OxySoft (olifilcon C) silicone hydrogel soft contact lens
K191397 · Visco Vision, Inc. · LPL · Oct 2, 2019 · Ophthalmic
Device Facts
Record ID
K191397
Device Name
OxySoft (olifilcon C) silicone hydrogel soft contact lens
Applicant
Visco Vision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Oct 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The OxySoft (olifileon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.
Device Story
OxySoft is a daily disposable silicone hydrogel soft contact lens (olifilcon C, 54% water) containing a UV-blocking benzotriazole monomer and Reactive Blue 19 handling tint. Available in spherical, toric, and multifocal designs. Prescribed by eye care practitioners for daily wear; discarded after single use, requiring no cleaning or disinfection. Device corrects refractive errors (myopia, hyperopia, astigmatism, presbyopia) in phakic or aphakic patients. Benefits include vision correction and convenience of a daily disposable modality. Clinical performance evaluated via visual acuity, slit lamp examination, and symptom monitoring.
Clinical Evidence
One-month prospective clinical study (n=36; 24 trial, 12 control) comparing OxySoft to Si-Hy (olifilcon B). Subjects (22-59 years) wore lenses daily for up to 9 hours. Primary endpoints: visual acuity, slit lamp findings, symptoms, and adverse reactions. Results: Mean VA similar between groups (corrected to 0.1 LogMAR or better); no VA decrease >2 lines; no significant slit lamp findings (Grade 3/4); no adverse events reported.
Technological Characteristics
Material: 46% olifilcon C, 54% water. UV blocker: benzotriazole monomer. Handling tint: Reactive Blue 19. Manufacturing: Molded. Dimensions: 13.0-15.0 mm diameter, 8.0-9.2 mm base curve. Oxygen permeability (Dk): 100 (Fatt method). Refractive index: 1.396. Sterile, buffered saline packaging. Standards: ISO 18369.
Indications for Use
Indicated for phakic or aphakic persons with non-diseased eyes requiring correction of myopia, hyperopia, astigmatism (up to 2.25D), or presbyopia. Age range 22-59 years. Contraindicated in patients with ocular disease.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lens (K160344)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2019
Visco Vision Inc. % Ms. Jennifer Ting Manager Jens Medical Consulting Ltd. 3F, No. 364, Jixian Rd., Luzhou Dist. 247 New Taipei City, Taiwan R.O.C.
Re: K191397
Trade/Device Name: OxySoft (olifilcon C) silicone hydrogel soft contact lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 26, 2019 Received: August 27, 2019
Dear Ms. Ting:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191397
#### Device Name
OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens
#### Indications for Use (Describe)
The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.
The OxySoft (olifileon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.
The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.
Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: auto;"> <span style="text-align: start;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: auto;"> <span style="word-spacing: 0px;"> <span style="-webkit-text-stroke-width: 0px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> | <span style="font-family: sans-serif;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: auto;"> <span style="text-align: start;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: auto;"> <span style="word-spacing: 0px;"> <span style="-webkit-text-stroke-width: 0px;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Preparation Date: July 16, 2019
## 5.1 Establishment Information:
| Name | Visco Vision Inc. |
|-----------|-------------------------------------------------------------|
| Address | No. 1, Xingye St., Guishan Dist., Taoyuan City 33341 Taiwan |
| Contact | Ted TT Huang |
| Phone No. | 886-3-3596868 |
| Fax No. | 886-3-3490202 |
## 5.2 Contact Person:
| Name | Evan Huang |
|-----------|-------------------------------|
| Phone No. | 886-3-3596868 |
| Fax No. | 886-3-3490202 |
| e-mail: | evan.huang@viscovision.com.tw |
# 5.3 Device Identification:
| Proprietary Name | OxySoft (olifilcon C) Silicone Hydrogel Soft Contact<br>Lens |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Soft (hydrophilic) Contact Lenses |
| Classification Name | Lenses, Soft Contact, Daily Wear,<br>(21 CFR 886.5925, Product Code LPL)<br>Lenses, Soft Contact, Daily Wear (Disposable),<br>(21 CFR 886.5925, Product Code MVN) |
| Classification | II |
# 5.4 Legally Marketed Equivalent Device:
| Predicate Device Name | Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact<br>Lens |
|-----------------------|------------------------------------------------------------|
| Manufacturer | Visco Vision Inc. |
| 510(k) Number | K160344 |
| Product Code | MVN |
{4}------------------------------------------------
## 5.5 Device Description
The OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 46% olifilcon C and 54 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 1.9% (< 5%) in the UVB range of 280-315nm and 14.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.
### 5.6 Indication for Use:
The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.
The OxySoft (olifilcon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.
The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.
Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is required.
{5}------------------------------------------------
## 5.7 Technological characteristic
OxySoft SPHERICAL Soft Contact Lenses characteristics:
- Diameter Range:13.0 to 15.0 mm .
- Base Curve:8.0 to 9.2 mm .
- . Center Thickness:0.08 mm for -3.00D (varies with power)
- Power:+20.00 to -20.00 D o
OxySoft TORIC Soft Contact Lenses characteristics:
- . Diameter Range:13.0 to 15.0 mm
- . Base Curve:8.0 to 9.2 mm
- Center Thickness:0.08 mm for -3.00D (varies with power) .
- Power:+20.00 to -20.00 D .
- Cylinder: -0.75D ~ -2.25D .
- . Axis: 10° to 180° (in 10° increments)
OxySoft MULTIFOCAL Soft Contact Lenses characteristics:
- Diameter Range:13.0 to 15.0 mm
- . Base Curve:8.0 to 9.2 mm
- Center Thickness:0.08 mm for -3.00D (varies with power) .
- Power:+20.00 to -20.00 D .
- Additional Powers: +0.75D ~ +1.25D (LOW) .
- +1.50D ~ +2.00D (MID)
- +2.25D ~ +2.75D (HIGH)
## 5.8 Comparison table:
The characteristic comparison to predicate device is summarized in the following table.
| Similarities and differences | | |
|------------------------------|--------------------------------------------------------------|------------------------------------------------------------|
| Item | Device | Predicate (K160344) |
| Product Name | OxySoft (olifilcon C) Silicone<br>Hydrogel soft contact lens | Si-Hy (olifilcon B) Silicone<br>Hydrogel soft contact lens |
| Manufacturer | VISCO VISION Inc. | VISCO VISION Inc. |
| Intended Use | Myopia, Hyperopia,<br>astigmatism, Presbyopia | The same |
| Lens Design | Spherical, toric, or multifocal | The same |
| Replacement Schedule | Daily Disposable (Single use) | The same |
| Chemical composition | olifilcon C | olifilcon B |
{6}------------------------------------------------
| Classification | Group II<br>(Nonionic, High water) | Group I (Nonionic, Low water) |
|------------------------------------|-----------------------------------------------------|--------------------------------------------------|
| Water Content | 54 % | 47 % (<50%), |
| Oxygen Permeability<br>(DK, 35°C ) | 100<br>(Fatt method) | 120<br>(Fatt method) |
| Base Curve Range (mm) | 8.0~9.2 | 8.0~9.2 |
| Diameter (mm) | 13.0~15.0 | 13.0~15.0 |
| Center Thickness | Varies with design and power<br>(0.08 mm at -3.00D) | Varies with design and power (0.08 mm at -3.00D) |
| Powers | -20.00D to +20.00D | -20.00D to +20.00D |
| Refractive Index | 1.396 | 1.410 |
| Light Transmittance | 94% | 94% |
| Blue handling tint | Reactive Blue19 | Reactive Blue19 |
| Method of Manufacture | Molded | Molded |
#### 5.9 Nonclinical Tests Performed
- Physiochemical studies were conducted according to ISO 18369 First 5.9.1 edition 2006-08-15, Ophthalmic optics - Contact lenses (Ophthalmic). The physical, optical and chemical properties of the lens are within established specifications for the lenses.
- 5.9.2 Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment.
#### 5.10 Clinical Studies
A one-month clinical study was conducted to demonstrate the safety and efffectiveness of the OxySoft (olifilcon C) silicone hydrogel soft contact lens by comparison with Si-Hy (olifilcon B) silicone hydrogel soft contact lens when worn on a daily wear basis. At least 30 evaluable subjects were participated. Parameters measured include visual acuities, slit lamp findings, symptom, problem and complaints, adverse reactions, as well as lens wearing time.
The study evaluated thirty six (36) male and female subjects who were randomized and dispensed lenses in a 2:1 ratio with 24 subjects dispensed into the Trial lenses and 12 Subjects dispensed into the Control lenses. Of the 36 subjects who were enrolled in the study, 35 subjects were female and 1 subject was male. All of them are Asia people. The subject ranged in age from 22 years to 59 years. The gender ratio is 1 male to 35 females. One subject is new contact lens wearer (2.78%) and 35 subjects (97.22%) have successful experience with daily wear soft contact lens.
{7}------------------------------------------------
In trial group, 24 subjects range in age from 22 years to 59 years with average 33.29±10.02 years. All of the subjects are female. The gender ratio is 24 to 0 (24 female/0 male). In control group, 12 subjects range in age from 25 years to 46 years with average 35.58±7.09 years. One subject is male and 11 subjects female. The gender ratio was 11 to 1 (11 female/1 male).
It was found that the OxySoft silicone hydrogel soft contact lens is safe as a daily disposable single use lens to wear up to 9 hours a day during 30 days follow up by comparing to Si-Hy (olifilcon B) silicone hydrogel soft contact lens. There is no symptom or syndrome that requires treatment for all subjects during study period.
The mean visual acuity (VA) for both lenses was similar at each visit. The visual acuity could be corrected to 0.1 LogMAR or better. None of the subjects had VA decreased more than 2 lines. No significant slit lamp findings (i.e. Grade 3 or Grade 4) were found. No adverse event (AE) occurred.
## 5.11 Conclusion
Comparison to the predicate device for chemical composition, physical and optical properties, it shows that "OxySoft silicone hydrogel soft contact lens" is as safe, as effective and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.