Vave Personal Ultrasound System
K191541 · Vave Health, Inc. · IYO · Jul 10, 2019 · Radiology
Device Facts
| Record ID | K191541 |
| Device Name | Vave Personal Ultrasound System |
| Applicant | Vave Health, Inc. |
| Product Code | IYO · Radiology |
| Decision Date | Jul 10, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
| Attributes | Pediatric, 3rd-Party Reviewed |
Intended Use
The Vave Personal Ultrasound is intended for diagnostic ultrasound imaging in B-Mode. It is indicated for diagnostic ultrasound imaging in the following applications: Fetal/Obstetric, Abdominal (includes Gynecology, Renal, and Urology), Pediatric, Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric, Peripheral Vessel and procedural guidance of needles into the body. The Vave Personal Ultrasound is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
Device Story
Handheld, wireless ultrasound system comprising an ultrasound probe, display application, and battery charger. Probe acquires ultrasound data via piezoelectric transducer and transmits wirelessly to a commercial off-the-shelf mobile device (iOS or Android) running the display application. Used by healthcare professionals in clinical, home, and emergency settings (ambulance/aircraft) for diagnostic imaging and procedural needle guidance. Healthcare providers view real-time B-Mode images on the mobile device to assist in clinical assessment and interventional procedures. Benefits include portability and ease of use for point-of-care diagnostics.
Clinical Evidence
No clinical data. Bench testing only. Safety and effectiveness of ultrasound systems with these characteristics are well-accepted.
Technological Characteristics
Handheld phased array ultrasound probe; piezoelectric sensing; wireless data transmission to mobile device; Li-ion battery powered. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-12, IEC 60601-2-37, IEC 62366-1, IEC 62133-2, ANSI AAMI IEC 62304, ISO 10993-1. Non-sterile.
Indications for Use
Indicated for diagnostic ultrasound imaging in B-Mode for fetal/obstetric, abdominal (gynecology, renal, urology), pediatric, thoracic/pleural, cardiac (adult/pediatric), peripheral vessel, and needle guidance. Intended for use by healthcare professionals in clinical, home, and emergency medical environments.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Vscan with Dual Probe (K140693)
Reference Devices
- Clarius Ultrasound Scanner (K172385)
Related Devices
- K241051 — Vave Wireless Ultrasound System · Vave Health, Inc. · Aug 30, 2024
- K250791 — ASUS Ultrasound Imaging System (LU800 series) · Asustek Computer, Inc. · Dec 4, 2025
- K222365 — Leltek Ultrasound Imaging System (Model: LU700 Series) · Leltek, Inc. · Oct 25, 2022
- K243226 — Ultrasound Imaging System (LU700 Series); Ultrasound Imaging System (128 Series); Ultrasound Imaging System (LX Series) · Leltek, Inc. · May 8, 2025
- K241161 — Ultrasound Imaging System (LK128L); Ultrasound Imaging System (LK128LH); Ultrasound Imaging System (LK128C); Ultrasound Imaging System (LK128M); Ultrasound Imaging System (LK128PA); Ultrasound Imaging System (LK128E); Ultrasound Imaging System (LU700C); Ultrasound Imaging System (LU700L); Ultrasound Imaging System (LU710L); Ultrasound Imaging System (LU710LH); Ultrasound Imaging System (LU710C); Ultrasound Imaging System (LU710M); Ultrasound Imaging System (LU710PA); Ultrasound Imaging · Leltek, Inc. · Sep 11, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
July 10, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vave Health, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114
Re: K191541
Trade/Device Name: Vave Personal Ultrasound Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX Dated: June 7, 2019 Received: June 11, 2019
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
K191541
Device Name Vave Personal Ultrasound
Indications for Use (Describe)
The Vave Personal Ultrasound is intended for diagnostic ultrasound imaging in B-Mode.
It is indicated for diagnostic ultrasound imaging in the following applications: Fetal/Obstetric, Abdominal (includes Gynecology, Renal, and Urology), Pediatric, Thoracic/Pleural, Cardiac Pediatric, Peripheral Vessel and procedural guidance of needles into the body.
The Vave Personal Ultrasound is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
| Clinical Application | | Mode of Operation | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Other |
| Ophthalmic | Ophthalmic | | | | | | |
| | Fetal/OB | N | | | | | |
| | Abdominal (1) | N | | | | | |
| | Intra-operative (Specify) | | | | | | |
| | Intra-operative (Neuro) | | | | | | |
| | Laparoscopic | | | | | | |
| | Pediatric | N | | | | | |
| | Small Organ (Specify) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | |
| | Intravascular | | | | | | |
| | Other (Specify) (2) | N | | | | | |
| | Cardiac Adult | N | | | | | |
| | Cardiac Pediatric | N | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | |
| | Intra-cardiac | | | | | | |
| | Other (Specify) | | | | | | |
| Peripheral | Peripheral vessel (3) | N | | | | | |
| Vessel | Other (Non-vascular) (4) | N | | | | | |
System: Vave Personal Ultrasound Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Notes:
N = new indication; P = previously cleared by FDA; E = added under this appendix
(1) Abdominal includes Gynecology, Renal and Urology
(2) Other use includes Thoracic/Pleural detection of fluid and pleural motion/sliding
- (3) Peripheral Vessel includes arteries and veins
(4) Non-vascular is image guidance for freehand needle/catheter placement
{4}------------------------------------------------
| Clinical Application | | Mode of Operation | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Other |
| Ophthalmic | Ophthalmic | | | | | | |
| | Fetal/OB | N | | | | | |
| | Abdominal (1) | N | | | | | |
| | Intra-operative (Specify) | | | | | | |
| | Intra-operative (Neuro) | | | | | | |
| | Laparoscopic | | | | | | |
| | Pediatric | N | | | | | |
| | Small Organ (Specify) | | | | | | |
| | Neonatal Cephalic | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | |
| | Trans-rectal | | | | | | |
| | Trans-vaginal | | | | | | |
| | Trans-urethral | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | |
| | Intravascular | | | | | | |
| | Other (Specify) (2) | N | | | | | |
| | Cardiac Adult | N | | | | | |
| | Cardiac Pediatric | N | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | |
| | Intra-cardiac | | | | | | |
| | Other (Specify) | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel (3) | N | | | | | |
| | Other (Non-vascular) (4) | N | | | | | |
Device Name: Vave Phased VP3 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Notes:
N = new indication; P = previously cleared by FDA; E = added under this appendix
(1) Abdominal includes Gynecology, Renal and Urology
(2) Other use includes Thoracic/Pleural detection of fluid and pleural motion/sliding
- (3) Peripheral Vessel includes arteries and veins
(4) Non-vascular is image guidance for freehand needle/catheter placement
{5}------------------------------------------------
# 510(K) SUMMARY
191541 510(k) Notification K_
#### GENERAL INFORMATION
Applicant:
Vave Health, Inc. 2955 Campus Drive, Suite 110 San Mateo, CA U.S.A. Phone: (415)264 -5284
Contact Person:
Joseph Ko Director of Quality Assurance and Regulatory Vave Health, Inc. Phone: (415) 264 -5284 Email: Regulatory@vavehealth.com
Date Prepared:
May 13, 2019
## DEVICE INFORMATION
Trade Name: Vave Personal Ultrasound
Generic/Common Name:
Ultrasound Imaging System
Classification:
Class II
Classification Name:
| 21 CFR Section | Classification Name | Product Code |
|----------------|---------------------------------------|--------------|
| 892.1560 | Ultrasonic Pulsed Echo Imaging System | 90 IYO |
| 892.1570 | Diagnostic Ultrasonic Transducer | 90 ITX |
{6}------------------------------------------------
#### SUBSTANTIALLY EQUIVALENT DEVICES
- . Primary Predicate Device: Vscan with Dual Probe (K140693)
- . Reference Device: Clarius Ultrasound Scanner (K172385)
### DEVICE DESCRIPTION
The Vave Personal Ultrasound is a handheld ultrasound imaging system consisting of the following components: Ultrasound Probe, Display Application, Battery Charger.
The ultrasound probe, including its embedded software, is designed to acquire ultrasound image data and to wirelessly transmit the data to the display application software. The display application software runs on a commercial off-the-shelf mobile device. The operating system of the mobile device can be either iOS or Android. The ultrasound probe is completely wireless and is powered by a removable Li-ion battery is charged with a proprietary charging system.
The Vave Personal Ultrasound is intended for use in environments where healthcare is provided by healthcare professionals, including home healthcare environments and emergency medical services, ambulance, or aircraft environments.
## INDICATIONS FOR USE
The Vave Personal Ultrasound is intended for diagnostic ultrasound imaging in B-Mode. It is indicated for diagnostic ultrasound imaging in the following applications: Fetal/Obstetric. Abdominal (includes Gynecology, Renal, and Urology), Pediatric, Thoracic/Pleural, Cardiac Adult, Cardiac Pediatric, Peripheral Vessel and procedural guidance of needles into the body.
The Vave Personal Ultrasound is a transportable ultrasound system that is intended for use in environments where healthcare is provided by healthcare professionals.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The intended use including all indications for use of the Vave Personal Ultrasound are the same as the primary predicate device (K140693). The fundamental technology of the Vave Personal Ultrasound is the same as the predicate (K140693) and reference (K172385) devices. Vave Personal Ultrasound introduces no new indications for use, modes, features, or technologies relative to the predicate devices. The Vave Personal Ultrasound is substantially equivalent to the predicate devices.
{7}------------------------------------------------
| Comparison Parameter | Subject Device: Vave Personal Ultrasound | Primary Predicate Device: Vscan with Dual Probe with Transducer G3S (K140693) | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Intended Use | Diagnostic ultrasound imaging or fluid flow analysis of the human body | Diagnostic ultrasound imaging or fluid flow analysis of the human body | |
| Indications for use | Fetal/OB | Yes | |
| | Abdominal | Yes | |
| | Intra-operative (Specify) | No | |
| | Pediatric | Yes | |
| | Small Organ | No | |
| | Neonatal Cephalic | No | |
| | Adult Cephalic | No | |
| | Musculoskeletal (Conventional) | No | |
| | Musculoskeletal (Superficial) | No | |
| | Urology | Yes | |
| | Gynecology | Yes | |
| | Renal | Yes | |
| | Thoracic/Pleural | Yes | |
| | Cardiac adult | Yes | |
| | Cardiac pediatric | Yes | |
| | Fetal echo | No | |
| | Peripheral vessel | Yes | |
| | Carotid | No | |
| | Needle Guidance | Yes | |
| Environment of Use | Professional | Yes | |
| | Home | Yes | |
| | Emergency Medical Services | Yes | |
| 510(k) Track | | Track 3 | |
| Patient-Contacting Materials | | All materials with patient contact are biocompatible and can be disinfected | |
| Transducer Technology | | Piezoelectric | |
| Transducer Types | Phased Array | Yes | |
| | Linear Array | No | |
| | Convex Array | No | |
| Comparison Parameter | Subject Device:<br>Vave Personal Ultrasound | Primary Predicate<br>Device: Vscan with Dual<br>Probe with Transducer<br>G3S<br>(K140693) | |
| Principle/Method of Operation | Piezoelectric material in<br>the transducer is used as an<br>ultrasound source to<br>transmit sound waves into<br>the body. Sound waves are<br>reflected back to the<br>transducer and converted to<br>electrical signals that are<br>processed and displayed as<br>images of anatomic<br>structures. | Piezoelectric material in<br>the transducer is used as an<br>ultrasound source to<br>transmit sound waves into<br>the body. Sound waves are<br>reflected back to the<br>transducer and converted to<br>electrical signals that are<br>processed and displayed as<br>images of anatomic<br>structures. | |
| Modes of<br>Operation | B-Mode | Yes | Yes |
| | Color Doppler | No | Yes |
| | M-Mode | No | No |
| | Combined (B+CD) | No | Yes |
{8}------------------------------------------------
### SUMMARY OF NONCLINICAL TESTING
All necessary performance testing was conducted on the Vave Personal Ultrasound to support a determination of substantial equivalence to the predicate devices. The nonclinical testing conducted on the Vave Personal Ultrasound includes:
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 ● And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment -Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment Part ● 1-2: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
- IEC 60601-1-11 Edition 2.0 2015-01 Medical Electrical Equipment Part 1-● 11: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
- IEC 60601-1-12 Edition 1.0 2014-06 Medical Electrical Equipment Part 1-● 12: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Intended For Use In The Emergency Medical Services Environment
{9}------------------------------------------------
- IEC 60601-2-37 Edition 2.1 2015 Medical Electrical Equipment Part 2-37: . Particular Requirements For The Basic Safety And Essential Performance Of Ultrasonic Medical Diagnostic And Monitoring Equipment
- IEC 62366-1 Edition 1.0 2015-02 Medical Devices Part 1: Application Of ● Usability Engineering To Medical Devices [Including CORRIGENDUM 1 (2016)]
- IEC 62133-2 Edition1.0 2017-02 Secondary Cells And Batteries Containing ● Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells. And For Batteries Made From Them. For Use In Portable Applications - Part 2: Lithium Systems
- ANSI AAMI IEC 62304:2006/A1:2016 Medical Device Software -● Software Life Cycle Processes [Including Amendment 1 (2016)]
- ISO 10993-1 Fifth Edition 2018-08 Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process
The collective results of the non-clinical testing demonstrate that the Vave Personal Ultrasound meets the established specifications necessary for consistent performance for its intended use.
## SUMMARY OF CLINICAL TESTING
The Vave Personal Ultrasound introduces no new indications for use, modes, features or technologies relative to the predicate devices that require clinical testing. The clinical safety and effectiveness of ultrasound systems with these characteristics are well accepted for both predicate and subject devices.
### QUALITY ASSURANCE MEASURES
Quality assurance measures applied to the system design and development include, but are not limited to:
- Risk Analysis
- . Product Specifications
- Design Reviews ●
- . Verification and Validation
### CONCLUSION
The Vave Personal Ultrasound is substantially equivalent to the identified predicate devices. The Vave Personal Ultrasound has the same intended use and fundamental technological characteristics as the identified predicate devices.
### PRESCRIPTION STATUS
This is a prescription device. The prescription device statement appears in the labeling.
{10}------------------------------------------------
## STERILIZATION SITE(S)
Not applicable. There are no components in the Vave Personal Ultrasound supplied sterile.
# TRACK
This is a Track 3 system.