← Product Code FRO · K191536

# Biatain Silicone Ag (K191536)

_Coloplast A/S · FRO · Feb 21, 2020 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191536

## Device Facts

- **Applicant:** Coloplast A/S
- **Product Code:** FRO
- **Decision Date:** Feb 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

Biatain® Silicone Ag is intended to provide a moist wound environment and exudate management of acute and chronic wounds.

## Device Story

Biatain Silicone Ag is a sterile, single-use wound dressing designed to maintain a moist wound environment and manage exudate. The device consists of a semi-permeable polyurethane (PU) top film, a 3.0 mm PU foam absorbing pad containing a silver sodium hydrogen zirconium phosphate antimicrobial complex, and a perforated silicone adhesive bi-layer. The dressing is applied to the wound site by healthcare professionals in hospitals, clinics, or home care settings. The silver complex provides antimicrobial activity against various bacteria, yeast, and mold for up to 7 days. The dressing absorbs exudate while the silicone adhesive facilitates secure attachment and atraumatic removal. By managing exudate and providing an antimicrobial barrier, the device supports optimal wound healing conditions and protects the wound from microbial contamination.

## Clinical Evidence

No clinical data was required to support substantial equivalence. Bench testing included: real-time and accelerated aging (ASTM F1980-07), packaging integrity (ASTM D4169-16), biocompatibility (ISO 10993-1), fluid handling capacity/absorptive capacity/moisture vapor transmission rate (EN 13726-1), waterproofness (EN 13726-3), dynamic friction (EN ISO 8295), conformability (EN 13726-4), peel adhesion (ASTM D3330), and antimicrobial effectiveness (modified AATCC 100:2012). All tests passed.

## Technological Characteristics

Materials: Polyurethane (PU) top film, 3.0 mm PU foam with silver sodium hydrogen zirconium phosphate, silicone adhesive gel. Dimensions: Various sizes/shapes. Connectivity: None. Sterilization: Ethylene oxide (EO) to SAL 10^-6. Antimicrobial: Ionic silver (0.95 mg/cm2). Standards: ASTM F1980-07, ASTM D4169-16, ISO 10993-1, EN 13726-1/3/4, EN ISO 8295, ASTM D3330, AATCC 100.

## Predicate Devices

- Biatain® Ag ([K100218](/device/K100218.md))

## Reference Devices

- Biatain® Silicone Ag ([K120828](/device/K120828.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 21, 2020

Coloplast A/S Lykke Forchhammer Director of RA Holtedam 1 Humlebaek, 3050 Dk

Re: K191536

Trade/Device Name: Biatain Silicone Ag Regulatory Class: Unclassified Product Code: FRO Dated: January 16, 2020 Received: January 21, 2020

Dear Lykke Forchhammer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kimberly M. Ferlin, Ph.D. Assistant Director (Acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K191536

Device Name Biatain Silicone Ag

Indications for Use (Describe)

Biatain® Silicone Ag is intended to provide a moist wound environment and exudate management of acute and chronic wounds.

Biatain® Silicone Ag is indicated for the management of exuding leg and foot ulcers, diabetic foot ulcers, superficial and partial thickness burns, donor sites, and, traumatic and post-operative wounds.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
|------------------------------------------------------------------------------------------------|---------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 5. TRADITIONAL 510(K) SUMMARY

| Submitted by:                 | Coloplast A/S<br>Holtedam 1<br>3050 Humlebaek<br>Denmark                                                                                                                                  |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:               | Lykke Forchhammer<br>Director of RA Innovation and Market Expansion<br>Coloplast A/S<br>Holtedam 1<br>3050 Humlebaek<br>Denmark<br>Telephone: 0045-49111271<br>Email: dklfo@coloplast.com |
| Date of Summary:              | 21 February 2020                                                                                                                                                                          |
| Trade or Proprietary<br>Name: | Biatain® Silicone Ag                                                                                                                                                                      |
| Common or Usual Name:         | Dressing, Wound, Drug                                                                                                                                                                     |
| Classification and CFR:       | Unclassified                                                                                                                                                                              |
| Product Code:                 | FRO                                                                                                                                                                                       |
| Review Panel:                 | General and Plastic Surgery                                                                                                                                                               |
| Predicate Device:             | K100218, Biatain® Ag (Coloplast A/S)                                                                                                                                                      |
| Reference Device:             | K120828, Biatain® Silicone Ag (Coloplast A/S)                                                                                                                                             |

{4}------------------------------------------------

Device Description: The device is available as sterile, single use only. The device is intended to provide a moist environment and exudate management of acute and chronic wounds. The device maintains a moist wound providing the optimal environment for wound healing.

The device consists of:

- Polyurethane (PU) top film - a 25 um semi-permeable barrier, printed with grey dots.
- . Absorbing pad - composed of a 3.0 mm PU foam containing an anti-bacterial silver complex (silver sodium hydrogen zirconium phosphate).
- Silicone adhesive bi-layer composed of a middle PU film . with a silicone adhesive gel. The PU top film and the middle PU film are laminated together at the border.
- . Protective film - composed of three or five parts; the center part and the remaining side parts.

The device:

- . contains a silver compound (silver sodium hydrogen zirconium phosphate). In vitro testing demonstrated that Biatain® Silicone Ag has an antimicrobial effect in the dressing against the following strains of 3 gram positive bacteria, 3 gram negative bacteria, 1 yeast, and 1 mold: Staphylococcus aureus, Enterococcus faecalis, Streptococcus pyogenes, Proteus mirabilis, Klebsiella pneumoniae, Escherichia coli, Aspergillus brasiliensis and Meyerozyma guilliermondii. The dressing sustains antimicrobial activity for up to 7 days.
The device may be used in hospitals, healthcare facilities, and home care.

Sterilized using ethylene oxide (EO).

Prescription Use Only.

{5}------------------------------------------------

Biatain® Silicone Ag is intended to provide a moist wound Indications for Use: environment and exudate management of acute and chronic wounds. Biatain® Silicone Ag is indicated for the management of exuding leg and foot ulcers, pressure ulcers, diabetic foot ulcers, superficial and partial thickness burns, donor sites, and, traumatic and postoperative wounds. Technological The subject device is substantially equivalent to the predicate Characteristics: device based upon the information below.

{6}------------------------------------------------

|                                                            | Subject Device<br>Biatain® Silicone Ag                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Biatain® Ag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device:<br>Biatain® Silicone Ag                                                                                                                                                                                                                           |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter                                           | Coloplast A/S                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| 510(k) Number                                              | K191536                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K100218                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K120828                                                                                                                                                                                                                                                             |
| Regulatory Class                                           | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Classification Name                                        | Dressing, Wound, Drug                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Product Code                                               | FRO                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Intended Use                                               | Biatain® Silicone Ag is intended to<br>provide a moist wound environment and<br>exudate management.                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
|                                                            | Subject Device<br>Biatain® Silicone Ag                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Biatain® Ag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device:<br>Biatain® Silicone Ag                                                                                                                                                                                                                           |
| Indications for<br>Use                                     | Biatain® Silicone Ag is intended to<br>provide a moist wound environment and<br>exudate management of acute and<br>chronic wounds.<br>Biatain® Silicone Ag is indicated for<br>the management of exuding leg and foot<br>ulcers, pressure ulcers, diabetic foot<br>ulcers, superficial and partial thickness<br>burns, donor sites, and, traumatic and<br>post-operative wounds.                                                                                 | Biatain Ag foam dressings are indicated for<br>wounds with moderate to high amounts of<br>exudate, including leg ulcers and Category II-<br>IV pressure ulcers with delayed healing due to<br>bacteria, or where there is a risk of infection.<br>Biatain Ag foam dressings may be used for<br>second-degree burns, donor sites,<br>postoperative wounds and skin abrasions with<br>delayed healing due to bacteria, or where<br>there is a risk of infection. Biatain Ag non-<br>adhesive foam dressings are additionally<br>indicated for diabetic foot ulcers. Biatain Ag<br>foam dressings may reduce odor caused by<br>micro-organisms in the wound. Biatain Ag<br>foam dressings may be used to support moist<br>wound healing on patients who are in<br>treatment for a local or systemic infection<br>under the discretion of a healthcare<br>professional. Depending on the prognosis of<br>the wound, Biatain Ag foam dressings may be<br>used throughout the healing process to<br>provide protection for the indicated types of<br>wounds. Biatain Ag foam dressings are<br>suitable for use in combination with<br>compression therapy. Biatain Ag foam<br>filler/cavity is indicated for deep exuding<br>wounds, including leg ulcers, pressure ulcers,<br>diabetic foot ulcers and acute wounds. | Biatain® Silicone Ag Foam<br>Dressings are indicated for the<br>management of moderately to<br>highly exuding leg ulcers and<br>pressure sores. The dressing can<br>also be used for 2nd degree burns,<br>donor sites, postoperative wounds,<br>and skin abrasions. |
| Prescription<br>Device                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Single Use                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Sterile Device                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Method of                                                  | EO                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Irradiated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same (EO)                                                                                                                                                                                                                                                           |
| Sterilization                                              | SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                         | SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | SAL 10-6                                                                                                                                                                                                                                                            |
|                                                            | Subject Device<br>Biatain® Silicone Ag                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Biatain® Ag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device:<br>Biatain® Silicone Ag                                                                                                                                                                                                                           |
| Recommended<br>Wear Time                                   | Up to 7 days. Repeated use of the<br>device should not exceed 30 days                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Shelf Life                                                 | 2-years                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 3-years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 3-years                                                                                                                                                                                                                                                             |
| Silver (Ag)                                                | Silver Compound (Ionic)<br>[silver sodium hydrogen zirconium<br>phosphate]                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Silver Content                                             | 0.95 mg/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Silver Release<br>(within the<br>dressing)                 | 24 h: 207 – 367 µg/cm2<br>48 h: 350 – 589 µg/cm2<br>72 h: 439 – 726 µg/cm2<br>96 h: 500 – 813 µg/cm2<br>168 h: 591 – 937 µg/cm2                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
|                                                            | Subject Device<br>Biatain® Silicone Ag                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Biatain® Ag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device:<br>Biatain® Silicone Ag                                                                                                                                                                                                                           |
| Antimicrobial<br>Effectiveness<br>within the<br>Dressing   | Fulfills both ≥4 log reduction (AATCC<br>100) as described below:<br><br>Fulfills ≥4 log reduction (AATCC 100)<br>for antimicrobial effect against 3 gram<br>positive bacteria, 3 gram negative<br>bacteria, 1 yeast, and 1 mold:<br><i>Staphylococcus aureus, Enterococcus<br/>faecalis, Streptococcus pyogenes,<br/>Proteus mirabilis, Klebsiella<br/>pneumoniae, Escherichia coli,<br/>Aspergillus brasiliensis</i> and<br><i>Meyerozyma guilliermondii</i> . | Effective antimicrobial activity within<br>the dressing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Effective antimicrobial activity<br>within the dressing.                                                                                                                                                                                                            |
| Primary<br>Packaging                                       | Paper/Polyester Film                                                                                                                                                                                                                                                                                                                                                                                                                                             | Biatain® Ag Non-adhesive: White<br>Polyester Film/Transparent Polyester<br>Film<br>Biatain® Ag Adhesive: Foil                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Tyvek/Transparent Foil                                                                                                                                                                                                                                              |
|                                                            | Subject Device<br>Biatain® Silicone Ag                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Biatain® Ag                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device:<br>Biatain® Silicone Ag                                                                                                                                                                                                                           |
| Top Film                                                   | Vapor permeable polyurethane (PU) top<br>film                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                |
| Pigment:<br>Top Film<br>Colorant                           | CI Pigment Black 7<br>Titanium dioxide<br>Aluminum hydroxide                                                                                                                                                                                                                                                                                                                                                                                                     | Black iron oxide<br>Titanium dioxide<br>C.I Pigment Black 7                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Black iron oxide<br>Titanium dioxide                                                                                                                                                                                                                                |
| Pigment:<br>Top Film Dots                                  | C.I. Pigment White 6<br>C.I. Pigment Black 7<br>Titanium chelates<br>Aluminum oxide                                                                                                                                                                                                                                                                                                                                                                              | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Not Applicable                                                                                                                                                                                                                                                      |
| Pigment:<br>Top Film Logo                                  | CI Pigment Black 7                                                                                                                                                                                                                                                                                                                                                                                                                                               | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Not Applicable                                                                                                                                                                                                                                                      |
| Absorbaent<br>Polyurethane<br>Foam with Silver<br>Compound | Polyurethane (PU) foam with silver<br>complex (silver sodium hydrogen<br>zirconium phosphate)                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same and contains hotmelt<br>Pressure Sensitive Adhesive (PSA)                                                                                                                                                                                                      |
| Polyurethane<br>Foam Thickness                             | 3.0 mm PU foam with silver complex                                                                                                                                                                                                                                                                                                                                                                                                                               | 3.0 mm and 4.4 mm PU foam with silver<br>complex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same (3.0 mm PU foam with<br>silver complex)                                                                                                                                                                                                                        |
| Adhesive Bilayer                                           | Silicone adhesive bi-layer across the<br>device with perforations across the<br>absorbent pad                                                                                                                                                                                                                                                                                                                                                                    | Biatain® Ag Non-adhesive Dressings<br>have no adhesive layer.<br>Biatain® Ag Adhesive: Hydrocolloid<br>adhesive border                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Perforated Silicone adhesive bi-<br>layer on the border                                                                                                                                                                                                             |
| Protective Film<br>(Release Liner)                         | Polypropylene (PP)<br>3 and 5-piece design                                                                                                                                                                                                                                                                                                                                                                                                                       | Polyethylene (PE)<br>2 and 3-piece design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Polyethylene (PE)<br>3-piece design                                                                                                                                                                                                                                 |
| Device Sizes and<br>Shapes.                                | Square:<br>7.5 x 7.5 cm<br>10 x 10 cm<br>12.5 x 12.5 cm<br>15 x 15 cm<br>17.5 x 17.5 cm                                                                                                                                                                                                                                                                                                                                                                          | Square:<br>10 x 10 cm (non-adhesive)<br>12.5 x 12.5 cm (adhesive)<br>15 x 15 cm (non-adhesive)<br>18 x 18 cm (adhesive)<br>20 x 20 cm (non-adhesive)…

---

**Source:** [https://fda.innolitics.com/device/K191536](https://fda.innolitics.com/device/K191536)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
