K191531 · Shina Med Corporation · FMF · Sep 6, 2019 · General Hospital
Device Facts
Record ID
K191531
Device Name
Sure-Fine Insulin Syringes
Applicant
Shina Med Corporation
Product Code
FMF · General Hospital
Decision Date
Sep 6, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
Device Story
Sure-Fine Insulin Syringes are single-use, sterile, disposable piston syringes designed for subcutaneous U100 insulin delivery. The device consists of a graduated polypropylene barrel, plunger rod, isoprene rubber piston, needle cap, protective end cap, and a stainless steel needle (STS304) permanently affixed with epoxy. The device is operated manually by the patient or healthcare provider to draw and inject a specific dose of insulin. It functions via standard piston syringe mechanics. The device is provided sterile via ethylene oxide gas. The modification adds 27G needle gauge and 1/4" needle length options to the existing product line. The device is intended for over-the-counter use.
Clinical Evidence
Bench testing only. The device complies with ISO 7864 (sterile hypodermic needles), ISO 9626 (stainless steel needle tubing), and ISO 8537 (sterile single-use insulin syringes). Biocompatibility conforms to ISO 10993-1.
Technological Characteristics
Materials: STS304 needle, polypropylene barrel/plunger, isoprene rubber piston, polyethylene caps, polydimethylsiloxane lubricant. Principle: Piston syringe. Energy: Manual. Connectivity: None. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 7864, ISO 9626, ISO 8537, ISO 10993-1.
Indications for Use
Indicated for subcutaneous injection of U100 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
September 6, 2019
Shina Med Corporation Park Sung-Soon Manager 455-30 Bogaewonsam-ro, Bogae-myun Anseong-si, 456-871 KOREA
Re: K191531
Trade/Device Name: Sure-Fine Insulin Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: August 6, 2019 Received: August 9, 2019
Dear Park Sung-Soon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191531
Device Name Sure-Fine Insulin Syringes
Indications for Use (Describe)
Sure-fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for SHINA MED CORPORATION. The logo consists of a stylized circle with a wavy line at the bottom left, followed by the company name in bold, sans-serif font. The text is all capitalized and the logo is in black and white.
# 510(K) SUMMARY - K191531 September 6, 2019
# 1. Submitted by :
Sung-soon, Park / Quality Management Representative
# SHINA MED CORPORATION
455-30, Bogaewonsam-ro Bogae-myeon, Anseong-si, Gyeonggi-do, 17509, Rep. of Korea
Phone : +82 31 8057 2125 Fax : + 82 31 8057 2150
### 2. Device Name :
| - Trade Name | : | Sure-Fine Insulin Syringes |
|-----------------------|---|----------------------------|
| - Classification | : | Class II |
| - Classification Name | : | Piston syringe |
| - Product Code | : | FMF |
| - Regulation Number | : | 21 CFR 880.5860 |
| - Review Panel | : | General Hospital |
### 3. 510(k) Premarket Notification: K152879
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for SHINA MED CORPORATION. The logo consists of a stylized graphic to the left of the company name. The graphic appears to be a circle with a wavy line at the bottom. The company name is written in a bold, sans-serif font.
## 510(k) Premarket Notification
| | | CFR Title 21 Radiation-Emitting Products X-Ray Assembler Medsun Reports CLIA TPLC | |
|------------|------------------------------|---------------------------------------------------------------------------------------------|------------------------|
| New Search | | | Back To Search Results |
| | Device Classification Name | Syringe, Piston | |
| | 510(K) Number | K152879 | |
| | Device Name | Sure-Fine Insulin Syringes | |
| | Applicant | SHINA MED CORPORATION<br>455-30, BOGAWONSAM-RO BOGAE-MYUN<br>Anseong-Si, KR 456-871 | |
| | Applicant Contact | Sung-Soon Park | |
| | Correspondent | LK CONSULTING GROUP USA, INC.<br>2651 E CHAPMAN AVE STE 110<br>Fullerton, CA 92831 | |
| | Correspondent Contact | Priscilla Chung | |
| | Regulation Number | 880.5860 | |
| | Classification Product Code | FMF | |
| | Date Received | 09/30/2015 | |
| | Decision Date | 09/09/2016 | |
| | Decision | Substantially Equivalent (SESE) | |
| | Regulation Medical Specialty | General Hospital | |
| | 510k Review Panel | General Hospital | |
| | Summary | Summary | |
| | Type | Traditional | |
| | Reviewed By Third Party | No | |
#### Device Description: 4
### 4.1 Predicate Device Description(K152879)
Sure-fine Insulin syringes are designed for the subcutaneous injection of a desired dose of insulin.
The syringe has a graduated barrel, a plunger rod, needle cap, protective end cap and needle permanently affixed to the tip of the syringe with epoxy. The syringes are available in the following sizes and cap color.
| Category | Insulin syringe | Needle Gauge | Needle Length | Cap color | |
|----------|-----------------------|--------------|---------------|------------|-----------------|
| | | | | Needle cap | Protective cap |
| U-100 | 1/2cc and 1cc | 28Gauge | 1/2" | Orange | Orange or white |
| | 3/10cc, 1/2cc and 1cc | 29Gauge | 1/2" | Orange | Orange or white |
| | 3/10cc, 1/2cc and 1cc | 30Gauge | 1/2" | Orange | Orange or white |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for SHINA MED CORPORATION. The logo consists of a circle with a wavy line at the bottom left, followed by the text "SHINA MED CORPORATION" in bold, sans-serif font. The logo is simple and professional, and the text is easy to read.
| 3/10cc, 1/2cc and 1cc | 30Gauge | 5/16" | Orange | Orange or white |
|-----------------------|---------|-------|--------|-----------------|
| 3/10cc, 1/2cc and 1cc | 31Gauge | 5/16" | Orange | Orange or white |
### 4.2 Device description for Special 510(k), K191531
The modified Sure-Fine Insulin syringe is substantially equivalent to the predicate device (Sure-Fine Insulin Syringes, K152879). The biggest difference between the modified device and predicate device is the addition of needle gauge (27G) and the addition of needle length (1/4").
The design features and sizing of the components were compared and the Sure-Fine Insulin Syringe is found to be substantially equivalent to the predicate device. The similarity in design between modified Sure-Fine Insulin Syringes and the previous device supports the substantial equivalence of Sure-Fine Insulin Syringe for the indicated use.
Therefore, modified Sure-Fine Insulin Syringe is substantially equivalent to the previous device in terms of indications, compositions, material, design, safety and effectiveness. The syringes are available in the following sizes and cap color.
| Category | Insulin syringe | Needle<br>Gauge | Needle<br>Length | Cap color | |
|----------|------------------|-----------------|------------------|------------|-----------------|
| | | | | Needle cap | Protective cap |
| U-100 | 3/10cc and 1/2cc | 30Gauge | 1/4" | Orange | Orange or white |
| U-100 | 3/10cc and 1/2cc | 31Gauge | 1/4" | Orange | Orange or white |
| U-100 | 1cc | 27Gauge | 1/2" | Orange | Orange or white |
The devices operate in the principles of a piston syringe fluid path is sterile (EO gas sterilization), Non-toxic, Non-pyrogenic and single use disposable.
### 4.3 Substantial Equivalence Discussion
Modified Sure-Fine Insulin syringe is substantially equivalent to the predicate Sure-Fine Insulin Syringes device(K152879).
The design features and sizing of the components were also compared and the Sure-Fine
{6}------------------------------------------------
Insulin Syringe is substantially the same as the predicate device product. The similarity in design between modified Sure-Fine Insulin Syringes and the previous device supports the safety and effectiveness of Sure-Fine Insulin Syringe for the indicated use. The biggest difference between modified device and previous device is the addition of needle gauge (27G) and the addition of needle length (1/4").
Therefore, modified Sure-Fine Insulin Syringe is substantially equivalent to the previous device in terms of indications, compositions, material, design, safety and effectiveness.
| Device Name | | Subject Device<br>(Special 510(k)) | Predicate Device#1<br>(Existing Traditional<br>510(k)) |
|--------------------|--------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Manufacturer | | SHINA MED CORPORATION | SHINA MED<br>CORPORATION |
| 510(k) Number | | N/A | K152879 |
| Product Code | | FMF | FMF |
| Intended Use | | Sure-Fine Insulin Syringes are<br>hypodermic insulin syringes for<br>subcutaneous injection of U100<br>insulin. | Sure-Fine Insulin<br>Syringes are hypodermic<br>insulin syringes for<br>subcutaneous injection of<br>U100 insulin. |
| Design (Syringe) | Needle<br>/ Barrel | Image: Needle / Barrel | Image: Needle / Barrel |
| | Gasket<br>/ Plunger | Image: Gasket / Plunger | Image: Gasket / Plunger |
| | Needle<br>Cap | Image: Needle Cap | Image: Needle Cap |
| | Protective<br>end<br>cap | Image: Protective end cap | Image: Protective end cap |
| | Volume | 0.3cc, 0.5cc, 1.0cc | 0.3cc, 0.5cc, 1.0cc |
| Design<br>(Needle) | Needle<br>Tip<br>shape | Image: Needle Tip shape | Image: Needle Tip shape |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SHINA MED CORPORATION. The logo consists of a stylized graphic to the left of the company name. The company name is written in a bold, sans-serif font. The logo is simple and professional.
## Special 510(K) Submission
| | Gauge | 27, 30, 31 Gauge | 28, 29, 30,31Gauge |
|--------------------------------|--------------------------------------------------|--------------------------------------------------|--------------------------------------------------|
| | Length | 1/2", 1/4" | 1/2", 5/16" |
| Materials | Needle | STS304 | STS304 |
| | Barrel | Polypropylene | Polypropylene |
| | Plunger | Polypropylene | Polypropylene |
| | Piston | Isoprene Rubber | Isoprene Rubber |
| | Needle<br>Cap | Polyethylene | Polyethylene |
| | Protective end cap | Polyethylene | Polyethylene |
| | Silicon | Polydimethylsiloxane | Polydimethylsiloxane |
| Biocompatibility | Conform ISO10993 -1 | Conform ISO10993-1 | Conform ISO10993-1 |
| Sterilization method and S.A.L | Sterilized by ethylene oxide gas SAL = $10^{-6}$ | Sterilized by ethylene oxide gas SAL = $10^{-6}$ | Sterilized by ethylene oxide gas SAL = $10^{-6}$ |
#### ട് Indication for Use :
Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
( The indication for use is substantially equivalent to the predicate device.)
{8}------------------------------------------------
### Special 510(K) Submission
Image /page/8/Picture/1 description: The image shows the logo for SHINA MED CORPORATION. The logo consists of a circle with a wavy line at the bottom left, followed by the text "SHINA MED CORPORATION" in bold, sans-serif font. The logo is simple and professional, and it is likely used to represent the company's brand.
### 6. Technological Characteristics :
Sure-fine insulin syringe and predicate device have the substantially equivalent technological characteristics and perform as piston syringes. Risks associated with the changes were identified and appropriate design controls implemented to mitigate the risks.
Based on the change assessment, it is concluded that the risks are associated with the mechanical performance of the needle. To mitigate these risks, appropriate testing was completed to demonstrate that the syringes comply with the following FDA recognized standards:
- ISO7864 Sterile hypodermic needles for single use Requirements and test methods ●
- . ISO9626 - Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
- ISO8537 Sterile single-use syringes, with or without needle, for insulin .
### 8. Conclusion
Based on the information provided in this special 510(k), the Sure-fine insulin syringes are substantially equivalent to the previous devices (K152879).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.