ERBEFLO 2 Endo Quick Connect Pentax Scope Port Connector

K191438 · Erbe USA, Inc. · OCX · Jun 27, 2019 · Gastroenterology, Urology

Device Facts

Record IDK191438
Device NameERBEFLO 2 Endo Quick Connect Pentax Scope Port Connector
ApplicantErbe USA, Inc.
Product CodeOCX · Gastroenterology, Urology
Decision DateJun 27, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ERBEFLO 2 Disposable Tubing System is intended to provide sterile water from a water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures.

Device Story

The ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector is a single-use, sterile accessory for the ERBEFLO® 2 Disposable Tubing System. It functions as a mechanical interface between an irrigation tubing set and a Pentax gastrointestinal video endoscope equipped with a forward water jet channel. The device features a standard female luer lock connection, a backflow (check) valve, and a scope-specific housing. During endoscopic procedures, it facilitates the delivery of sterile water from an irrigation pump through the endoscope to the surgical site. The connector is disposable and intended for single-patient use. It is operated by clinical staff in an endoscopic setting. The device ensures a secure, leak-resistant connection to maintain irrigation flow, aiding visualization and procedural efficacy.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility evaluation, 2X sterilization functional testing (visual inspection, connection, leak, and flow testing), and sterilization validation (EO residuals).

Technological Characteristics

Materials: Medical-grade plastics and silicone. Connectivity: Standard female luer lock. Sterilization: Ethylene Oxide (EO). Standards: AAMI/ANSI/ISO 10993-1, ISO 594-1, ISO 594-2, AAMI/ANSI/ISO 15223-1, AAMI/ANSI/ISO 11607-1/2, AAMI/ANSI/ISO 11135(-1), AAMI/ANSI/ISO 10993-7.

Indications for Use

Indicated for use in endoscopic procedures requiring irrigation via an endoscope, specifically for patients undergoing procedures where the ERBEFLO 2 Disposable Tubing System and Pentax gastrointestinal video endoscopes are utilized.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 27, 2019 ERBE USA, Inc. John Tartal Director of Quality and Regulatory Affairs 2225 Northwest Parkway Marietta, GA 30067 K191438 Re: > Trade/Device Name: ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCX Dated: May 29, 2019 Received: May 30, 2019 Dear John Tartal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191438 Device Name ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector Indications for Use (Describe) The ERBEFLO 2 Disposable Tubing System is intended to provide sterile water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### May 2019 ### 510(k) SUMMARY | Submitted By: | Erbe USA, Inc.<br>2225 Northwest Parkway<br>Marietta, GA 30067<br>Tel: 770-955-4400<br>Fax: 770-955-2577 | |---------------------------------------|----------------------------------------------------------------------------------------------------------| | Contact Person: | John Tartal<br>Director of Quality and Regulatory Affairs | | Date Prepared: | May 29, 2019 | | Common Name: | Endoscopic Irrigation Tubing System Accessory/Scope Port Connector | | Trade/Proprietary Name: | ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector | | Classification Name: | Endoscopes and accessories (21 CFR Part 876.1500) | | Regulatory Class: | II | | Product Code: | OCX | | Legally Marketed<br>Predicate Device: | ERBEFLO® 2 Endo QuickConnect Scope Port Connectors, 510(k) Number<br>K143186 | ### Device Description: In general; the ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector will be manufactured with materials or agents that can be used in the medical device industry such as plastics, silicone, solvent, etc. The device adjoins an irrigation line of an of an ERBEFLO Tubing/Cap Set designated for a Pentax Scope to a Pentax Gastrointestinal Video Endoscope with a forward water jet channel for irrigation in endoscopic procedures. The Port Connector has a standard female luer lock connection that attaches to a designated ERBEFLO Tubing/Cap Set for Pentax Scopes, back flow (check) valve, and housing (a scope specific connector for a Pentax Scope). The Port Connector is provided sterile, is single patient use, and is disposable. ### Intended Use: The ERBEFLO 2 Disposable Tubing System is intended to provide sterile water from a water source through an irrigation pump and an endoscope (or to an endoscope) for endoscopic procedures. Note: The ERBEFLO 2 Endo Quick Connect Pentax Scope Port Connector is an accessory/single use connector for the ERBEFLO® 2 Disposable Tubing System. {4}------------------------------------------------ ### Similarities and Differences of the Proposed Device to the Current Device (Predicate Comparison/Substantial Equivalence): ## Similarities The proposed ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector is dimensionally the same from an exterior point of view as the current Connector with the same backflow valve, packaged the same, sterilized the same via Ethylene Oxide, single use (i.e., use for a patient), and is disposable like the predicate device. ### Differences The proposed ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector is different than the predicate in that its Port Connector housing is manufactured with a different plastic resin. Additionally since the last submission, there were some minor internal changes in which the core depth was increased along with a slight increase to the ring seal. Evaluations and testing as described below demonstrated the safety and efficacy of the Port Connector. ### Evaluations and Testing: The following evaluations and tests demonstrate safety and efficacy. ### Bioloqical Evaluation The evaluation was performed per the current recognized standard and demonstrated that there were no biocompatibility issues with the materials used for the proposed Port Connector. ### 2X Sterilization Functional Testing Visual inspection, connections testing, as well as leak and flow testing demonstrated that the proposed device upon 2X sterilization met established performance specifications. ### Sterilization Evaluation The evaluation was performed using current recognized standards and demonstrates product sterility as well as the Connector meets ethylene oxide residual requirements. Applied Standards (or as applicable adoption of previous work regarding a standard) AAMI / ANSI / ISO 10993-1, ISO 594-1, ISO 594-2, AAMI / ANSI / ISO 15223-1, AAMI / ANSI / ISO 11607-1, AAMI / ANSI / ISO 11607-2, AAMI / ANSI / ISO 11135(-1), AAMI / ANSI / ISO 10993-7 ### Conclusion: The proposed device has the same intended use (i.e., single patient use), principles of operation, and technological characteristics as the predicate device. As compared to the proposed Connector is constructed with the same type of materials as well as has the same performance characteristics. In conclusion, the ERBEFLO® 2 Endo Quick Connect Pentax® Scope Port Connector is safe and efficacious.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...