← Product Code [NAY](/productcode/NAY) · K191482

# Senhance Surgical System (K191482)

_Transenterix, Inc. · NAY · Jul 11, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K191482

## Device Facts

- **Applicant:** Transenterix, Inc.
- **Product Code:** [NAY](/productcode/NAY.md)
- **Decision Date:** Jul 11, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inquinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

## Device Story

The Senhance Surgical System is a computer-controlled instrument system for laparoscopic surgery. It provides visualization and enables manipulation of tissue (grasping, cutting, dissection, suturing, electrocautery) via robotic control of instruments. The system is operated by a physician in an operating room. This submission adds 5 mm Monopolar L-Hook Electrode instruments and an adapter to the existing system. These instruments are multi-use, requiring cleaning and steam sterilization between procedures. The system integrates with third-party electrosurgical units (ESUs) for cautery functions. The device assists surgeons in performing minimally invasive procedures with precision, potentially reducing patient trauma and recovery time compared to traditional manual laparoscopy.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included mechanical verification of cleaning effectiveness, biocompatibility (ISO 10993-1), electrical safety (IEC 60601-2-2, IEC 60601-2-18), and ESU compatibility. Results confirmed mechanical integrity, effective reprocessing, and compatibility with the Senhance system and third-party ESUs.

## Technological Characteristics

Multi-use 5 mm monopolar L-hook electrodes and adapter. Materials assessed for biocompatibility per ISO 10993-1. Connectivity includes RFID programming in the adapter. Electrical safety compliant with IEC 60601-2-2 and IEC 60601-2-18. Sterilization via steam. System is computer-controlled for laparoscopic instrument manipulation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- TransEnterix Senhance™ Surgical System ([K181517](/device/K181517.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name written out in a smaller font next to it.

July 11, 2019

TransEnterix Inc. Kaitlyn Alexander Senior Regulatory Affairs Specialist 635 Davis Drive, Suite 300 Morrisville, North Carolina 27560

Re: K191482

Trade/Device Name: Senhance Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: June 3. 2019 Received: June 4, 2019

Dear Kaitlyn Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Dr. Long Chen Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191482

#### Device Name

TransEnterix® Senhance™ Surgical System

#### Indications for Use (Describe)

The Senhance™ Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inguinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span style="font-size: 20px;">☐</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K191482

## 510(k) Summary

## Senhance™ 5 mm Monopolar Hook Instruments and Adapter

[In accordance with 21CFR 807.92]

## 1. Submitter

| 510(k) Sponsor:                       | TransEnterix, Inc.                                                                         |
|---------------------------------------|--------------------------------------------------------------------------------------------|
| Address:                              | 635 Davis Drive, Suite 300<br>Morrisville, NC 27560                                        |
| Contact Person:                       | Kaitlyn Alexander<br>Senior Regulatory Affairs Specialist                                  |
| Contact Information:                  | Email: kalexander@transenterix.com<br>Phone: 919-765-8400 x8505<br>Facsimile: 919.765.8459 |
| Date Summary Prepared:                | 7/10/2019                                                                                  |
| 2. Device                             |                                                                                            |
| Proprietary (Trade) Name:             | Senhance™ Surgical System                                                                  |
| Common Name:                          | System, Surgical, Computer Controlled Instrument                                           |
| Classification:                       | Class II                                                                                   |
| Classification Advisory<br>Committee: | General and Plastic Surgery                                                                |
| Regulation Number:                    | 21 CFR 876.1500, Endoscope and Accessories                                                 |
| Product Codes:                        | NAY (System, Surgical, Computer Controlled Instrument)                                     |
| 3. Predicate Device                   |                                                                                            |
| Predicate Device:                     | TransEnterix Senhance™ Surgical System (K181517)                                           |

{4}------------------------------------------------

# 4. Device Description:

The Senhance™ 5 mm Monopolar L-Hook Electrode instruments and adapter are intended as additions to the suite of monopolar hook instruments and adapters previously cleared for use with the TransEnterix Senhance™ Surgical System. Three monopolar electrodes and one monopolar electrode adapter are being added. The instruments and adapter are modifications of the predicate TransEnterix® Senhance™ 5 mm Monopolar L-Hook Electrode instruments and adapters FDA cleared through K181517 (October 9, 2018). Like the predicate devices, the subject devices are multi-use surgical devices that are cleaned and steam sterilized by the end user before the first use and after each use.

# 5. Intended Use/ Indications for Use:

The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in laparoscopic gynecological surgery, colorectal surgery, cholecystectomy, and inquinal hernia repair. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use.

## Comparison with Predicate Device Intended Use/ Indications for Use:

The Senhance Monopolar L-Hook Electrode instruments and adapter have the same intended use/ indications for use as previously cleared for the Senhance system through K181517. There are no differences in how the subject devices are used that alter the Senhance system's therapeutic effect or raise different questions of safety or effectiveness.

# 6. Technological Characteristics:

The Senhance Monopolar L-Hook Electrode instruments are very similar to the predicate devices cleared through K181517, with differences in manufacturer, instrument length, and materials of construction. The subject adapter is equivalent to the predicate monopolar hook adapter in overall design characteristics. The only differences for the new adapter are RFID programming and the content of the laser marking.

## Performance Data:

Performance testing of the Senhance Monopolar L-Hook Electrode instruments and adapter was conducted to support substantial equivalence to the predicate devices and demonstrated that the technological differences between the subject monopolar hook instruments and adapter and the predicate devices do not raise any different questions of safety or effectiveness.

Performance testing included mechanical verification of cleaning effectiveness, biocompatibility evaluation, electrical safety, and electrosurgical unit (ESU) compatibility. Due to the similarity of the subject devices to the predicate devices, no other testing was necessary.

The following performance testing was conducted for the devices in this submission.

{5}------------------------------------------------

### K191482

Biocompatibility testing: The Senhance Monopolar L-Hook Electrode instruments are categorized as tissue contacting for a limited duration of less than 24 hours for contact with tissue or bone. The instruments were assessed in accordance with the FDA Guidance for Industry and FDA Staff "Use of International Standard ISO 10993-1. Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" issued on June 16, 2016, and International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Testing demonstrated that the patient-contacting portions of the Senhance Monopolar L-Hook Electrode instruments are non-toxic, non-irritating, and do not result in an unacceptable adverse biological response resulting from contact of the devices' materials with the body.

Reprocessing, Cleaning, and Sterilization: The reusable Senhance Monopolar L-Hook Electrode instruments were tested for cleaning effectiveness to confirm the overall effectiveness of the prescribed cleaning procedures. The cleaning procedures and test protocols for the testing are consistent with the procedures and protocols utilized for the previously cleared monopolar hooks used with the Senhance Surgical System. The test results demonstrated that the cleaning procedures for the subject instruments allow them to be effectively cleaned according to the processing instructions provided in the labeling.

Performance Testing: Testing evaluated the performance of the Senhance Monopolar L-Hook Electrode instruments as well as compatibility of the subject devices when used with the Senhance Surgical System. Mechanical verification testing confirmed that the subject devices perform as intended after tests of mechanical integrity under conditions of simulated use and that the Senhance system performs as intended when used with the subject devices. Additionally, ESU compatibility testing demonstrated that all third-party ESUs that are compatible with the Senhance system are also compatible with the subject devices.

Electrical Safety and Compatibility: The monopolar hook instruments fulfilled all applicable requirements to demonstrate that the instruments are compliant with the current electrical safety standard, IEC 60601-2-2, Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, and IEC 60601-2-18, Medical Electrical Equipment - Part 2-18: Particular Requirements For The Basic Safety And Essential Performance Of Endoscopic Equipment.

## 7. Conclusions/ Substantial Equivalence:

The data acquired from the performance testing of the Senhance 5 mm Monopolar L-Hook Electrode instruments and adapter, as summarized herein, demonstrate that the devices are as safe and effective and perform similarly to the predicate devices cleared through K181517. The intended use/ indications for use for the subject devices are identical to those cleared for the Senhance Surgical System under K181517. The subject instruments and adapter do not raise any new issues of safety or effectiveness when compared to the predicate devices, thus, they are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K191482](https://fda.innolitics.com/device/K191482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com