← Product Code [DQX](/productcode/DQX) · K191502

# Heraeus Peripheral Guidewire (K191502)

_Heraeus Medical Components, LLC · DQX · Nov 26, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191502

## Device Facts

- **Applicant:** Heraeus Medical Components, LLC
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Nov 26, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal angioplasty (PTA), in peripheral arteries including but not limited to femoral, popliteal arteries. This guide wire may also be used with compatible interventional devices during therapeutic procedures. The guide wire may also be used to reach and cross a target lesion, provide a pathway within the vessel structure, facilitate the substitution of one diagnostic or interventional device for another, and to distinguish the vasculature.

## Device Story

Heraeus Peripheral Guidewire is a manually operated device used to facilitate placement of balloon dilation catheters and other interventional devices in peripheral arteries. It consists of a steel core with an optional nitinol alloy tip, encapsulated by polymer and a lubricious outer layer. The distal section features radiopaque loading and a hydrophilic coating. Operated by physicians in clinical settings, the guidewire is inserted into the vasculature through a guiding catheter and advanced to a target lesion. It provides a pathway for device exchange and helps distinguish vasculature. By enabling navigation to and crossing of target lesions, the device assists in therapeutic procedures, potentially improving patient outcomes in peripheral vascular interventions.

## Clinical Evidence

Bench testing only. Performance testing included biocompatibility (ISO 10993-1:2018), corrosion resistance, dimensional inspection, sterile package integrity, tensile strength, tip flexibility, torque strength, torqueability, kink/fracture resistance, radiopacity, coating integrity/lubricity, track/simulated use, and particulate evaluation. All predetermined acceptance criteria were met.

## Technological Characteristics

Steel core with optional nitinol alloy tip; encapsulated by polymer with lubricious outer layer. Distal section includes radiopaque loading and hydrophilic coating. Manual operation. Sterilization method equivalent to predicate.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Abbott Hi-Torque Command guidewire ([K122573](/device/K122573.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 26, 2019

Heraeus Medical Components, LLC Chelsea Pioske Regulatory Affairs Manager 2605 Fernbrook Lane North, Suite J Plymouth, Minnesota 55447

Re: K191502

Trade/Device Name: Heraeus Peripheral Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: October 25, 2019 Received: October 28, 2019

Dear Chelsea Pioske:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191502

Device Name Heraeus Peripheral Guidewire

#### Indications for Use (Describe)

Intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal angioplasty (PTA), in peripheral arteries including but not limited to femoral, popliteal arteries. This guide wire may also be used with compatible interventional devices during therapeutic procedures. The guide wire may also be used to reach and cross a target lesion, provide a pathway within the vessel structure, facilitate the substitution of one diagnostic or interventional device for another, and to distinguish the vasculature.

| Type of Use (Select one or both, as applicable)                                                                                           |                                                                                                                                          |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px"> <b> </b> </span>           Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size:16px"> <b> </b> </span>           Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Submitter:                                | Heraeus Medical Components, LLC<br>1805 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                           | Chelsea Becker, Regulatory Manager<br>1805 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 USA<br>763-225-6500 (voice)<br>763-559-7676 (fax)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Date Prepared:                            | November 25th, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:                               | Heraeus Peripheral Guidewire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name:                              | Guidewire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification:                           | Class II<br>Regulation Number: 21 CFR 870.1330<br>Catheter guide wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code:                             | DQX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device:                         | The subject device is substantially equivalent to K122573;<br>Abbott Hi-Torque Command guidewire.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device Description:                       | The Heraeus Peripheral Guidewire is a guidewire constructed<br>using a steel core with an optional nitinol alloy tip<br>encapsulated by polymer and lubricious outer layer. The<br>distal polymer includes a radiopaque loading, and the distal<br>polymer section has a hydrophilic coating.                                                                                                                                                                                                                                                                                                                                              |
| Indication for Use:                       | The peripheral guidewire is intended to facilitate the<br>placement of balloon dilatation catheters during percutaneous<br>transluminal angioplasty (PTA), in peripheral arteries<br>including but not limited to femoral, popliteal and infra-<br>popliteal arteries. This guide wire may also be used with<br>compatible interventional devices during therapeutic<br>procedures. The guide wire may also be used to reach and<br>cross a target lesion, provide a pathway within the vessel<br>structure, facilitate the substitution of one diagnostic or<br>interventional device for another, and to distinguish the<br>vasculature. |
| Contraindications:                        | Not intended for use in the coronary or cerebral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Principle of<br>Operation:                | The Heraeus Peripheral Guidewire is manually inserted into<br>vasculature through a compatible guiding catheter and<br>advanced to the target region.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Substantial<br>Equivalence<br>Comparison: | The Heraeus Peripheral Guidewire is substantially equivalent<br>to the Abbott Hi-Torque Command guidewire (K122573).<br>Substantial equivalence is based on indications for use,<br>physical and technological characteristics, and comparative<br>device testing.                                                                                                                                                                                                                                                                                                                                                                         |
|                                           | Indications for use:<br>• The indications for use for the Heraeus Peripheral<br>Guidewire are equivalent to the predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                           | Fundamental scientific technology<br>• Dimensional characteristics – the length and diameter of the device are similar to the dimensions of the predicate devices<br>• Materials – the device is constructed of equivalent materials to provide equivalent performance characteristics and coating properties<br>• Operating principle - equivalent to the predicate device<br>• Packaging materials - equivalent to the predicate device<br>• Sterility assurance level and method of sterilization - equivalent to the predicate devices                                                                                                 |
| Performance<br>Testing:                   | In vitro bench tests were utilized to demonstrate equivalence<br>with reference to FDA Guidance - Coronary and<br>Cerebrovascular Guidewire Guidance - January 1995.<br>The performance testing assessment supports that the<br>biocompatibility, shelf life, and functional specifications of<br>the proposed guidewire device were met.                                                                                                                                                                                                                                                                                                  |
|                                           | The Heraeus Peripheral Guidewire device test data supports<br>the claims of substantial equivalence to the predicate devices.<br>Biological Safety of the predicate device has been established<br>through biocompatibility testing carried out in compliance<br>with ISO 10993-1:2018 and the 2016 FDA guidance, "Use of<br>International Standard ISO 10993-1, “Biological evaluation of<br>medical devices – Part 1: Evaluation and testing within a risk<br>management process."                                                                                                                                                       |
|                                           | The following bench tests were conducted or evaluated to<br>support the proposed device:<br>• Biocompatibility testing:<br>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{4}------------------------------------------------

{5}------------------------------------------------

- Cytotoxicity o
- Sensitization o
- Intracutaneous Reactivity O
- Acute Systemic Toxicity O
- Pyrogenicity Testing O
- Hemolysis O
- In-Vivo Thrombogenicity O
- Coagulation O
- 0 Platelet Activation
- o Haematology
- o Complement Activation
- Corrosion resistance ●
- Dimensional inspection
- Sterile package integrity testing ●
- Tensile strength
- Tip flexibility ●
- Torque strength
- Torqueability
- Kink/Fracture Resistance
- Kink resistance
- Radiopacity ●
- Coating Integrity, Lubricity ●
- Track/Simulated Use ●
- Particulate Evaluation ●

## Conclusion:

The Heraeus Peripheral Guidewire met all predetermined acceptance criteria and compared favorably with the predicate

Heraeus considers the Heraeus Peripheral Guidewire to be equivalent to the predicate device. This conclusion is based upon the fact that the device has an equivalent intended use, and there are no differences that raise new types of questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K191502](https://fda.innolitics.com/device/K191502)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
