company Name: | aqua-tools |
|------------------------|------------------------------------------------------------|
| Contact Person : | Ms. Virginie S. Grondin<br>virginie.grondin@aqua-tools.com |
| Date Prepared: | May 28th, 2019 |
| II. | DEVICE |
| Trade Name: | FILT'RAY 2G Disposable Water Filters |
| Common Name: | System, Water purification, General medical use |
| Device Classification: | 21 CFR 876.5665, Product code NHV |
| Classification Name: | Water purification system for hemodialysis |
| Device Class: | Class II |
| Panel: | Gastroenterology/Urology |
#### III. PREDICATE DEVICES
| Primary Predicate Device: | Pall-Aquasafe <sup>TM</sup> Water Filters (K153434) 21 CFR 876.5665, Class II, NHV |
|-------------------------------|--------------------------------------------------------------------------------------|
| Additional Predicate Devices: | Evoqua Nosogard Filters (K153784) Nephros, Inc. S100 Point of Use Filters (K153084) |
#### IV. DEVICE DESCRIPTION
The FILT'RAY 20 Disposable Water Filters are intended to produce bacteriologically controlled, purified water for general medical uses (e.g., washing of surgeon's hands, surgical instruments, and cleansing). These devices retain bacteria including opportunistic waterborne pathogenic microorganisms via a hollow fiber microfiltration technology with a corresponding pore size of 0.1 µm nominal / 0.2 µm absolute.
{4}------------------------------------------------
This microfiltration technology ensures that the purified water will meets the criteria of bacteriologically controlled water that presents a higher bacteriological quality than the water of the distribution network by retaining bacteria, including Legionella spp., L. pneumophila. Pseudomonas aeruginosa and other opportunistic waterborne pathogenic microorganisms. Of note, the subject FILT'RAY 20 devices do not contain latex, nor tissues or by-product animal origins, nor derived of blood or any substance considered as a medicine. The subject FILT' RAY 20 devices are provided sterile and are designed to be used for 1 month up to 4 months after initial connection and depending on the specific model.
#### V. INDICATIONS FOR USE
The aqua-tools FILT' RAY 20 Disposable Water Filter is intended to be used to filter EPA (Environmental Protection Agency in USA) quality drinking water. By retaining bacteria, the filters may aid in infection control. The filters produce water that is suitable for washing and drinking, superficial wound cleansing (minor cuts, scrapes, or abrasions), cleaning of equipment used in medical procedures and washing of surgeon's hands. The filters are not intended to provide water that can be used as a substitute for USP grade sterile water.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVICE
The subject and predicate devices have identical or similar indications for use and yield water that is suitable for drinking and washing. The subject FILT'RAY 20 Disposable Water Filter device is also similar in its system configuration (e.g., shower, faucet, in-line filters), use specifications (use life, pressure rating, water source/network), and in other significant design and technological features. The subject and predicate devices utilize the same fundamental technology that consists of a sterilizing grade filter1 membrane that has been validated to retain and remove microorganisms in the water supply. The characteristics of the subject device and the predicate device were assessed and compared via performance testing, as described below. Any minor differences between the subject and predicate devices do not raise new questions of safety or effectiveness. Rather, these differences can and were evaluated through performance testing to determine substantial equivalence.
#### PERFORMANCE DATA VII.
Non-clinical performance testing was conducted to characterize the subject FILT RAY 2G Disposable Water Filters and is summarized as follows:
{5}------------------------------------------------
## Microbial Retention
The ability of the subject FILT'RAY 20 Disposable Water Filter to retain a bacterial load greater than 107 colony-forming units (CFU)/cm² was captured via performance bench testing per ASTM F838-15a: Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration. This standard testing was used on water filters that have underwent worst case use. The Brevundimonas diminuta test microorganism was not found in any of the filtrates for all tested filters.
### Evaluation of Bacteriostatic Agent Effectiveness
The addition of a silver ion-based bacteriostatic agent into the material used to manufacture the housing components was evaluated per ISO 22196: Measurement of antibacterial activity on plastics and other non-porous surfaces. These tests were performed on new filters and on aged filters. Per the results of these tests, this agent allows to reduce the external surface contamination by more than 99.9% for Pseudomonas aeruginosa, Escherichia coli, Staphylococcus aureus and Methicillin Resistant Staphylococcus aureus (MRSA). Ageing of the samples had no impact on the efficacy of the treated samples.
### Silver Ion Testing
The presence of any silver-based ions in the water filtered through the FILT RAY 20 Disposable Water Filters was quantified via inductively coupled plasma mass spectrometry (ICP/MS) and found to be less than the quantification limit (i.e., <1.0ug/L) and are in compliance with the World Health Organization (WHO) guidelines without risk to health.
### Maximum Operating Temperature and Pressure Rating
Testing was conducted at up to 5 bar as well as up to 70℃ (for a continuous period of 30 minutes) to ensure that the subject filters were able to not only withstand these conditions, but also adequately retain bacteria at a greater than 7 log reductions after filtration of water at these conditions.
### Flow Rate Testing
All filter models were subjected to a flow rate/pressure test with water pressures varying from 1 to 5 bar (on the filter input) and were found to have adequate flow for use at these various inputs.
Sterilizing grade filter is defined in "American standard Test Method ASTM F838-15a and in FDA - Guidance for Industry - Sterile Drug Products Produced by Aseptic Processing - Current Good Manufacturing Practice" as a filter ability to sustain a bacterial challenge of 107 CFU of Brevundimonas diminuta per cm2 of filtration surface.
{6}------------------------------------------------
# Biocompatibility
The subject FILT'RAY 20 Disposable Water Filter device is proposed to be used for washing, drinking, superficial wound cleansing, cleaning of equipment used in medical procedures and washing of surgeon's hands. Therefore, given the intended use, nature and degree of tissue contact and exposure duration during clinical use, the subject device is considered a surface device, contacting breached or compromised surfaces repeatedly for limited duration of contact (< 24 hour-exposure duration is considered as worst-case cumulative patient-contacting hours, even though there is potential for repeat use of the same or a new device during its intended use. Therefore, per the FDA Guidance on the "Use of International Standard ISO 10993-1," the following tests for biocompatibility evaluation of the subject FILT'RAY 20 Disposable Water Filter device have been performed: cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, and material mediated pyrogenicity testing. The NAMSA tests reports indicate that it is safe for use in medical devices, as evidenced by the passing results of cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, and material mediated pyrogenicity testing, which are required for this classification.
### Sterilization Testing
The FILT'RAY 20 Disposable Water Filters are available sterile with a use life of 1-month (31 days), 2-month (62 days), 3-month (93 days) and 4-month (124 days). Bioburden was assessed per ISO 11737-1 and 11737-2, and sterilization was conducted per ISO 11137-1, parts 1 and 2. The VDMax® method was utilized to characterize a minimum dose of 25kGy to ensure a sterility assurance limit (SAL) of 10-6
### Shelf Life Testing
The packaging and sterilizing grade performance for the subject FILT'RAY 20 Disposable Water Filters was conducted via 3-year accelerated aging tests and a 2 years and 9 months real-time aging test. Packaging integrity testing was performed. All results passed the acceptance criteria, and the packaging integrity (and sterility) was determined to be acceptable after sterilization, transportation simulation, and accelerated aging testing. Therefore, the subject FILT RAY 20 Filters remain sterile and produce bacteriologicallycontrolled water with an expiration date of up to 3 years.
## VIII. CONCLUSIONS
Comparison of the indications for use, technical characteristics, as well as performance of the subject and predicate water filters provided in the 510(k) submission, is sufficient to demonstrate substantial equivalence. The information contained in this 510(k) demonstrates that the FILT' RAY 20 Disposable Water Filters are as safe and effective as the predicate device. With identical indications for use and similar technological characteristics that were evaluated by performance testing, the aqua-tools FILT RAY 2G Disposable Water Filters are substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.