AcQMap High Resolution Imaging and Mapping System

K191392 · Acutus Medical, Inc. · DQK · Sep 6, 2019 · Cardiovascular

Device Facts

Record IDK191392
Device NameAcQMap High Resolution Imaging and Mapping System
ApplicantAcutus Medical, Inc.
Product CodeDQK · Cardiovascular
Decision DateSep 6, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2

Indications for Use

The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. OR When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.

Device Story

Diagnostic recording system for EP labs; images, navigates, and maps atrial chambers. Inputs: ultrasound data (via AcQMap catheter), ECG/EGM signals, and impedance-based localization signals. Transforms inputs into 3D cardiac chamber reconstructions, catheter position displays, and electrical activity maps (charge density or voltage). Operated by physicians; used in EP labs. Output: visual display of cardiac anatomy and electrical activation on workstation. Assists physicians in identifying arrhythmia sources and guiding procedures. Benefits: enables visualization of complex arrhythmias difficult to identify with conventional systems; allows remapping during procedures; supports traditional contact mapping without AcQMap catheter.

Clinical Evidence

Clinical data incorporated by reference from the original AcQMap System (K170948). DDRAMATIC-SVT study: prospective, non-randomized, open-label, 84 patients at 8 international sites. Demonstrated safety and effectiveness for intended use. No new clinical testing required for this modification.

Technological Characteristics

Console, workstation, and auxiliary interface box. Ultrasound imaging (10 MHz), ECG/EGM recording, and impedance-based localization. Dimensions: 99cm x 58cm x 76cm; Weight: 80kg. Power: 100-127/220-230 VAC. Safety: IEC 60601-1, Class I, Type CF/BF. Connectivity: wired catheter/electrode inputs. Software: diagnostic recording and mapping algorithm.

Indications for Use

Indicated for patients undergoing prescribed electrophysiology procedures. Used for chamber reconstruction, visualization of anatomy, and mapping of complex arrhythmias (charge density or voltage-based) using AcQMap catheters, or for displaying catheter position and cardiac electrical activity using conventional EP catheters.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 6, 2019 Acutus Medical, Inc. Karla Schaffner Principal Regulatory Affairs Specialist 2210 Faraday Ave., Suite 100 Carlsbad, California 92008 Re: K191392 Trade/Device Name: AcQMap High Resolution Imaging and Mapping System, Model 900100 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, IYO, ITX Dated: August 12, 2019 Received: August 13, 2019 Dear Karla Schaffner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191392 #### Device Name AcQMap ® High Resolution Imaging and Mapping System, Model 900100 #### Indications for Use (Describe) The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. #### AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. OR When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Notification K191392 ### General Information [807.92(a)(1)] Date Prepared: 23 May 2019 ### Applicant: Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6080 Fax: 1-442-232-6081 ### Contact Person: Karla Schaffner Principal Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6161 FAX: 1-442-232-6081 Email: karla.schaffner@acutus.com ## DEVICE INFORMATION [807.92(a)(2)] #### Trade Name: AcQMap® High Resolution Imaging and Mapping System, Model 900100 ### Generic/Common Name: Programable diagnostic computer and Ultrasonic pulsed echo imaging system ### Classification: Class 11/21 CFR § 870.1425 and Class 11/21 CFR § 892.1560 # Product Code: DQK, IYO, ITX {4}------------------------------------------------ ## PREDICATE DEVICE(S) [807.92(a)(3)] Predicate Device AcQMap High Resolution Imaging and Mapping System, Model 900100 (K190131) Reference Device CARTO® 3 EP Navigation System, Version 6.0 And Accessories (K170600) ## DEVICE DESCRIPTION [807.92(a)(4)] The AcQMap High Resolution Imaging and Mapping System, Model 900100 operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box. The AcQMap High Resolution Imaging and Mapping System, Model 900100 ("AcQMap System Model, 900100") is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart. The AcQMap System hardware consists of three functional subsystems: - . Ultrasound imaging, - . ECG and EGM recording; and - o Impedance based electrode Localization. The AcQMap System, Model 900100 is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter (cleared under K170819). The AcQMap System provides: - . 3-D cardiac chamber reconstruction - Contact and non-contact (ultrasound), - . Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters, - o Cardiac electrical activity as waveform traces, - . Contact LAT and voltage amplitude maps - . Remapping of the chamber at any time during the procedure; and - Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction . to show chamber-wide electrical activation. The AcQMap System, Model 900100 is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia. Additionally, the modified AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one another within the coording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter. {5}------------------------------------------------ ### INDICATIONS FOR USE [807.92(a)(5)] The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. ### AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. #### OR When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure. ## Comparison of Technological Characteristics with the Predicate Devices [807.92(A)(6)] Tables I provides a comparison of the modified AcQMap System classifications for use against the predicate and reference devices. Table 2 provides a comparison of the technological characteristics for the modified AcQMap System against the predicate device. | Table I. Comparison of Classification with the Predicate and Reference Devices | | | | | |-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Subject Device | Predicate Device | Reference Device | Rationale for Substantial Equivalence | | Characteristics | AcQMap® High<br>Resolution Imaging and<br>Mapping System, Model<br>900100 | AcQMap® High<br>Resolution Imaging and<br>Mapping System, Model<br>900100 (K190131) | CARTO® 3 EP<br>Navigation System,<br>Version 6.0 And<br>Accessories | | | 510(k) Number | TBD | K190131 | K170600 | -- | | Classification/<br>Regulation<br>Number/<br>Regulation Name/<br>Product Code | Class II/ 21 CFR §<br>870.1425/ Programable<br>diagnostic computer/<br>DQK<br>Class II/ 21 CFR §<br>892.1560/ Ultrasonic<br>pulsed echo imaging<br>system/ IYO, ITX | Class II/ 21 CFR §<br>870.1425/ Programable<br>diagnostic computer/<br>DQK<br>Class II/ 21 CFR §<br>892.1560/ Ultrasonic<br>pulsed echo imaging<br>system/ IYO, ITX | Class II/ 21 CFR §<br>870.1425/<br>Programable<br>diagnostic<br>computer/DQK | The subject device is<br>identical to the<br>predicate. The<br>additional product code<br>and regulation number<br>for the subject and<br>predicate device versus<br>the reference device<br>are required. | | Table 2. Comparison of Indications for Use with the Predicate and Reference Devices | | | | | | | Subject Device | Predicate Device | Reference Device | | | Characteristics | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 (K190131) | CARTO® 3 EP<br>Navigation System,<br>Ver. 6.0 And<br>Accessories | Rationale for<br>Substantial<br>Equivalence | | 510(k) Number | TBD | K190131 | K170600 | | | Indications for<br>Use | The AcQMap System is intended for use in<br>patients for whom electrophysiology<br>procedures have been prescribed.<br><br>When used with the AcQMap Catheters, the<br>AcQMap System is intended to be used to<br>reconstruct the selected chamber from<br>ultrasound data for purposes of visualizing the<br>chamber anatomy and displaying electrical<br>impulses as either charge density-based or<br>voltage-based maps of complex arrhythmias that<br>may be difficult to identify using conventional<br>mapping systems alone.<br><br>AND<br><br>When used with the specified Patient System<br>Electrodes, the AcQMap System is intended to<br>display the position of AcQMap Catheters and<br>conventional electrophysiology (EP) catheters in<br>the heart.<br><br>OR<br><br>When used with conventional electrophysiology<br>catheters, the AcQMap System provides<br>information about the electrical activity of the<br>heart and about catheter location during the<br>procedure. | The AcQMap System is intended for use in<br>patients for whom electrophysiology<br>procedures have been prescribed.<br><br>When used with the AcQMap Catheters, the<br>AcQMap System is intended to be used to<br>reconstruct the selected chamber from<br>ultrasound data for purposes of visualizing the<br>chamber anatomy and displaying electrical<br>impulses as either dipole density-based or<br>voltage-based maps of complex arrhythmias that<br>may be difficult to identify using conventional<br>mapping systems alone.<br><br>AND<br><br>When used with the specified Patient<br>Electrodes, the AcQMap System is intended to<br>display the position of AcQMap Catheters and<br>conventional electrophysiology (EP) catheters in<br>the heart. | The intended use of<br>the CARTO® 3<br>System is catheter-<br>based cardiac<br>electrophysiological<br>(EP) procedures. The<br>CARTO® 3 System<br>provides information<br>about the electrical<br>activity of the heart<br>and about catheter<br>location during the<br>procedure. The<br>system can be used on<br>patients who are<br>eligible for a<br>conventional<br>electrophysiological<br>procedure. The<br>system has no special<br>contraindications. | All devices are<br>intended for<br>electrophysiology<br>procedures. The<br>additional paragraph<br>in the modified<br>AcQMap System<br>indications for use is<br>aligned with the<br>indications for use<br>of the reference<br>device. The<br>indications for use<br>in the modified<br>AcQMap System are<br>otherwise identical<br>to the predicate.<br>The additional<br>capability raises no<br>new questions of<br>safety or<br>effectiveness as<br>demonstrated by<br>the AcQMap<br>performance testing. | | Table 3: Comparison of Technological Characteristics Against the Predicate Device | | | | | | Characteristics | Subject Device | Predicate Device | Rationale for Substantial Equivalence | | | | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 (K190131) | | | | Patient Anatomy | Intracardiac Structures | Intracardiac Structures | Identical | | | Testing to Support<br>Substantial<br>Equivalence | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, and Animal Testing | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, Animal Testing; and Clinical Testing | Complete performance<br>testing conducted by<br>Acutus demonstrates that<br>the AcQMap System<br>performs as intended and<br>that there are no different<br>questions of safety or<br>effectiveness. | | | System Safety<br>Standards | IEC 60601-1:2005 /A1:2012 IEC 60601-1-2:2014 IEC 60601-1-6:2010/A1:2013 IEC 60601-2-25:2015 IEC 60601-2-37:2015 | IEC 60601-1:2005 /A1:2012 IEC 60601-1-2:2014 IEC 60601-1-6:2010/A1:2013 IEC 60601-2-25:2015 IEC 60601-2-37:2015 | Identical. There are no<br>changes to hardware. | | | System<br>Components | Console Workstation Workstation Cable Auxiliary Interface Box ECG Input Cable Ampere Ablation Catheter Adapter Cable Ampere RF Generator Adapter Cable ECG Output Cable Ablation Reference Cable Ablation Electrogram Cable ECG w/Snaps Cable ECG POST Cable 2mm Pin Jumper Set | Console Workstation Workstation Cable Auxiliary Interface Box ECG Input Cable Ampere Ablation Catheter Adapter Cable Ampere RF Generator Adapter Cable ECG Output Cable Ablation Reference Cable Ablation Electrogram Cable ECG w/Snaps Cable ECG POST Cable 2mm Pin Jumper Set | Identical | | | Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) | | | | | | Characteristics | Subject Device | Predicate Device | Rationale for Substantial<br>Equivalence | | | Visual/Mapping<br>Characteristics | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 3-D cardiac chamber reconstructions -<br>Contact and non-contact (ultrasound); Three-dimensional position of the AcQMap<br>Catheter and conventional electrophysiology<br>catheters; Cardiac electrical activity as waveform traces; Contact LAT and voltage amplitude maps; Remapping of the chamber at any time during<br>the procedure; Dynamic, three-dimensional, Charge Density<br>maps overlaid on the cardiac chamber<br>reconstruction to show chamber-wide<br>electrical activation. | AcQMap® High Resolution Imaging and Mapping<br>System, Model 900100 (K190131) 3-D cardiac chamber reconstructions imaging; Three-dimensional position of the AcQMap<br>Catheter and conventional electrophysiology<br>catheters; Cardiac electrical activity as waveform traces; Remapping of the chamber at any time during<br>the procedure; and Dynamic, three-dimensional, Dipole Density<br>maps overlaid on the cardiac chamber<br>reconstruction to show chamber-wide<br>electrical activation. | The additional capability to<br>display cardiac electrical<br>activity and catheter<br>location without the<br>AcQMap Catheter present<br>does not raise different<br>questions of safety or<br>effectiveness as<br>demonstrated through<br>performance testing. | | | Visualization<br>Device/Catheter | AcQMap Catheter (electrodes & transducers)<br>or<br>Conventional electrophysiology catheters | AcQMap Catheter (electrodes & transducers) | | | | Physical Characteristics - Console/Amplifier Comparison | | | | | | Dimensions | 99 cm L x 58 cm W x 76 cm D | 99 cm L x 58 cm W x 76 cm D | Identical | | | Weight Maximum | 80 kg | 80 kg | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...